Doug, Greg, Alfio, Enrico

NCPA measurement on LUCI1 (FeII filter) with ARGOS calib source

08:15 ROT=330deg

Create disturbance for the optical gain on KL_v20

useing the procedure -> sinusmode_generate

Amplitude = 1E-8 m; Freq = 30Hz; loop=990Hz, /EVEN
ModalDiturb_20160312_153051.fits

The ARGOS calib source provides R= 9.4 (on CCD39 display).

Autofocus

Autofocus procedure with LUCI subwindow

Best focus = 54.6mm

Best focus results really close to the stage limit (Zstage SW limit = 56.0mm)

09:20 Start NCPA measurement

By eye we spot a trifoil that get better with Z9=-10nm

Now we start the automatic procedure -> luci_ncpa.py

9:45 No dark

We forgot the DARKFILE for LUCI

We turn off the ARGOS source and we acquire the dark using the WFS Arbitrator GUI

The procedure finds Z5=-12nm

Together with the Z9=-10nm we found SR=0.6

End of the first angle NCPA measurement (performed twice for cross-check)

NCPA vector -> 20160312_330deg.fits

TN = 201603012_182952... 183943

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160312_182952 ??? 9.5 323994 13 NaN 0.07 184300 1 400 990 0.5 0.2 -1.0 3 95 ON 59.0 FeII clear 172 20 ONSKY zero.fits Warning: Selected luci dark +2h old!
20160312_183916 ??? 9.5 323994 6 NaN 0.07 184300 1 400 990 0.5 0.2 -1.0 3 99 ON 56.7 FeII clear 172 20 ONSKY zero.fits Warning: Selected luci dark +2h old!
20160312_183943 ??? 9.4 323994 9 NaN 0.07 184300 1 400 990 0.5 0.2 -1.0 3 99 ON 57.1 FeII clear 172 20 ONSKY zero.fits Warning: Selected luci dark +2h old!

IDL string: ['20160312_182952','20160312_183916','20160312_183943']
SR 0 (NaN)
SEEING (WFS) 0.0676763 (0.000917317)
SR = 59%

THE FOCUS WAS NOT SCAN AS NCPA: WRONG!!!!!!! We will do it after angle 300deg

11:40 ROT=300deg

12:30 problem getting frames from LUCI with the automatic procedure, soved restarting the WFS arbitrator

NCPA vector -> 20160312_300deg.fits

TN = 201603012_200851... 200941 .... 201005

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturbSorted ascending SN notes
20160312_200851 ??? 9.6 323994 6 NaN 0.07 184300 1 400 990 0.5 0.2 -1.0 3 84 ON 52.7 FeII clear 172 20 ONSKY zero.fits - long stop of 67 frames at frame 3342
20160312_200941 ??? 9.6 323994 9 NaN 0.07 184300 1 400 990 0.5 0.2 -1.0 3 82 ON 53.8 FeII clear 172 20 ONSKY zero.fits  
20160312_201005 ??? 9.6 323994 8 NaN 0.07 184300 1 400 990 0.5 0.2 -1.0 3 84 ON 51.5 FeII clear 172 20 ONSKY zero.fits - long stop of 70 frames at frame 493

IDL string: ['20160312_200851','20160312_200941','20160312_201005']
SR 0 (NaN)
SEEING (WFS) 0.0717626 (0.00201626)

13:15 ROT=330deg to check Z4

Rifatto Z4 otteniamo -8nm (compatibile con il -10nm di ROT=300)

TN = 202508 - 202556

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160312_202508 ??? 9.6 323994 11 NaN 0.07 184300 1 400 990 0.5 0.2 -1.0 3 88 ON 54.0 FeII clear 172 20 ONSKY zero.fits  
20160312_202535 ??? 9.6 323994 11 NaN 0.07 184300 1 400 990 0.5 0.2 -1.0 3 88 ON 53.4 FeII clear 172 20 ONSKY zero.fits - long stop of 68 frames at frame 304
20160312_202556 ??? 9.6 323994 12 NaN 0.07 184300 1 400 990 0.5 0.2 -1.0 3 88 ON 55.6 FeII clear 172 20 ONSKY zero.fits - long stop of 75 frames at frame 2582
IDL string: ['20160312_202508','20160312_202535','20160312_202556']
SR 0 (NaN)
SEEING (WFS) 0.0674196 (0.000983398)

-- EnricoPinna - 12 Mar 2016
Topic revision: r3 - 12 Mar 2016, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.