H08:30 Retro On

H08:45 New intmat Bin1 acquisition

INTMAT -> 20150125_155941

Pup = 20130919_004544

Modal basis = KL_v16

Disturbo =20150125_153750

Bin = 1

Freq = 600HZ

Modulation = 3

n modes = 500

REC -> 20150125_161348

INTMAT= 20150125_155941

# modes = 400

cloop con rec vecchio

INTMAT -> 20150125_16258

Pup = 20130919_004544

Modal basis = KL_v16

Disturbo =20150125_153750

Bin = 1

Freq = 600HZ

Modulation = 4

n modes = 500

REC -> 20150125_163456

INTMAT= 20150125_16258

# modes = 400

cloop con rec vecchio

H09:45 Test reconstructor

Test Rec Mod 4

TN164539

Test Rec Mod 3 nuovo

TN165328

Per confronto prendiamo un TN con il vecchio REC TN165721

Rifacciamo l'autogain e ne prendiamo un altro con il vecchio REC TN

Il nuovo REC da un modalplot decisamente migliore.!!

Riprendiamo un TN con il nuovo rec TN170642

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150125_165328 ??? 7.0 90 9 NaN -1.00 161348 1 400 900 0.6 0.7 0.5 3 1000 ON -1.0 ? -1 -1 ON NO  
20150125_165721 ??? 6.9 90 6 NaN -1.00 014135 1 400 900 0.5 0.3 0.2 3 1048 ON -1.0 ? -1 -1 ON NO  
20150125_170122 ??? 6.9 90 5 NaN -1.00 014135 1 400 900 0.4 0.3 0.3 3 1036 ON -1.0 ? -1 -1 ON NO  
20150125_170642 ??? 6.9 90 8 NaN -1.00 161348 1 400 900 0.6 0.6 0.5 3 1081 ON -1.0 ? -1 -1 ON NO  


H10:10 check PSF new REC

ROT a 205deg

Subwindow LUCI > subwin sw 1 974 1050 128 128

NCPA Scan OK 2

Test con turbolenza 0.8 -> TN173240...173532

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150125_173240 ??? 8.3 90 4 NaN -1.00 161348 1 400 990 0.4 0.3 0.2 3 261 ON 66.0 FeII clear 172 20 ON NO - Background correction failed - No compatible luci dark found (i.e. same exposure time or filter or dimensions)
20150125_173407 ??? 8.3 90 5 NaN -1.00 161348 1 400 990 0.4 0.3 0.2 3 261 ON 65.4 FeII clear 172 20 ON NO - Background correction failed
20150125_173532 ??? 8.3 90 5 NaN -1.00 161348 1 400 990 0.4 0.3 0.2 3 262 ON 65.9 FeII clear 172 20 ON NO - long stop of 62 frames at frame 2469 - Background correction failed
IDL string: ['20150125_173240','20150125_173407','20150125_173532']

  • Screenshot-2.png:
    Screenshot-2.png


H11:30 Procedura autogain + macchina infernale

Step 1 -> chiusura loop con gain basso (0.1), Gopt =1 e macchina off

Step 2 -> autogain e macchinetta off

Step 3 -> Renormalize loop gain che: moltiplica i gain del loop * measured Gopt e applica il Gopt al loop

Step 4 -> Macchinetta ON

H11:55 Ridefinizione gruppi autogain

Da un plot di Fernando sui guadagni ottici della piramide sembra che i Gopt siano costanti fino a circa il modo 40 (mod = 3) e poi salgono.

Spostiamo taglio fra HO1 e HO2 dal modo 120 al modo 50.

H12:05 Test nuovo rec + procedura + taglio a 50

Nuovo REC NCPA ON rot205-> TN190940 ... TN191016

Vecchio REC NCPA On rot 205 -> TN181527 TN191602

Nuovo REC NCPA OFF -> TN193450.. TN193526

TrackNo RefStar MagSorted ascending El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150125_190940 ??? 8.3 90 5 NaN -1.00 161348 1 400 990 0.4 0.4 0.3 3 267 ON 63.2 FeII clear 172 20 ON NO  
20150125_190959 ??? 8.3 90 8 NaN -1.00 161348 1 400 990 0.4 0.4 0.3 3 267 ON 62.0 FeII clear 172 20 ON NO  
20150125_191016 ??? 8.3 90 5 NaN -1.00 161348 1 400 990 0.4 0.4 0.3 3 267 ON 63.4 FeII clear 172 20 ON NO - long stop of 71 frames at frame 1850
20150125_191527 ??? 8.3 90 3 NaN -1.00 014135 1 400 990 0.3 0.3 0.2 3 270 ON 60.1 FeII clear 172 20 ON NO  
20150125_191545 ??? 8.3 90 2 NaN -1.00 014135 1 400 990 0.3 0.3 0.2 3 270 ON 59.9 FeII clear 172 20 ON NO  
20150125_191602 ??? 8.3 90 5 NaN -1.00 014135 1 400 990 0.3 0.3 0.2 3 269 ON 60.4 FeII clear 172 20 ON NO - long stop of 66 frames at frame 1425
20150125_193450 ??? 8.3 90 5 NaN -1.00 161348 1 400 990 0.4 0.3 0.3 3 267 ON 50.0 FeII clear 172 20 ON NO  
20150125_193509 ??? 8.3 90 6 NaN -1.00 161348 1 400 990 0.4 0.3 0.3 3 267 ON 49.4 FeII clear 172 20 ON NO - long stop of 78 frames at frame 2199
20150125_193526 ??? 8.3 90 7 NaN -1.00 161348 1 400 990 0.4 0.3 0.3 3 267 ON 50.1 FeII clear 172 20 ON NO - long stop of 63 frames at frame 2514
IDL string: ['20150125_190940','20150125_190959','20150125_191016','20150125_191527','20150125_191545','20150125_191602','20150125_193450','20150125_193509','20150125_193526']


H12:40 Acquisiamo immagini per stabilire orientazione KL vs PSF

Z2=+100 TN194009

KL0 =+100 TN194243... TN194319

KL1 =+100 TN194416 ... TN194451

KL7= +50 TN194642... TN194717

H12:50 Check PSF ROT NO NCPA no TURB

ROT 180 -> TN195456

ROT 135 ->TN200135 Luci2_2_DX> subwin sw 1 1024 1102 128 128

ROT 90 -> TN200332 subwin sw 1 1024 1102 128 128

ROT 45 -> TN200551 subwin sw 1 1024 1040 128 128

ROT 0 -> TN200806 subwin sw 1 1024 1040 128 128

ROT 225 ->TN201633 subwin sw 1 940 1060 128 128

ROT 270 -> TN201918 subwin sw 1 940 1060 128 128

ROT 315 -> TN202131 subwin sw 1 1000 1000 128 128

ROT 360 -> TN202317 subwin sw 1 1000 1000 128 128

La PSF su LUCI fa un gran salto (60 pix in verticale) vicino a ROT=380del, gli stage sembrano fermi (encoder costanti)

H13:45 Scan NCPA @270deg

Continuiamo lo scan a 270 interrotto ieri

Luci2_2_DX> subwin sw 1 968 1022 128 128

No NCPA -> TN211355

NCPA di ieri ON -> TN211612

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150125_211355 ??? 8.2 90 7 NaN -1.00 161348 1 400 990 0.3 0.2 -1.0 3 295 ON 39.5 FeII clear 172 20 ON NO  
20150125_211515 ??? 8.2 90 6 NaN -1.00 161348 1 400 990 0.3 0.2 -1.0 3 292 ON 43.4 FeII clear 172 20 ON NO - long stop of 61 frames at frame 282
20150125_211612 ??? 8.2 90 7 NaN -1.00 161348 1 400 990 0.3 0.2 -1.0 3 298 ON 49.7 FeII clear 172 20 ON NO  
IDL string: ['20150125_211355','20150125_211515','20150125_211612']

Per editare lo script di NCPA -> wfsdx: ~/sources/FLAO2/scripts/luci_ncpa.py

Gui della macchinetta -> wfsdx

NCPA vect [Z5=-24, Z6=-25, Z7=-9, Z8=-20, Z9=-5, Z10=+5 , Z11=-10] -> ncpaLuci270deg.sav

NCPA ON TN221238 -> SR60% (no diturb)

NCPA ON disturb on TN221936 (abbassiamo un po' a mano i gain medi e alti) TN222025

NCPA OFF disturb on TN222110

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150125_221238 ??? 8.2 90 5 NaN -1.00 161348 1 400 990 0.4 0.2 -1.0 3 287 ON 59.7 FeII clear 172 20 ON NO - long stop of 58 frames at frame 1603
20150125_221936 ??? 8.2 90 1 NaN -1.00 161348 1 400 990 0.4 0.3 0.2 3 297 ON 58.1 FeII clear 172 20 ON NO - Warning: Selected luci dark +1h old!
20150125_222025 ??? 8.2 90 5 NaN -1.00 161348 1 400 990 0.4 0.3 0.2 3 296 ON 59.5 FeII clear 172 20 ON NO - Warning: Selected luci dark +1h old!

                                         
IDL string: ['20150125_221238','20150125_221936','20150125_222025','20150125_222110']

This topic: FLAO > WebHome > LbtTel > Flao1andFlao2DetailedLogAtTelescope > FLAO1TelLog20150125