Last night were were able to checkout the SX AO system.

Everything worked as well as it did last Spring with the exception

of the Optical Gain. I tried all the fixes I know (restarting processed,

creating the c2m matrix in IDL) but the optical gain calculated was

100 times smaller than expected (0.001).

The DX fastlink alignment error occurred many more times tonight than

on the previous night. 18 failures. This problem wasted probably

30 minutes tonight and it seems to be getting worse.

**Guido, Armando: Should we try to reseat the fastlink fibers at the switch

BCU and at the AdSec?

We were able to test both the CentOS 6 and 7 versions of the AO Supervisor.

Progress was made with both.

The CentOS version seemed to work properly with the exceptions of

applying NPCA amplitudes. The Optical Gains seemed to work and

the Loop keep the Gain at a reasonable value (~0.3). However, when

I applied 50 nm of Z5, the value of Z5 coming back from the AdSec

remained 0.0

For the CentOS 7 version, the Save Optical Loop Data GUI crashed

when you clicked save. Xianyu will look into it.

I seemed to be able to get the Optical Gain loop to run under CentOS 7, with reasonable values,

but twice the Loop opened with Safe Skip Activated soon after the Optical

Gain was applied. I think this may be a generic Optical Gain problem and will

require input from Alfio.

There are also a few minor changes to the CentOS 7 version of the

AO Supervisor.

You can see more details in

http://wiki.lbto.org/bin/view/AdaptiveOptics/20160910_DX_Wakeup

http://wiki.lbto.org/bin/view/AdaptiveOptics/20160910_DX_Wakeup

I think are next AO testing of 6/7 is planned for Monday night.

Doug



SX LOG_TWIKI

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #modSorted ascending freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160910_250000 ??? NaN 323976 10 NaN -1.00 063432 1 12 100 0.1 -1.0 -1.0 3 -1 -1 -1.0 ? -1 -1 OFF NO - M2C used to generate sinusmode disturbance is NOT the same as current M2C - Pyramid modulation mismatch - Pupils mismatch
20160911_114939 ??? 16.3 262142 4 0.86 *** 205357 4 36 100 3.4 0.5 0.9 6 36 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_115900 ??? 16.7 256702 3 0.86 1.46 205357 4 36 100 3.4 1.7 2.5 6 24 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_120510 ??? 16.1 253010 5 0.77 0.62 205357 4 36 100 3.4 0.3 0.4 6 43 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_121136 ??? 16.4 249081 4 0.80 0.80 205357 4 36 100 3.4 0.9 -1.0 6 32 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160910_062053 ??? 13.6 288568 6 NaN 0.85 203720 3 66 200 1.9 -1.0 -1.0 6 92 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160910_062311 ??? 13.6 289492 4 NaN 0.76 203720 3 66 200 1.9 -1.0 -1.0 6 93 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_173504 ??? 9.5 323996 1 NaN 0.02 062006 1 300 990 0.8 0.1 0.1 3 92 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_173633 ??? 9.5 323996 2 NaN 0.02 062006 1 300 990 0.8 0.1 0.1 3 93 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_174606 ??? 9.5 323996 2 NaN 0.02 062006 1 300 990 0.8 0.1 0.1 3 93 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_180320 ??? 9.5 323996 3 NaN 0.02 062006 1 300 990 0.8 0.1 0.1 3 92 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_180339 ??? 9.5 323996 3 NaN 0.02 062006 1 300 990 0.8 0.1 0.1 3 92 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_182410 ??? 9.5 323996 1 NaN 0.02 062006 1 300 990 0.8 0.1 0.1 3 95 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160910_044727 ??? 9.3 225219 7 NaN 0.81 184300 1 400 990 1.7 0.6 0.4 3 112 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_065037 ??? 9.4 213549 6 0.89 0.67 184300 1 400 990 1.7 0.6 0.3 3 102 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_065300 ??? 9.4 211872 6 0.87 0.57 184300 1 400 990 1.5 0.6 0.3 3 104 ON -1.0 ? -1 -1 ONSKY zero.fits - long stop of 55 frames at frame 2249
20160911_065501 ??? 9.4 210447 6 0.86 0.53 184300 1 400 990 1.2 0.6 0.3 3 103 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_065718 ??? 9.4 208835 5 0.80 0.39 184300 1 400 990 1.2 0.5 0.2 3 105 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_070703 ??? 9.4 201962 5 0.80 0.43 184300 1 400 990 1.1 0.3 0.3 3 102 ON -1.0 ? -1 -1 ONSKY zero.fits  

IDL string: [,'20160910_044727','20160910_062053','20160910_062311','20160910_250000','20160911_065037','20160911_065300','20160911_065501','20160911_065718','20160911_070703','20160911_114939','20160911_115900','20160911_120510','20160911_121136','20160911_173504','20160911_173633','20160911_174606','20160911_180320','20160911_180339','20160911_182410']
SR 0 (NaN)
SEEING (WFS) -0.542913 (4.32448)
SEEING (DIMM) 0.834853 (0.0416192)



DX log_twiki

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160910_044831 ??? 9.2 224635 6 NaN 0.83 014135 1 400 990 0.0 -1.0 -1.0 3 115 ON -1.0 ? -1 -1 ONSKY zero.fits - Pupils not on target
20160910_062148 ??? 13.4 288947 6 1.40 1.08 221257 2 153 400 2.7 1.1 -1.0 3 22 ON -1.0 ? -1 -1 ONSKY zero.fits - long stop of 52 frames at frame 1583
20160910_062308 ??? 13.3 289473 5 NaN 0.88 221257 2 153 400 2.7 1.1 -1.0 3 24 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_065020 ??? 9.2 213739 6 0.90 0.68 014135 1 400 990 1.8 0.3 0.3 3 112 ON -1.0 ? -1 -1 ONSKY zero.fits - long stop of 106 frames at frame 3611
20160911_065301 ??? 9.2 211850 6 0.85 0.48 014135 1 400 990 1.5 0.3 0.3 3 115 ON -1.0 ? -1 -1 ONSKY zero.fits - long stop of 86 frames at frame 1633
20160911_065459 ??? 9.2 210470 6 0.86 0.60 014135 1 400 990 1.2 0.3 0.3 3 112 ON -1.0 ? -1 -1 ONSKY zero.fits - long stop of 101 frames at frame 303
20160911_065849 ??? 9.2 207761 6 0.74 0.54 014135 1 400 990 1.2 0.3 0.2 3 116 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160911_070756 ??? 9.2 201341 6 0.83 0.64 014135 1 400 990 1.4 0.6 0.5 3 111 ON -1.0 ? -1 -1 ONSKY zero.fits - long stop of 55 frames at frame 3500

IDL string: [,'20160910_044831','20160910_062148','20160910_062308','20160911_065020','20160911_065301','20160911_065459','20160911_065849','20160911_070756']
SR 0 (NaN)
SEEING (WFS) 0.715986 (0.199004)
SEEING (DIMM) 0.930977 (0.23498)