DAYTIME activity

tutto fatto con lo spot di destra sulla IRTC

we decided yesterday night to measure again the BIN11 recisntructor because in January the pupil hottization was done before appliing the optimal the pupil shift. This lead to loosing some part of the mirror because the optimal shift was 0.7 pixel in one axis.

15:30 RR installed & spot acquired start the measurement.

Closed loop with 400 modes we see that the pupil are not centered and even the central obstruction seems misplaced. We take an optical loop data to get a snapshot of the situation.

TN 20120507_224840

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120507_224840 ??? 7.4 90 13 NaN -1.00 180658 1 400 990 0.1 -1.0 -1.0 3 677 ON -1.0 ? -1 -1 OFF NO  

Center separation: 36.0802 OK!!!

we start the procedure to acquire the new pupil using the routine pupOptimizeGui

1) use acqpupil to find the pupil and define large masks centered.

2) We set a threshold to 1e-8 (the number is set by experience) the we look at the trimmed signal

we identified the new hottizzed pupils having 689 subaps (the DX side has 696 subaps) the central obstruction as now two missing pixel

data acquired:

the pupil acquisition has been done using as pupil centers the data obatined in January when we optimized the signals amplitude by rotating and displacing the pupils over the CCD39 using the pre arranged patterns tronco di coda verticale ecc ecc. The pupil data are listed below:

TN 20120507_225700 pupille allargateprima della hottizazione

TN 20120507_230705 pupille hottizzate a bin1

We start the computation of the new reconstructor BIN1

Step1) acquired a 10 modes reconstructor in open loop REC TN 234513 (high modulation is used)

Step2) acquirig a 50 modes reconstructor REC TN 235018 (high modulation is used)

Step3) acquiring 400 modes in closed loop with the 50 modes. (modulation is set to 3lamda/D)

we use the disturbance optimized_amano_3l_scalettato

acquisition frequency 600Hz

REC TN 000322

Step4) acquire the 600 modes reconstructor in closed loop with the 400 modes rec acquired above.

interrotti da un intervento distrutore di david.

Si prende un nuovo flat a 400 modi. 20120507_400modes

17:41 si lancia una nuova acquisizione di 600 modi.

acquisita IM 600 modi TN 20120508_004107

calcoliamo 4 ricostruttoridalla stessa

400 modi 005456

fatto test con immagini

TN 010616 - 010855 SR J band attorno a 30%

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_010616 ??? 7.3 90 6 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 695 ON 29.7 J 50 200 ON NO  
20120508_010641 ??? 7.3 90 10 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 696 ON 29.9 J 50 200 ON NO  
20120508_010705 ??? 7.3 90 3 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 696 ON 30.0 J 50 200 ON NO  
20120508_010806 ??? 7.4 90 8 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 681 ON 26.7 J 50 200 ON NO  
20120508_010831 ??? 7.3 90 1 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 690 ON 27.0 J 50 200 ON NO  
20120508_010855 ??? 7.3 90 9 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 697 ON 27.4 J 50 200 ON NO  

Center separation: 36.0215 36.0309 36.0215 36.0683 36.0562 36.0616

TN 011614- 011852

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_011614 ??? 7.3 90 11 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 694 ON 26.7 J 50 200 ON NO  
20120508_011638 ??? 7.3 90 4 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 694 ON 27.2 J 50 200 ON NO  
20120508_011703 ??? 7.3 90 9 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 695 ON 26.4 J 50 200 ON NO  
20120508_011803 ??? 7.4 90 10 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 683 ON 26.2 J 50 200 ON NO  
20120508_011828 ??? 7.3 90 11 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 692 ON 25.2 J 50 200 ON NO  
20120508_011852 ??? 7.3 90 11 NaN -1.00 005456 1 400 990 0.5 0.5 0.4 3 691 ON 25.9 J 50 200 ON NO  
IDL string: ['20120508_011614','20120508_011638','20120508_011703','20120508_011803','20120508_011828','20120508_011852']

Center separation: 36.0675 36.0675 36.0800 36.0621 36.0621 36.0621

SR basso ma abbiamo vibrazione su un asse.

500 modi REC 005850 OK

550 modi REC 011809 con due modi un po bassi nella SVD

600 modi REC 013056

si fanno dei test con il 500 modi 005850

il loop chiude bene

si salva anche un flat 20120507_500modes.sav

si fanno dei test con il 550 modi 011809

il loop chiude bene

si fanno dei test con il 600 modi 013056

il loop perde un pezzo di specchio. NON VA !!!!

Confronto saturazione tra 180005 e la nuova 004107:

180005: plot_sat_180005

004107: plot_sat_004107

NIGHTTIME

start at 20:01

preset su SAO27969 mag 8.5 per i test a BIN1

ricorda mirror ventilation off

proviamo il nuovo ricostruttore a BIN1 400 modi 005456con pupille 20120507_230705

facciamo il 100 modi di questo REC che mancava..REC030418. Si aggiorna la tabella per usare le nuove pupille e i nuovi rec

20:06 telescope collimated.

20:07 mandato preset

seeing from DIMM 1.0

Il nuovo REC sembra skippare subito anche a basso gain. Rimettiamo il REC vecchio che skippa anche di piu.

Il seeing del DIMM e 1.0 ma la FWHM del guider dice Runa e piu grande 1.3, 1.7

Skippa anche il BIN2 vecchio. Perde un bordo dello specchio.

Il bin1 a 250 modi dopo essrestato chiuso per un po skippa anche lui.

misura del RON 9.1, 6.2, 3.8, 3.9, poi

Attenzione slope NaN in circolo. Si deve spengere il W per resettare i NaN a giro.

21:08 dopo la frittata si fa ripartire il sistema.

Si misura la distanza fra le pupille che risulta 35.5 sui frame presi a loop chiuso prima per test.

Abilitiamo lo acq_pup separate e si misura la interdistanza sul display del CCD39. Si ha 36.0 spaccato.

misuriamo sul display 36.02 di interdistanza

salviamo i dato epr un check del programma che calcola i centri TN041655 per test..

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_041655 ??? 9.0 68 5 1.18 0.80 185250 1 250 990 1.8 0.9 0.7 3 152 ON -1.0 ? -1 -1 ONSKY NO  

non vi sono skip rimettiamo il vecchio 400 modi per evdere che cosa succcede =>>> sjkippa in modo simile a ieri durate lo autogain

passo lo autogain senka skip

prendiamo optical loop data (separazione 35.83)

si ha uno skip istantaneo appena partiti con la acquisizione

non riusciamo a prendere le immagini perche vi sono sempre degli skip che allla fine non permettono di prenderei dati.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_042310 ??? 9.1 68 2 1.29 0.88 005456 1 400 990 1.6 0.9 0.6 3 143 ON 0.0 H 50 200 ONSKY NO  
20120508_042334 ??? 8.9 68 1 1.28 0.92 005456 1 400 990 1.6 0.9 0.6 3 163 ON 0.0 H 50 200 ONSKY NO  
20120508_042358 ??? 9.2 68 5 1.35 0.14 005456 1 400 990 0.0 -1.0 -1.0 3 131 ON 0.0 H 50 200 ONSKY NO  
20120508_042852 ??? 9.2 67 4 1.18 0.59 005456 1 400 990 1.8 0.9 0.7 3 129 ON 0.0 H 50 200 ONSKY NO - long stop of 56 frames at frame 2216

preset su SAO44572 mag 7.6

per levarsi dal vento anche se e basso.

