Preset Statistics

Target UT OK Notes
Jun12_Mrk231 03:56:00 FAIL No star in the guider
Jun12_Mrk231 03:58:00 FAIL No star in the guider
Jun12_Mrk231 04:00:00 FAIL No star in the guider
Jun12_Mrk231 04:06:00 OK Only preset no runao
Jun12_Mrk231 04:09:00 OK PRESET ACE + GCS=track
Jun12_Mrk231 04:19:00 OK PRESET ACE, no keyword GCS, failed shell not ready!
Jun12_Mrk231 04:23:00 OK PRESET ACE, no keyword GCS, RUNAO
Jun12_Mrk231 04:36:00 OK PRESET ACE, no keyword GCS, RUNAO
Jun12_Mrk231 04:52:00 OK PRESET ACE, no keyword GCS, RUNAO
HIP80414 05:41:00 FAIL WRONG AOS WFS NAME
HIP80414 05:43:00 FAIL PROBE FAILS
HIP80414 05:45:00 FAIL PROBE FAILS
HIP80414 06:23:00 OK PROBE BACK TO LIFE- Long collimation (vignetting)
HIP80414 06:27:00 FAIL RUNAO, BAYSIDE STAGE X
HIP80414 06:29:00 OK With the workaround moving by panel the stage X, but we found the cube in the optical path
HIP80414 06:33:00 OK  
       
       
AO578   OK 400 modes
AO578   OK 500 modes
AO578   OK 500 modes
AO646 09:40:00 OK ONLY PRESET
AO646 09:48:00 OK REC 100 modes
AO654   OK 400 PP
AO654   OK 400 SINUS
AO654   OK 400 PP
AO722   Fail Problema catalogo
AO723   OK 400 PP
AO723   OK 400 SINUS

Log

In control room: Carmelo Arcidiacono, Juan Carlos Guerra, Lorenzo Busoni e Geno (TO)

Tree Houses Off. We took the opportunity to inspect vibrations status on the LUCI platform (FRONT): the cooling refrigerator of LUCI is actually connected through two flexible pipes that are in contact with the gallery. See pictures

Luci_refr_sx.JPG

Problem on the pointing

In IRAF the software for the pointing had the option AOMODE set, but the preset was not ACE and the pointing failed to center the star for collimation.

IDL> agw_set_station,pisces

**********************************************
*
* => PISCES at RFBG Focal Station
*
**********************************************

IDL> iif_register

Starting at 03:45:00 (08:25 local time)

Right acquisition test on AGN Jun12_Mrk231

IDL> iif_preset,Jun12_Mrk231,GS=1,RS=0,MODE=ACE,PA=0

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 1
12:56:03.421 +56:52:47.26 RADEC 2000.0 J2000 22.0 -2.0 15.81 R 1.52 B_V 0.63
# AO Reference star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65

preset, Jun12_Mrk231, MODE=ACE, GS=1, RS=0

===> Error in irc_presetTelescope
err = Resmsg[0] = PresetTelescope result status: Error
err = Resmsg[1] = Stop[8] complete
err = Resmsg[2] = no guide star found in guide box after resumeGuiding
IDL> iif_preset,Jun12_Mrk231,GS=3,RS=0,MODE=ACE,PA=180

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 3
12:56:15.734 +56:48:17.81 RADEC 2000.0 J2000 -16.0 6.0 12.33 R 1.23 B_V 0.63
# AO Reference star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65

preset, Jun12_Mrk231, MODE=ACE, GS=3, RS=0, PA=180

===> Error in irc_presetTelescope
err = Resmsg[0] = PresetTelescope result status: Error
err = Resmsg[1] = couldn't find a star in image

STAR NOT IN THE PROBE POINTING FAILURE

IDL> iif_preset,Jun12_Mrk231,GS=0,RS=0,MODE=ACTIVE,PA=180

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 0.63

preset, Jun12_Mrk231, MODE=ACTIVE, GS=0, PA=180

===> Error in irc_presetTelescope
err = Resmsg[0] = PresetTelescope result status: Error err = Resmsg[1] = Stop[10] complete
err = Resmsg[2] = couldn't find a star in image
IDL> iif_preset,Jun12_Mrk231,GS=0,RS=0,MODE=ACE,PA=180,GCS =TRACK

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 0.63
# AO Reference star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65

preset, Jun12_Mrk231, MODE=ACE, GS=0, RS=0, PA=180

IDL> iif_preset,Jun12_Mrk231,GS=13,RS=0,MODE=ACE,PA=180,GCS =TRACK
IDL> iif_preset,Jun12_Mrk231,GS=3,RS=0,MODE=ACE,PA=180,GCS =TRACK
===> Error in iif_preset: Telescope mode not set (ie MODE=ACTIVE)
IDL> exit

AFTER AN EXCEPTION in the IIF PRESET reset of idl is needed.

RESTARTING IDL

IDL> agw_set_station,pisces

**********************************************
*
* => PISCES at RFBG Focal Station
*
**********************************************

IDL> agw_set_station,pisces

**********************************************
*
* => PISCES at RFBG Focal Station
*
**********************************************

IDL> iif_register
IDL> iif_preset,Jun12_Mrk231,GS=3,RS=0,MODE=ACE,PA=180,GCS =TRACK

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 3
12:56:15.734 +56:48:17.81 RADEC 2000.0 J2000 -16.0 6.0 12.33 R 1.23 B_V 0.63
# AO Reference star: 0

preset, Jun12_Mrk231, MODE=ACE, GS=3, RS=0, PA=180

IDL> iif_runao
IDL> iif_offset,0,-40,COORD=DETXY
=> OffsetPointing 0 0 -40 DETXY ABS right
IDL> iif_offset,0,40,COORD=DETXY
=> OffsetPointing 0 0 40 DETXY ABS right

FAILED! loop did not restarted

IDL> iif_offset,0,0,COORD=DETXY
=> OffsetPointing 0 0 0 DETXY ABS right

Nothing on the WFS no reaction!


IDL> iif_offset,0,0,COORD=DETXY
=> OffsetPointing 0 0 0 DETXY ABS right
===> Error in irc_offsetpointing
err = Resmsg[0] = OffsetPointing result status: Error
err = Resmsg[1] = AOARB:Illegal command for state LoopSuspended (Pause)
err = Resmsg[2] = Pause[26] failed - AOARB:Illegal command for state LoopSuspended (Pause)

ERROR MESSAGE. LOOP IS PAUSED. NO REACTION from the WFS

NEW PRESET

IDL> iif_preset,Jun12_Mrk231,GS=3,RS=0,MODE=ACE,PA=180,GCS =TRACK

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 3
12:56:15.734 +56:48:17.81 RADEC 2000.0 J2000 -16.0 6.0 12.33 R 1.23 B_V 0.63
# AO Reference star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65

preset, Jun12_Mrk231, MODE=ACE, GS=3, RS=0, PA=180

IDL> iif_runao
IDL> iif_offset,0,-40,COORD=DETXY
=> OffsetPointing 0 0 -40 DETXY ABS right

SKIP FRAME ONCE WE START to offset.

NEW PRESET

IDL> iif_preset,Jun12_Mrk231,GS=3,RS=0,MODE=ACE,PA=180

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 3
12:56:15.734 +56:48:17.81 RADEC 2000.0 J2000 -16.0 6.0 12.33 R 1.23 B_V 0.63
# AO Reference star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65

preset, Jun12_Mrk231, MODE=ACE, GS=3, RS=0, PA=180

===> Error in irc_presetTelescope
err = Resmsg[0] = PresetTelescope result status: Error err = Resmsg[1] = Stop[34] complete
err = Resmsg[2] = adsecarb:Illegal command for state Ready (Set focal station)
err = Resmsg[3] = PresetAO[35] failed - adsecarb:Illegal command for state Ready (Set focal station)


SHELL NOT SET

IDL> iif_preset,Jun12_Mrk231,GS=3,RS=0,MODE=ACE,PA=180

IDL> iif_runao
IDL> iif_offset,0,-40,COORD=DETXY
=> OffsetPointing 0 0 -40 DETXY ABS right
===> Error in irc_offsetpointing
err = Resmsg[0] = OffsetPointing result status: Error err = Resmsg[1] = Pause[41] complete
err = Resmsg[2] = Offset too large. Max offset in closed loop is 0.6 mm (combined XY), requested 23.9
err = Resmsg[3] = OffsetXY[42] failed - Offset too large. Max offset in closed loop is 0.6 mm (combined XY), requested 23.9
err = Resmsg[4] = guiding is not active
IDL> iif_preset,Jun12_Mrk231,GS=3,RS=0,MODE=ACE,PA=180

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 3
12:56:15.734 +56:48:17.81 RADEC 2000.0 J2000 -16.0 6.0 12.33 R 1.23 B_V 0.63
# AO Reference star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65

preset, Jun12_Mrk231, MODE=ACE, GS=3, RS=0, PA=180

IDL> iif_runao
IDL> iif_offset,0,0,COORD=DETXY
=> OffsetPointing 0 0 0 DETXY ABS right
IDL> iif_offset,0,-0.1,COORD=DETXY
=> OffsetPointing 0 0 -0.100000 DETXY ABS right
IDL> iif_offset,0,-10,COORD=DETXY
=> OffsetPointing 0 0 -10 DETXY ABS right
===> Error in irc_offsetpointing
err = Resmsg[0] = OffsetPointing result status: Error err = Resmsg[1] = Pause[54] complete
err = Resmsg[2] = Offset too large. Max offset in closed loop is 0.6 mm (combined XY), requested 5.9
err = Resmsg[3] = OffsetXY[55] failed - Offset too large. Max offset in closed loop is 0.6 mm (combined XY), requested 5.9
err = Resmsg[4] = guiding is not active

SHELL RIP after a safe skip due to the telescope moving for the offset but not placing the loop in "pause" .

IDL> exit

Restarting Adsec software (adsec_stop- adsec_start)

Restarting WFS software (w_stop- w_start)

closing and opening IDL

IDL> agw_set_station,pisces

**********************************************
*
* => PISCES at RFBG Focal Station
*
**********************************************

IDL> iif_register
IDL> iif_preset,Jun12_Mrk231,GS=3,RS=0,MODE=ACE,PA=180

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 3
12:56:15.734 +56:48:17.81 RADEC 2000.0 J2000 -16.0 6.0 12.33 R 1.23 B_V 0.63
# AO Reference star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65

preset, Jun12_Mrk231, MODE=ACE, GS=3, RS=0, PA=180

IDL> iif_runao
IDL> iif_offset,0,0,COORD=DETXY
=> OffsetPointing 0 0 0 DETXY ABS right

PAUSE-CLOSED LOOP

IDL> iif_offset,0,0,COORD=DETXY
=> OffsetPointing 0 0 0 DETXY ABS right

PAUSE-CLOSED LOOP

IDL> iif_offset,0,0.1,COORD=DETXY
=> OffsetPointing 0 0 0.100000 DETXY ABS right

PAUSE-MOVE-CLOSED LOOP

IDL> iif_offset,0,0.0,COORD=DETXY
=> OffsetPointing 0 0 0.00000 DETXY ABS right

PAUSE-MOVE-CLOSED LOOP

IDL> iif_offset,0,5.0,COORD=DETXY
=> OffsetPointing 0 0 5.00000 DETXY ABS right

PAUSE-MOVE-CLOSED LOOP (actually we saw in the 47 the star moving of ~5")

IDL> iif_offset,0,0.0,COORD=DETXY
=> OffsetPointing 0 0 0.00000 DETXY ABS right

PAUSE-MOVE-CLOSED LOOP

IDL> iif_offset,0,-40.0,COORD=DETXY
=> OffsetPointing 0 0 -40.0000 DETXY ABS right

PAUSE-MOVE-CLOSED LOOP

IDL> iif_offset,0,40.0,COORD=DETXY
=> OffsetPointing 0 0 40.0000 DETXY ABS right
IDL> iif_offset,0,0.0,COORD=DETXY
=> OffsetPointing 0 0 0.00000 DETXY ABS right
IDL> iif_preset,Jun12_Mrk231,GS=3,RS=0,MODE=ACE,PA=180

# Target star: 0 Jun12_Mrk231
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65
# Guide star: 3
12:56:15.734 +56:48:17.81 RADEC 2000.0 J2000 -16.0 6.0 12.33 R 1.23 B_V 0.63
# AO Reference star: 0
12:56:14.233 +56:52:25.23 RADEC 2000.0 J2000 0.0 0.0 14.10 R 3.84 B_V 1.65

preset, Jun12_Mrk231, MODE=ACE, GS=3, RS=0, PA=180

IDL> iif_runao
IDL> iif_offset,0,0.0,COORD=DETXY
=> OffsetPointing 0 0 0.00000 DETXY ABS right
IDL> iif_offset,0,10.0,COORD=DETXY
=> OffsetPointing 0 0 10.0000 DETXY ABS right
===> Error in irc_offsetpointing
err = Resmsg[0] = OffsetPointing result status: Error err = Resmsg[1] = Pause[40] complete
err = Resmsg[2] = Offset too large. Max offset in closed loop is 0.6 mm (combined XY), requested 6.0
err = Resmsg[3] = OffsetXY[41] failed - Offset too large. Max offset in closed loop is 0.6 mm (combined XY), requested 6.0
err = Resmsg[4] = guiding is not active
IDL> iif_offset,0,0.0,COORD=DETXY

BACK ON SX- HIP80414

Starting Adsec and Wfs sx.

TELESCOPE COLLIMATION. Some problem to authorize the left side bent rear.

PRESET
IDL> iif_preset,HIP80414,GS=5,RS=0,MODE=ACE,PA=0

# Target star: 0 HIP80414
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65
# Guide star: 5
16:24:28.3158 +44:43:44.134 RADEC 2000.0 J2000 -3.3 7.6 13.43 R 1.21 B_V 0.63
# AO Reference star: 0
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65

preset, HIP80414, MODE=ACE, GS=5, RS=0

===> Error in irc_presetTelescope
err = Resmsg[0] = PresetTelescope result status: Warning err = Resmsg[1] = Stop[25] complete
err = Resmsg[2] = guide star not stable in guide box
err = Resmsg[3] = Illegal WFS spec: None

FAILED on the AOS GUI WFS=none !! We correct the name and do the preset from the eng gui.

PRESET

IDL> iif_preset,HIP80414,GS=5,RS=0,MODE=ACE,PA=0

# Target star: 0 HIP80414
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65
# Guide star: 5
16:24:28.3158 +44:43:44.134 RADEC 2000.0 J2000 -3.3 7.6 13.43 R 1.21 B_V 0.63
# AO Reference star: 0
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65

preset, HIP80414, MODE=ACE, GS=5, RS=0

===> Error in irc_presetTelescope
err = Resmsg[0] = PresetTelescope result status: Error err = Resmsg[1] = Stop[28] complete
err = Resmsg[2] = AGw command status: AGW1: RPC Communication Error
err = Resmsg[3] = error: setProbePosition: couldn't move probe to position (SFP) (x,y,z): -126.192, 84.0643, 20.3353 mm.
err = Resmsg[4] = guidestar list successfully changed but probe couldn't be moved to position.

PROBE FAILS
IDL> iif_preset,HIP80414,GS=7,RS=0,MODE=ACE,PA=90

# Target star: 0 HIP80414
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65
# Guide star: 7
16:25:12.7660 +44:40:52.136 RADEC 2000.0 J2000 3.2 3.4 10.60 R 1.25 B_V 0.63
# AO Reference star: 0
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65

preset, HIP80414, MODE=ACE, GS=7, RS=0, PA=90

===> Error in irc_presetTelescope
err = Resmsg[0] = PresetTelescope result status: Error err = Resmsg[1] = Stop[30] complete
err = Resmsg[2] = AGw command status: AGW1: RPC Communication Error
err = Resmsg[3] = error: setProbePosition: couldn't move probe to position (SFP) (x,y,z): 18.9316, 157.642, 19.2573 mm.
err = Resmsg[4] = guidestar list successfully changed but probe couldn't be moved to position.

PROBE FAILS

GENO is checking with the software guy in Tucson about what to do.

RESET of 0, 1 and sandbox

PROBE BACK TO LIFE

IDL> agw_set_station,irtc

**********************************************
*
* => IRTC at LBBG Focal Station
*
**********************************************

IDL> iif_register
IDL> iif_preset,HIP80414,GS=7,RS=0,MODE=ACE,PA=90

# Target star: 0 HIP80414
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65
# Guide star: 7
16:25:12.7660 +44:40:52.136 RADEC 2000.0 J2000 3.2 3.4 10.60 R 1.25 B_V 0.63
# AO Reference star: 0
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65

preset, HIP80414, MODE=ACE, GS=7, RS=0, PA=90

IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] = RunAO result status: Error
err = Resmsg[1] = Timeout error on baysidey (-5001) TIMEOUT_ERROR
err = Resmsg[2] = AcquireRef[35] failed - Timeout error on baysidey (-5001) TIMEOUT_ERROR

BAYSIDE X STAGE FAILS. STOP AT 37.00mm.

IDL> iif_preset,HIP80414,GS=7,RS=0,MODE=ACE,PA=90

# Target star: 0 HIP80414
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65
# Guide star: 7
16:25:12.7660 +44:40:52.136 RADEC 2000.0 J2000 3.2 3.4 10.60 R 1.25 B_V 0.63
# AO Reference star: 0
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65

preset, HIP80414, MODE=ACE, GS=7, RS=0, PA=90

BAYSIDE X STAGE WORKAROUND, moving by the panel the stage

IDL> iif_runao

SUCCESS.

BUT AFTER THE RESTART OF THE WFS the CUBE is in the optical path

NEW PRESET

IDL> iif_preset,HIP80414,GS=7,RS=0,MODE=ACE,PA=90

# Target star: 0 HIP80414
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65
# Guide star: 7
16:25:12.7660 +44:40:52.136 RADEC 2000.0 J2000 3.2 3.4 10.60 R 1.25 B_V 0.63
# AO Reference star: 0
16:24:48.071 +44:41:23.79 RADEC 2000.0 J2000 6.3 5.7 7.65 R 0.00 B_V 1.65

preset, HIP80414, MODE=ACE, GS=7, RS=0, PA=90

IDL> iif_runao

DIMM 0.9, GUIDER 1.08.

Mentre discutiamo con Alfio il problema dell'offset sul lato destro, Carmelo prende una serie infinita di immagini a 10ms 200ms e 2s su quella che spera essere una tripla.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120606_063737 HIP80414 7.9 78 11 NaN 0.43 181313 1 400 990 1.1 0.7 0.5 3 423 ON 52.5 H 10 500 ONSKY NO  
20120606_063814 HIP80414 7.9 78 11 NaN 0.52 181313 1 400 990 1.1 0.7 0.5 3 417 ON 53.5 H 10 500 ONSKY NO  
20120606_063853 HIP80414 7.9 78 11 NaN 0.52 181313 1 400 990 1.1 0.7 0.5 3 417 ON 52.5 H 10 500 ONSKY NO  
20120606_064026 HIP80414 7.9 78 9 NaN 0.59 181313 1 400 990 1.1 0.7 0.5 3 406 ON 54.3 H 10 500 ONSKY NO  
20120606_064104 HIP80414 7.9 78 9 NaN 0.46 181313 1 400 990 1.1 0.7 0.5 3 412 ON 58.8 H 10 500 ONSKY NO  
20120606_064143 HIP80414 7.9 78 11 NaN 0.63 181313 1 400 990 1.1 0.7 0.5 3 418 ON 53.0 H 10 500 ONSKY NO  
20120606_064233 HIP80414 7.9 78 8 NaN 0.57 181313 1 400 990 1.1 0.7 0.5 3 415 ON 22.5 H 200 50 ONSKY NO  
20120606_064248 HIP80414 7.9 78 9 NaN 0.64 181313 1 400 990 1.1 0.7 0.5 3 413 ON 22.7 H 200 50 ONSKY NO  
20120606_064303 HIP80414 7.9 78 9 NaN 0.48 181313 1 400 990 1.1 0.7 0.5 3 420 ON 23.0 H 200 50 ONSKY NO  
20120606_064355 HIP80414 7.9 78 9 NaN 0.46 181313 1 400 990 1.1 0.7 0.5 3 418 ON 24.6 H 200 50 ONSKY NO  
20120606_064410 HIP80414 7.9 78 10 NaN 0.47 181313 1 400 990 1.1 0.7 0.5 3 417 ON 23.8 H 200 50 ONSKY NO  
20120606_064425 HIP80414 7.9 78 10 NaN 0.53 181313 1 400 990 1.1 0.7 0.5 3 418 ON 23.2 H 200 50 ONSKY NO  
20120606_064440 HIP80414 7.9 78 11 NaN 0.47 181313 1 400 990 1.1 0.7 0.5 3 422 ON 4.8 H 2000 5 ONSKY NO  
20120606_064453 HIP80414 7.9 78 10 NaN 0.52 181313 1 400 990 1.1 0.7 0.5 3 418 ON 4.9 H 2000 5 ONSKY NO  
20120606_064507 HIP80414 7.9 78 11 NaN 0.66 181313 1 400 990 1.1 0.7 0.5 3 414 ON 3.9 H 2000 5 ONSKY NO  
20120606_064556 HIP80414 7.9 78 10 NaN 0.46 181313 1 400 990 1.1 0.7 0.5 3 422 ON 4.8 H 2000 5 ONSKY NO  
20120606_064609 HIP80414 7.9 78 11 NaN 0.56 181313 1 400 990 1.1 0.7 0.5 3 419 ON 4.4 H 2000 5 ONSKY NO  
20120606_064623 HIP80414 7.9 78 11 NaN 0.52 181313 1 400 990 1.1 0.7 0.5 3 425 ON 4.3 H 2000 5 ONSKY NO  
20120606_064900 HIP80414 7.9 78 12 NaN 0.49 181313 1 400 990 1.1 0.5 0.5 3 420 ON 50.8 H 10 500 ONSKY NO  
20120606_064939 HIP80414 7.9 78 12 NaN 0.39 181313 1 400 990 1.1 0.5 0.5 3 420 ON 58.6 H 10 500 ONSKY NO  
20120606_065017 HIP80414 7.9 78 11 NaN 0.50 181313 1 400 990 1.1 0.5 0.5 3 413 ON 53.7 H 10 500 ONSKY NO  
20120606_065149 HIP80414 7.9 78 10 NaN 0.44 181313 1 400 990 1.1 0.5 0.5 3 420 ON 48.1 H 10 500 ONSKY NO  
20120606_065227 HIP80414 7.9 78 12 NaN 0.58 181313 1 400 990 1.1 0.5 0.5 3 415 ON 50.4 H 10 500 ONSKY NO  
20120606_065304 HIP80414 8.0 78 11 NaN 0.91 181313 1 400 990 1.1 0.5 0.5 3 389 ON 31.7 H 10 500 ONSKY NO  
20120606_065700 HIP80414 7.9 78 11 NaN 0.59 181313 1 400 990 1.1 0.5 0.5 3 418 ON 5.0 H 2000 5 ONSKY NO  
20120606_065712 HIP80414 7.9 78 10 NaN 0.42 181313 1 400 990 1.1 0.5 0.5 3 417 ON 5.1 H 2000 5 ONSKY NO  
20120606_065727 HIP80414 7.9 78 12 NaN 0.44 181313 1 400 990 1.1 0.5 0.5 3 416 ON 5.1 H 2000 5 ONSKY NO  
20120606_065813 HIP80414 7.9 78 12 NaN 0.58 181313 1 400 990 1.1 0.5 0.5 3 417 ON 5.5 H 2000 5 ONSKY NO  
20120606_065827 HIP80414 7.9 78 11 NaN 0.42 181313 1 400 990 1.1 0.5 0.5 3 420 ON 5.5 H 2000 5 ONSKY NO  
20120606_065841 HIP80414 7.9 78 11 NaN 0.47 181313 1 400 990 1.1 0.5 0.5 3 419 ON 5.5 H 2000 5 ONSKY NO  
20120606_065855 HIP80414 7.9 78 11 NaN 0.53 181313 1 400 990 1.1 0.5 0.5 3 419 ON 24.1 H 200 50 ONSKY NO
20120606_065910 HIP80414 7.9 78 12 NaN 0.51 181313 1 400 990 1.1 0.5 0.5 3 425 ON 24.1 H 200 50 ONSKY NO  
20120606_065926 HIP80414 7.9 78 11 NaN 0.78 181313 1 400 990 1.1 0.5 0.5 3 405 ON 24.5 H 200 50 ONSKY NO  
20120606_070320 HIP80414 7.9 77 11 0.87 0.51 181313 1 400 990 1.2 0.7 0.4 3 420 ON 26.8 J 10 500 ONSKY NO  
20120606_070359 HIP80414 7.9 77 11 1.08 0.58 181313 1 400 990 1.2 0.7 0.4 3 413 ON 17.2 J 10 500 ONSKY NO  
20120606_070437 HIP80414 7.9 77 11 NaN 0.58 181313 1 400 990 1.2 0.7 0.4 3 415 ON 27.4 J 10 500 ONSKY NO  
20120606_070613 HIP80414 7.9 77 11 0.93 0.36 181313 1 400 990 1.2 0.7 0.4 3 424 ON 9.3 J 10 500 ONSKY NO  
20120606_070652 HIP80414 7.9 77 11 0.89 0.63 181313 1 400 990 1.2 0.7 0.4 3 413 ON 32.4 J 10 500 ONSKY NO  
20120606_070730 HIP80414 7.9 77 11 NaN 0.36 181313 1 400 990 1.2 0.7 0.4 3 423 ON 9.7 J 10 500 ONSKY NO  
20120606_071037 HIP80414 7.9 77 11 NaN 0.48 181313 1 400 990 1.1 0.7 0.4 3 421 ON 33.3 J 15 500 ONSKY NO  
20120606_071118 HIP80414 7.9 77 11 NaN 0.53 181313 1 400 990 1.1 0.7 0.4 3 420 ON 28.6 J 15 500 ONSKY NO  
20120606_071157 HIP80414 7.9 77 11 NaN 0.44 181313 1 400 990 1.1 0.7 0.4 3 425 ON 33.8 J 15 500 ONSKY NO  
20120606_071337 HIP80414 7.9 77 10 NaN 0.53 181313 1 400 990 1.1 0.7 0.4 3 418 ON 29.5 J 15 500 ONSKY NO  
20120606_071417 HIP80414 7.9 77 9 NaN 0.70 181313 1 400 990 1.1 0.7 0.4 3 408 ON 27.2 J 15 500 ONSKY NO  
20120606_071456 HIP80414 7.9 77 10 NaN 0.71 181313 1 400 990 1.1 0.7 0.4 3 408 ON 23.9 J 15 500 ONSKY NO  
20120606_071551 HIP80414 7.9 77 11 0.83 0.47 181313 1 400 990 1.1 0.7 0.4 3 417 ON 3.2 J 2000 5 ONSKY NO  
20120606_071604 HIP80414 7.9 77 10 0.83 0.49 181313 1 400 990 1.1 0.7 0.4 3 417 ON 3.1 J 2000 5 ONSKY NO  
20120606_071618 HIP80414 7.9 77 9 0.88 0.55 181313 1 400 990 1.1 0.7 0.4 3 415 ON 3.1 J 2000 5 ONSKY NO  
20120606_071707 HIP80414 7.9 76 10 NaN 0.52 181313 1 400 990 1.1 0.7 0.4 3 416 ON 2.9 J 2000 5 ONSKY NO  
20120606_071720 HIP80414 7.9 76 9 NaN 0.47 181313 1 400 990 1.1 0.7 0.4 3 416 ON 3.0 J 2000 5 ONSKY NO  
20120606_071734 HIP80414 7.9 76 10 NaN 0.54 181313 1 400 990 1.1 0.7 0.4 3 411 ON 3.1 J 2000 5 ONSKY NO  
20120606_071748 HIP80414 7.9 76 9 NaN 0.57 181313 1 400 990 1.1 0.7 0.4 3 415 ON 22.7 J 200 50 ONSKY NO  
20120606_071804 HIP80414 7.9 76 10 NaN 0.49 181313 1 400 990 1.1 0.7 0.4 3 420 ON 21.7 J 200 50 ONSKY NO  
20120606_071819 HIP80414 7.9 76 10 NaN 0.46 181313 1 400 990 1.1 0.7 0.4 3 417 ON 22.1 J 200 50 ONSKY NO  
20120606_071911 HIP80414 7.9 76 8 NaN 0.63 181313 1 400 990 1.1 0.7 0.4 3 416 ON 21.3 J 200 50 ONSKY NO  
20120606_071926 HIP80414 7.9 76 9 NaN 0.49 181313 1 400 990 1.1 0.7 0.4 3 424 ON 21.1 J 200 50 ONSKY NO  
20120606_071941 HIP80414 7.9 76 9 NaN 0.47 181313 1 400 990 1.1 0.7 0.4 3 418 ON 21.2 J 200 50 ONSKY NO  
20120606_072147 HIP80414 7.9 76 9 NaN 0.48 181313 1 400 990 1.1 0.7 0.5 3 419 ON 34.3 J 15 500 ONSKY NO  
20120606_072227 HIP80414 7.9 76 8 NaN 0.49 181313 1 400 990 1.1 0.7 0.5 3 418 ON 34.3 J 15 500 ONSKY NO  
20120606_072307 HIP80414 7.9 76 8 NaN 0.49 181313 1 400 990 1.1 0.7 0.5 3 416 ON 35.9 J 15 500 ONSKY NO  
20120606_072447 HIP80414 7.9 76 8 NaN 0.57 181313 1 400 990 1.1 0.7 0.5 3 415 ON 34.8 J 15 500 ONSKY NO  
20120606_072527 HIP80414 7.9 76 9 0.83 0.53 181313 1 400 990 1.1 0.7 0.5 3 417 ON 35.6 J 15 500 ONSKY NO  
20120606_072607 HIP80414 7.9 76 8 0.91 0.49 181313 1 400 990 1.1 0.7 0.5 3 418 ON 33.7 J 15 500 ONSKY NO  
20120606_072702 HIP80414 7.9 75 8 0.89 0.50 181313 1 400 990 1.1 0.7 0.5 3 416 ON 3.1 J 2000 5 ONSKY NO  
20120606_072715 HIP80414 7.9 75 8 NaN 0.53 181313 1 400 990 1.1 0.7 0.5 3 418 ON 3.1 J 2000 5 ONSKY NO  
20120606_072729 HIP80414 7.9 75 8 NaN 0.47 181313 1 400 990 1.1 0.7 0.5 3 420 ON 3.1 J 2000 5 ONSKY NO  
20120606_072817 HIP80414 7.9 75 7 NaN 0.52 181313 1 400 990 1.1 0.7 0.5 3 417 ON 3.0 J 2000 5 ONSKY NO  
20120606_072831 HIP80414 7.9 75 8 NaN 0.57 181313 1 400 990 1.1 0.7 0.5 3 414 ON 2.9 J 2000 5 ONSKY NO  
20120606_072845 HIP80414 7.9 75 7 NaN 0.60 181313 1 400 990 1.1 0.7 0.5 3 413 ON 2.9 J 2000 5 ONSKY NO  
20120606_072859 HIP80414 7.9 75 7 NaN 0.47 181313 1 400 990 1.1 0.7 0.5 3 418 ON 19.2 J 200 50 ONSKY NO  
20120606_072915 HIP80414 7.9 75 8 NaN 0.47 181313 1 400 990 1.1 0.7 0.5 3 419 ON 19.3 J 200 50 ONSKY NO  
20120606_072930 HIP80414 7.9 75 8 NaN 0.47 181313 1 400 990 1.1 0.7 0.5 3 419 ON 19.1 J 200 50 ONSKY NO  
20120606_073021 HIP80414 7.9 75 8 NaN 0.46 181313 1 400 990 1.1 0.7 0.5 3 419 ON 20.6 J 200 50 ONSKY NO  
20120606_073036 HIP80414 7.9 75 8 NaN 0.48 181313 1 400 990 1.1 0.7 0.5 3 418 ON 20.8 J 200 50 ONSKY NO  
20120606_073051 HIP80414 7.9 75 8 NaN 0.47 181313 1 400 990 1.1 0.7 0.5 3 419 ON 20.4 J 200 50 ONSKY NO  
20120606_073351 HIP80414 7.9 75 8 NaN 0.52 181313 1 400 990 1.1 0.7 0.5 3 415 ON 3.2 J 2000 5 ONSKY NO  
20120606_073404 HIP80414 7.9 75 8 NaN 0.51 181313 1 400 990 1.1 0.7 0.5 3 416 ON 3.3 J 2000 5 ONSKY NO  
20120606_073418 HIP80414 7.9 75 7 NaN 0.56 181313 1 400 990 1.1 0.7 0.5 3 412 ON 3.0 J 2000 5 ONSKY NO  
20120606_073507 HIP80414 7.9 75 8 NaN 0.54 181313 1 400 990 1.1 0.7 0.5 3 414 ON 3.1 J 2000 5 ONSKY NO  
20120606_073520 HIP80414 7.9 74 8 NaN 0.46 181313 1 400 990 1.1 0.7 0.5 3 416 ON 3.2 J 2000 5 ONSKY NO  
20120606_073534 HIP80414 7.9 74 8 NaN 0.55 181313 1 400 990 1.1 0.7 0.5 3 412 ON 3.0 J 2000 5 ONSKY NO  
20120606_073548 HIP80414 7.9 74 8 NaN 0.61 181313 1 400 990 1.1 0.7 0.5 3 412 ON 21.1 J 200 50 ONSKY NO  
20120606_073604 HIP80414 7.9 74 8 NaN 0.53 181313 1 400 990 1.1 0.7 0.5 3 415 ON 21.0 J 200 50 ONSKY NO  
20120606_073619 HIP80414 7.9 74 7 NaN 0.51 181313 1 400 990 1.1 0.7 0.5 3 418 ON 20.5 J 200 50 ONSKY NO  
20120606_073709 HIP80414 7.9 74 8 NaN 0.46 181313 1 400 990 1.1 0.7 0.5 3 414 ON 20.6 J 200 50 ONSKY NO  
20120606_073724 HIP80414 7.9 74 8 NaN 0.50 181313 1 400 990 1.1 0.7 0.5 3 414 ON 20.8 J 200 50 ONSKY NO  
20120606_073740 HIP80414 7.9 74 8 NaN 0.44 181313 1 400 990 1.1 0.7 0.5 3 418 ON 20.0 J 200 50 ONSKY NO  
20120606_073930 HIP80414 7.9 74 9 NaN 0.48 181313 1 400 990 1.1 0.7 0.5 3 414 ON 37.3 J 15 500 ONSKY NO  
20120606_074010 HIP80414 7.9 74 9 NaN 0.55 181313 1 400 990 1.1 0.7 0.5 3 412 ON 35.3 J 15 500 ONSKY NO  
20120606_074049 HIP80414 7.9 74 8 NaN 0.50 181313 1 400 990 1.1 0.7 0.5 3 415 ON 34.5 J 15 500 ONSKY NO  
20120606_074226 HIP80414 7.9 74 8 NaN 0.52 181313 1 400 990 1.1 0.7 0.5 3 416 ON 37.3 J 15 500 ONSKY NO  
20120606_074306 HIP80414 7.9 73 8 NaN 0.50 181313 1 400 990 1.1 0.7 0.5 3 415 ON 36.6 J 15 500 ONSKY NO  
20120606_074346 HIP80414 7.9 73 8 NaN 0.53 181313 1 400 990 1.1 0.7 0.5 3 414 ON 34.1 J 15 500 ONSKY NO  
20120606_074458 HIP80414 7.9 73 9 NaN 0.41 181313 1 400 990 1.1 0.7 0.5 3 421 ON 3.4 J 2000 5 ONSKY NO  
20120606_074510 HIP80414 7.9 73 9 NaN 0.45 181313 1 400 990 1.1 0.7 0.5 3 420 ON 3.4 J 2000 5 ONSKY NO  
20120606_074524 HIP80414 7.9 73 9 NaN 0.45 181313 1 400 990 1.1 0.7 0.5 3 418 ON 3.5 J 2000 5 ONSKY NO  
20120606_074612 HIP80414 7.9 73 9 NaN 0.48 181313 1 400 990 1.1 0.7 0.5 3 417 ON 3.4 J 2000 5 ONSKY NO  
20120606_074625 HIP80414 7.9 73 9 NaN 0.43 181313 1 400 990 1.1 0.7 0.5 3 418 ON 3.7 J 2000 5 ONSKY NO  
20120606_074639 HIP80414 7.9 73 9 NaN 0.40 181313 1 400 990 1.1 0.7 0.5 3 419 ON 3.4 J 2000 5 ONSKY NO  
20120606_074653 HIP80414 7.9 73 9 NaN 0.44 181313 1 400 990 1.1 0.7 0.5 3 418 ON 20.2 J 200 50 ONSKY NO  
20120606_074708 HIP80414 7.9 73 9 NaN 0.43 181313 1 400 990 1.1 0.7 0.5 3 420 ON 20.2 J 200 50 ONSKY NO  
20120606_074723 HIP80414 7.9 73 9 NaN 0.39 181313 1 400 990 1.1 0.7 0.5 3 420 ON 19.9 J 200 50 ONSKY NO  
20120606_074812 HIP80414 7.9 73 9 NaN 0.42 181313 1 400 990 1.1 0.7 0.5 3 419 ON 20.2 J 200 50 ONSKY NO  
20120606_074828 HIP80414 7.9 73 9 NaN 0.44 181313 1 400 990 1.1 0.7 0.5 3 418 ON 19.6 J 200 50 ONSKY NO  
20120606_074843 HIP80414 7.9 73 9 NaN 0.41 181313 1 400 990 1.1 0.7 0.5 3 420 ON 19.3 J 200 50 ONSKY NO  
20120606_073534 HIP80414 7.9 74 8 NaN 0.55 181313 1 400 990 1.1 0.7 0.5 3 412 ON 3.0 J 2000 5 ONSKY NO  
20120606_073548 HIP80414 7.9 74 8 NaN 0.61 181313 1 400 990 1.1 0.7 0.5 3 412 ON 21.1 J 200 50 ONSKY NO  
20120606_073604 HIP80414 7.9 74 8 NaN 0.53 181313 1 400 990 1.1 0.7 0.5 3 415 ON 21.0 J 200 50 ONSKY NO  
20120606_073619 HIP80414 7.9 74 7 NaN 0.51 181313 1 400 990 1.1 0.7 0.5 3 418 ON 20.5 J 200 50 ONSKY NO  
20120606_073709 HIP80414 7.9 74 8 NaN 0.46 181313 1 400 990 1.1 0.7 0.5 3 414 ON 20.6 J 200 50 ONSKY NO  
20120606_073724 HIP80414 7.9 74 8 NaN 0.50 181313 1 400 990 1.1 0.7 0.5 3 414 ON 20.8 J 200 50 ONSKY NO  
20120606_073740 HIP80414 7.9 74 8 NaN 0.44 181313 1 400 990 1.1 0.7 0.5 3 418 ON 20.0 J 200 50 ONSKY NO  
20120606_073930 HIP80414 7.9 74 9 NaN 0.48 181313 1 400 990 1.1 0.7 0.5 3 414 ON 37.3 J 15 500 ONSKY NO  
20120606_074010 HIP80414 7.9 74 9 NaN 0.55 181313 1 400 990 1.1 0.7 0.5 3 412 ON 35.3 J 15 500 ONSKY NO  
20120606_074049 HIP80414 7.9 74 8 NaN 0.50 181313 1 400 990 1.1 0.7 0.5 3 415 ON 34.5 J 15 500 ONSKY NO  
20120606_074226 HIP80414 7.9 74 8 NaN 0.52 181313 1 400 990 1.1 0.7 0.5 3 416 ON 37.3 J 15 500 ONSKY NO  
20120606_074306 HIP80414 7.9 73 8 NaN 0.50 181313 1 400 990 1.1 0.7 0.5 3 415 ON 36.6 J 15 500 ONSKY NO  
20120606_074346 HIP80414 7.9 73 8 NaN 0.53 181313 1 400 990 1.1 0.7 0.5 3 414 ON 34.1 J 15 500 ONSKY NO  
20120606_074458 HIP80414 7.9 73 9 NaN 0.41 181313 1 400 990 1.1 0.7 0.5 3 421 ON 3.4 J 2000 5 ONSKY NO  
20120606_074510 HIP80414 7.9 73 9 NaN 0.45 181313 1 400 990 1.1 0.7 0.5 3 420 ON 3.4 J 2000 5 ONSKY NO  
20120606_074524 HIP80414 7.9 73 9 NaN 0.45 181313 1 400 990 1.1 0.7 0.5 3 418 ON 3.5 J 2000 5 ONSKY NO  
20120606_074612 HIP80414 7.9 73 9 NaN 0.48 181313 1 400 990 1.1 0.7 0.5 3 417 ON 3.4 J 2000 5 ONSKY NO  
20120606_074625 HIP80414 7.9 73 9 NaN 0.43 181313 1 400 990 1.1 0.7 0.5 3 418 ON 3.7 J 2000 5 ONSKY NO  
20120606_074639 HIP80414 7.9 73 9 NaN 0.40 181313 1 400 990 1.1 0.7 0.5 3 419 ON 3.4 J 2000 5 ONSKY NO  
20120606_074653 HIP80414 7.9 73 9 NaN 0.44 181313 1 400 990 1.1 0.7 0.5 3 418 ON 20.2 J 200 50 ONSKY NO  
20120606_074708 HIP80414 7.9 73 9 NaN 0.43 181313 1 400 990 1.1 0.7 0.5 3 420 ON 20.2 J 200 50 ONSKY NO  
20120606_074723 HIP80414 7.9 73 9 NaN 0.39 181313 1 400 990 1.1 0.7 0.5 3 420 ON 19.9 J 200 50 ONSKY NO  
20120606_074812 HIP80414 7.9 73 9 NaN 0.42 181313 1 400 990 1.1 0.7 0.5 3 419 ON 20.2 J 200 50 ONSKY NO  
20120606_074828 HIP80414 7.9 73 9 NaN 0.44 181313 1 400 990 1.1 0.7 0.5 3 418 ON 19.6 J 200 50 ONSKY NO  
20120606_074843 HIP80414 7.9 73 9 NaN 0.41 181313 1 400 990 1.1 0.7 0.5 3 420 ON 19.3 J 200 50 ONSKY NO  

Prova Offset DX

We switch back to DX to test the bug fixing done by Alfio on the AOArbitrator. This was releated to an operation mode during large offset that is now obsolete (TCS 2012C) but was still in the AO DX SW because the software was frozen to last year. Since it is critical for the forecoming PISCES run we decide to fix it and test it.

The test is succesfull. A series of small and large offsets is given through iifoffset and the system behaves as expected: pause loop, move stages and resume if the new position is in the AO acquisition field or just paure loop in case of large offset outside the AO acquisition field.

AO654 (KL_v10 new 400modi PP)

We switch again to SX

Si balla, c'e' buon seeing ma vento e siamo un poco contro vento

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120606_090131 AO654 9.1 82 8 NaN 0.45 181313 1 400 990 0.9 0.5 0.4 3 140 ON 40.5 H 10 500 ONSKY NO  
20120606_090210 AO654 9.1 82 12 NaN 0.49 181313 1 400 990 0.9 0.5 0.4 3 140 ON 56.1 H 10 500 ONSKY NO  
20120606_090248 AO654 9.1 82 11 NaN 0.48 181313 1 400 990 0.9 0.5 0.4 3 138 ON 50.8 H 10 500 ONSKY NO  
20120606_090425 AO654 9.1 81 13 0.78 0.43 181313 1 400 990 0.9 0.5 0.4 3 138 ON 35.9 H 10 500 ONSKY NO  
20120606_090501 AO654 9.1 81 14 NaN 0.46 181313 1 400 990 0.9 0.5 0.4 3 136 ON 40.7 H 10 500 ONSKY NO  
20120606_090538 AO654 9.0 81 14 NaN 0.38 181313 1 400 990 0.9 0.5 0.4 3 147 ON 36.8 H 10 500 ONSKY NO  

AO654 (KL_v10 new 400modi sinusoidale)

Si prova la sinusoidale a 400 modi fatta ieri. L'autogain non ha i range giusti per questa matrice: esagera! Ma lo specchio sopravvive. Fatto autogain 2 volte, e' venuto 0.34, 0.18, 0.34 entrambe.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120606_091120 AO654 9.1 80 13 NaN 0.59 110241 1 400 990 0.3 0.2 0.3 3 141 ON 65.2 H 10 500 ONSKY NO  
20120606_091157 AO654 9.1 80 11 0.87 0.48 110241 1 400 990 0.3 0.2 0.3 3 142 ON 71.7 H 10 500 ONSKY NO  
20120606_091233 AO654 9.1 80 12 0.76 0.53 110241 1 400 990 0.3 0.2 0.3 3 140 ON 65.9 H 10 500 ONSKY NO  
20120606_091407 AO654 9.1 79 13 0.78 0.65 110241 1 400 990 0.3 0.2 0.3 3 139 ON 53.9 H 10 500 ONSKY NO  
20120606_091443 AO654 9.1 79 12 0.93 0.64 110241 1 400 990 0.3 0.2 0.3 3 140 ON 53.2 H 10 500 ONSKY NO  
20120606_091520 AO654 9.1 79 13 0.98 0.52 110241 1 400 990 0.3 0.2 0.3 3 142 ON 69.5 H 10 500 ONSKY NO  

Aggiusto i guadagni a mano nell'assurda speranza di essere piu' furbo dell'autogain, ma il seeing migliora e parte una poderosa oscillazione di TT (vedere modalplot di 092133)

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120606_091735 AO654 9.1 79 12 NaN 0.55 110241 1 400 990 0.4 0.2 -1.0 3 143 ON 65.5 H 10 500 ONSKY NO  
20120606_091811 AO654 9.1 78 11 NaN 0.54 110241 1 400 990 0.4 0.2 -1.0 3 140 ON 70.8 H 10 500 ONSKY NO  
20120606_091848 AO654 9.1 78 10 NaN 0.50 110241 1 400 990 0.4 0.2 -1.0 3 140 ON 73.1 H 10 500 ONSKY NO  
20120606_092021 AO654 9.1 78 11 NaN 0.46 110241 1 400 990 0.4 0.2 -1.0 3 142 ON 73.3 H 10 500 ONSKY NO  
20120606_092057 AO654 9.1 78 12 NaN 0.44 110241 1 400 990 0.4 0.2 -1.0 3 141 ON 45.7 H 10 500 ONSKY NO  
20120606_092133 AO654 9.1 78 12 NaN 0.46 110241 1 400 990 0.4 0.2 -1.0 3 141 ON 34.7 H 10 500 ONSKY NO OSCILLA!

AO654 (KL_v10 new 400modi PP)

Controprova con la Push Pull.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120606_092911 AO654 9.0 76 11 NaN 0.45 181313 1 400 990 1.2 0.5 0.4 3 148 ON 75.2 H 10 500 ONSKY NO  
20120606_092949 AO654 9.0 76 7 NaN 0.50 181313 1 400 990 1.2 0.5 0.4 3 150 ON 75.9 H 10 500 ONSKY NO  
20120606_093028 AO654 9.1 76 9 NaN 0.52 181313 1 400 990 1.2 0.5 0.4 3 142 ON 74.2 H 10 500 ONSKY NO  
20120606_093206 AO654 9.1 75 8 NaN 0.54 181313 1 400 990 1.2 0.5 0.4 3 142 ON 75.1 H 10 500 ONSKY NO  
20120606_093244 AO654 9.1 75 13 NaN 0.67 181313 1 400 990 1.2 0.5 0.4 3 139 ON 65.4 H 10 500 ONSKY NO  
20120606_093322 AO654 9.1 75 12 NaN 0.53 181313 1 400 990 1.2 0.5 0.4 3 140 ON 73.6 H 10 500 ONSKY NO  

AO723 (KL_v10 new 400modi PP)

Si riprova su una altra stella per controprova. Ma JCG me la ripiglia di m=8.5 (catalogo 7.3). Ora siamo fuori vento.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120606_094812 AO723 8.7 70 10 0.85 0.46 181313 1 400 990 1.1 0.7 0.5 3 191 ON 74.6 H 10 500 ONSKY NO  
20120606_094850 AO723 8.7 70 8 0.89 0.53 181313 1 400 990 1.1 0.7 0.5 3 191 ON 74.0 H 10 500 ONSKY NO  
20120606_094928 AO723 8.8 70 9 0.84 0.57 181313 1 400 990 1.1 0.7 0.5 3 188 ON 67.8 H 10 500 ONSKY NO  
20120606_095102 AO723 8.8 71 9 NaN 0.62 181313 1 400 990 1.1 0.7 0.5 3 186 ON 68.9 H 10 500 ONSKY NO  
20120606_095139 AO723 8.8 71 10 NaN 0.59 181313 1 400 990 1.1 0.7 0.5 3 188 ON 66.1 H 10 500 ONSKY NO  
20120606_095217 AO723 8.8 71 10 NaN 0.58 181313 1 400 990 1.1 0.7 0.5 3 187 ON 69.4 H 10 500 ONSKY NO  

AO723 (KL_v10 new 400modi sinusoidale)

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120606_095901 AO723 8.7 72 8 NaN 0.62 110241 1 400 990 0.3 0.2 0.4 3 195 ON 60.5 H 10 500 ONSKY NO  
20120606_095938 AO723 8.7 72 11 NaN 0.52 110241 1 400 990 0.3 0.2 0.4 3 193 ON 67.9 H 10 500 ONSKY NO  
20120606_100015 AO723 8.8 72 11 NaN 0.64 110241 1 400 990 0.3 0.2 0.4 3 189 ON 57.7 H 10 500 ONSKY NO  
20120606_100152 AO723 8.7 73 9 NaN 0.53 110241 1 400 990 0.3 0.2 0.4 3 193 ON 64.9 H 10 500 ONSKY NO  
20120606_100230 AO723 8.8 73 11 NaN 0.60 110241 1 400 990 0.3 0.2 0.4 3 189 ON 62.4 H 10 500 ONSKY NO  
20120606_100309 AO723 8.8 73 9 NaN 0.79 110241 1 400 990 0.3 0.2 0.4 3 183 ON 39.1 H 10 500 ONSKY NO  

E' vero che la sinusoidale fa peggio, ma e' anche vero che pesca sistematicamente i valori minimi di autogain, quindi forse potrebbe fare meglio con dei range di autogain piu' adatti

Fun Time : Nettuno

Forse... si, alla fine ce la fa

Il bin 2 e' nervosetto, aumentando i guadagni dei modi alti pare l'orecchia allo specchio. Si mette il bin 3 e si va in souplesse.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120606_111457 Neptune 10.2 42 5 NaN 0.98 201828 3 66 990 2.6 1.8 -1.0 6 410 ON 7.9 H 1000 15 ONSKY NO H grande
20120606_111516 Neptune 10.2 42 9 NaN 0.80 201828 3 66 990 2.6 1.8 -1.0 6 440 ON 7.1 H 1000 15 ONSKY NO H grande
20120606_111534 Neptune 10.2 42 8 NaN 0.93 201828 3 66 990 2.6 1.8 -1.0 6 428 ON 6.7 H 1000 15 ONSKY NO H grande
20120606_111629 Neptune 10.4 42 7 NaN 1.19 201828 3 66 990 2.6 1.8 -1.0 6 373 ON 0.4 H 1000 15 ONSKY NO H grande
20120606_111647 Neptune 10.3 42 10 NaN 1.18 201828 3 66 990 2.6 1.8 -1.0 6 393 ON 0.5 H 1000 15 ONSKY NO H grande
20120606_111705 Neptune 10.2 42 9 NaN 1.04 201828 3 66 990 2.6 1.8 -1.0 6 415 ON 7.4 H 1000 15 ONSKY NO H grande
20120606_111956 Neptune 10.1 42 12 NaN 0.74 201828 3 66 990 2.6 1.8 -1.0 6 452 ON 2.0 J 1000 15 ONSKY NO J grande
20120606_112014 Neptune 10.2 42 7 NaN 0.77 201828 3 66 990 2.6 1.8 -1.0 6 446 ON 2.0 J 1000 15 ONSKY NO J grande
20120606_112032 Neptune 10.1 42 10 NaN 0.66 201828 3 66 990 2.6 1.8 -1.0 6 458 ON 2.0 J 1000 15 ONSKY NO J grande
20120606_112125 Neptune 10.2 43 7 NaN 0.83 201828 3 66 990 2.6 1.8 -1.0 6 444 ON 1.9 J 1000 15 ONSKY NO J grande
20120606_112143 Neptune 10.2 43 5 NaN 0.79 201828 3 66 990 2.6 1.8 -1.0 6 447 ON 2.0 J 1000 15 ONSKY NO J grande
20120606_112201 Neptune 10.1 43 5 NaN 0.71 201828 3 66 990 2.6 1.8 -1.0 6 452 ON 2.0 J 1000 15 ONSKY NO J grande
20120606_112552 Neptune 10.3 43 9 NaN 1.31 201828 3 66 990 2.6 1.8 -1.0 6 390 ON 0.3 H 4000 30 ONSKY NO H medio fuori fuoco?
20120606_112915 Neptune 10.3 43 10 NaN 1.16 201828 3 66 990 2.5 2.3 -1.0 6 396 ON ** H 4000 30 ONSKY NO H medio aperto e richiuso
20120606_113126 Neptune 10.2 43 9 NaN 0.91 201828 3 66 990 2.5 2.1 -1.0 6 434 ON 0.6 H 4000 30 ONSKY NO H medio ok
20120606_113337 Neptune 10.3 44 10 NaN 1.33 201828 3 66 990 2.5 2.1 -1.0 6 397 ON 0.5 H 4000 30 ONSKY NO H medio ok
20120606_113704 Neptune 11.3 44 9 NaN 1.60 201828 3 66 990 2.5 2.1 -1.0 6 158 ON ** J 4000 30 ONSKY NO J medio rip