You are here: Foswiki>FLAO Web>LbtTel>LabLog20100611 (11 Jun 2010, EnricoPinna)Edit Attach

Work in progress on HBS, HBS off all the time

LC Offet @BIN1 di oggi Dx=21.2, Dy=-7.0

Prendiamo loop BIN1 per vedere le vibrazioni che ci sono Cloop no disturb tree house accesa
TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100610_231802 ??? 6.4 90 0 185716 1 400 1000 0.3 0.4 2 1668 47 H 3 500  

Closed loop TreeHous Fan OFF

Open loop con subframe a freq limitata a 270Hz. HBS off, TreeHouse Off
20100611_000146

Closed loop 1kHz con subframe a freq limitata a 270Hz. HBS off, TreeHouse Off
20100611_001144

Closed loop 500Hz con subframe a freq limitata a 270Hz. HBS off, TreeHouse Off
20100611_001749

Closed loop 200Hz con subframe a freq limitata a 270Hz. HBS off, TreeHouse Off
20100611_002146

Closed loop 100Hz con subframe a freq limitata a 270Hz. HBS off, TreeHouse Off
20100611_002525

TEST BIN4 @mag 13.3

La lampada fa allo scatto 1 16.5, al 2 ca 13.0
Mettiamo mag 13.0 per confrontarci con il BIN3 dell'altra volta.

Facciamo diverse frequenze con il disturbo per vedere lo SR come cambia e poi paragonarlo con quelle del BIN3 anche se oggi le vibrazioni sono molto maggiori.
300Hz
TrackNo RefStar Mag El Wind Rec binSorted ascending #mod freq gain mod nph SR band exp #frames notes
20100611_004901 ??? 13.3 90 0 081806 4 36 300 0.6 1.8 6 179 10 H 1000 20 disturbo all frames 1000
20100611_005124 ??? 13.3 90 0 081806 4 36 300 0.6 1.8 6 188 9 H 1000 20 disturbo all frames 1000
20100611_005151 ??? 13.2 90 0 081806 4 36 300 0.6 1.8 6 199 9 H 1000 20 disturbo all frames 1000

Passiamo a 500Hz
TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100611_005526 ??? 13.2 90 0 081806 4 36 501 0.6 1.4 5 117 12 H 1000 20  
20100611_005559 ??? 13.2 90 0 081806 4 36 501 0.6 1.4 5 119 13 H 1000 20  
20100611_005625 ??? 13.2 90 0 081806 4 36 501 0.6 1.4 5 121 12 H 1000 20  

Passiamo a 700Hz
TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100611_005946 ??? 10.4 90 0 081806 4 36 701 0.6 0.7 5 1112 10 H 1000 20  
20100611_010011 ??? 13.2 90 0 081806 4 36 701 0.6 0.7 5 86 9 H 1000 20  
20100611_010036 ??? 13.2 90 0 081806 4 36 701 0.6 0.7 5 86 11 H 1000 20  

Test BIN4 @mag 15.7

200Hz
TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100611_010644 ??? 15.7 90 0 081806 4 36 200 0.6 2.4 6 29 5 H 1000 20  
20100611_010712 ??? 15.7 90 0 081806 4 36 200 0.6 2.4 6 29 12 H 1000 20  
20100611_010751 ??? 15.7 90 0 081806 4 36 200 0.6 2.4 6 29 17 H 1000 20  
20100611_010843 ??? 15.7 90 0 081806 4 36 200 0.6 2.4 6 30 5 H 1000 40  
20100611_010944 ??? 15.7 90 0 081806 4 36 200 0.6 2.4 6 31 8 H 1000 40  

Prendiamo una senza disturbo per vedere a differenza
TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100611_011234 ??? 15.7 90 0 081806 4 36 200 0.6 2.0 6 29 12 H 1000 40 no disturb

400Hz
TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100611_011629 ??? 15.2 90 0 081806 4 36 401 0.6 2.0 6 23 6 H 1000 40  
20100611_011731 ??? 15.3 90 0 081806 4 36 401 0.6 2.0 6 22 -2147483648 H 1000 40  
20100611_011821 ??? 15.3 90 0 081806 4 36 401 0.6 2.0 6 22 8 H 1000 40  

-- EnricoPinna - 10 Jun 2010
Topic revision: r1 - 11 Jun 2010, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback