H08:30 Retro On

Test del SW di LUCI per leggere le immagini dal SW FLAO

H09:15 Debug Zernike to KL per NCPA

H09:30 Subframe reading of LUCI2

Luci2_2_DX> subwin clear
DEBUG make_itables call
33 740 0 1
Luci2_2_DX> subwin sw 1 800 800 256 256
DEBUG make_itables call
33 740 0 1
Luci2_2_DX> subwin auto on
INFO onoff (after setting): SW=on,HW=on, HWENA=on, usedflag=xB
INFO onoff (after auto): SW=on,HW=on
DEBUG make_itables call
33 740 0 1
Luci2_2_DX> WARNING WARNING: physical unitsize limit defines new unit size 0x400000 (phsybuffers=8) 
WARNING WARNING: physical unitsize limit defines new unit size 0x200000 (phsybuffers=16) 
WARNING WARNING: physical unitsize limit defines new unit size 0x100000 (phsybuffers=32) 
INFO starting ROE .. (idle break frame timeout 0.2 s)
total ctime (measure not available) / preset-ctime=686.0487576
read: time used: 0.899 (sec)
read: terminated ok

Da qui possiamo ora semplicementre switchare subwin on and Off dalla GUI dove si acquisisce il frame di LUCI

Tempo minimo di lettura per 256X256 e' 0.34s

Il tempo di integrazione minimo scala linearmente con il lato della subwin

Definizione della Subwindow:

>subwin sw #, Xs*, Ys, Xsize, Ysize
dove:
# = un intero che definisce la subwin (possiamo definirne molte diverse)
Xs, Ys = starting point in pix della subwin (angolo sx basso del subframe*)
Xsize, Ysize = dimensione x e y della subwin in pix
NOTABENE: c'e' un'anomalia nella definizione della subwin e in realta' le Xs sono flippate. Quindi se voglio iniziare la subwin da Xs, devo mettere
Xs* = 2048 - Xs - Xsize

H10:00 Misure SR in HeI e BrG

HeI filter

Definiamo la subwin: Luci2_2_DX> subwin sw 1 986 1195 128 128

Test di acquisizione della LUCI IMA e del dark dal SW FLAO e stima dello SR(HeI) = 24% TN171449

Facciamo un autogain

Misura no NCPA e senza disturbo TN172123 (SR23%)

Inseriamo la correzione in KL di ieri z2=-10 z6=-5 z7=+10 z8=-10 z11=-20

NCPA ON no disturbo TN172402 (SR 30.5%) TN172442 TN172510 (30.6)

Facciamo autofocus con le NCPA ON -> best focus z=46.8mm

NCPA ON no disturb TN174745 (SR 36.2) TN174830

OC nel calcolo dello SR

Scopriamo che in realta' lo SR veniva calcolato considerando l'ostruzione di 0.1 e non di 0.3. Cambiamo in ao_global_data_init dove W2 era associato a OC=0.1. OC era stato cambiato nella config LUCI.

Cancelliamo tutte le psf_dl a tutte le WL e glie le facciamo ricalcolare.

Inseriamo come parametro richiesto alla psf_dl anche l'ostruzione oltre alla lunghezza d'onda, in modo che poi si possano rianalizzare dati vecchi e nuovi senza falsare i valori di SR. Ricalcolo TN174745 SR(HeI) 41%

TrackNo RefStar Mag El Wind DIMM/OL Rec binSorted ascending #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150122_172123 ??? 7.6 90 1 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 519 ON 25.0 HeI clear 3000 1 ON NO - Background correction failed
20150122_172402 ??? 7.6 90 -2147483648 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 485 ON 34.0 HeI clear 3000 1 ON NO - Background correction failed
20150122_172442 ??? 7.6 90 -2147483648 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 500 ON 34.2 HeI clear 3000 1 ON NO - Background correction failed
20150122_172510 ??? 7.6 90 -2147483648 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 506 ON 34.4 HeI clear 3000 1 ON NO  
20150122_173655 ??? 7.7 90 3 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 476 ON 37.9 HeI clear 3000 1 ON NO  
20150122_174745 ??? 7.6 90 6 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 511 ON 41.0 HeI clear 3000 1 ON NO - Background correction failed
20150122_174830 ??? 7.6 90 2 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 523 ON 38.8 HeI clear 3000 1 ON NO  
IDL string: ['20150122_172123','20150122_172402','20150122_172442','20150122_172510','20150122_173022','20150122_173655','20150122_174745','20150122_174830']

