On the mountain Carmelo and Lorenzo

Restoring operation

Start at 11.30 AM AZ time (in the following we will use UT TIME)

Modified the file /lbt/tcs/2012C/aos/.../SXOffload... to disable offload of high order modes to M1 to prevente M1 panic in daytime. Remember to revert for the nights. Restart AOS after that

Spots are found soon on the IRTC camera with M2 global offset [2.6, 0.13, 4.5, 317, 574,0]

M3 is in the standard leftback bent gregorian position

Acqpupil mode has to be changed to the single circle fit instead of 4, Alfio takes care of it.

Best cube rotation position found at 240.33

Hard working on the concrete pier of the telescope to install PEPSI platform: big oscillation present.

Test of KL_v9 bin1 reconstructor to be sure that we are still ok

TrackNo RefStar Mag Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120530_210405 Retro 7.4 005456 1 400 1000 0.5 0.4 0.3 3 677 ON 32.9 EMPTY 10 100 ON NO SR WRONG
20120530_210432 Retro 7.4 005456 1 400 1000 0.5 0.4 0.3 3 668 ON 31.6 EMPTY 10 1000 ON NO SR WRONG
Ho fatto un po' di casino a salvare questi dati, l'immagine IRTC è senza senso (no filtro) ma il modalplot è significativo e sembra che non ci siano problemi. Il loop a 400modi con il disturbo sta chiuso normalmente con il disturbo da 0.8".

Save a flat 20120530_400modi.

KL_v10 calibration

La KL_V10 è fatta con slaving degli attuatori dell'anello interno del secondario. Per dettagli sentire AR e FQP

Si comincia a fare l'acquisizione della IM per la KL_v10, più per identificare eventuali problemi che per avere buone calibrazioni visto il lavoro martellante sul telescopio.

Si mette comunque tutto al meglio: HBS on, treehouse fans sx e dx off, elevation brakes disengaged. Non serve a molto.

Disturb Basis No modes Type Frames/Movement Push-pull cycle Amplitude fileSorted ascending Notes
20120530_211410 KL_v10 10 PP 3 60 600_optimized_3l.fits  
20120530_222319 KL_v10 600 PP 3 1 600_optimized_3l.fits  
20120530_225852 KL_v10 600 PP 3 1 optimized_amano_3l_scalettato.fits  

IM Basis Disturb Freq Mod Pupils Iteration CL basis CL Rec CL gain Notes
20120530_213853 KL_v10 20120530_211410 600 3 20120507-230757 4 KL_v9 030418 0.05 Martellate
20120530_222434 KL_v10 20120530_222319 600 3 20120507-230757 4 KL_v10 20120530_214131 0.05 Martellate
20120530_225948 KL_v10 20120530_225852 600 3 20120507-230757 4 KL_v10 20120530_214131 0.05 Martellate
20120530_233533 KL_v10 20120530_222319 600 3 20120507-230757 48 KL_v10 20120530_214131 0.05 Martellate

REC Basis IM Skip Frame Avg Frame Remove TipTilt No modes Cutoff Threshold Notes
20120530_214131 KL_v10 20120530_213853 2 2 Y 10 0  
20120530_222710 KL_v10 20120530_222434 2 2 Y 400 0 from 4 iter only
20120530_234954 KL_v10 20120530_233533 2 2 Y 400 0  

Si vorrebbe provare i ricostruttori ma c'è l'handover. Nella fretta si prendono 3 misure inutili con il rec a 10 modi. In realtà il 20120530_222710 è stato brevemente provato con successo ma non ci sono dati. Domani li prendiamo

TrackNo RefStar Mag Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120531_000007 Retro 7.4 214131 1 10 990 0.5 0.2 -1.0 3 645 ON 32.7 J 50 200 OFF NO  
20120531_000038 Retro 7.4 214131 1 10 990 0.5 0.2 -1.0 3 648 ON 35.6 J 50 200 OFF NO  
20120531_000125 Retro 7.4 214131 1 10 990 0.5 0.2 -1.0 3 649 ON 27.5 J 50 200 OFF NO  

Due immaginette che Carmelo ha voluto assolutamente mettere relative alla IM 20120530_222434:

IMtest.jpg

Im_slope.jpg

How to switch off treehouse fan and open the vent doors

1. Open the box in the treehouse with 5 eurocent

2. Turn off the fan rotating to the left the labeled switch (up-right)

3. Move to off the labeled switch in the box (low). The window opens

4. Move the switch back to on (the fan starts and a beep rings)

5. Push the reset the button from the panel OUTSIDE of the treehouse. The beep stops

We tried to test different elevation brakes/stow pin combinations.

At first glance, the best operating condition seems to be with elevation on and all the brakes off. (thanks to John Little)
I Attachment Action Size Date Who Comment
IMtest.jpgjpg IMtest.jpg manage 202 K 30 May 2012 - 22:33 CarmeloArcidiacono 400 modes all brakes off, notice the dark (dust?) spot on the TT
Im_slope.jpgjpg Im_slope.jpg manage 106 K 30 May 2012 - 22:35 CarmeloArcidiacono 400 modes