-- RunaBriguglio - 13 Sep 2013

System restart

Spot found, coma and focus corrected with IIF

HEXAPOD POSITION

x=2
y=1.9
z=3.76
rx=-383
ry=254
rz=0

Astigmatism corrected with AdSec command:

the default flat has some problems with the shell edge. much better results with 20130829_060533

some features on the CCD39 removed moving the cube

New intmat calibrated

20130913_2

possibly problematic because of low amplitude?

the Closed Loop test however diverged (showing some low order mode diverging...)

New intmat calibrated

20130913_231905

TT =9V, 150Hz

OVS=125

the Closed Loop test diverged again. there is a strange cycling pattern when testing 120 modes. when testing more modes, we have a skip frame.

new intmat 10 modes