You are here: Foswiki>FLAO Web>LbtLab>LabLog20100728 (28 Jul 2010, EnricoPinna)Edit Attach
Simone Enrico

AGW restarted

Pup check
Board setup Pup_check_internal (freq CCD39 200Hz)
self.flowerpot.setOperatingState(0)
self.fw1.moveTo(-0.000004)
self.fw2.moveTo(0.999997)
self.rerot.moveTo(1.830000)
self.cuberot.moveTo(181.633767)
self.cubestage.moveTo(-4.400000)
self.stagex.moveTo(-75.049100)
self.stagey.moveTo(35.257400)
self.stagez.moveTo(51.086400)
self.lens.moveTo(9.700000,25.200000)
self.tt.set(amp=3.200000,freq=0.000000,offx=0.000000, offy=0.000000)
self.adc._motor1.moveTo(-0.036177)
self.adc._motor2.moveTo(-0.009044)

CCD39 Acquired: Pup_internal_kprism_check.fits

-- EnricoPinna - 27 Jul 2010
  • Pup_Cehck_kprism.jpg:
    Pup_Cehck_kprism.jpg
Controlliamo anche il CCD47

Frames acquisiti:CCD47 Test.fits

  • CCD47_check.jpg:
    CCD47_check.jpg
ADC TEST OK

CCD39 RON TEST

Stages off ma elettronica calda (45C invece dei 30C soliti)
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 22.59 - RON (e-): 11.29 (estimated)
Speed: 2500 kpix/sec - RON (ADU): 22.59 - RON (e-): 11.29 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.18 - RON (e-): 6.59 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.18 - RON (e-): 6.59 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.16 - RON (e-): 4.08 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.16 - RON (e-): 4.08 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.22 - RON (e-): 4.61 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.22 - RON (e-): 4.61 (estimated)

Altra misura
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.40 - RON (e-): 11.70 (estimated)
Speed: 2500 kpix/sec - RON (ADU): 23.40 - RON (e-): 11.70 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.35 - RON (e-): 6.68 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.35 - RON (e-): 6.68 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.22 - RON (e-): 4.11 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.22 - RON (e-): 4.11 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.48 - RON (e-): 4.74 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.48 - RON (e-): 4.74 (estimated)

Muovi leggi rirotatore

Minmax pup X (pix): 2.6657
Minmax pup Y (pix): 3.0428
Detailed data saved to: /home/aoacct/testw2/measurements/MuoviLeggiRotatoreTT/20100727_165542

  • muovi_leggi_rerot.jpg:
    muovi_leggi_rerot.jpg
Seconda prova

Minmax pup X (pix): 2.8398
Minmax pup Y (pix): 2.8871
Detailed data saved to: /home/aoacct/testw2/measurements/MuoviLeggiRotatoreTT/20100727_171220

  • muovi_leggi_rerot_2.jpg:
    muovi_leggi_rerot_2.jpg
Campo con FW thorlabs installata

X(Y=35) = -71; -80

Y(X-75)= 32; 41

RON TEST CON FW thorlabs accesa e stages spenti, tutto il resto ON


SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.22 - RON (e-): 11.61 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.66 - RON (e-): 6.83 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.40 - RON (e-): 4.20 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.81 - RON (e-): 4.91 (estimated)

Richiuduamo le scatole e facciamo il test finale

con stages spenti, fioriera spenta, Thorlabs FW accesa


SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 21.54 - RON (e-): 10.77 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.22 - RON (e-): 6.61 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.18 - RON (e-): 4.09 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.71 - RON (e-): 4.85 (estimated)
I Attachment Action Size Date Who Comment
CCD47_check.jpgjpg CCD47_check.jpg manage 121 K 27 Jul 2010 - 16:38 UnknownUser  
Pup_Cehck_kprism.jpgjpg Pup_Cehck_kprism.jpg manage 156 K 27 Jul 2010 - 16:30 UnknownUser  
muovi_leggi_rerot.jpgjpg muovi_leggi_rerot.jpg manage 130 K 27 Jul 2010 - 17:00 UnknownUser  
muovi_leggi_rerot_2.jpgjpg muovi_leggi_rerot_2.jpg manage 120 K 27 Jul 2010 - 17:15 UnknownUser  
Topic revision: r3 - 28 Jul 2010, EnricoPinna