DOME CLOSED - HUMIDITY 100%

Using ARGOS calibration unit

Hexapod position: X=-1.6 Y=-2.06 Z=4.24 RX=-674 RY=-1.3 RZ=0

11:40pm achieved closed loop using reconstructor 161348

TN20150128_064241 Frames with ARGOS calibration unit: we can see a better illuminated pupil, the only drawback is that the swing arm makes a significant shadow. Cubo off

TN20150128_064845 Come prima ma con cubo IN

rotator angle 205

applying NCPA, we have to multiply by 2 because we are in single pass... z5-z11 -> 2,60,-24,4,8,-15,-24 (ncpaLuci205degARGOS.fits)

Loop instabile sui modi alti. Riaggiustiamo l'offset del rerot (in cupola ci sono 10C) -> nuovo offset 0.8deg

Acquisizione PSF su LUCI2 No disturb, NCPA ON

TN074036 ... TN074519

Problema di DARK -> ora il ns SW spegne la ns lampada ma non quella da ARGOS
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_074036 ??? 9.0 90 10 NaN 0.05 161348 1 400 990 0.4 0.3 0.2 3 134 ON 47.1 FeII clear 172 60 ONSKY zero.fits - long stop of 57 frames at frame 1194
20150128_074102 ??? 9.0 90 10 NaN 0.05 161348 1 400 990 0.4 0.3 0.2 3 134 ON 46.5 FeII clear 172 60 ONSKY zero.fits - long stop of 55 frames at frame 1252
20150128_074519 ??? 9.0 90 10 NaN 0.05 161348 1 400 990 0.4 0.3 0.2 3 134 ON 47.8 FeII clear 172 60 ONSKY zero.fits  
IDL string: ['20150128_074036','20150128_074102','20150128_074519']

Acquisizione PSF su LUCI2 No disturb, NO NCPA
TN080141 TN080234

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_080141 ??? 9.0 90 10 NaN 0.05 161348 1 400 990 0.4 0.2 0.2 3 135 ON 33.3 FeII clear 172 60 ONSKY zero.fits - long stop of 78 frames at frame 682
20150128_080207 ??? 9.0 90 9 NaN 0.04 161348 1 400 990 0.4 0.2 0.2 3 135 ON 33.2 FeII clear 172 60 ONSKY zero.fits - long stop of 55 frames at frame 1631
20150128_080234 ??? 9.0 90 10 NaN 0.04 161348 1 400 990 0.4 0.2 0.2 3 134 ON 32.2 FeII clear 172 60 ONSKY zero.fits  
IDL string: ['20150128_080141','20150128_080207','20150128_080234']

Nuove ncpa a Autofocus Aggiunto z22 ottimizzando a mano ncpaLuci205degARGOSv2.fits Z5=+5 Z6=+55 Z7=-25 Z8=-5 Z9=+10 Z10=-15 Z11=-30 Z22=-50

Test PSF TN090158 ... TN090251

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_090158 ??? 9.0 90 8 NaN 0.04 161348 1 400 990 0.4 0.2 0.2 3 133 ON 67.6 FeII clear 172 60 ONSKY zero.fits - long stop of 84 frames at frame 1136
20150128_090225 ??? 9.1 90 9 NaN 0.04 161348 1 400 990 0.4 0.2 0.2 3 133 ON 67.8 FeII clear 172 60 ONSKY zero.fits  
20150128_090251 ??? 9.1 90 9 NaN 0.04 161348 1 400 990 0.4 0.2 0.2 3 132 ON 66.6 FeII clear 172 60 ONSKY zero.fits  
Queste NCPA sono al limite del loop. Vanno applicate poco alla volta e comunque il loop non e' molto stabile

Riproviamo con le NCPA di ieri semplicemente moltiplicate x 2 (ncpaLuci205degARGOS.fits con il fuoco appena ottimizzato

TN091729 ... TN091819
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_091729 ??? 9.0 90 10 NaN 0.05 163459 1 400 990 0.3 0.2 -1.0 4 134 ON 69.5 FeII clear 172 60 ONSKY zero.fits - long stop of 75 frames at frame 472
20150128_091754 ??? 9.0 90 10 NaN 0.05 163459 1 400 990 0.3 0.2 -1.0 4 134 ON 65.5 FeII clear 172 60 ONSKY zero.fits  
20150128_091819 ??? 9.0 90 10 NaN 0.05 163459 1 400 990 0.3 0.2 -1.0 4 134 ON 69.7 FeII clear 172 60 ONSKY zero.fits - long stop of 77 frames at frame 2669
IDL string: ['20150128_091729','20150128_091754','20150128_091819']

H02:50 PSF IN&OUT FOCUS @205deg

NO NCPA

Z=+53.0mm TN095039

Z=+54.3mm TN095225

Z=+55.6mm TN095309

Z=51.7mm TN095400

Z=50.4mm TN095457

NCPA On ncpaLuci205degARGOS.fits

Z=+53.0mm TN095825

Z=+54.3mm TN095903

Z=+55.6mm TN095942

Z=+51.7mm TN100029

Z=+50.4mm TN100109

H03:00 NCPA @ROT320deg

Partimao dalle NCPA dei 270 moltiplicate per 2 ncpaLuci270degARGOS.fits

Scan con funzione di merito a occhio

Z5=+35 Z6=-60, Z7=-10, Z8=-45, Z9=+20, Z10=-10, Z11=-20, Z12=0, Z13=-5, Z14=20 -> ncpaLuci320degARGOSv2.fits

TN101939 ... TN102033

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_101939 ??? 8.9 90 8 NaN 0.03 163459 1 400 990 0.3 0.2 0.2 4 149 ON 63.2 FeII clear 172 60 ONSKY zero.fits  
20150128_102007 ??? 8.9 90 9 NaN 0.04 163459 1 400 990 0.3 0.2 0.2 4 148 ON 62.9 FeII clear 172 60 ONSKY zero.fits  
20150128_102033 ??? 8.9 90 8 NaN 0.04 163459 1 400 990 0.3 0.2 0.2 4 148 ON 61.4 FeII clear 172 60 ONSKY zero.fits - long stop of 74 frames at frame 3726
IDL string: ['20150128_101939','20150128_102007','20150128_102033']

ncpaLuci270degARGOS.fits

TN102244 TN102337
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_101939 ??? 8.9 90 8 NaN 0.03 163459 1 400 990 0.3 0.2 0.2 4 149 ON 63.2 FeII clear 172 60 ONSKY zero.fits  
20150128_102007 ??? 8.9 90 9 NaN 0.04 163459 1 400 990 0.3 0.2 0.2 4 148 ON 62.9 FeII clear 172 60 ONSKY zero.fits  
20150128_102033 ??? 8.9 90 8 NaN 0.04 163459 1 400 990 0.3 0.2 0.2 4 148 ON 61.4 FeII clear 172 60 ONSKY zero.fits - long stop of 74 frames at frame 3726
IDL string: ['20150128_101939','20150128_102007','20150128_102033']

No NCPA

TN102509... TN102600
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_102509 ??? 8.9 90 7 NaN 0.04 163459 1 400 990 0.3 0.2 0.2 4 149 ON 40.5 FeII clear 172 60 ONSKY zero.fits - long stop of 64 frames at frame 472
20150128_102533 ??? 8.9 90 8 NaN 0.04 163459 1 400 990 0.3 0.2 0.2 4 148 ON 39.7 FeII clear 172 60 ONSKY zero.fits  
20150128_102600 ??? 8.9 90 8 NaN 0.04 163459 1 400 990 0.3 0.2 0.2 4 148 ON 39.3 FeII clear 172 60 ONSKY zero.fits  
IDL string: ['20150128_102509','20150128_102533','20150128_102600']

Autofocus Z=+52.8

Check PSF NCPA ON ncpaLuci320degARGOSv2.fits TN103902 TN103957
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_103902 ??? 8.9 90 8 NaN 0.04 163459 1 400 990 0.3 0.2 0.2 4 150 ON 61.0 FeII clear 172 60 ONSKY zero.fits - long stop of 60 frames at frame 2366
20150128_103930 ??? 8.9 90 8 NaN 0.06 163459 1 400 990 0.3 0.2 0.2 4 150 ON 64.5 FeII clear 172 60 ONSKY zero.fits - long stop of 61 frames at frame 1709
20150128_103957 ??? 8.9 90 7 NaN 0.04 163459 1 400 990 0.3 0.2 0.2 4 149 ON 63.4 FeII clear 172 60 ONSKY zero.fits  
IDL string: ['20150128_103902','20150128_103930','20150128_103957']

H03:40 NCPA @ROT100deg

No NCPA Z=52.8 TN105023 ... TN105118
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_105023 ??? 8.9 90 9 NaN 0.05 163459 1 400 990 0.3 0.2 -1.0 4 149 ON 29.7 FeII clear 172 60 ONSKY zero.fits  
20150128_105048 ??? 8.9 90 10 NaN 0.05 163459 1 400 990 0.3 0.2 -1.0 4 148 ON 29.2 FeII clear 172 60 ONSKY zero.fits  
20150128_105118 ??? 8.9 90 9 NaN 0.05 163459 1 400 990 0.3 0.2 -1.0 4 150 ON 29.5 FeII clear 172 60 ONSKY zero.fits - long stop of 72 frames at frame 507
IDL string: ['20150128_105023','20150128_105048','20150128_105118']

Scan con funzione di merito a occhio

Z5=-20 Z6=-100, Z7=20, Z8=10, Z9=10, Z10=-10, Z11=-20, Z12=0, Z13=-5, Z14=20 -> ncpaLuci100degARGOSv2.fits

Autofocus -> confermato 52.8 mm

Con NCPA on (a 4 l/D):

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_111329 ??? 8.9 90 8 NaN 0.05 163459 1 400 990 0.3 0.2 -1.0 4 148 ON 65.6 FeII clear 172 60 ONSKY zero.fits - long stop of 74 frames at frame 1619
20150128_111356 ??? 8.9 90 8 NaN 0.05 163459 1 400 990 0.3 0.2 -1.0 4 148 ON 64.5 FeII clear 172 60 ONSKY zero.fits  
20150128_111422 ??? 8.9 90 8 NaN 0.06 163459 1 400 990 0.3 0.2 -1.0 4 148 ON 65.0 FeII clear 172 60 ONSKY zero.fits - long stop of 74 frames at frame 2663
Verifichiamo che regge anche a 3l/D.

Test atmosfera a 3 l/D:

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150128_112000 ??? 8.9 90 9 NaN 0.38 161348 1 400 990 0.3 0.3 0.2 3 147 ON 62.1 FeII clear 172 60 ONSKY zero.fits  
20150128_112026 ??? 8.9 90 8 NaN 0.39 161348 1 400 990 0.3 0.3 0.2 3 147 ON 62.1 FeII clear 172 60 ONSKY zero.fits - long stop of 64 frames at frame 1074
20150128_112051 ??? 8.9 90 8 NaN 0.39 161348 1 400 990 0.3 0.3 0.2 3 147 ON 59.1 FeII clear 172 60 ONSKY zero.fits  

Summary of NCPA

NCPA vs angle with sinusoidal fitting.

NCPAvsAngle.PNG

NCPA vs Zernike

NCPAvsZern.PNG