2012 Sep 19 UT

Simone, Lorenzo, Armando, Juan-Carlos

03:00UT DIMM 1.0

AO734 03:00UT (mag 15.3 con ND)

03:18UT DIMM 0.8

Siamo ancora sulla versione del oftware SRC4

vari tentativi di usare il codice di Alfio con la modifica che stima la mag in loop chiuso: problemi con i file installati e bug vari. Tentativo FALLITO e torniamo alla versione originale FLAO2-20120129.

04:33UT DIMM 0.9

torniamo indietro alla versione originale

05:34UT DIMM 0.9

BIN4 a 200Hz (nuovo ricostruttore)

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_053625 AO734 15.3 74 2 0.83 0.55 223524 4 36 200 1.8 0.5 -1.0 6 44 ON 15.5 H 200 50 ONSKY zero.fits  
20120919_053647 AO734 15.3 74 3 0.93 0.57 223524 4 36 200 1.8 0.5 -1.0 6 44 ON 13.5 H 200 50 ONSKY zero.fits  
20120919_053748 AO734 15.3 73 2 0.92 0.57 223524 4 36 200 1.8 0.5 -1.0 6 41 ON 22.2 H 200 50 ONSKY zero.fits  
20120919_053810 AO734 15.3 73 2 0.88 0.61 223524 4 36 200 1.8 0.5 -1.0 6 41 ON 20.0 H 200 50 ONSKY zero.fits  
20120919_053832 AO734 15.4 73 3 0.95 0.62 223524 4 36 200 1.8 0.5 -1.0 6 40 ON 16.1 H 200 50 ONSKY zero.fits  
20120919_053901 AO734 15.3 73 2 0.98 0.70 223524 4 36 200 1.8 0.5 -1.0 6 42 ON 12.6 H 200 50 ONSKY zero.fits  
20120919_053923 AO734 15.3 73 2 0.99 0.68 223524 4 36 200 1.8 0.5 -1.0 6 42 ON 15.0 H 200 50 ONSKY zero.fits  
20120919_053946 AO734 15.3 73 3 1.05 0.68 223524 4 36 200 1.8 0.5 -1.0 6 43 ON 16.5 H 200 50 ONSKY zero.fits  
20120919_054048 AO734 15.3 73 2 0.96 0.59 223524 4 36 200 1.8 0.5 -1.0 6 43 ON 18.6 H 200 50 ONSKY zero.fits  
20120919_054110 AO734 15.3 73 2 0.89 0.64 223524 4 36 200 1.8 0.5 -1.0 6 43 ON 14.8 H 200 50 ONSKY zero.fits  
20120919_054132 AO734 15.3 73 2 0.89 0.63 223524 4 36 200 1.8 0.5 -1.0 6 43 ON 18.7 H 200 50 ONSKY zero.fits  
20120919_054951 AO734 15.3 71 3 0.86 0.55 223524 4 36 200 1.8 0.5 -1.0 6 41 ON 22.5 H 200 50 ONSKY zero.fits  
20120919_055013 AO734 15.4 71 3 0.93 0.56 223524 4 36 200 1.8 0.5 -1.0 6 41 ON 23.7 H 200 50 ONSKY zero.fits  
20120919_055035 AO734 15.4 71 3 0.94 0.68 223524 4 36 200 1.8 0.5 -1.0 6 41 ON 19.5 H 200 50 ONSKY zero.fits  
20120919_055137 AO734 15.4 71 2 0.88 0.56 223524 4 36 200 1.8 0.5 -1.0 6 41 ON 23.2 H 200 50 ONSKY zero.fits - Pupils not on target
20120919_055159 AO734 15.4 71 2 0.90 0.58 223524 4 36 200 1.8 0.5 -1.0 6 41 ON 19.6 H 200 50 ONSKY zero.fits  
20120919_055221 AO734 15.4 71 2 0.88 0.56 223524 4 36 200 1.8 0.5 -1.0 6 41 ON 23.0 H 200 50 ONSKY zero.fits  
IDL string: ['20120919_053625','20120919_053647','20120919_053748','20120919_053810','20120919_053832','20120919_053901','20120919_053923','20120919_053946','20120919_054048','20120919_054110','20120919_054132','20120919_054951','20120919_055013','20120919_055035','20120919_055137','20120919_055159','20120919_055221']

SR average (stdev) = 0.185294

SR std_dev = 0.0361118

Seeing average = 0.607526

Seeing std_dev = 0.0517466

OPEN LOOP

Ora un OPEN LOOP o meglio un CL in TT con guadagno piccolo (0.1)

