On Sky

da 20100531_032000 GMT a 20100531_123000 GMT


Puntata WREF1 10:59:27.97 40:25:48.98

HIP 53721 R=4.8


  • Ricostruttore 200 modi bin 1
  • 3.30? shell ripped
  • 3.37 shell ripped
  • 03.53 shell ripped
  • 04.11 shell ripped


TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames
20100531_034258 Wref1 NaN 65 2 185716 1 400 1000 0.5 0.8 2 -1 -1 ? -1 -1
20100531_034607 Wref1 NaN 65 1 185716 1 400 1000 0.5 0.8 2 -1 -1 ? -1 -1
20100531_034625 Wref1 NaN 65 2 185716 1 400 1000 0.5 0.8 2 -1 62 H 2 1000
20100531_034856 Wref1 NaN 64 1 185716 1 400 1000 0.5 0.8 2 -1 59 H 2 1000
20100531_035038 Wref1 NaN 64 2 185716 1 400 1000 0.5 0.8 2 -1 62 H 2 1000
20100531_042018 Wref1 5.5 58 3 185716 1 400 1000 0.5 0.8 2 3599 58 H 2 1000
20100531_042254 Wref1 5.6 58 2 185716 1 400 1000 0.5 1.3 2 3516 57 H 2 1000
20100531_042529 Wref1 5.5 57 2 185716 1 400 1000 0.5 1.3 2 3587 42 H 2 1000
20100531_042935 Wref1 5.5 56 2 174206 1 200 1000 0.6 0.9 2 3637 59 H 2 1000
20100531_043133 Wref1 5.5 56 2 174206 1 200 1000 0.6 0.9 2 3633 58 H 2 1000


Kalman filter

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nphSorted ascending SR band exp #frames notes
20100531_044353 Wref1 5.6 54 3 174206 1 200 1000 0.4 0.4 2 3421 22 H 2 1000  
20100531_044710 Wref1 5.6 53 2 174206 1 200 1000 0.5 0.5 2 3517 28 H 2 1000  
20100531_044902 Wref1 5.5 53 2 174206 1 200 1000 0.8 0.8 2 3545 32 H 2 1000  
20100531_044156 Wref1 5.5 54 2 174206 1 200 1000 0.8 0.8 2 3606 45 H 2 1000  
20100531_045123 Wref1 5.5 52 2 174206 1 200 1000 0.5 0.5 2 3626 37 H 2 1000  
20100531_045356 Wref1 5.5 52 3 174206 1 200 1000 0.8 0.8 2 3638 40 H 2 1000  
20100531_043755 Wref1 5.5 55 3 174206 1 200 1000 0.4 0.4 2 3651 41 H 2 1000  
20100531_043618 Wref1 5.5 55 2 174206 1 200 1000 0.4 0.4 2 3654 45 H 2 1000  


Puntata WREF93 14:13:46.14 +37:34:23.8

HD 124624 - Star - R 7.11 H 4.4

problemi di puntamento di stella guida. errore nel file


TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames
20100531_054007 Wref93 7.6 80 2 185716 1 400 1000 0.4 0.9 2 546 -1 ? -1 -1
20100531_054055 Wref93 7.6 80 1 185716 1 400 1000 0.4 0.9 2 546 -1 ? -1 -1
20100531_054115 Wref93 7.6 80 2 185716 1 400 1000 0.4 0.9 2 546 71 H 3 1000
20100531_054853 Wref93 7.6 79 3 185716 1 400 1000 0.4 0.9 2 549 72 H 3 1000
20100531_055033 Wref93 7.6 78 2 185716 1 400 1000 0.4 0.9 2 548 65 H 3 1000
20100531_055236 Wref93 7.6 78 3 185716 1 400 1000 0.4 0.9 2 550 70 H 3 1000
20100531_055445 Wref93 7.6 78 3 185716 1 400 1000 0.4 0.9 2 550 71 H 3 1000


Spenta la ventilazione nella tree-house. Ripuntata la Wref93


TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames
20100531_062043 Wref93 7.6 73 2 185716 1 400 1000 0.4 1.1 2 546 -1 ? -1 -1
20100531_062113 Wref93 7.6 73 2 185716 1 400 1000 0.4 1.1 2 547 75 H 3 1000
20100531_062252 Wref93 7.6 72 2 185716 1 400 1000 0.4 1.1 2 547 72 H 3 1000
20100531_062427 Wref93 7.6 72 2 185716 1 400 1000 0.4 1.1 2 551 58 H 3 1000
20100531_062602 Wref93 7.6 72 2 185716 1 400 1000 0.4 1.0 2 548 75 H 3 1000
20100531_062934 Wref93 7.6 71 2 185716 1 400 1000 0.4 1.0 2 551 40 J 10 1000
20100531_063404 Wref93 7.6 70 2 185716 1 400 1000 0.4 1.0 2 553 47 J 10 1000
20100531_063937 Wref93 7.6 69 2 185716 1 400 1000 0.4 1.0 2 552 45 J 10 1000
20100531_064130 Wref93 7.6 69 2 185716 1 400 1000 0.4 1.0 2 548 40 J 10 1000


Serie di loop fatti per sbaglio a 10modi. Volevano essere open loop

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames
20100531_065326 Wref93 7.8 66 2 005930 2 10 100 0.0 0.0 29 19875 0 J 500 60
20100531_065414 Wref93 8.0 66 1 005930 2 10 100 0.0 0.0 29 15206 0 J 500 60
20100531_065919 Wref93 7.6 65 2 005930 2 10 100 0.1 0.1 29 22714 1 H 200 60
20100531_070017 Wref93 7.7 65 3 005930 2 10 100 0.1 0.1 29 21563 0 H 200 60
20100531_070124 Wref93 7.7 65 3 005930 2 10 100 0.0 0.0 29 21284 1 H 200 60
20100531_070210 Wref93 7.6 65 3 005930 2 10 100 0.0 0.0 29 22138 1 H 200 60


prove con ricostruttore a Bin 2

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames notes
20100531_071036 Wref93 7.6 63 2 003208 2 153 991 0.7 1.2 3 2355 10 H 10 60  
20100531_071124 Wref93 7.5 63 2 003208 2 153 991 0.7 1.2 3 2426 21 H 10 1000  
20100531_071333 Wref93 7.5 62 2 003208 2 153 991 0.7 1.2 3 2574 27 H 10 1000  
20100531_071512 Wref93 7.5 62 3 003208 2 153 991 0.7 1.2 3 2582 36 H 10 1000  
20100531_072135 Wref93 7.5 61 2 003208 2 153 991 0.7 1.2 3 2606 41 H 5 1000  
20100531_072359 Wref93 7.5 60 2 003208 2 153 991 0.7 1.2 3 2605 42 H 5 1000  
20100531_072709 Wref93 7.5 60 3 003208 2 153 991 0.7 1.2 3 2639 55 H 5 1000  
20100531_072850 Wref93 7.5 59 1 003208 2 153 991 0.7 1.2 3 2632 48 H 5 1000  
20100531_073227 Wref93 7.5 58 1 003208 2 153 991 0.7 1.2 3 2636 26 J 10 1000 dark vecchio


Puntata WREF63 17:10:41.280 +39:53:32.30

TYC 3077-1681-1 R 9.4 H 6.4


TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames
20100531_075529 Wref63 9.7 83 1 003208 2 153 991 0.7 1.0 3 323 13 J 100 1000
20100531_075915 Wref63 9.7 83 0 003208 2 153 991 0.7 1.0 3 323 54 H 30 1000
20100531_080117 Wref63 9.7 83 1 003208 2 153 991 0.7 1.0 3 320 46 H 30 1000
20100531_080312 Wref63 9.7 83 1 003208 2 153 991 0.7 1.0 3 324 41 H 30 1000
20100531_081244 Wref63 9.7 82 1 003208 2 153 991 0.7 1.0 3 321 41 H 30 1000
20100531_081436 Wref63 9.7 82 1 003208 2 153 991 0.7 1.0 3 325 45 H 30 1000
20100531_081630 Wref63 9.7 82 1 003208 2 153 991 0.7 1.0 3 324 46 H 30 1000
20100531_081826 Wref63 9.7 82 1 003208 2 153 991 0.7 1.0 3 326 49 H 30 1000


