Wref94 17:10:18.53 +40:41:23.8 0.017 0.030 R=7.180 J=1.832 H=0.752

Automatic setup BIN1 Full start AO andato liscio per la prima volta (in 2-3min)

Wref95 18:54:29.17 +42:31:43.4 -0.004 -0.010 R=7.120 J=3.447 H=2.528

Automatic setup eseguito correttamente in 2m 34s

loop con setup automatico e gain uguale a quello delle sere precedenti con questo ricostruttore
TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100626_082929 ??? 6.5 78 1 185716 1 400 1000 0.5 1.0 3 1519 77.2 H 1 1000  
20100626_083109 ??? 6.5 78 1 185716 1 400 1000 0.5 1.0 3 1505 64.2 H 1 1000  
20100626_083250 ??? 6.5 78 1 185716 1 400 1000 0.5 1.0 3 1497 66.2 H 1 1000  
20100626_083834 ??? 6.5 77 1 185716 1 400 1000 0.5 1.0 3 1521 85.9 H 1 1000 10^4 di contrasto
20100626_084224 ??? 6.5 77 1 185716 1 400 1000 0.5 1.0 3 1512 79.1 H 1 1000  
20100626_084411 ??? 6.5 76 2 185716 1 400 1000 0.5 1.0 3 1518 81.1 H 1 1000  
20100626_084620 ??? 6.5 76 1 185716 1 400 1000 0.5 1.0 3 1515 24.4 H 10 1000 saturata
20100626_084829 ??? 6.5 76 1 185716 1 400 1000 0.5 1.0 3 1518 -1.0 ? -1 -1 saturata
20100626_084849 ??? 6.5 76 1 185716 1 400 1000 0.5 1.0 3 1519 25.5 H 10 1000 saturata
20100626_085357 ??? 6.5 75 1 185716 1 400 1000 0.5 1.0 3 1511 0.6 H 1000 30 super-saturata
20100626_085437 ??? 6.5 75 1 185716 1 400 1000 0.5 1.0 3 1515 0.7 H 1000 30 super-saturata

DIMM Seeing 0.9" @8:45 UT
DIMM Seeing 0.9" @9:08 UT
DIMM Seeing 0.8" @9:19 UT

Test Nuovo ricotruttore 500 modi 2l/D non giunstato
TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100626_090946 ??? 6.5 73 1 050509 1 500 1000 0.7 1.0 2 1500 83.1 H 1 1000  
20100626_091121 ??? 6.5 72 1 050509 1 500 1000 0.7 1.0 2 1499 85.2 H 1 1000  
20100626_091255 ??? 6.5 72 1 050509 1 500 1000 0.7 1.0 2 1494 85.7 H 1 1000  
20100626_091713 ??? 6.5 71 0 050509 1 500 1000 0.7 1.0 2 1495 76.3 H 1 1000  
20100626_091904 ??? 6.5 71 1 050509 1 500 1000 0.7 1.0 2 1492 49.8 H 1 1000 save skip permanente

DIMM Seeing 0.8" @9:19 UT

Mettimao il ricostruttore 500 modi giuntato Sbagliat al modulazione: rec a 2l/D e loop 3l/D
TrackNo RefStar Mag El Wind RecSorted ascending bin #mod freq gain mod nph SR band exp #frames notes
20100626_093721 ??? 6.5 68 1 062918 1 500 1000 0.7 1.0 3 1466 82.2 H 1 1000  
20100626_093856 ??? 6.5 68 1 062918 1 500 1000 0.7 1.0 3 1498 84.1 H 1 1000  
20100626_094040 ??? 7.3 67 1 062918 1 500 1000 0.7 1.0 3 718 85.2 H 1 1000  
20100626_094212 ??? 6.9 67 1 062918 1 500 1000 0.7 1.0 3 1036 74.8 H 1 1000  

Con altri dark dava SR piu alti

-- CarmeloArcidiacono - 26 Jun 2010