il REC vecchio a 250 modi va a posto. separazione 35.9

si mette il nuovo 400 modi.autogain OK (separazione 35.82)

si prendono 3 tracking numbers poi al dark ce lo skip quando richiude il loop

044536-044634 SR fra 60 e 77

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_044326 ??? 8.7 76 5 1.34 0.94 005456 1 400 990 1.6 1.1 0.7 3 200 ON 0.0 H 30 300 ONSKY NO - Unknown IRTC pixelscale
20120508_044355 ??? 8.7 76 5 1.32 0.13 005456 1 400 990 0.0 -1.0 -1.0 3 198 ON 0.0 H 30 300 ONSKY NO - Unknown IRTC pixelscale
20120508_044536 ??? 8.7 76 5 NaN 0.96 005456 1 400 990 1.6 0.9 0.7 3 206 ON 65.7 H 30 300 ONSKY NO  
20120508_044605 ??? 8.7 76 5 NaN 1.22 005456 1 400 990 1.6 0.9 0.7 3 196 ON 58.9 H 30 300 ONSKY NO  
20120508_044634 ??? 8.7 76 6 1.17 0.90 005456 1 400 990 1.6 0.9 0.7 3 207 ON 77.8 H 30 300 ONSKY NO  
IDL string: ['20120508_044536','20120508_044605','20120508_044634']

Center Separation: 35.7977 35.7449 35.7877

facciamo una prova per registrare dei segnali ad anello e vedere che slopes li generano.

si salva il TN 045718 dove ce uno skip ad anello

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_045718 ??? 8.9 77 5 1.42 1.12 005456 1 400 990 0.0 -1.0 -1.0 3 171 ON -1.0 ? -1 -1 ONSKY NO - long stop of 80 frames at frame 15849
20120508_045812 ??? 8.8 77 5 1.42 0.17 005456 1 400 990 0.0 -1.0 -1.0 3 174 ON -1.0 ? -1 -1 ONSKY NO  

si analizzache cosa e successso. Il risultato e il plot allegato sotto.

altro TN che skippa duramte l'autogain

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_061643 ??? 8.6 77 4 NaN 0.77 005456 1 400 990 0.1 -1.0 -1.0 3 217 ON -1.0 ? -1 -1 ONSKY NO - long stop of 80 frames at frame 14544
20120508_061736 ??? 8.7 77 5 NaN 0.11 005456 1 400 990 0.0 -1.0 -1.0 3 191 ON -1.0 ? -1 -1 ONSKY NO - Pupils not on target

Centers separation: 35.5412 35.6658

si cerca di preneder una sequenza che skippa dopo lo autogain.

autogain passato

controlliamo che la separazione delle pupille non cambi con l'orientamento dell'agw rispetto alla gravita.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_065536 ??? 8.6 71 3 1.34 0.74 005456 1 400 990 1.6 0.9 0.6 3 221 ON -1.0 ? -1 -1 ONSKY NO  
20120508_065627 ??? 8.6 71 4 1.33 0.85 005456 1 400 990 1.6 0.9 0.6 3 211 ON -1.0 ? -1 -1 ONSKY NO - long stop of 79 frames at frame 15843
20120508_065719 ??? 8.7 71 6 1.56 0.16 005456 1 400 990 0.0 -1.0 -1.0 3 200 ON -1.0 ? -1 -1 ONSKY NO - Pupils not on target
20120508_065811 ??? 8.7 71 3 NaN 0.12 005456 1 400 990 0.0 -1.0 -1.0 3 207 ON -1.0 ? -1 -1 ONSKY NO - Pupils not on target
IDL string: ['20120508_065536','20120508_065627','20120508_065719','20120508_065811']

Centers separation: 35.6246 35.5457 35.1616 34.8892

si fa un preset sulla stessa stella ma con PA diverso di -90

La separazione misurata da acq pupil sulle pupille live e 35.5 porco zio !!!!!!

riacquisiamo altri TN senza PSF: da 072552 a 072734

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_072552 ??? 8.6 66 2 1.13 0.74 005456 1 400 990 1.6 0.9 0.7 3 216 ON -1.0 ? -1 -1 ONSKY NO  
20120508_072643 ??? 8.6 66 5 1.16 0.74 005456 1 400 990 1.6 0.9 0.7 3 218 ON -1.0 ? -1 -1 ONSKY NO  
20120508_072734 ??? 8.6 66 4 1.50 0.71 005456 1 400 990 1.6 0.9 0.7 3 218 ON -1.0 ? -1 -1 ONSKY NO  
20120508_072826 ??? 8.6 65 4 1.25 0.71 005456 1 400 990 1.6 0.9 0.7 3 218 ON -1.0 ? -1 -1 ONSKY NO  
IDL string: ['20120508_072552','20120508_072643','20120508_072734','20120508_072826']

Estimation of pupil center separation: 35.5563 35.5513 35.5494 35.5424

prendiamo le PSF: 072926 to 073229

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_072926 ??? 8.6 65 3 1.24 0.75 005456 1 400 990 1.6 0.9 0.7 3 214 ON 71.3 H 20 300 ONSKY NO  
20120508_072952 ??? 8.6 65 3 1.24 0.77 005456 1 400 990 1.6 0.9 0.7 3 216 ON 73.5 H 20 300 ONSKY NO  
20120508_073019 ??? 8.6 65 5 1.24 0.64 005456 1 400 990 1.6 0.9 0.7 3 221 ON 75.7 H 20 300 ONSKY NO  
20120508_073136 ??? 8.6 65 2 NaN 0.74 005456 1 400 990 1.6 0.9 0.7 3 216 ON 74.9 H 20 300 ONSKY NO - Pupils not on target
20120508_073203 ??? 8.6 65 4 NaN 0.78 005456 1 400 990 1.6 0.9 0.7 3 212 ON 73.0 H 20 300 ONSKY NO  
20120508_073229 ??? 8.6 65 2 NaN 0.77 005456 1 400 990 1.6 0.9 0.7 3 213 ON 74.8 H 20 300 ONSKY NO  
IDL string: ['20120508_072926','20120508_072952','20120508_073019','20120508_073136','20120508_073203','20120508_073229']

Estimation of pupil center separation: 35.5685 35.5421 35.5572 35.5601 35.5649 35.5618

si fa un preset sulla stessa stella ma con PA diverso di -180

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_074646 ??? 8.6 62 4 1.37 0.77 005456 1 400 990 0.1 -1.0 -1.0 3 224 ON -1.0 ? -1 -1 ONSKY NO - Pupils not on target
20120508_074738 ??? 8.6 62 2 1.25 0.91 005456 1 400 990 1.4 0.9 0.7 3 214 ON -1.0 ? -1 -1 ONSKY NO  
IDL string: ['20120508_074646','20120508_074 738']

Estimation of pupil center separation: 35.4696 35.4935

la separazione delle pupille misurata sul display e 35.5 completa l'autogain ma poi skippa subito aloop chiuso.

si fa un preset sulla stessa stella ma con PA diverso di 0

skippa subito

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_074829 ??? 8.6 62 1 1.30 0.70 005456 1 400 990 1.4 0.9 0.7 3 220 ON 74.6 H 20 300 ONSKY NO  
20120508_074856 ??? 8.6 62 2 1.30 0.77 005456 1 400 990 1.4 0.9 0.7 3 216 ON 76.1 H 20 300 ONSKY NO  
20120508_074922 ??? 8.6 62 4 1.31 0.83 005456 1 400 990 1.4 0.9 0.7 3 215 ON 75.6 H 20 300 ONSKY NO  
20120508_075348 ??? 8.6 61 4 NaN 0.81 005456 1 400 990 1.4 0.9 0.7 3 210 ON 74.3 H 20 300 ONSKY NO  
20120508_075414 ??? 8.6 61 1 NaN 0.81 005456 1 400 990 1.4 0.9 0.7 3 215 ON 73.7 H 20 300 ONSKY NO  
20120508_075440 ??? 8.6 61 1 NaN 0.80 005456 1 400 990 1.4 0.9 0.7 3 216 ON 73.6 H 20 300 ONSKY NO  
20120508_075556 ??? 8.7 60 4 NaN 0.96 005456 1 400 990 1.4 0.9 0.7 3 205 ON 68.4 H 20 300 ONSKY NO  
20120508_075623 ??? 8.6 60 5 NaN 0.83 005456 1 400 990 1.4 0.9 0.7 3 210 ON 70.6 H 20 300 ONSKY NO  
20120508_075650 ??? 8.6 60 3 NaN 0.82 005456 1 400 990 1.4 0.9 0.7 3 212 ON 70.6 H 20 300 ONSKY NO  

Estimation of center separation: 35.5093 35.4920 35.4900 35.5057 35.5182 35.4883 35.5170 35.4870 35.5206

Se porta il telescopio al orizonte e poi allo zenith per muovere un po' il mostro.

Preset SAO66660

UT08:13 Preset fallisce. La stella del technical viewer non c'e.... Pointing problem.

UT08:16 Altro preset. RunAO fallisce: Skip in autogain...

PA=0.

UT08:18 RunAO e' andato male: skippa al centro con PA=90.

UT08.34 Altro preset con PA=90.

UT08.40 Altro preset questa volta con PA= - 90.

Fallisce subito. Anellone dei attuatori interni

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_084459 ??? 9.1 70 5 0.94 0.30 005456 1 400 990 0.1 -1.0 -1.0 3 139 ON -1.0 ? -1 -1 ONSKY NO - long stop of 80 frames at frame 2735 - Pupils not on target
20120508_084551 ??? 16.6 70 5 0.94 *** 005456 1 400 100 0.0 -1.0 -1.0 3 1 ON -1.0 ? -1 -1 ONSKY NO Loop Schiatatto

UT08:46 Altro preset con PA=180

eRRORe: no star found (CCD47 vuoto). Puntamento sbagliato

Si trova che la stella e' fuori del campo.

UT08:50 Altro preset, questa volta la luce si vede nel 47.

Telescopio e' andato in TimeOut.

Slitta Y non si muove bene.... sucede a certi angoli...

Alfio ci abbandona. Simone prende il timone.

UT09:05 Altro preset

il TN 090835 va in skip:

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_090835 ??? 9.1 74 4 1.04 0.73 005456 1 400 990 0.2 0.1 -1.0 3 142 ON -1.0 ? -1 -1 ONSKY NO - long stop of 79 frames at frame 19937
20120508_090928 ??? 9.3 74 4 1.07 0.12 005456 1 400 990 0.0 -1.0 -1.0 3 113 ON -1.0 ? -1 -1 ONSKY NO - Pupils not on target

UT9:19 Si manda altro preset.

Si mette il 250 modi perche il 400 non c'e la fa.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120508_100915 ??? 9.1 83 3 0.93 0.41 185250 1 250 990 2.0 1.4 1.1 3 137 ON -1.0 ? -1 -1 ONSKY NO  
20120508_100926 ??? 9.1 83 4 0.93 0.37 185250 1 250 990 2.0 1.4 1.1 3 138 ON 76.3 H 20 300 ONSKY NO - SR may be in error - Warning: Selected IRTC dark +2h old!
20120508_100959 ??? 9.1 83 3 0.47 0.39 185250 1 250 990 2.0 1.4 1.1 3 141 ON 378.2 H 20 300 ONSKY NO - Pupils not on target - PSF fuori FOV
20120508_101032 ??? 9.1 83 3 NaN 0.47 185250 1 250 990 2.0 1.4 1.1 3 139 ON 1367.2 H 20 300 ONSKY NO - Pupils not on target - PSF fuori FOV
20120508_101104 ??? 9.1 83 4 1.02 0.38 185250 1 250 990 2.0 1.4 1.1 3 138 ON ** H 20 300 ONSKY NO PSF fuori FOV
20120508_101137 ??? 9.1 83 3 1.02 0.40 185250 1 250 990 2.0 1.4 1.1 3 139 ON 353.0 H 20 300 ONSKY NO PSF fuori FOV
20120508_101210 ??? 9.1 83 3 NaN 0.36 185250 1 250 990 2.0 1.4 1.1 3 142 ON 51.8 H 20 300 ONSKY NO - SR may be in error - Warning: Selected IRTC dark +2h old!
20120508_101437 ??? 9.0 84 4 NaN 0.93 185250 1 250 990 2.0 1.4 1.1 3 145 ON -1.0 ? -1 -1 ONSKY NO  
20120508_101528 ??? 9.0 84 4 NaN 0.90 185250 1 250 990 2.0 1.4 1.1 3 146 ON -1.0 ? -1 -1 ONSKY NO  
20120508_101619 ??? 9.1 84 3 NaN 0.99 185250 1 250 990 2.0 1.4 1.1 3 140 ON -1.0 ? -1 -1 ONSKY NO  
20120508_101710 ??? 9.1 84 3 NaN 0.98 185250 1 250 990 2.0 1.4 1.1 3 142 ON -1.0 ? -1 -1 ONSKY NO  
20120508_101801 ??? 9.1 84 3 NaN 1.01 185250 1 250 990 2.0 1.4 1.1 3 139 ON -1.0 ? -1 -1 ONSKY NO  
20120508_101852 ??? 9.0 84 3 NaN 0.86 185250 1 250 990 2.0 1.4 1.1 3 145 ON -1.0 ? -1 -1 ONSKY NO - long stop of 58 frames at frame 19178
20120508_101944 ??? 9.1 84 2 NaN 0.91 185250 1 250 990 2.0 1.4 1.1 3 142 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102038 ??? 9.1 84 4 NaN 0.93 185250 1 250 990 2.0 1.4 1.1 3 141 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102130 ??? 9.0 84 4 NaN 0.93 185250 1 250 990 2.0 1.4 1.1 3 143 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102224 ??? 9.0 84 4 NaN 0.84 185250 1 250 990 2.0 1.4 1.1 3 146 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102314 ??? 9.1 84 3 NaN 1.04 185250 1 250 990 2.0 1.4 1.1 3 141 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102406 ??? 9.1 84 3 NaN 0.99 185250 1 250 990 2.0 1.4 1.1 3 139 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102456 ??? 9.1 84 3 NaN 1.04 185250 1 250 990 2.0 1.4 1.1 3 141 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102547 ??? 9.0 84 3 NaN 1.00 185250 1 250 990 2.0 1.4 1.1 3 143 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102638 ??? 9.0 84 3 0.63 0.93 185250 1 250 990 2.0 1.4 1.1 3 147 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102728 ??? 9.0 84 3 0.63 0.92 185250 1 250 990 2.0 1.4 1.1 3 148 ON -1.0 ? -1 -1 ONSKY NO  
20120508_102820 ??? 9.0 83 3 NaN 0.84 185250 1 250 990 2.0 1.4 1.1 3 147 ON -1.0 ? -1 -1 ONSKY NO  
IDL string: ['20120508_100915','20120508_100926','20120508_100959','20120508_101032','20120508_101104','20120508_101137','20120508_101210','20120508_101437', $ '20120508_101528','20120508_101619','20120508_101710','20120508_101801','20120508_101852','20120508_101944','20120508_102038','20120508_102130','20120508_102224', $ '20120508_102314','20120508_102406','20120508_102456','20120508_102547','20120508_102638','20120508_102728','20120508_102820']

Center separation analysis from datasets above: 35.30+/- 0.04 (Sara' un errore di stima con il 250 modi, oppure un vero spostamento ulteriore?????)

TN 102911 al 103520: Da buttare.... Il telescopio si e' spostato al zenith mentre se continuava a prendere dati... i CCD frames sono vuoti...

UT10:30 Simone e Runa salgono al telescopio per verificare che non ci sia un pezzo dentro il W che sia lento.

Mentre rotavano il AGW c'e stato un ROTATOR FAULT!!!

Risultato della inspezzione: nessun pezzo e' lento.... quindi non cappiamo ancora cosa succede con il nostro sistema..... frown, sad smile

-- SimoneEsposito - 07 May 2012-- SimoneEsposito - 07 May 2012
Topic revision: r12 - 11 Sep 2012, EnricoPinna