Puntata WREF111 17:10:41.280 +39:53:32.30

TYC 3077-1681-1 , R 11.450 , H 8.63


TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames
20100531_085035 Wref111 11.9 77 3 003208 2 153 625 0.6 1.0 2 72 33 H 500 36
20100531_085116 Wref111 11.9 77 3 003208 2 153 625 0.6 1.0 2 73 33 H 500 100
20100531_085226 Wref111 11.9 77 3 003208 2 153 625 0.6 1.0 2 73 31 H 500 100
20100531_085331 Wref111 11.8 76 3 003208 2 153 625 0.6 1.0 2 74 30 H 500 100
20100531_090112 Wref111 11.8 75 3 003208 2 153 625 0.6 1.0 2 74 26 H 500 100
20100531_090219 Wref111 11.8 75 3 003208 2 153 625 0.6 1.0 2 74 24 H 500 100
20100531_090326 Wref111 11.9 75 3 003208 2 153 625 0.6 1.0 2 67 29 H 500 100
20100531_090447 Wref111 11.9 74 3 003208 2 153 625 0.6 1.0 2 67 26 H 500 100
20100531_090948 Wref111 11.9 73 3 003208 2 153 625 0.6 1.2 2 67 33 H 500 100
20100531_091120 Wref111 11.9 73 3 003208 2 153 625 0.7 1.2 2 67 33 H 500 100
20100531_091302 Wref111 11.9 73 2 003208 2 153 625 0.7 1.2 2 67 33 H 500 100


Misure a bin 3

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames
20100531_093514 Wref111 11.9 68 2   3 -1 200 -1.0 -1.0 6 479 -1 ? -1 -1
20100531_093542 Wref111 11.9 68 3   3 -1 200 -1.0 -1.0 6 478 7 H 1000 60
20100531_093739 Wref111 12.0 68 3   3 -1 200 -1.0 -1.0 6 470 5 H 1000 60
20100531_093859 Wref111 12.0 68 3   3 -1 200 -1.0 -1.0 6 464 6 H 1000 60

MANCANO I RICOSTRUTTORI NEI DATI SALVATI (?)

Usare ee=getaoelab('20100531_093514', modal_rec='adsec_calib/adsec/M2C/KL_v4/RECs/Rec_20100529_004214.fits')


Puntata WREF114 17:17:12.17 +43:08:00.3

M92 - R 11.5


Spianata pupilla con 10 modi

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames
20100531_102153 Wref114 11.8 61 2 005930 2 10 100 0.1 0.1 29 472 -1 ? -1 -1

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames
20100531_102705 Wref114 11.8 60 1 003208 2 153 625 0.6 1.0 2 79   H 200 100
20100531_102743 Wref114 11.8 60 1 003208 2 153 625 0.6 1.0 2 78 35 H 200 100
20100531_102821 Wref114 11.8 60 1 003208 2 153 625 0.6 1.0 2 79 42 H 200 300
20100531_102948 Wref114 11.7 59 1 003208 2 153 625 0.6 1.0 2 80 37 H 200 300
20100531_103137 Wref114 11.7 59 2 003208 2 153 625 0.6 1.0 2 82 44 H 200 300
20100531_103313 Wref114 11.7 59 2 003208 2 153 625 0.6 1.0 2 82 39 H 200 300
20100531_103453 Wref114 11.7 58 2 003208 2 153 625 0.6 1.0 2 83 38 H 200 300
20100531_103631 Wref114 11.7 58 2 003208 2 153 625 0.6 1.0 2 84 44 H 200 300
20100531_103811 Wref114 11.7 58 2 003208 2 153 625 0.6 1.0 2 84 38 H 200 300
20100531_103946 Wref114 11.7 58 2 003208 2 153 625 0.6 1.0 2 85 39 H 200 300
20100531_104134 Wref114 11.7 57 1 003208 2 153 625 0.6 1.0 2 86 46 J 500 45
20100531_104213 Wref114 11.7 57 1 003208 2 153 625 0.6 1.0 2 86 44 J 500 200
20100531_104442 Wref114 11.7 57 1 003208 2 153 625 0.6 1.0 2 86 38 J 500 200
20100531_104708 Wref114 11.7 56 2 003208 2 153 625 0.6 1.0 2 87 40 J 500 200
20100531_104925 Wref114 11.7 56 1 003208 2 153 625 0.6 1.0 2 86 ? J 500 200
20100531_105144 Wref114 11.7 55 1 003208 2 153 625 0.6 1.0 2 86 ? J 500 200
20100531_105709 Wref114 11.7 54 1 003208 2 153 625 0.6 1.0 2 86 ? H 60 500
20100531_105849 Wref114 11.7 54 2 003208 2 153 625 0.6 1.0 2 87 ? H 60 500
20100531_110020 Wref114 11.7 54 1 003208 2 153 625 0.6 1.0 2 87 ? H 60 500
20100531_110139 Wref114 11.7 54 1 003208 2 153 625 0.6 1.0 2 87 ? H 60 500
20100531_110300 Wref114 11.7 53 1 003208 2 153 625 0.6 1.0 2 87 ? H 60 500
20100531_110434 Wref114 11.7 53 1 003208 2 153 625 0.6 1.0 2 87 ? H 60 500
20100531_110557 Wref114 11.6 53 2 003208 2 153 625 0.6 1.0 2 88 ? H 60 500
20100531_111108 Wref114 11.6 52 2 003208 2 153 625 0.6 1.0 2 89 ? ? -1 -1
20100531_111130 Wref114 11.6 52 1 003208 2 153 625 0.6 1.0 2 90 ? J 600 35
20100531_111209 Wref114 11.6 52 2 003208 2 153 625 0.6 1.0 2 89 ? J 600 100
20100531_111345 Wref114 11.6 51 1 003208 2 153 625 0.6 1.0 2 90 ? J 600 100
20100531_111511 Wref114 11.6 51 1 003208 2 153 625 0.6 1.0 2 91 ? J 600 100
20100531_111636 Wref114 11.6 51 2 003208 2 153 625 0.6 1.0 2 91 ? J 600 100
20100531_111752 Wref114 11.6 51 2 003208 2 153 625 0.6 1.0 2 91 ? J 600 100
20100531_111910 Wref114 11.6 50 1 003208 2 153 625 0.6 1.0 2 92 ? J 600 100
20100531_112033 Wref114 11.6 50 1 003208 2 153 625 0.6 1.0 2 92 ? J 600 100
20100531_112159 Wref114 11.6 50 2 003208 2 153 625 0.6 1.0 2 92 ? J 600 100
20100531_112321 Wref114 11.6 50 2 003208 2 153 625 0.6 1.0 2 93 ? J 600 100
20100531_112443 Wref114 11.6 49 1 003208 2 153 625 0.6 1.0 2 93 ? J 600 100
20100531_112609 Wref114 11.6 49 1 003208 2 153 625 0.6 1.0 2 94 ? J 600 100
20100531_112728 Wref114 11.6 49 2 003208 2 153 625 0.6 1.0 2 94 ? J 600 100
20100531_112848 Wref114 11.6 49 2 003208 2 153 625 0.6 1.0 2 94 ? J 600 100
20100531_113004 Wref114 11.6 48 1 003208 2 153 625 0.6 1.0 2 95 ? J 600 100
20100531_113236 Wref114 11.6 48 1 003208 2 153 625 0.6 1.0 2 95 ? H 60 100
20100531_113300 Wref114 11.6 48 2 003208 2 153 625 0.6 1.0 2 96 ? H 60 500
20100531_113426 Wref114 11.5 48 1 003208 2 153 625 0.6 1.0 2 98 ? H 60 500
20100531_113555 Wref114 11.5 47 1 003208 2 153 625 0.6 1.0 2 101 ? H 60 500
20100531_113804 Wref114 11.4 47 2 003208 2 153 625 0.6 1.0 2 107 ? H 60 500
20100531_113943 Wref114 11.4 47 1 003208 2 153 625 0.6 1.0 2 114 ? H 60 500

-- LorenzoBusoni - 31 May 2010
Topic revision: r4 - 31 May 2010, LorenzoBusoni
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback