Calcolate per notte 06 Febbr 2012
Rmag    alt >60 da a    Cat_name         RA        DEC        GS    PA
07.2     19:00    23:00    AO1071      05 38 12     +51 26 44    3 
07.2    21.40    02.30    AO305        08 48 01    +39 35 50    2
07.2    01:15    05:30    AO420        12 00 38     +18 20 05    1

09.8    18:15    23:00    AO1073       05 14 58     +44 31 44    11
09.1    21:00    01:30    AO327        08 04 22    +43 57 42    7
09.5    00:30    05:00    AO396        11 24 59    +40 00 12    3    0
09.1    01:30    06:30    AO470        12 29 39    +41 59 32    3

10.7    18:00    22:30    AO842        05 08 51    +27 19 25    8
10.6    20:45    01:00    AO850        07 34 39    +22 20 15    6
10.4    01:15    06:00    AO860        12 22 21     +25 10 12    1
10.7    01:15    05:30    AO981        12 03 26    +19 54 03    1 (GS debole)

12.3    18:00    22:30    AO182        04 57 10    +24 35 47    3
12.4     21:15    01:30    AO312        08 10 54    +25 01 22    3
12.5    01:50    06:00    AO472        12 34 07    +50 37 07    4 (GS a 5')

Nuvole

H20:30 Puntata AO1071

Preset via IIF non riesce a partire

Preset via AOS -> c'e' un problema di tabella, con mag 10 va a bin 2 , probabilmente bisogna rivedere la conversione nount vs mag

Seleziona il bin2 ma mancava il rec 10 modi

Il filtro del 47 e' piu' 10nm invece sul dc sono a 40nm. Quindi il 47 e' un po' cieato.

Riproviamo, la nuvola abbassa l'intensita' e rimette il bin2

UT 04:01 Preset OK, Acquire_Ref OK, Start_AO OK (augtogain lento xche ancora col flag retroreflector che divide i gain x 2)

Cloop 20120207_041002
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120207_041001 ??? 9.4 71 11 NaN 0.86 222220 2 153 801 2.2 1.8 1.1 3 525 ON 63.3 H 50 200 ONSKY NO  
20120207_041028 ??? 9.3 71 12 NaN 0.87 222220 2 153 801 2.2 1.8 1.1 3 612 ON 59.1 H 50 200 ONSKY NO  
20120207_041054 ??? 9.3 71 11 NaN 0.75 222220 2 153 801 2.2 1.8 1.1 3 590 ON 66.1 H 50 200 ONSKY NO  
20120207_041431 ??? 9.3 71 11 NaN 0.93 222220 2 153 801 2.2 1.8 1.1 3 582 ON 46.6 H 50 200 ONSKY NO  
20120207_041458 ??? 8.4 71 10 NaN 0.84 222220 2 153 801 2.2 1.8 1.1 3 1356 ON 53.5 H 50 200 ONSKY NO - long stop of 65 frames at frame 2269
20120207_041524 ??? 8.9 71 14 NaN 0.92 222220 2 153 801 2.2 1.8 1.1 3 832 ON 48.5 H 50 200 ONSKY NO  
IDL string: ['20120207_041001','20120207_041028','20120207_041054','20120207_041431','20120207_041458','20120207_041524']

Prendiamo qualche OPEN LOOP
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120207_041706 ??? 8.7 71 15 NaN 0.10 222220 2 153 801 0.0 -1.0 -1.0 3 1088 ON 0.6 H 1000 10 ONSKY NO Open Loop
20120207_041721 ??? 8.5 71 14 NaN 0.11 222220 2 153 801 0.0 -1.0 -1.0 3 1217 ON 0.7 H 1000 10 ONSKY NO Open Loop
20120207_041736 ??? 8.5 71 14 NaN 0.11 222220 2 153 801 0.0 -1.0 -1.0 3 1262 ON 0.8 H 1000 10 ONSKY NO OL: FWHM 0.5"
IDL string: ['20120207_041706','20120207_041721','20120207_041736']

Scambiamo il gain della LC (forse ieri erano stati cambiati invertiti).

Problema di autogain che manda in skip frame l'ADSEC

Facciamo autogain a manoe prendiamo un set di dati, le misure vengono interrotte da una nuvolache ci fa aprore il loop.
TrackNo RefStar MagSorted ascending El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120207_044314 ??? 8.2 69 13 NaN 0.88 181039 1 400 990 1.4 0.9 -1.0 3 307 ON 80.9 H 10 1000 ONSKY NO  
20120207_044429 ??? 8.2 69 12 NaN 0.65 181039 1 400 990 1.4 0.9 -1.0 3 312 ON 90.5 H 10 1000 ONSKY NO  
20120207_044545 ??? 8.2 69 10 NaN 0.67 181039 1 400 990 1.4 0.9 -1.0 3 313 ON 84.2 H 10 1000 ONSKY NO  
20120207_045550 ??? 8.4 68 6 NaN 0.94 181039 1 400 990 1.4 0.9 -1.0 3 256 ON 76.5 H 10 1000 ONSKY NO  
IDL string: ['20120207_044314','20120207_044429','20120207_044545','20120207_044823','20120207_045550']
  • 20120207_044429_PSF.jpg:
    20120207_044429_PSF.jpg

Proviamo 600 e 500 modi

Non riusciamo a chiudere bene con questi recs...

Test autogain

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120207_050537 ??? 8.2 67 11 NaN 0.75 180658 1 400 990 1.2 0.9 0.7 3 300 ON 74.9 H 10 1000 ONSKY NO  
20120207_050653 ??? 8.2 67 11 NaN 0.64 180658 1 400 990 1.2 0.9 0.7 3 308 ON 74.4 H 10 1000 ONSKY NO  
20120207_050809 ??? 8.2 67 14 NaN 1.03 180658 1 400 990 1.2 0.9 0.7 3 300 ON 76.3 H 10 1000 ONSKY NO  
20120207_051306 ??? 8.1 66 9 NaN 0.62 180658 1 400 990 1.2 0.9 0.7 3 347 ON 75.8 H 10 1000 ONSKY NO  
20120207_051422 ??? 8.1 66 10 NaN 0.78 180658 1 400 990 1.2 0.9 0.7 3 339 ON 80.0 H 10 1000 ONSKY NO  
20120207_051538 ??? 8.1 66 11 NaN 0.59 180658 1 400 990 1.2 0.9 0.7 3 342 ON 82.9 H 10 1000 ONSKY NO  
20120207_051821 ??? 8.1 66 12 NaN 0.11 180658 1 400 990 0.0 -1.0 -1.0 3 334 ON 1.0 H 1000 10 ONSKY NO Open Loop
IDL string: ['20120207_050537','20120207_050653','20120207_050809','20120207_051306','20120207_051422','20120207_051538','20120207_051821']

H23:00 Spento il fan della LUTH, qullo della RUTH era gia' spenta

H23:05 Puntiamo AO305 GS=2 R7.2 (velo di nubi uniforme su tutto il cielo)

Checking the pointing...

H23:10 preset active sulla stella

H23:19 Preset ACE -> OK

H23:20 RUNAO -> OK

H23:30 Loop chiuso su entrambi i rai SX e DX contemporaneamente

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120207_062910 ??? 7.9 81 12 NaN 0.56 180658 1 400 990 1.2 0.7 0.7 3 403 ON 82.7 H 5 2000 ONSKY NO  
20120207_063147 ??? 7.9 81 10 NaN 0.62 180658 1 400 990 1.2 0.7 0.7 3 397 ON 78.8 H 5 2000 ONSKY NO  
20120207_063424 ??? 7.9 81 12 NaN 0.58 180658 1 400 990 1.2 0.7 0.7 3 406 ON 77.1 H 5 2000 ONSKY NO  
IDL string: ['20120207_062910','20120207_063147','20120207_063424']

H23:50 si chiude per nuvole

.....

H04:30 Si riapre

H04:42 Preset ACE su AO470 R9.1

Sparisce la stella dal campo appena mandato il preset da IDL

UT11:48 Rimandiamo il preset con l'active (da IRAF di David) e centra la stella ma quando mandiamo la stessa stella modo ACE da IDL manda il telescopio in altre coordinate.

Dopo qualche tentativo fallito decidiamo di fare il preset Active e poi chiudere dall'AOS GUI

Preset -> OK
Acquire REF -> OK
Start AO -> OK MA l'ADC e' fuori posizionee fallisce l'autogain.

Sistemiamo l'ADC e ripartiamo con l'autogain a rampa -> OK

Abbiamo anche il CD47 veloce. Ora e' troppo veloce

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20120207_123246 ??? 9.4 66 9 NaN 0.60 180658 1 400 990 1.1 0.7 0.5 3 99 ON 81.4 H 15 700 ONSKY NO - Pupils mismatch
20120207_123342 ??? 9.4 66 9 NaN 0.53 180658 1 400 990 1.1 0.7 0.5 3 102 ON 77.9 H 15 700 ONSKY NO - Pupils mismatch
20120207_123438 ??? 9.4 66 11 NaN 0.50 180658 1 400 990 1.1 0.7 0.5 3 103 ON 79.0 H 15 700 ONSKY NO - Pupils mismatch
IDL string: ['20120207_123246','20120207_123342','20120207_123438']

H05:41 Puntiamo AO860 R10.4 (BIN2)

Preset ACE da IDL, fallito perche' la BCU39 era bloccata: le pupille nella LUT del BIN2 erano sbagliate

Rimandiamo il Preset da IIF IDL -> OK

Run AO da IIF IDL -> misura la stella R14 e quindi sbaglia...

Rimandiamo il Preset ACE ->OK

Run AO -> Fallisce ancora l'acquisizione della stella che pero' si vede bene nel CCD47

-- EnricoPinna - 07 Feb 2012

  • 20120206_plottone-sky2.jpg:
    20120206_plottone-sky2.jpg
Topic revision: r4 - 11 Sep 2012, EnricoPinna