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_055544 AO734 15.3 70 2 0.93 0.38 223524 4 36 200 0.1 -1.0 -1.0 6 42 ON 1.0 H 2000 5 ONSKY zero.fits  
20120919_055606 AO734 15.3 70 2 0.91 0.37 223524 4 36 200 0.1 -1.0 -1.0 6 43 ON 0.8 H 2000 5 ONSKY zero.fits  
20120919_055628 AO734 15.3 70 2 0.92 0.40 223524 4 36 200 0.1 -1.0 -1.0 6 43 ON 0.8 H 2000 5 ONSKY zero.fits - Background correction failed
20120919_055729 AO734 15.3 70 2 1.00 0.39 223524 4 36 200 0.1 -1.0 -1.0 6 41 ON 0.5 H 2000 5 ONSKY zero.fits  
20120919_055751 AO734 15.3 70 1 1.00 0.35 223524 4 36 200 0.1 -1.0 -1.0 6 42 ON 0.8 H 2000 5 ONSKY zero.fits  
20120919_055813 AO734 15.3 70 1 0.96 0.36 223524 4 36 200 0.1 -1.0 -1.0 6 41 ON 0.9 H 2000 5 ONSKY zero.fits - Background correction failed
IDL string: ['20120919_055544','20120919_055606','20120919_055628','20120919_055729','20120919_055751','20120919_055813']

SR average (stdev) = 0.00803640

SR std_dev = 0.00180960

Seeing average = 0.374718

Seeing std_dev = 0.0179737

  • ol_psf_profile.png: profilo radiale della PSF: FWHM 0.5as
    ol_psf_profile.png

BIN4 a 200Hz (Ricostruttore vecchio)

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_061032 AO734 15.3 67 1 0.94 0.65 200610 4 36 200 2.0 0.9 0.5 6 40 ON 15.1 H 400 25 ONSKY zero.fits  
20120919_061054 AO734 15.3 67 2 0.93 0.70 200610 4 36 200 2.0 0.9 0.5 6 41 ON 16.5 H 400 25 ONSKY zero.fits  
20120919_061116 AO734 15.3 67 1 0.93 0.64 200610 4 36 200 2.0 0.9 0.5 6 41 ON 17.7 H 400 25 ONSKY zero.fits  
20120919_061216 AO734 15.3 67 2 0.90 0.59 200610 4 36 200 2.0 0.9 0.5 6 41 ON 18.6 H 400 25 ONSKY zero.fits  
20120919_061238 AO734 15.3 67 2 0.94 0.61 200610 4 36 200 2.0 0.9 0.5 6 40 ON 15.6 H 400 25 ONSKY zero.fits  
20120919_061300 AO734 15.3 67 2 0.99 0.71 200610 4 36 200 2.0 0.9 0.5 6 40 ON 13.5 H 400 25 ONSKY zero.fits  
20120919_061357 AO734 15.3 67 2 0.96 0.64 200610 4 36 200 2.0 0.9 0.5 6 41 ON 16.5 H 400 25 ONSKY zero.fits  
20120919_061419 AO734 15.3 67 2 0.97 0.60 200610 4 36 200 2.0 0.9 0.5 6 40 ON 16.4 H 400 25 ONSKY zero.fits  
20120919_061441 AO734 NaN 66 2 0.99 -1.00 200610 4 36 200 2.0 0.9 0.5 6 -1 -1 16.4 H 400 25 ONSKY NO  
20120919_061540 AO734 15.3 66 2 0.93 0.49 200610 4 36 200 2.0 0.9 0.5 6 41 ON 24.7 H 400 25 ONSKY zero.fits  
20120919_061602 AO734 15.3 66 2 0.89 0.51 200610 4 36 200 2.0 0.9 0.5 6 44 ON 19.8 H 400 25 ONSKY zero.fits  
20120919_061624 AO734 15.3 66 2 0.82 0.52 200610 4 36 200 2.0 0.9 0.5 6 40 ON 21.0 H 400 25 ONSKY zero.fits  
IDL string: [,'20120919_061032','20120919_061054','20120919_061116','20120919_061216','20120919_061238','20120919_061300','20120919_061357','20120919_061419','20120919_061441','20120919_061540','20120919_061602','20120919_061624']

SR average (stdev) = 0.176603

SR std_dev = 0.0302827

Seeing average = 0.606966

Seeing std_dev = 0.0739295

AO1066 (magR 11.1) 06:22UT (mag 16 con ND)

BIN4 a 100Hz (Ricostruttore vecchio)

