AGW in bent gregorian platform.

Ambient temperature -2"C

Rimettiamo il board setup TestPup6 _bis.png:

-- AlfioPuglisi - 11 Mar 2010
  • Pupil screenshot:
    TestPup6_bis.png
Test readout noise, devices OFF: ccd47


Speed: 2500 kpix/sec - RON (ADU): 21.34 - RON (e-): 10.67 (estimated)
Speed: 890 kpix/sec - RON (ADU): 15.95 - RON (e-): 7.97 (estimated)
Speed: 400 kpix/sec - RON (ADU): 11.86 - RON (e-): 5.93 (estimated)
Speed: 150 kpix/sec - RON (ADU): 18.68 - RON (e-): 9.34 (estimated)

Test readout noise, devices OFF: ccd47, Flowerpot, LJ fans, filterwheels, stages 110V.

Speed: 2500 kpix/sec - RON (ADU): 20.50 - RON (e-): 10.25 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.56 - RON (e-): 7.28 (estimated)
Speed: 400 kpix/sec - RON (ADU): 11.02 - RON (e-): 5.51 (estimated)
Speed: 150 kpix/sec - RON (ADU): 16.14 - RON (e-): 8.07 (estimated)

We see that there is a small amount of light coming though.

With a better cover, now no light on ccd:

Speed: 2500 kpix/sec - RON (ADU): 20.50 - RON (e-): 10.25 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.32 - RON (e-): 7.16 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.60 - RON (e-): 5.30 (estimated)
Speed: 150 kpix/sec - RON (ADU): 15.08 - RON (e-): 7.54 (estimated)

Speed: 2500 kpix/sec - RON (ADU): 20.31 - RON (e-): 10.16 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.36 - RON (e-): 7.18 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.59 - RON (e-): 5.30 (estimated)
Speed: 150 kpix/sec - RON (ADU): 15.01 - RON (e-): 7.51 (estimated)

RESUME OF RON RESULTS

Lbt Lab measurement (from 20100217):

AG on:

Speed: 2500 kpix/sec - RON (ADU): 21.09 - RON (e-): 10.54 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.32 - RON (e-): 6.66 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.78 - RON (e-): 4.39 (estimated)

AG off:

Speed: 2500 kpix/sec - RON (ADU): 21.46 - RON (e-): 10.73 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.31 - RON (e-): 6.66 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.61 - RON (e-): 4.31 (estimated)

LBT telescope measurements:

AG on:

Speed: 2500 kpix/sec - RON (ADU): 20.31 - RON (e-): 10.16 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.36 - RON (e-): 7.18 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.59 - RON (e-): 5.30 (estimated)

Arcetri lab:

EVENT at 2010-03-11 22:21:01.049768

PIC controller either re-setted or lost power for a short time. Log from RelayCtrl:

RelayCtrl01        |TRA|     64692|2010-03-11 22:20:54.038140|             MAIN > Sending command "7"







RelayCtrl01        |TRA|     64693|2010-03-11 22:20:55.038016|             MAIN > SerialInString received (len 0)







RelayCtrl01        |TRA|     64694|2010-03-11 22:20:55.038037|             MAIN > SerialInString buffer (len 0)







RelayCtrl01        |TRA|     64695|2010-03-11 22:20:57.039023|             MAIN > SerialInString received (len 0)







RelayCtrl01        |TRA|     64696|2010-03-11 22:20:57.039049|             MAIN > SerialInString buffer (len 0)







RelayCtrl01        |TRA|     64697|2010-03-11 22:20:59.040061|             MAIN > SerialInString received (len 0)







RelayCtrl01        |TRA|     64698|2010-03-11 22:20:59.040087|             MAIN > SerialInString buffer (len 0)







RelayCtrl01        |TRA|     64699|2010-03-11 22:21:01.049768|             MAIN > SerialInString received (len 129) À^M







RelayCtrl01        |TRA|     64699|2010-03-11 22:21:01.049768|             MAIN >  .    ----------    Test Pic18F97J60  -  Com1-------^M







RelayCtrl01        |TRA|     64699|2010-03-11 22:21:01.049768|             MAIN >  .                Mauro Sozzi @ Gennaio 2008^M







RelayCtrl01        |TRA|     64699|2010-03-11 22:21:01.049768|             MAIN >  .     ---------- Per Help Comandi







RelayCtrl01        |TRA|     64700|2010-03-11 22:21:01.049813|             MAIN > SerialInString buffer (len 129) À^M







RelayCtrl01        |TRA|     64700|2010-03-11 22:21:01.049813|             MAIN >  .    ----------    Test Pic18F97J60  -  Com1-------^M







RelayCtrl01        |TRA|     64700|2010-03-11 22:21:01.049813|             MAIN >  .                Mauro Sozzi @ Gennaio 2008^M







RelayCtrl01        |TRA|     64700|2010-03-11 22:21:01.049813|             MAIN >  .     ---------- Per Help Comandi







RelayCtrl01        |TRA|     64701|2010-03-11 22:21:01.049830|             MAIN > EmptySerial(): received string: À^M







RelayCtrl01        |TRA|     64701|2010-03-11 22:21:01.049830|             MAIN >  .







RelayCtrl01        |TRA|     64702|2010-03-11 22:21:01.050019|             MAIN > SerialInString buffer (len 126)    ----------    Test Pic18F97J60  -  Com1-------^M







RelayCtrl01        |TRA|     64702|2010-03-11 22:21:01.050019|             MAIN >  .                Mauro Sozzi @ Gennaio 2008^M







RelayCtrl01        |TRA|     64702|2010-03-11 22:21:01.050019|             MAIN >  .     ---------- Per Help Comandi







RelayCtrl01        |TRA|     64703|2010-03-11 22:21:01.050030|             MAIN > EmptySerial(): received string:    ----------    Test Pic18F97J60  -  Com1-------^M







RelayCtrl01        |TRA|     64703|2010-03-11 22:21:01.050030|             MAIN >  .







RelayCtrl01        |TRA|     64704|2010-03-11 22:21:01.050125|             MAIN > SerialInString buffer (len 75)                Mauro Sozzi @ Gennaio 2008^M







RelayCtrl01        |TRA|     64704|2010-03-11 22:21:01.050125|             MAIN >  .     ---------- Per Help Comandi







RelayCtrl01        |TRA|     64705|2010-03-11 22:21:01.050141|             MAIN > EmptySerial(): received string:                Mauro Sozzi @ Gennaio 2008^M







RelayCtrl01        |TRA|     64705|2010-03-11 22:21:01.050141|             MAIN >  .







RelayCtrl01        |TRA|     64706|2010-03-11 22:21:01.063281|             MAIN > SerialInString received (len 23) premere ?  ----------^M







RelayCtrl01        |TRA|     64706|2010-03-11 22:21:01.063281|             MAIN >  .







RelayCtrl01        |TRA|     64707|2010-03-11 22:21:01.063308|             MAIN > SerialInString buffer (len 55)     ---------- Per Help Comandi premere ?  ----------^M







RelayCtrl01        |TRA|     64707|2010-03-11 22:21:01.063308|             MAIN >  .







RelayCtrl01        |TRA|     64708|2010-03-11 22:21:01.063324|             MAIN > EmptySerial(): received string:     ---------- Per Help Comandi premere ?  ----------^M







RelayCtrl01        |TRA|     64708|2010-03-11 22:21:01.063324|             MAIN >  .







RelayCtrl01        |TRA|     64709|2010-03-11 22:21:02.063096|             MAIN > SerialInString received (len 0)







RelayCtrl01        |TRA|     64710|2010-03-11 22:21:02.063110|             MAIN > SerialInString buffer (len 0)  







RelayCtrl01        |TRA|     64711|2010-03-11 22:21:04.064072|             MAIN > SerialInString received (len 0)







RelayCtrl01        |TRA|     64712|2010-03-11 22:21:04.064098|             MAIN > SerialInString buffer (len 0)







RelayCtrl01        |TRA|     64713|2010-03-11 22:21:06.266095|             MAIN > SerialInString received (len 0)







RelayCtrl01        |TRA|     64714|2010-03-11 22:21:06.266110|             MAIN > SerialInString buffer (len 0)







RelayCtrl01        |INF|     64715|2010-03-11 22:21:06.266121|             MAIN > State change from Operating to No connection







RelayCtrl01        |TRA|     64716|2010-03-11 22:21:06.266309|             MAIN > Sending command "7"







RelayCtrl01        |TRA|     64717|2010-03-11 22:21:06.461096|             MAIN > SerialInString received (len 89) Temp_board1:31.7   Temp_board2:1.7   Temp_board3:2.1   Temp_CCD39:2.1   Temp_water:-1.6^M







RelayCtrl01        |TRA|     64717|2010-03-11 22:21:06.461096|             MAIN >  .







RelayCtrl01        |TRA|     64718|2010-03-11 22:21:06.461124|             MAIN > SerialInString buffer (len 89) Temp_board1:31.7   Temp_board2:1.7   Temp_board3:2.1   Temp_CCD39:2.1   Temp_water:-1.6^M







RelayCtrl01        |TRA|     64718|2010-03-11 22:21:06.461124|             MAIN >  .







RelayCtrl01        |TRA|     64719|2010-03-11 22:21:06.461155|             MAIN > EmptySerial(): received string: Temp_board1:31.7   Temp_board2:1.7   Temp_board3:2.1   Temp_CCD39:2.1   Temp_water:-1.6^M







RelayCtrl01        |TRA|     64719|2010-03-11 22:21:06.461155|             MAIN >  .







RelayCtrl01        |DEB|     64720|2010-03-11 22:21:06.461187|             MAIN > Temperature sensor 0: 31.700000 (with offset  0.00)







RelayCtrl01        |DEB|     64721|2010-03-11 22:21:06.461209|             MAIN > Temperature sensor 1: 1.700000 (with offset  0.00)







RelayCtrl01        |DEB|     64722|2010-03-11 22:21:06.461229|             MAIN > Temperature sensor 2: 2.100000 (with offset  0.00)







RelayCtrl01        |DEB|     64723|2010-03-11 22:21:06.461248|             MAIN > Temperature sensor 3: 2.100000 (with offset  0.00)







RelayCtrl01        |DEB|     64724|2010-03-11 22:21:06.461267|             MAIN > Temperature sensor 4: -1.600000 (with offset  0.00)







RelayCtrl01        |TRA|     64725|2010-03-11 22:21:06.634565|             MAIN > SerialInString received (len 86)  Temp_lj39:12.5   Temp_bcu39:9.1   Temp_lj47:8.9   Temp_bcu47:5.1   Temp_copley:16.6^M







RelayCtrl01        |TRA|     64725|2010-03-11 22:21:06.634565|             MAIN >  .







RelayCtrl01        |TRA|     64726|2010-03-11 22:21:06.634592|             MAIN > SerialInString buffer (len 86)  Temp_lj39:12.5   Temp_bcu39:9.1   Temp_lj47:8.9   Temp_bcu47:5.1   Temp_copley:16.6^M







RelayCtrl01        |TRA|     64726|2010-03-11 22:21:06.634592|             MAIN >  . RelayCtrl01        |TRA|     64727|2010-03-11 22:21:06.634609|             MAIN > EmptySerial(): received string:  Temp_lj39:12.5   Temp_bcu39:9.1   Temp_lj47:8.9   Temp_bcu47:5.1   Temp_copley:16.6^M







RelayCtrl01        |TRA|     64727|2010-03-11 22:21:06.634609|             MAIN >  .







RelayCtrl01        |DEB|     64728|2010-03-11 22:21:06.634642|             MAIN > Temperature sensor 5: 12.500000 (with offset  0.00)







RelayCtrl01        |DEB|     64729|2010-03-11 22:21:06.634665|             MAIN > Temperature sensor 6: 9.100000 (with offset  0.00)







RelayCtrl01        |DEB|     64730|2010-03-11 22:21:06.634684|             MAIN > Temperature sensor 7: 8.900000 (with offset  0.00)







RelayCtrl01        |DEB|     64731|2010-03-11 22:21:06.634703|             MAIN > Temperature sensor 8: 5.100000 (with offset  0.00)







RelayCtrl01        |DEB|     64732|2010-03-11 22:21:06.634723|             MAIN > Temperature sensor 9: 16.600000 (with offset  0.00)







RelayCtrl01        |TRA|     64733|2010-03-11 22:21:07.634009|             MAIN > SerialInString received (len 0)







RelayCtrl01        |TRA|     64734|2010-03-11 22:21:07.634035|             MAIN > SerialInString buffer (len 0)







RelayCtrl01        |INF|     64735|2010-03-11 22:21:10.637053|             MAIN > Connecting to ts839dx, 2104







RelayCtrl01        |INF|     64736|2010-03-11 22:21:10.644998|             MAIN > Network reconfigured OK

home/aoacct/testw2/measurements/ccd47/condensa_bin1.fits