04---+ Alfio, Enrico, Runa, Mauro

07:45 Montato il retro

Spente THs

08:20 Startup ROT@180deg OK homing slitte

FW#2 Homing fallace: va in posizione con filtro in mezzo al fascio

Oggi vuole un altro offset di homing....

UT 15:47Slopes non arrivano all'Adsec

Reset BCU39, slopes non arrivano ancora.

Spegnamo il lineare -> le slopes ripartono

09:20 Test cloop sinusoidal 500 modes

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120430_162607 ??? 7.5 90 4 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 572 ON 24.8 J 100 100 ON NO  
20120430_162633 ??? 7.6 90 6 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 568 ON 26.9 J 100 100 ON NO  
20120430_162651 ??? 7.6 90 3 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 565 ON 23.9 J 100 100 ON NO  
20120430_162739 ??? 7.6 90 6 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 566 ON 24.4 J 100 100 ON NO - Pupils not on target
20120430_162757 ??? 7.6 90 5 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 567 ON 22.7 J 100 100 ON NO  
20120430_162815 ??? 7.5 90 6 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 578 ON 23.6 J 100 100 ON NO  
20120430_163113 ??? 7.5 90 4 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 580 ON 27.3 J 100 100 ON NO  
20120430_163131 ??? 7.5 90 6 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 576 ON 27.1 J 100 100 ON NO  
20120430_163150 ??? 7.5 90 8 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 574 ON 25.2 J 100 100 ON NO  
20120430_163238 ??? 7.6 90 6 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 548 ON 25.4 J 100 100 ON NO  
20120430_163256 ??? 7.6 90 6 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 558 ON 22.8 J 100 100 ON NO  
20120430_163314 ??? 7.5 90 6 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 580 ON 23.6 J 100 100 ON NO  
IDL string: ['20120430_162607','20120430_162633','20120430_162651','20120430_162739','20120430_162757','20120430_162815','20120430_163113','20120430_163131','20120430_163150','20120430_163238','20120430_163256','20120430_163314']

Facciamo un check cloo con il REC 500 modi standard
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #framesSorted ascending disturb SN notes
20120430_165100 ??? 7.5 90 4 NaN -1.00 181326 1 500 990 0.7 0.4 0.4 3 581 ON 31.2 J 50 200 ON NO  
20120430_165127 ??? 7.5 90 2 NaN -1.00 181326 1 500 990 0.7 0.4 0.4 3 589 ON 33.1 J 50 200 ON NO  
20120430_165151 ??? 7.5 90 2 NaN -1.00 181326 1 500 990 0.7 0.4 0.4 3 587 ON 32.9 J 50 200 ON NO  
20120430_165251 ??? 8.1 90 5 NaN -1.00 181326 1 500 990 0.7 0.4 0.4 3 329 ON 3.8 J 50 200 ON NO - long stop of 79 frames at frame 1138 - Pupils not on target
20120430_165316 ??? 7.5 90 5 NaN -1.00 181326 1 500 990 -1.0 0.0 0.0 3 593 ON 4.9 J 50 200 OFF NO - Pupils not on target
20120430_165536 ??? 7.5 90 5 NaN -1.00 181326 1 500 990 0.6 0.4 0.3 3 591 ON 31.9 J 50 200 ON NO  
20120430_165601 ??? 7.5 90 4 NaN -1.00 181326 1 500 990 0.6 0.4 0.3 3 588 ON 34.4 J 50 200 ON NO  
20120430_165625 ??? 7.5 90 4 NaN -1.00 181326 1 500 990 0.6 0.4 0.3 3 580 ON 34.9 J 50 200 ON NO  
20120430_165724 ??? 7.6 90 6 NaN -1.00 181326 1 500 990 0.6 0.4 0.3 3 558 ON 22.3 J 50 200 ON NO - Pupils not on target
20120430_165748 ??? 7.5 90 5 NaN -1.00 181326 1 500 990 0.6 0.4 0.3 3 584 ON 27.9 J 50 200 ON NO  
20120430_165813 ??? 7.5 90 5 NaN -1.00 181326 1 500 990 0.6 0.4 0.3 3 589 ON 28.8 J 50 200 ON NO  
IDL string: ['20120430_165100','20120430_165127','20120430_165151','20120430_165251','20120430_165316','20120430_165408','20120430_165536','20120430_165601','20120430_165625','20120430_165724','20120430_165748','20120430_165813']

rimettiamo il REC 500 sinusoidale
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120430_171110 ??? 7.5 90 4 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 616 ON 24.6 J 50 200 ON NO  
20120430_171135 ??? 7.5 90 4 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 598 ON 26.3 J 50 200 ON NO  
20120430_171159 ??? 7.5 90 5 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 597 ON 27.3 J 50 200 ON NO  
20120430_171259 ??? 7.5 90 4 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 585 ON 26.6 J 50 200 ON NO  
20120430_171323 ??? 7.5 90 6 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 592 ON 25.7 J 50 200 ON NO  
20120430_171348 ??? 7.5 90 6 NaN -1.00 000500 1 500 990 0.4 0.2 0.2 3 599 ON 25.6 J 50 200 ON NO  
IDL string: ['20120430_171110','20120430_171135','20120430_171159','20120430_171259','20120430_171323','20120430_171348']

IL REC SINUSOIDALE DA SR SISTEMATICAMENTE PEGGIORI

10:18 Sin IM Amp Optimization wide range

Script per l'acquisizione dati

imsinus_ampcheck3_tt -> Misura solo TT frequ 210Hz amp = [2, 4, 6, 8, 20, 28, 40, 52, 68, 80, 100, 120 ]nm

20120430_172519 -> 20120430_173129

imsinus_ampcheck3 -> 16 modi per 16 diveri ordini radiali. amp_2-100=30, amp_101-300=20, amp_301-450=15, amp_451-629=10nm.
Queste amp vengono moltiplicate per un array di fattori [0.05, 0.1, 0.15, 0.2, 0.5, 1.0, 1.3, 1.7, 2.0, 2.5, 3.0]

Riduzione dati al volo con sinfreq_check.pro

Riprendiamo il TT per aumentare ancora l'ampiezza [2, 4, 6, 8, 20, 28, 40, 52, 68, 80, 92, 104, 116, 128, 132, 144, 160,200,240]nm,
nella prima misura, la demodulazione a 120nm aveva invertito il segno e

20120430 19:48:34 FAILURE: Tubo della master fermo

Si fa ripartire la master diagnosit e la fast diagnostic (lo specchio rippa).

Misura TT grande range: 20120430_195548 -> 20120430_200547

Il TT non cambia piu' il segno

Giudichiamo l'optimal ange osservando il SNR(C) e il P2V della IM del modo (Segnale/C) demodulata in funzione di C
Modo Optimal
amp (nm)
TT 80
20 10-20
28 10-20
37 15-20
88 15-20
117 10-20
136 15-25
256 15-20
315 15-30
346 10-20
375 7-15
409 10-15
472 7-15
581 3-7
620 3-10
   
   
   
   
   

13:10 Test occultazione Coro

Il filo del coro attraversa il centro del CCD in 565 512, quindi pass a ca 50 pix dal centro del frame del CCD47

Cloop 400 mnodi no disturb FW1 700-100, FW2 700-10

Dietro il coro no sat TT: -1.6; .1.7 -> 20120430_201329

Fuori dal filo no sat TT:0;0 -> 20120430_201453

Fuori dal filo sat TT:0;0 -> 20120430_201931

Dietro il coro sat TT: -1.6; .1.7 -> 20120430_202018

Coro flat field -> Flat 20120430_ 202803

14:10 Acquisizione IM Sinusoidale con freq e amp ottimizzate

Acquisizione 3 modi in multiplex

FREQ: 209, 241,263 (forse era meglio Numeri primi:211, 241, 271 HZ)

Ampiezze:

tt->80nm

2-346-> 20nm

347-472-> 10nm

473-629-> 7nm

20120430 21:12:28 FAILURE Message daemon non comunicano e quello del secondario non se ne accorge

Matrice IM fino al modo 627

20120430_213241-> 20120430_222914

acquisito senza interuzioni

Sembra che i modi sotto il 20 siano un po' rumorosi.

Ci sono dei modi bacati come il 21, 267, 437, 475

Da ca 520 in poi i modi sono meno convincenti

Dal 2 al 25 riacquisiamo mettendo 40nm 230514-> 230654

Riacquisiti stessa ampiezza 267; 437: 475 230906 -> 230941

REC -> 20120430_000400 400 modi

20120430_000500 500 modi

20120430_000600 600 modi

Ancora non chiude, ci erano sfufggiti 2 modi bacati: 59 e 229

Non abbiamo tempo di riprenderli.

Li uccidiamo dall'intmat e generiamo il nuovo

REC -> 20120430_000400 400 modi

ora non abbiamo piu' i 2 momdi viggliacchi negli autovaloei della SVD: tutto in un fattore 10!!

-- EnricoPinna - 30 Apr 2012
Topic revision: r3 - 11 Sep 2012, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback