The last night of this week. We start as usual with a bright star

Tonight's conditions: Sky partially clouded.

Star: Wref162, Rmag=9.5.

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100604_033641 wref162 9.3 79 1 185716 1 400 1000 0.5 0.9 2 110 -1 ? -1 -1  
20100604_033658 wref162 9.3 79 1 185716 1 400 1000 0.5 0.9 2 110 73 H 50 200  
20100604_033728 wref162 9.3 79 1 185716 1 400 1000 0.5 0.9 2 110 81 H 50 200  
20100604_033759 wref162 9.3 79 1 185716 1 400 1000 0.5 0.9 2 110 78 H 50 200  
20100604_034111 wref162 9.3 78 1 185716 1 400 1000 0.5 1.2 2 115 75 H 50 200  
20100604_034149 wref162 9.3 78 1 185716 1 400 1000 0.5 1.2 2 115 76 H 50 200  
20100604_034224 wref162 9.3 78 1 185716 1 400 1000 0.5 1.1 2 115 84 H 50 200 RECORD!
20100604_034333 wref162 9.3 78 2 185716 1 400 1000 0.5 1.1 2 114 83 H 50 200  
20100604_034452 wref162 9.3 78 1 185716 1 400 1000 0.5 1.1 2 115 83 H 50 200  
20100604_034533 wref162 9.3 77 1 185716 1 400 1000 0.5 1.1 2 114 82 H 50 200  
20100604_034756 wref162 9.3 77 1 185716 1 400 1000 0.5 1.1 2 115 -1 ? -1 -1 irtc.fits missing
20100604_034818 wref162 9.3 77 1 185716 1 400 1000 0.5 1.1 2 115 81 H 50 200  
20100604_034850 wref162 9.3 77 2 185716 1 400 1000 0.5 1.1 2 115 81 H 50 500  
20100604_035035 wref162 9.3 77 1 185716 1 400 1000 0.5 1.1 2 114 79 H 50 500  
20100604_035327 wref162 9.3 76 1 185716 1 400 1000 0.0 0.0 2 111 0 H 2000 10 OL ITRTC fwhm 1.23";1.17"

20100604_035327 Open-loop image showing a seeing of about 0.8 arcsec:

We take saturated images to see cleary the rings:

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100604_035732 wref162 9.3 75 1 185716 1 400 1000 0.5 1.1 2 116 35 H 500 50  
20100604_040051 wref162 9.3 75 1 185716 1 400 1000 0.5 1.1 2 115 35 H 500 50  
20100604_040130 wref162 9.3 75 1 185716 1 400 1000 0.5 1.1 2 115 36 H 500 50  

A cloud passing by makes us get less photons!

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100604_041951 wref162 10.1 80 1 185716 1 400 1000 0.5 1.1 2 54 78 H 50 50  
20100604_042035 wref162 10.1 80 1 185716 1 400 1000 0.5 1.1 2 54 75 H 30 50  
20100604_042055 wref162 10.1 80 1 185716 1 400 1000 0.5 1.1 2 54 66 H 30 200  
20100604_042155 wref162 10.1 80 1 185716 1 400 1000 0.5 1.1 2 52 74 H 30 200  
20100604_042236 wref162 10.2 80 1 185716 1 400 1000 0.5 1.1 2 48 70 H 30 200  
20100604_042855 wref162 10.4 79 1 185716 1 400 1000 0.5 1.1 2 39 -1 ? -1 -1  
20100604_042915 wref162 10.4 79 1 185716 1 400 1000 0.5 1.1 2 41 73 H 30 200  
20100604_042949 wref162 10.2 79 1 185716 1 400 1000 0.5 1.1 2 50 69 H 30 200  

The cloud is gone! we get all photons again! yuppi

Note: Last two tracknum of the following set contain saturated images.

TrackNo RefStar Mag El WindSorted ascending Rec bin #mod freq gain mod nph SR band exp #frames notes
20100604_044506 wref162 9.9 76 1 185716 1 400 1000 0.5 1.0 2 65 83 H 30 200  
20100604_044646 wref162 9.9 76 1 185716 1 400 1000 0.5 1.0 2 64 77 H 30 500  
20100604_044853 wref162 9.9 75 1 185716 1 400 1000 0.5 1.0 2 63 34 H 200 50 Saturated image
20100604_044918 wref162 9.9 75 1 185716 1 400 1000 0.5 1.0 2 66 43 H 200 100 Saturated image
20100604_044540 wref162 9.9 76 2 185716 1 400 1000 0.5 1.0 2 64 76 H 30 500  
20100604_044828 wref162 9.9 75 2 185716 1 400 1000 0.5 1.0 2 65 -1 ? -1 -1  

Star: Wref72, Rmag = 13.5

We change to bin2

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100604_054114 wref72 14.2 76 2 054335 2 153 200 0.5 0.9 6 27 -1 ? -1 -1 IRTC missing
20100604_054156 wref72 14.2 76 1 054335 2 153 200 0.5 0.9 6 26 11 H 2000 20  
20100604_054251 wref72 14.3 76 1 054335 2 153 200 0.5 0.9 6 25 12 H 2000 20  
20100604_054348 wref72 14.3 76 1 054335 2 153 200 0.5 0.9 6 25 10 H 2000 20  

We change to bin3

Clouds preclude us from testing binning modes 2 and 3.... photons vanish interceded by the pass of clouds over our line of sight.

We switch back to bin1

Star: Wref5, Rmag=5

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100604_064115 wref5 4.8 78 1 185716 1 400 1000 0.6 0.9 2 7284 78 H 0 500  
20100604_064208 wref5 4.8 78 1 185716 1 400 1000 0.6 0.9 2 7338 85 H 0 500 RECORD!
20100604_064302 wref5 4.8 78 2 185716 1 400 1000 0.6 0.9 2 7306 82 H 0 500  
20100604_064351 wref5 4.8 78 1 185716 1 400 1000 0.6 0.9 2 7167 80 H 0 500  
20100604_064806 wref5 4.8 79 1 185716 1 400 1000 0.6 0.9 2 6992 40 J 1 500  
20100604_064924 wref5 4.9 79 1 185716 1 400 1000 0.6 0.9 2 6379 15 J 1 500  
20100604_065018 wref5 4.8 79 2 185716 1 400 1000 0.6 0.9 2 7175 40 J 1 500  
20100604_065114 wref5 4.8 79 1 185716 1 400 1000 0.6 0.9 2 7223 40 J 1 500  
20100604_065653 wref5 4.7 80 1 185716 1 400 1000 0.0 0.0 2 7504 1 H 5 500 OL IRTC FWHM 0.94";0.96"
20100604_065813 wref5 4.8 80 1 185716 1 400 1000 0.0 0.0 2 7322 1 H 5 500 OL IRTC FWHM 0.96";0.97"
20100604_071232 wref5 4.7 80 1 185716 1 400 1000 0.4 0.4 2 7499 86 H 0 500

kalman: ammazza1c

RECORD ASSOLUTO!!!!

20100604_071329 wref5 4.7 80 1 185716 1 400 1000 0.4 0.4 2 7493 82 H 0 500 ammazza1c
20100604_071420 wref5 4.7 80 1 185716 1 400 1000 0.3 0.3 2 7533 83 H 0 500 ammazza1c
20100604_072310 wref5 4.7 80 1 185716 1 400 1000 0.5 0.5 2 7502 77 H 0 500 ammazza1c
20100604_072402 wref5 4.7 80 1 185716 1 400 1000 0.5 0.5 2 7490 70 H 0 500 ammazza1c
20100604_072500 wref5 4.8 80 1 185716 1 400 1000 0.5 0.5 2 7214 32 H 0 500 ammazza1c
20100604_073226 wref5 4.7 80 1 185716 1 400 1000 0.5 0.5 2 7485 80 H 0 500 ammazza1a
20100604_073314 wref5 4.7 80 1 185716 1 400 1000 0.5 0.5 2 7460 81 H 0 500 ammazza1a
20100604_073402 wref5 4.7 80 1 185716 1 400 1000 0.5 0.5 2 7437 72 H 0 500 ammazza1a
20100604_074339 wref5 4.7 79 1 185716 1 400 1000 0.1 0.5 2 7517 18 H 0 500  
20100604_074448 wref5 4.7 79 1 185716 1 400 1000 0.1 0.5 2 7565 16 H 0 500  
20100604_074538 wref5 4.7 79 1 185716 1 400 1000 0.1 0.5 2 7546 18 H 0 500  
20100604_074933 wref5 4.7 78 1 185716 1 400 1000 0.0 0.0 2 7499 1 H 5 500 OL IRTC FWHM 1.06";1.00"
20100604_075023 wref5 4.8 78 2 185716 1 400 1000 0.0 0.0 2 7325 1 H 5 500 OL IRTC FWHM 1.07";0.98"

Wref203 R=5.0

Si cerca il pianetino. Alcune delle immagini sotto sono saturate

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100604_084119 wref203 5.9 63 3 185716 1 400 1000 0.6 1.0 2 2457 -1 ? -1 -1  
20100604_084138 wref203 5.9 63 3 185716 1 400 1000 0.6 1.0 2 2518 53 H 10 1000  
20100604_084322 wref203 6.0 62 3 185716 1 400 1000 0.6 1.0 2 2428 50 H 10 1000  
20100604_085058 wref203 5.9 61 3 185716 1 400 1000 0.6 1.0 2 2458 20 H 40 1000  
20100604_085317 wref203 6.0 60 3 185716 1 400 1000 0.6 1.0 2 2419 19 H 40 1000  
20100604_085536 wref203 6.0 60 4 185716 1 400 1000 0.6 1.0 2 2298 18 H 40 1000  
20100604_085742 wref203 5.9 59 2 185716 1 400 1000 0.6 1.0 2 2500 20 H 40 1000  
20100604_085956 wref203 5.9 59 4 185716 1 400 1000 0.6 1.0 2 2523 21 H 40 1000  
20100604_090424 wref203 5.9 58 2 185716 1 400 1000 0.6 0.9 2 2518 10 J 80 1000  
20100604_090706 wref203 5.9 57 3 185716 1 400 1000 0.6 0.9 2 2526 11 J 80 1000  
20100604_090947 wref203 5.9 57 3 185716 1 400 1000 0.6 0.9 2 2528 10 J 80 1000  
20100604_091336 wref203 5.9 56 3 185716 1 400 1000 0.6 0.9 2 2499 -1 ? -1 -1  
20100604_091401 wref203 5.9 56 2 185716 1 400 1000 0.6 0.9 2 2519 32 J 8 1000  

Wref84 R=13.6

non va. problemi di telescopio.

Problemi al W: stages in timeout, qualche problema di rete. Troviamo il cubo in posizione pennina...

Wref11 aka Wref32 R 5.8

alla caccia del pianetino

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph SR band exp #frames disturb notes
20100604_110255 wref11 6.8 54 0 0.00 0.96 185716 1 400 1000 0.8 1.0 3 1094 41.4 H 10 30 ONSKY  
20100604_110318 wref11 6.8 54 0 0.00 0.76 185716 1 400 1000 0.8 1.0 3 1130 58.9 H 10 1000 ONSKY  
20100604_110824 wref11 6.8 55 0 0.00 1.04 185716 1 400 1000 0.8 1.0 3 1080 4.3 H 1000 60 ONSKY  
20100604_111106 wref11 6.8 56 0 0.00 0.81 185716 1 400 1000 0.8 1.0 3 1133 4.1 H 1000 60 ONSKY  
20100604_111219 wref11 6.8 56 0 0.00 0.71 185716 1 400 1000 0.8 1.0 3 1146 4.5 H 1000 60 ONSKY  
20100604_111334 wref11 6.8 56 0 0.00 0.74 185716 1 400 1000 0.8 1.0 3 1140 4.2 H 1000 60 ONSKY  
20100604_111717 wref11 6.8 57 0 0.00 0.77 185716 1 400 1000 0.8 1.0 3 1134 ** H 1000 60 ONSKY  
20100604_112313 wref11 6.8 58 0 0.00 1.00 185716 1 400 1000 0.8 1.0 3 1093 4.1 H 1000 60 ONSKY  
20100604_112501 wref11 6.8 59 0 0.00 0.74 185716 1 400 1000 0.8 1.0 3 1139 4.3 H 1000 60 ONSKY  
20100604_112651 wref11 6.8 59 0 0.00 0.76 185716 1 400 1000 0.8 1.0 3 1139 4.4 H 1000 60 ONSKY  
20100604_112817 wref11 6.8 59 0 0.00 0.82 185716 1 400 1000 0.8 1.0 3 1131 4.3 H 1000 60 ONSKY  
20100604_112933 wref11 6.8 60 0 0.00 0.66 185716 1 400 1000 0.8 1.0 3 1157 4.4 H 1000 60 ONSKY  
20100604_113440 wref11 6.8 61 0 0.00 0.65 185716 1 400 1000 0.7 1.0 3 1155 4.4 H 1000 60 ONSKY  
20100604_113551 wref11 6.8 61 0 0.00 0.70 185716 1 400 1000 0.7 1.0 3 1151 4.3 H 1000 60 ONSKY  
20100604_113706 wref11 6.8 61 0 0.00 0.77 185716 1 400 1000 0.7 1.0 3 1136 4.3 H 1000 60 ONSKY  
20100604_113922 wref11 6.8 62 0 0.00 0.69 185716 1 400 1000 0.7 1.0 3 1150 2.7 J 1000 60 ONSKY  
20100604_114304 wref11 6.8 62 0 0.00 0.65 185716 1 400 1000 0.7 1.0 3 1161 3.0 J 1000 60 ONSKY  
20100604_114418 wref11 6.7 63 0 0.00 0.57 185716 1 400 1000 0.7 1.0 3 1176 2.9 J 1000 60 ONSKY  
20100604_114533 wref11 6.7 63 0 0.00 0.60 185716 1 400 1000 0.7 1.0 3 1176 2.9 J 1000 60 ONSKY  

-- FernandoQuiros - 04 Jun 2010