Bin 2 TEST Gopt aaplied up to mode 50 autogain with DIMM not avaiable TN 20160423_060847 060858 060909 MISSING SLOPES and PSF

TN 061003 061016 061027 MISSING SLOPES and PSF

TN 061203 061214 061225 MISSING SLOPES and PSF

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160423_060847 ??? 9.8 286174 8 1.66 -1.00 203208 2 153 991 0.3 0.2 0.1 3 296 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160423_060858 ??? 9.9 286111 8 1.66 -1.00 203208 2 153 991 0.3 0.2 0.1 3 286 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160423_060909 ??? 9.9 286047 7 1.60 -1.00 203208 2 153 991 0.3 0.2 0.1 3 286 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160423_061003 ??? 9.9 285727 8 1.55 -1.00 203208 2 153 991 0.3 0.2 0.1 3 271 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160423_061016 ??? 9.9 285646 8 1.60 -1.00 203208 2 153 991 0.3 0.2 0.1 3 276 ON -1.0 ? -1 -1 ONSKY zero.fits - long stop of 67 frames at frame 990
20160423_061027 ??? 9.9 285580 9 1.72 -1.00 203208 2 153 991 0.3 0.2 0.1 3 286 ON -1.0 ? -1 -1 ONSKY zero.fits - long stop of 65 frames at frame 3915
20160423_061203 ??? 9.9 284985 8 1.75 -1.00 203208 2 153 991 0.3 0.2 0.1 3 272 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160423_061214 ??? 9.9 284916 9 1.76 -1.00 203208 2 153 991 0.3 0.2 0.1 3 284 ON -1.0 ? -1 -1 ONSKY zero.fits  
20160423_061225 ??? 9.8 284846 8 1.73 -1.00 203208 2 153 991 0.3 0.2 0.1 3 294 ON -1.0 ? -1 -1 ONSKY zero.fits  
IDL string: ['20160423_060847','20160423_060858','20160423_060909','20160423_061003','20160423_061016','20160423_061027','20160423_061203','20160423_061214','20160423_061225']
SR 0 (NaN)
SEEING (WFS) -1 (NaN)
SEEING (DIMM) 1.67158 (0.074699)

LUCI Online - Dark acquired

AUTOGAIN (DIMM 1.6) GOPT OFF NCPA OFF ->TN 062403 062427 062450 GOPT ON NCPA OFF ->TN 062549 062615 062637 GOTP ON NCPA ON ->TN 062752 062814 062835 GOTP ON NCPA ON ->TN 063330 063353 063416

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160423_062403 ??? 9.8 279848 7 1.42 1.01 203208 2 153 991 2.5 1.6 0.9 3 297 ON 34.0 FeII clear 172 50 ONSKY zero.fits - long stop of 52 frames at frame 3190 - SR may be in error
20160423_062427 ??? 9.9 279649 7 1.40 1.27 203208 2 153 991 2.5 1.6 0.9 3 281 ON 31.0 FeII clear 172 50 ONSKY zero.fits - long stop of 55 frames at frame 2133 - SR may be in error
20160423_062450 ??? 9.9 279471 7 1.44 1.15 203208 2 153 991 2.5 1.6 0.9 3 284 ON 31.8 FeII clear 172 50 ONSKY zero.fits - SR may be in error
20160423_062549 ??? 9.8 278997 8 1.57 1.16 203208 2 153 991 0.3 0.2 0.1 3 300 ON 42.0 FeII clear 172 50 ONSKY zero.fits - long stop of 58 frames at frame 3786 - SR may be in error
20160423_062615 ??? 9.8 278787 8 1.58 1.05 203208 2 153 991 0.3 0.2 0.1 3 311 ON 41.7 FeII clear 172 50 ONSKY zero.fits - SR may be in error
20160423_062637 ??? 9.8 278608 8 1.50 1.08 203208 2 153 991 0.3 0.2 0.1 3 303 ON 39.0 FeII clear 172 50 ONSKY zero.fits - long stop of 69 frames at frame 1982 - SR may be in error
20160423_062752 ??? 9.9 277998 7 1.41 1.32 203208 2 153 991 0.3 0.2 0.1 3 290 ON 36.7 FeII clear 172 50 ONSKY zero.fits - long stop of 66 frames at frame 3357 - SR may be in error
20160423_062814 ??? 9.8 277812 7 1.41 1.14 203208 2 153 991 0.3 0.2 0.1 3 299 ON 37.4 FeII clear 172 50 ONSKY zero.fits - SR may be in error
20160423_062835 ??? 9.8 277632 8 1.46 1.25 203208 2 153 991 0.3 0.2 0.1 3 293 ON 41.1 FeII clear 172 50 ONSKY zero.fits - long stop of 65 frames at frame 3910 - SR may be in error
IDL string: ['20160423_062403','20160423_062427','20160423_062450','20160423_062549','20160423_062615','20160423_062637','20160423_062752','20160423_062814','20160423_062835']
SR 0 (NaN)
SEEING (WFS) 1.1594 (0.105815)
SEEING (DIMM) 1.46567 (0.0702545)