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_063651 AO1066 16.0 82 2 NaN 0.64 200610 4 36 100 2.9 1.6 0.9 6 46 ON 13.0 H 400 25 ONSKY zero.fits  
20120919_063734 AO1066 16.0 82 2 NaN 0.49 200610 4 36 100 2.9 1.6 0.9 6 45 ON 12.8 H 400 25 ONSKY zero.fits  
20120919_063816 AO1066 15.9 82 3 NaN 0.54 200610 4 36 100 2.9 1.6 0.9 6 46 ON 16.2 H 400 25 ONSKY zero.fits  
20120919_063937 AO1066 16.0 82 2 NaN 0.50 200610 4 36 100 2.9 1.6 0.9 6 45 ON 16.7 H 400 25 ONSKY zero.fits  
20120919_064019 AO1066 16.0 82 2 NaN 0.44 200610 4 36 100 2.9 1.6 0.9 6 45 ON 18.8 H 400 25 ONSKY zero.fits  
20120919_064101 AO1066 16.0 82 2 NaN 0.52 200610 4 36 100 2.9 1.6 0.9 6 45 ON 18.6 H 400 25 ONSKY zero.fits  
20120919_064153 AO1066 16.0 82 2 0.82 0.47 200610 4 36 100 2.9 1.6 0.9 6 44 ON 17.4 H 400 25 ONSKY zero.fits  
20120919_064235 AO1066 16.0 82 2 0.83 0.46 200610 4 36 100 2.9 1.6 0.9 6 44 ON 19.3 H 400 25 ONSKY zero.fits  
20120919_064317 AO1066 NaN 82 2 0.82 -1.00 200610 4 36 100 2.9 1.6 0.9 6 -1 -1 18.6 H 400 25 ONSKY NO - CRC unknown - FLTimeout unknown - PendingCounter unknown - SkipCounter unknown - TimeStamp unknown - WFSGlobalTimeout unknown
20120919_064439 AO1066 15.9 82 2 0.86 0.44 200610 4 36 100 2.9 1.6 0.9 6 47 ON 19.3 H 400 25 ONSKY zero.fits  
20120919_064521 AO1066 16.0 82 3 0.85 0.45 200610 4 36 100 2.9 1.6 0.9 6 45 ON 18.2 H 400 25 ONSKY zero.fits  
20120919_064603 AO1066 16.0 82 2 NaN 0.41 200610 4 36 100 2.9 1.6 0.9 6 45 ON 18.1 H 400 25 ONSKY zero.fits  
IDL string: [,'20120919_063651','20120919_063734','20120919_063816','20120919_063937','20120919_064019','20120919_064101','20120919_064153','20120919_064235','20120919_064317','20120919_064439','20120919_064521','20120919_064603']

SR average (stdev) = 0.172352

SR std_dev = 0.0225018

Seeing average = 0.487051

Seeing std_dev = 0.0625201

BIN4 a 100Hz (Ricostruttore nuovo)

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_064603 AO1066 16.0 82 2 NaN 0.41 200610 4 36 100 2.9 1.6 0.9 6 45 ON 18.1 H 400 25 ONSKY zero.fits  
20120919_065937 AO1066 16.0 82 2 NaN 0.59 223524 4 36 100 2.7 0.5 0.9 6 45 ON 11.8 H 400 25 ONSKY zero.fits  
20120919_070019 AO1066 16.0 81 2 0.85 0.53 223524 4 36 100 2.7 0.5 0.9 6 44 ON 18.6 H 400 25 ONSKY zero.fits  
20120919_070101 AO1066 16.0 81 3 0.77 0.51 223524 4 36 100 2.7 0.5 0.9 6 44 ON 18.7 H 400 25 ONSKY zero.fits  
20120919_070221 AO1066 16.0 81 2 0.93 0.52 223524 4 36 100 2.7 0.5 0.9 6 45 ON 20.2 H 400 25 ONSKY zero.fits  
20120919_070303 AO1066 16.0 81 2 0.90 0.53 223524 4 36 100 2.7 0.5 0.9 6 45 ON 18.3 H 400 25 ONSKY zero.fits  
20120919_070348 AO1066 16.0 81 3 0.85 0.49 223524 4 36 100 2.7 0.5 0.9 6 46 ON 21.1 H 400 25 ONSKY zero.fits  
20120919_070523 AO1066 16.0 81 2 0.00 0.52 223524 4 36 100 2.7 0.5 0.9 6 44 ON 19.1 H 400 25 ONSKY zero.fits  
20120919_070605 AO1066 16.0 81 2 0.85 0.46 223524 4 36 100 2.7 0.5 0.9 6 43 ON 24.1 H 400 25 ONSKY zero.fits  
20120919_070647 AO1066 16.0 81 1 0.85 0.46 223524 4 36 100 2.7 0.5 0.9 6 45 ON 21.1 H 400 25 ONSKY zero.fits  
20120919_070807 AO1066 16.0 81 2 0.86 0.48 223524 4 36 100 2.7 0.5 0.9 6 46 ON 20.1 H 400 25 ONSKY zero.fits  
20120919_070849 AO1066 16.0 81 1 0.85 0.59 223524 4 36 100 2.7 0.5 0.9 6 44 ON 20.3 H 400 25 ONSKY zero.fits  
20120919_070930 AO1066 16.0 81 2 0.92 0.50 223524 4 36 100 2.7 0.5 0.9 6 44 ON 20.6 H 400 25 ONSKY zero.fits  
IDL string: ['20120919_064603','20120919_065937','20120919_070019','20120919_070101','20120919_070221','20120919_070303','20120919_070348','20120919_070523','20120919_070605','20120919_070647','20120919_070807','20120919_070849','20120919_070930']

SR average (stdev) = 0.193938

SR std_dev = 0.0279058

Seeing average = 0.507154

Seeing std_dev = 0.0498880

Rotatore strumento 212deg: lente di camara a saturazione (-100um)

OPEN LOOP

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_071144 AO1066 16.0 80 2 0.00 0.40 223524 4 36 100 0.1 -1.0 -1.0 6 44 ON 3.2 H 2000 5 ONSKY zero.fits - Warning: Selected IRTC dark +1h old!
% AOELAB::MAG: impossible to compute the magnitude
20120919_071226 AO1066 NaN 80 2 0.00 -1.00 223524 4 36 100 0.1 -1.0 -1.0 6 -1 -1 2.0 H 2000 5 ONSKY NO - CRC unknown - FLTimeout unknown - PendingCounter unknown - SkipCounter unknown - TimeStamp unknown - WFSGlobalTimeout unknown - Background correction failed - Warning: Selected IRTC dark +1h old!
20120919_071309 AO1066 16.0 80 2 NaN 0.38 223524 4 36 100 0.1 -1.0 -1.0 6 43 ON 1.2 H 2000 5 ONSKY zero.fits - Warning: Selected IRTC dark +1h old!
20120919_071428 AO1066 16.0 80 2 0.93 0.34 223524 4 36 100 0.1 -1.0 -1.0 6 44 ON 0.9 H 2000 5 ONSKY zero.fits - Warning: Selected IRTC dark +1h old!
20120919_071509 AO1066 16.1 80 2 0.92 0.34 223524 4 36 100 0.1 -1.0 -1.0 6 43 ON 1.4 H 2000 5 ONSKY zero.fits - Warning: Selected IRTC dark +1h old!
20120919_071551 AO1066 16.1 80 2 0.87 0.36 223524 4 36 100 0.1 -1.0 -1.0 6 42 ON 0.7 H 2000 5 ONSKY zero.fits  
IDL string: ['20120919_071144','20120919_071226','20120919_071309','20120919_071428','20120919_071509','20120919_071551']

SR average (stdev) = 0.0156421

SR std_dev = 0.00887535

Seeing average = 0.364303

Seeing std_dev = 0.0242070

AO1062 (mag con ND 16.0)

Risulta essere di mag 16.0 come la precedente, interrompiamo l'acquisizione e passiamo ad un'altra stella

AO41 (mag con ND 16.4)

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_080310 AO41 16.4 84 3 1.05 0.67 223524 4 36 100 1.8 1.2 0.5 6 33 ON 16.8 H 700 13 ONSKY zero.fits  
20120919_080352 AO41 16.3 84 3 1.05 0.67 223524 4 36 100 1.8 1.2 0.5 6 33 ON 13.3 H 700 13 ONSKY zero.fits  
20120919_080434 AO41 16.3 84 3 0.98 0.65 223524 4 36 100 1.8 1.2 0.5 6 33 ON 15.7 H 700 13 ONSKY zero.fits  
20120919_080553 AO41 16.3 84 2 1.17 0.62 223524 4 36 100 1.8 1.2 0.5 6 33 ON 14.2 H 700 13 ONSKY zero.fits  
20120919_080635 AO41 16.3 85 3 0.99 0.64 223524 4 36 100 1.8 1.2 0.5 6 33 ON 15.4 H 700 13 ONSKY zero.fits  
20120919_080717 AO41 16.3 85 3 1.02 0.60 223524 4 36 100 1.8 1.2 0.5 6 34 ON 16.1 H 700 13 ONSKY zero.fits  
20120919_080811 AO41 NaN 85 3 0.78 -1.00 223524 4 36 100 1.8 1.2 0.5 6 -1 -1 15.2 H 700 13 ONSKY NO  
20120919_080853 AO41 16.3 85 3 NaN 0.69 223524 4 36 100 1.8 1.2 0.5 6 33 ON 16.3 H 700 13 ONSKY zero.fits  
20120919_080936 AO41 16.4 85 3 0.93 0.66 223524 4 36 100 1.8 1.2 0.5 6 31 ON 15.2 H 700 13 ONSKY zero.fits  
20120919_081056 AO41 16.4 85 3 0.31 0.64 223524 4 36 100 1.8 1.2 0.5 6 31 ON 14.3 H 700 13 ONSKY zero.fits  
20120919_081138 AO41 16.4 86 3 0.93 0.81 223524 4 36 100 1.8 1.2 0.5 6 31 ON 12.4 H 700 13 ONSKY zero.fits  
20120919_081220 AO41 16.4 86 2 1.01 0.64 223524 4 36 100 1.8 1.2 0.5 6 31 ON 12.6 H 700 13 ONSKY zero.fits  
IDL string: ['20120919_080310','20120919_080352','20120919_080434','20120919_080553','20120919_080635','20120919_080717','20120919_080811','20120919_080853','20120919_080936','20120919_081056','20120919_081138','20120919_081220']

SR average (stdev) = 0.147953

SR std_dev = 0.0143799

Seeing average = 0.662562

Seeing std_dev = 0.0542678

AO40 (mag 16.5 con ND)

BIN2 a 100Hz

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SNSorted ascending notes
20120919_083458 AO40 16.5 75 3 0.82 0.60 223524 4 36 100 1.6 0.5 -1.0 6 29 ON -1.0 ? -1 -1 ONSKY zero.fits  
20120919_083540 AO40 16.5 76 2 0.85 0.59 223524 4 36 100 1.6 0.5 -1.0 6 29 ON 21.2 H 700 13 ONSKY zero.fits - SR may be in error
20120919_083622 AO40 16.5 76 3 0.86 0.63 223524 4 36 100 1.6 0.5 -1.0 6 29 ON 21.3 H 700 13 ONSKY zero.fits - SR may be in error
20120919_083740 AO40 16.5 76 3 0.81 0.58 223524 4 36 100 1.6 0.5 -1.0 6 28 ON 22.2 H 700 13 ONSKY zero.fits  
20120919_083822 AO40 16.5 76 2 0.87 0.61 223524 4 36 100 1.6 0.5 -1.0 6 28 ON 23.0 H 700 13 ONSKY zero.fits  
20120919_083905 AO40 16.5 76 3 0.85 0.59 223524 4 36 100 1.6 0.5 -1.0 6 28 ON 21.9 H 700 13 ONSKY zero.fits - SR may be in error
20120919_084042 AO40 16.5 76 3 0.85 0.55 223524 4 36 100 1.6 0.5 -1.0 6 29 ON 25.4 H 700 13 ONSKY zero.fits  
20120919_084125 AO40 16.5 76 2 0.88 0.54 223524 4 36 100 1.6 0.5 -1.0 6 29 ON 19.7 H 700 13 ONSKY zero.fits  
20120919_084210 AO40 16.5 76 3 0.86 0.55 223524 4 36 100 1.6 0.5 -1.0 6 29 ON 22.0 H 700 13 ONSKY zero.fits  
20120919_084328 AO40 16.5 77 3 NaN 0.56 223524 4 36 100 1.6 0.5 -1.0 6 28 ON 21.5 H 700 13 ONSKY zero.fits  
20120919_084410 AO40 16.5 77 3 NaN 0.99 223524 4 36 100 1.6 0.5 -1.0 6 28 ON 20.0 H 700 13 ONSKY zero.fits  
20120919_084452 AO40 16.5 77 3 NaN 0.59 223524 4 36 100 1.6 0.5 -1.0 6 28 ON 23.6 H 700 13 ONSKY zero.fits  
IDL string: ['20120919_083458','20120919_083540','20120919_083622','20120919_083740','20120919_083822','20120919_083905','20120919_084042','20120919_084125','20120919_084210','20120919_084328','20120919_084410','20120919_084452']

SR average (stdev) = 0.219764

SR std_dev = 0.0161539

Seeing average = 0.615699

Seeing std_dev = 0.121804

centriamo meglio la PSF nella IRTC

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_084817 AO40 16.5 77 3 NaN 0.54 223524 4 36 100 1.6 0.5 -1.0 6 30 ON 17.8 H 700 13 ONSKY zero.fits  
20120919_084859 AO40 16.5 77 3 0.86 0.52 223524 4 36 100 1.6 0.5 -1.0 6 29 ON 24.5 H 700 13 ONSKY zero.fits  
20120919_084941 AO40 16.5 77 3 0.85 1.47 223524 4 36 100 1.6 0.5 -1.0 6 29 ON 24.0 H 700 13 ONSKY zero.fits  
20120919_085101 AO40 16.6 77 3 0.82 0.96 223524 4 36 100 1.6 0.5 -1.0 6 27 ON 21.9 H 700 13 ONSKY zero.fits  
20120919_085143 AO40 16.6 77 2 0.82 0.58 223524 4 36 100 1.6 0.5 -1.0 6 27 ON 20.4 H 700 13 ONSKY zero.fits  
20120919_085225 AO40 16.6 78 3 0.86 0.56 223524 4 36 100 1.6 0.5 -1.0 6 27 ON 21.7 H 700 13 ONSKY zero.fits  
IDL string: ['20120919_084817','20120919_084859','20120919_084941','20120919_085101','20120919_085143','20120919_085225'] SR average (stdev) = 0.217260

SR std_dev = 0.0246036

Seeing average = 0.771143

Seeing std_dev = 0.381522

AO111 (mag 16.9 con ND)

BIN4 a 100Hz

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_090909 AO111 16.9 84 4 0.82 0.69 223524 4 36 100 0.6 0.5 -1.0 6 21 ON 27.6 H 700 13 ONSKY zero.fits  
20120919_090951 AO111 16.9 84 3 0.88 1.46 223524 4 36 100 0.6 0.5 -1.0 6 19 ON 22.5 H 700 13 ONSKY zero.fits  
20120919_091033 AO111 16.8 84 5 0.80 0.66 223524 4 36 100 0.6 0.5 -1.0 6 22 ON 22.4 H 700 13 ONSKY zero.fits  
20120919_091153 AO111 16.8 84 3 0.86 0.67 223524 4 36 100 0.6 0.5 -1.0 6 21 ON 22.5 H 700 13 ONSKY zero.fits  
20120919_091236 AO111 16.7 84 3 0.83 1.08 223524 4 36 100 0.6 0.5 -1.0 6 23 ON 20.7 H 700 13 ONSKY zero.fits  
20120919_091318 AO111 16.8 85 3 0.86 0.71 223524 4 36 100 0.6 0.5 -1.0 6 21 ON 22.9 H 700 13 ONSKY zero.fits  
IDL string: ['20120919_090909','20120919_090951','20120919_091033','20120919_091153','20120919_091236','20120919_091318']

SR average (stdev) = 0.231116

SR std_dev = 0.0233813

Seeing average = 0.876506

Seeing std_dev = 0.325432

OPEN LOOP

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_091653 AO111 16.9 85 6 0.86 0.66 223524 4 36 100 0.1 -1.0 -1.0 6 20 ON 1.9 H 2000 5 ONSKY zero.fits - SR may be in error
20120919_091735 AO111 17.0 85 4 0.76 0.72 223524 4 36 100 0.1 -1.0 -1.0 6 18 ON 1.7 H 2000 5 ONSKY zero.fits - Pupils not on target
20120919_091817 AO111 17.2 85 3 0.84 0.90 223524 4 36 100 0.1 -1.0 -1.0 6 15 ON 2.0 H 2000 5 ONSKY zero.fits  
20120919_091939 AO111 17.5 86 3 0.92 1.32 223524 4 36 100 0.1 -1.0 -1.0 6 11 ON ** H 2000 5 ONSKY zero.fits - Pupils not on target - SR may be in error
20120919_092021 AO111 17.7 86 6 0.83 *** 223524 4 36 100 0.1 -1.0 -1.0 6 10 ON 32.8 H 2000 5 ONSKY zero.fits - Pupils not on target
20120919_092104 AO111 18.2 86 5 0.81 *** 223524 4 36 100 0.1 -1.0 -1.0 6 6 ON 34.3 H 2000 5 ONSKY zero.fits - Pupils not on target - SR may be in error
IDL string: ['20120919_091653','20120919_091735','20120919_091817','20120919_091939','20120919_092021','20120919_092104']

Seeing average = -6.96480

Seeing std_dev = 12.9800

AO109

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_093455 AO109 17.0 78 3 0.92 0.85 223524 4 36 100 1.2 0.5 0.9 6 17 ON 3.8 H 2000 5 ONSKY zero.fits  
20120919_093537 AO109 17.1 78 3 0.85 0.89 223524 4 36 100 1.2 0.5 0.9 6 17 ON 8.4 H 2000 5 ONSKY zero.fits  
20120919_093619 AO109 17.1 78 3 0.90 0.91 223524 4 36 100 1.2 0.5 0.9 6 17 ON 4.1 H 2000 5 ONSKY zero.fits  
20120919_093740 AO109 17.1 78 3 0.77 0.88 223524 4 36 100 1.2 0.5 0.9 6 17 ON 5.4 H 2000 5 ONSKY zero.fits  
20120919_093822 AO109 17.1 79 3 NaN 0.86 223524 4 36 100 1.2 0.5 0.9 6 17 ON 6.7 H 2000 5 ONSKY zero.fits - Background correction failed
20120919_093905 AO109 17.0 79 3 NaN 0.82 223524 4 36 100 1.2 0.5 0.9 6 18 ON 9.6 H 2000 5 ONSKY zero.fits  
20120919_094940 AO109 17.2 81 3 0.89 0.86 223524 4 36 100 0.3 0.5 -1.0 6 15 ON 5.6 H 2000 5 ONSKY zero.fits  
20120919_095022 AO109 17.1 81 3 0.85 1.38 223524 4 36 100 0.3 0.5 -1.0 6 17 ON 5.5 H 2000 5 ONSKY zero.fits  
20120919_095104 AO109 17.0 81 3 0.82 0.89 223524 4 36 100 0.3 0.5 -1.0 6 17 ON 6.1 H 2000 5 ONSKY zero.fits  
20120919_095223 AO109 17.1 82 3 0.79 0.85 223524 4 36 100 0.3 0.5 -1.0 6 17 ON 11.7 H 2000 5 ONSKY zero.fits  
20120919_095305 AO109 17.3 82 3 0.86 *** 223524 4 36 100 0.3 0.5 -1.0 6 14 ON 6.7 H 2000 5 ONSKY zero.fits - Pupils not on target
20120919_095348 AO109 -NaN 75 3 0.62 *** 223524 4 36 100 0.0 -1.0 -1.0 6 -4 ON 330.3 H 2000 5 ONSKY zero.fits - Pupils not on target - SR may be in error
IDL string: ['20120919_093455','20120919_093537','20120919_093619','20120919_093740','20120919_093822','20120919_093905','20120919_094940','20120919_095022','20120919_095104','20120919_095223','20120919_095305','20120919_095348'] SR average (stdev) = 0.336584 SR std_dev = 0.934557 Seeing average = -11.0288 Seeing std_dev = 30.3253

AO90

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_102654 AO90 17.6 64 3 0.64 0.81 223524 4 36 100 0.3 0.2 -1.0 6 10 ON 13.7 H 1000 10 ONSKY zero.fits - SR may be in error
20120919_102736 AO90 17.4 64 3 0.64 0.74 223524 4 36 100 0.3 0.2 -1.0 6 13 ON 18.3 H 1000 10 ONSKY zero.fits - SR may be in error
20120919_102818 AO90 17.3 64 3 0.64 0.71 223524 4 36 100 0.3 0.2 -1.0 6 13 ON 13.7 H 1000 10 ONSKY zero.fits - SR may be in error
20120919_102937 AO90 17.3 64 3 0.68 0.71 223524 4 36 100 0.3 0.2 -1.0 6 13 ON 21.4 H 1000 10 ONSKY zero.fits - SR may be in error
20120919_103019 AO90 17.6 63 3 0.62 0.77 223524 4 36 100 0.3 0.2 -1.0 6 10 ON 16.4 H 1000 10 ONSKY zero.fits - SR may be in error
20120919_103101 AO90 17.4 63 4 0.64 0.89 223524 4 36 100 0.3 0.2 -1.0 6 13 ON 18.0 H 1000 10 ONSKY zero.fits - Pupils not on target - SR may be in error
20120919_103310 AO90 17.5 63 3 0.62 0.84 223524 4 36 100 0.3 0.2 -1.0 6 11 ON 12.7 H 1000 10 ONSKY zero.fits - Pupils not on target - SR may be in error
20120919_103354 AO90 17.4 63 3 0.66 0.71 223524 4 36 100 0.2 -1.0 -1.0 6 13 ON 6.4 H 1000 10 ONSKY zero.fits - SR may be in error
20120919_103436 AO90 17.1 63 4 0.70 1.35 223524 4 36 100 0.3 0.2 -1.0 6 16 ON 17.9 H 1000 10 ONSKY zero.fits - Pupils not on target - SR may be in error
20120919_103554 AO90 17.4 63 3 0.66 1.15 223524 4 36 100 0.3 0.2 -1.0 6 13 ON 17.8 H 1000 10 ONSKY zero.fits - SR may be in error
20120919_103636 AO90 17.4 63 3 0.67 1.03 223524 4 36 100 0.3 0.2 -1.0 6 12 ON 14.0 H 1000 10 ONSKY zero.fits - SR may be in error
20120919_103718 AO90 17.0 63 3 0.74 0.97 223524 4 36 100 0.3 0.2 -1.0 6 17 ON 14.1 H 1000 10 ONSKY zero.fits - SR may be in error
IDL string: ['20120919_102654','20120919_102736','20120919_102818','20120919_102937','20120919_103019','20120919_103101','20120919_103310','20120919_103354','20120919_103436','20120919_103554','20120919_103636','20120919_103718'] SR average = 0.153685 SR std_dev = 0.0383119 Seeing average = 0.889678 Seeing std_dev = 0.201754

AO112

AO244

Tolta la densita' e torniamo BIN1, reimpostata la slitta z a 45.5mm

chiuso il loop e eseguiro autofocus. Confermato 45.5mm

BIN1 400 modi vecchio

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_115757 AO244 7.4 71 2 0.92 0.39 201326 1 400 990 0.9 0.6 0.6 3 642 ON 88.2 H 2 1000 ONSKY zero.fits  
20120919_115913 AO244 7.4 71 3 0.84 0.33 201326 1 400 990 0.9 0.6 0.6 3 643 ON 89.2 H 2 1000 ONSKY zero.fits  
20120919_120030 AO244 7.4 71 4 0.82 0.32 201326 1 400 990 0.9 0.6 0.6 3 645 ON 89.5 H 2 1000 ONSKY zero.fits  
20120919_120327 AO244 7.4 72 3 0.84 0.33 201326 1 400 990 0.9 0.6 0.6 3 642 ON 89.0 H 2 1000 ONSKY zero.fits  
20120919_120444 AO244 7.4 72 3 0.94 0.34 201326 1 400 990 0.9 0.6 0.6 3 645 ON 86.6 H 2 1000 ONSKY zero.fits  
20120919_120600 AO244 7.4 73 3 0.79 0.33 201326 1 400 990 0.9 0.6 0.6 3 644 ON 88.7 H 2 1000 ONSKY zero.fits  
IDL string: ['20120919_115757','20120919_115913','20120919_120030','20120919_120327','20120919_120444','20120919_120600']

SR average (stdev) = 0.885283

SR std_dev = 0.0104433

Seeing average = 0.340683

Seeing std_dev = 0.0242302

BIN1 500 modi vecchio (filtro J)

CAMBIATO FILTRO: J

Si vedono chiaramente oscillazioni di tilt a peggiorare dall'inizio alla fine della serie di acquisizioni (troppo guadagno di TT?)

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120919_121939 AO244 7.4 75 3 0.85 0.34 182144 1 500 990 0.8 0.6 0.4 3 653 ON 60.2 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_122059 AO244 7.4 76 4 0.81 0.31 182144 1 500 990 0.8 0.6 0.4 3 651 ON 60.7 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_122216 AO244 7.4 76 3 0.89 0.41 182144 1 500 990 0.8 0.6 0.4 3 651 ON 53.3 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_122511 AO244 7.4 77 3 0.88 0.41 182144 1 500 990 0.8 0.6 0.4 3 649 ON 53.2 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_122627 AO244 7.4 77 2 0.89 0.44 182144 1 500 990 0.8 0.6 0.4 3 650 ON 47.0 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_122744 AO244 7.4 77 2 0.94 0.47 182144 1 500 990 0.8 0.6 0.4 3 646 ON 41.6 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_122949 AO244 7.4 78 3 0.80 0.35 182144 1 500 990 0.8 0.6 0.4 3 652 ON 36.6 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_123106 AO244 7.4 78 3 0.85 0.37 182144 1 500 990 0.8 0.6 0.4 3 651 ON 45.3 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_123222 AO244 7.4 78 1 1.00 0.44 182144 1 500 990 0.8 0.6 0.4 3 649 ON 36.6 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_123517 AO244 7.4 79 3 0.91 0.37 182144 1 500 990 0.8 0.6 0.4 3 655 ON 52.5 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_123633 AO244 7.4 79 2 0.89 0.37 182144 1 500 990 0.8 0.6 0.4 3 656 ON 35.0 J 4 1000 ONSKY zero.fits - Pupils mismatch
20120919_123753 AO244 7.4 79 2 0.83 0.39 182144 1 500 990 0.8 0.6 0.4 3 654 ON 36.9 J 4 1000 ONSKY zero.fits - Pupils mismatch
IDL string: ['20120919_121939','20120919_122059','20120919_122216','20120919_122511','20120919_122627','20120919_122744','20120919_122949','20120919_123106','20120919_123222','20120919_123517','20120919_123633','20120919_123753']

SR average (stdev) = 0.465967

SR std_dev = 0.0935743

Seeing average = 0.390678

Seeing std_dev = 0.0445975

-- ArmandoRiccardi - 19 Sep 2012

This topic: FLAO > WebHome > LbtTel > Flao1andFlao2DetailedLogAtTelescope > FLAO2LabLog20120919
Topic revision: 20 Sep 2012, LorenzoBusoni
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback