Best focus IRTC Hband Z= 48.0mm

REC 20100926_214405 Camera lens X=-61.6 Y=-20.2

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20101013_223603 rr 6.3 90 0 214405 1 399 1000 0.3 0.4 3 1716 64.6 H 15 1000  
20101013_223734 rr 6.3 90 0 214405 1 399 1000 0.3 0.4 3 1734 65.8 H 15 1000  
20101013_223913 rr 6.3 90 0 214405 1 399 1000 0.3 0.4 3 1742 64.1 H 15 1000  
20101013_224101 rr 6.3 90 0 214405 1 399 1000 0.3 0.4 3 1740 63.6 H 15 1000  

In queste condizioni veniva il 70% di SR il 26 settembre

Si sospetta un fuorifuoco della IRTC. Si mette a mano Z=47.7

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20101013_225614 rr 6.4 90 0 214405 1 399 1000 0.3 0.4 3 1628 58.6 H 15 1000  
20101013_225750 rr 6.4 90 0 214405 1 399 1000 0.3 0.4 3 1685 59.2 H 15 1000  
20101013_225929 rr 6.4 90 0 214405 1 399 1000 0.3 0.4 3 1667 59.4 H 15 1000  

NO sembra che il best focus a 48 sia meglio!

Proviamo a Z=48.1

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20101013_230653 rr 6.3 90 0 214405 1 399 1000 0.3 0.4 3 1734 62.7 H 15 1000  
20101013_230833 rr 6.3 90 0 214405 1 399 1000 0.3 0.4 3 1696 63.4 H 15 1000  
20101013_230959 rr 6.3 90 0 214405 1 399 1000 0.3 0.4 3 1716 62.9 H 15 1000  

Quindi il massimo dello SR e' davvero dove lo trova la procedura AutoFocusIrtcGui e lo SR in questa condizione cambia di 2% con 100um di slitta Z
Topic revision: r2 - 11 Sep 2012, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.