AUTOGAIN DIMM=1.6-1.7 GOPT ON NCPA ON ->TN 063900 063923 063947 GOPT ON NCPA ON ->TN 064443 064506 064528

I modal plot hanno ancora un ritorno di fiamma ai modi >100, ma esiste anche senza Gopt, quindi sospettiamo che sia il ricostruttore.

Star for BIN1 test bad seeing mode AUTOGAIN Seeing 1.6 High seeing variability GOPT ON NCPA ON -> TN 070151 070216 070248 070317 070341 070403 070442 070615 070717
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160423_070151 ??? 8.4 259126 7 1.37 1.25 062006 1 300 990 0.4 0.3 0.3 3 272 ON 183.2 FeII clear 172 50 ONSKY zero.fits - long stop of 59 frames at frame 1370 - Could not test SR quality
20160423_070216 ??? 8.4 259404 7 2.14 1.31 062006 1 300 990 0.4 0.3 0.2 3 259 ON 380.6 FeII clear 172 50 ONSKY zero.fits - long stop of 68 frames at frame 1193 - SR may be in error
20160423_070248 ??? 8.3 259775 7 1.51 1.44 062006 1 300 990 0.4 0.3 0.2 3 291 ON 1603.6 FeII clear 172 50 ONSKY zero.fits - long stop of 59 frames at frame 1254 - Could not test SR quality
20160423_070317 ??? 8.3 260104 7 1.35 1.13 062006 1 300 990 0.4 0.3 0.2 3 292 ON 0.0 FeII clear 172 50 ONSKY zero.fits - long stop of 74 frames at frame 2613 - Center of PSF estimation error
20160423_070341 ??? 8.3 260378 8 1.15 1.00 062006 1 300 990 0.4 0.3 0.2 3 295 ON 0.0 FeII clear 172 50 ONSKY zero.fits - long stop of 68 frames at frame 2134 - Center of PSF estimation error
20160423_070403 ??? 8.3 260628 8 1.18 0.97 062006 1 300 990 0.4 0.3 0.2 3 296 ON 66.0 FeII clear 172 50 ONSKY zero.fits - Could not test SR quality
20160423_070442 ??? 8.3 261066 7 1.23 0.86 062006 1 300 990 0.4 0.3 0.2 3 296 ON 0.0 FeII clear 172 50 ONSKY zero.fits - long stop of 65 frames at frame 1184 - Center of PSF estimation error
20160423_070615 ??? 8.3 262114 8 1.16 0.91 062006 1 300 990 0.4 0.3 0.2 3 297 ON 221.7 FeII clear 172 50 ONSKY zero.fits - Could not test SR quality
IDL string: ['20160423_070151','20160423_070216','20160423_070248','20160423_070317','20160423_070341','20160423_070403','20160423_070442','20160423_070615']
SR 0 (NaN)
SEEING (WFS) 1.10913 (0.208854)
| SEEING (DIMM) | 1.38672 (0.329634) |รน

NOW we are overestimating high order when seein is going lower than the one we had during autogain.

Change gain renormalization: no gain compensation (pure Gopt) modes #0:50 partial gain compensation (0.9 *Gopt) modes #50:200 partial gain compensation (0.7 Gopt) modes #200:

AUTOGAIN seeing 1.1 GOPT ON NCPA ON -> TN 074913 075220....

NEW Gain change algorythm (implemented log dependence on the mode #) Now is fix up to mode 50 than is applied a correction 0.16*(log(mode#)-log(50)). the slope has been computed from the Fernando's E2E plot for seeing 0.7"

AUTOGAIN DIMM=1.6" GoptON NCPA OFF -> TN 084819 084931 084953 085146 GoptON NCPA ON -> TN 085257 085334

LOOP opens

AUTOGAIN DIMM=1.1-1.2" GOPT ON NCPA ON ->TN 090636 090701 090724

-- EnricoPinna - 26 Apr 2016
Topic revision: r1 - 26 Apr 2016, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback