Bin3 INTMAT con KL_V6 e 270deg

Decidiamo di tenerci le pupille vecchie, visto che a bin 1 la differenza fra pup vecchie e nuove e' di poche subap.

Rotazione copiata dal BIN 1, ovvero: 270deg di offset

=======================================

DISTURBOIM 20100928_193153 KL v6
Modal basis = kl v6
type = Push-pull
Frames per mov = 3
Cycle = 60
#Modes = 10
Amp = optimized a mano 6l
Note =

=======================================

INTMATACQ 20100928_195346 10 modi clock 270
Pup = 20100528_235150
Dist = 20100928_193153
Iteration = 4
Loop speed = 600
bin = 3
Modul = 6 l/d
Rec.Close loop = none
Gain vector close loop = none

=======================================

REC 20100928_195547 10 modi TTremoval off
Modal Basis:Modes = kl_V6
intmatacq = 20100928_195346
n modes = 10
cutted = 0
Note: ...............

=======================================

===================================

DISTURBOIM 20100928_200951
Modal basis = kl v6
type = Push-pull
Frames per mov = 3
Cycle = 5
#Modes = 66
Amp = optimized a mano 6l
Note =

===================================

INTMATACQ 20100928_ 66 modi clock 270
Pup = 20100528_235150
Dist = 20100928_193153
Iteration = 16
Loop speed = 600
bin = 3
Modul = 6 l/d
Rec.Close loop = 20100928_195547
Gain vector close loop = 0.05

===================================

REC #
Modal Basis:Modes = kl_v6
intmatacq = #
n modes = 66
TT removal = ON/OFF
cutted = #
Note: ...............

=======================================

PROBLEMA: SBAGLIA a ricombinare i modi! E' il secondario che si era impippato, dopo un reset e' ripartito bene

===============================

INTMATACQ 20100928_211628 66 modi clock 270
Pup = 20100528_235150
Dist = 20100928_193153
Iteration = 16
Loop speed = 600
bin = 3
Modul = 6 l/d
Rec.Close loop = 20100928_195547
Gain vector close loop = 0.05
===================================

OK, ma ampiezze un po' bassine

Proviamo ad aumentare le ampiezze

===============================

DISTURBOIM 20100928_212328
Modal basis = kl v6
type = Push-pull
Frames per mov = 3
Cycle = 5
#Modes = 66
Amp = optimized a mano 6l per 2
Note =

===============================

INTMATACQ 20100928_212346 66 modi clock 270
Pup = 20100528_235150
Dist = 20100928_212328
Iteration = 16
Loop speed = 600
bin = 3
Modul = 6 l/d
Rec.Close loop = 20100928_195547
Gain vector close loop = 0.05
=====================

REC 20100928_212832

Modal Basis: kl_v6
intmatacq = 20100928_212346
n modes = 65
TT removal = OFF
cutted = #
Note: OK chiude 0.5;05 @700Hz

===================================

Test REC 20100928_212832 no disturb

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100928_214453 ??? 7.7 90 0 212832 3 65 699 0.6 0.7 6 6985 52.2 H 50 200  
20100928_214950 ??? 7.7 90 0 212832 3 65 699 0.6 0.7 6 6892 52.5 H 50 200  
20100928_215018 ??? 7.7 90 0 212832 3 65 699 0.6 0.7 6 6891 52.9 H 50 200  
20100928_215100 ??? 7.7 90 0 212832 3 65 699 0.6 0.7 6 6883 54.7 H 50 200  

=================

REC 20100928_215420

Modal Basis: kl_v6
intmatacq = 20100928_212346
n modes = 65
TT removal = ON
cutted = #
Note: Regge gain:0.8; 0.6 @700Hz

===============================

Test REC 20100928_215420 no disturb

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100928_215628 ??? 7.7 90 0 215420 3 65 699 0.6 0.8 6 6868 46.2 H 50 200  
20100928_215700 ??? 7.7 90 0 215420 3 65 699 0.6 0.8 6 6880 47.4 H 50 200  
20100928_215729 ??? 7.7 90 0 215420 3 65 699 0.6 0.8 6 6884 50.7 H 50 200  
20100928_215815 ??? 7.7 90 0 215420 3 65 699 0.6 0.7 6 6885 -1.0 ? -1 -1  
20100928_215833 ??? 7.7 90 0 215420 3 65 699 0.6 0.7 6 6884 53.5 H 50 200  

Test REC 20100928_212832 disturb 0.8"

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100928_220754 ??? 7.7 90 0 212832 3 65 699 0.7 0.9 6 7009 20.9 H 50 200  
20100928_220829 ??? 7.7 90 0 212832 3 65 699 0.7 0.9 6 7009 22.6 H 50 200  
20100928_220855 ??? 7.7 90 0 212832 3 65 699 0.7 0.9 6 7007 21.8 H 50 200  
20100928_220926 ??? 7.7 90 0 212832 3 65 699 0.7 0.9 6 7014 21.1 H 50 200  

Per paragone proviamo il REC vecchio 20100529_004214

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100928_221210 ??? 7.7 90 0 004214 3 66 699 0.6 0.7 6 7008 19.8 H 50 200  
20100928_221241 ??? 7.7 90 0 004214 3 66 699 0.6 0.7 6 7007 20.1 H 50 200  
20100928_221309 ??? 7.7 90 0 004214 3 66 699 0.6 0.7 6 7006 19.2 H 50 200  

Come check delle condizioni di vibrazioni testiamo il BIN 1 20100926_214405

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100928_222021 ??? 6.4 90 0 214405 1 399 1000 0.3 0.4 3 1617 65.7 H 20 200  
20100928_222046 ??? 6.4 90 0 214405 1 399 1000 0.3 0.4 3 1598 65.3 H 20 200  
20100928_222109 ??? 6.4 90 0 214405 1 399 1000 0.3 0.4 3 1640 27.0 H 10 200  
20100928_222134 ??? 6.4 90 0 214405 1 399 1000 0.3 0.4 3 1640 66.8 H 10 400  
20100928_222318 ??? 6.3 90 0 214405 1 399 1000 0.3 0.4 3 1838 66.0 H 10 400  

BIN 4 nuova intmat con KL v6 e 270deg

Facciamo il 10 modi

===================================

DISTURBOIM 20100928_223711 KL v6
Modal basis = kl v6
type = Push-pull
Frames per mov = 3
Cycle = 60
#Modes = 10
Amp = optimized a mano 6l per 2
Note =

===========================

INTMATACQ 20100928_223736 10 modi clock 270
Pup = 20100528_075347
Dist = 20100928_212328
Iteration = 16
Loop speed = 600
bin = 4
Modul = 6 l/d
Rec.Close loop = 20100928_195547
Gain vector close loop = 0.05
=====================

REC 20100928_223910 10 modi

Modal Basis: kl_v6
intmatacq = 20100928_223736
n modes = 10
TT removal = OFF
cutted = #
Note:

===============================

===================================

DISTURBOIM 20100928_224107 KL v6
Modal basis = kl v6
type = Push-pull
Frames per mov = 3
Cycle = 18
#Modes = 37
Amp = optimized a mano 6l per 2
Note =

===========================

===========================

INTMATACQ 20100928_224337 kl_v6 clock 270
Pup = 20100528_075347
Dist = 20100928_224107
Iteration = 16
Loop speed = 600
bin = 4
Modul = 6 l/d
Rec.Close loop = 20100928_223910
Gain vector close loop = 0.05
=====================

REC 20100928_224817 36 modi

Modal Basis: kl_v6
intmatacq = 20100928_224337
n modes = 10
TT removal = OFF
cutted = #
Note:

===========================

Test REC 20100928_224817 cloop no disturb

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100928_225532 ??? 8.4 90 0 224817 4 36 600 0.4 1.1 6 8056 65.2 H 25 400  
20100928_225901 ??? 8.4 90 0 224817 4 36 701 0.3 0.7 6 6838 62.4 H 25 400  
20100928_230034 ??? 8.4 90 0 224817 4 36 701 0.4 0.7 6 6831 64.5 H 25 400  
20100928_230201 ??? 8.4 90 0 224817 4 36 701 0.4 0.7 6 6812 63.2 H 25 400  
20100928_230249 ??? 8.4 90 0 224817 4 36 701 0.4 0.7 6 6788 63.7 H 25 400  

Riproviamo il REC a bin 3 20100928_212832

Test no disturb

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100928_230647 ??? 7.8 90 0 212832 3 65 699 0.5 0.7 6 6023 61.3 H 25 400  
20100928_230736 ??? 7.8 90 0 212832 3 65 699 0.5 0.7 6 6020 58.1 H 25 400  
20100928_230821 ??? 7.8 90 0 212832 3 65 699 0.5 0.7 6 6018 59.7 H 25 400  
20100928_230914 ??? 7.8 90 0 212832 3 65 699 0.5 0.7 6 6016 59.8 H 25 400  

Riprendiamo il REC a bin 3

===========================

INTMATACQ 20100928_231203 66 modi clock 270
Pup = 20100528_235150
Dist = 20100928_212328
Iteration = 16
Loop speed = 700
bin = 3
Modul = 6 l/d
Rec.Close loop = 20100928_195547
Gain vector close loop = 0.05
=====================

REC 20100928_231743 65 modi TTremoval off
Modal Basis:Modes = kl_V6
intmatacq = 20100928_231203
n modes = 65
cutted = 0
Note: ...............

===================================

Test cloop REC 20100928_231743
TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100928_232104 ??? 7.8 90 0 231743 3 65 699 0.4 0.7 6 5993 54.3 H 25 400  
20100928_232146 ??? 7.8 90 0 231743 3 65 699 0.4 0.7 6 5992 53.6 H 25 400  
20100928_232312 ??? 7.8 90 0 231743 3 65 699 0.4 0.7 6 5986 54.9 H 25 400  

Riproviamo ancora il rec vecchio 20100928_212832

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100928_232446 ??? 7.8 90 0 212832 3 65 699 0.4 0.7 6 5983 61.4 H 25 400  
20100928_232533 ??? 7.8 90 0 212832 3 65 699 0.4 0.7 6 5978 61.0 H 25 400  
20100928_232614 ??? 7.8 90 0 212832 3 65 699 0.4 0.7 6 5976 59.5 H 25 400  