h11:00 Proviamo a migliorare le NCPA in KL manuale

z2=-10 z6=-5 z7=+10 z8=-10 z11=-20

Scan di Z2 a step di 2nm usando lo SR come funzione di merito ma non troviamo una posizione migliore. Siamo poco sensibili

h11:20 Check NCPA in Zernike modes

Reference TN183920 (Zernike tutti a 0)

Z7=20nm TN184057

Guardiamo le posizioni dello specchio, le riproiettiamo sugli Z e si vedono effettivamente 20nm di Z7 e pochi nanometri dei modi superiori.

z9=20nm TN14400 TN184537, anche qui stesso risultato positivo

z11=20nm TN184655 TN184801 anche qui stesso risultato positivo

h11:50 Test autoscan NCPA con Zernike

- Scan di prova (5 punti sul range -20 , +20) con Z5 e Z6 trova Z5=+2.8 e +14.7

- Ripetiamo come check, trova Z5=+4.6 e Z6=+17.6

- Proseguiamo con Z7 e Z8 tenendo Z5=+3.5 e Z6=+16. Trova Z7=-12 e Z8=-10

- Ripetiamo Z7 e Z8. Z7=-6.5 e Z8=-13.2 (top SR 34)

- Proseguiamo con Z9 e Z10 tenendo [Z5=+3.5, Z6=+16, Z7=-9, Z8=-11]. Trova Z9=+3.2 e Z10=-6 (top SR 37.5)

- Ripetiamo Z9 e Z10. Z9=+3 e Z10=-8

- Passiamo alla sferica Z11 tenendo [Z5=+3.5, Z6=+16, Z7=-9, Z8=-11, Z9=+3.1, Z10=-7.5]. Dobbiamo aumentare il range in negativo. Z11=-9.4 (SR 39.6)

- Ripetiamo Z11=-4.8

Autofocus con [Z5=+3.5, Z6=+16, Z7=-9, Z8=-11, Z9=+3.1, Z10=-7.5, Z11=-7.5] -> Z=46.8 come prima

Check PSF HeI

Facciamo PSF con 6s e prendiamo un dark

NCPA correction on TN205707 TN205753 TN205820, riaggiustiamo il gain TN205846

NCPA correction off TN205938 TN210040 TN210123

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150122_205707 ??? 7.4 90 1 NaN -1.00 014135 1 400 990 0.4 0.1 0.2 3 608 ON 35.3 HeI clear 6000 1 ON NO - Background correction failed
20150122_205753 ??? 7.4 90 1 NaN -1.00 014135 1 400 990 0.4 0.1 0.2 3 600 ON 34.6 HeI clear 6000 1 ON NO  
20150122_205820 ??? 7.4 90 1 NaN -1.00 014135 1 400 990 0.4 0.1 0.2 3 599 ON 37.4 HeI clear 6000 1 ON NO  
20150122_205846 ??? 7.4 90 2 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 598 ON 34.0 HeI clear 6000 1 ON NO - Background correction failed
20150122_205938 ??? 7.4 90 1 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 595 ON 25.7 HeI clear 6000 1 ON NO  
20150122_210040 ??? 7.4 90 2 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 589 ON 25.8 HeI clear 6000 1 ON NO  
20150122_210123 ??? 7.4 90 2 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 586 ON 25.3 HeI clear 6000 1 ON NO - Background correction failed
IDL string: ['20150122_205707','20150122_205753','20150122_205820','20150122_205846','20150122_205938','20150122_210040','20150122_210123']

H14:00 CHeck PSF BraG

Usiamo il solito fuoco Z=53.3mm

NCPA correction On TN211044 TN211546

NCPA correction Off TN211852

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150122_211044 ??? 7.4 90 3 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 612 ON 59.3 Br_gam clear 150000 1 ON NO - Background correction failed
20150122_211546 ??? 7.5 90 3 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 576 ON 58.4 Br_gam clear 150000 1 ON NO - Background correction failed
20150122_211852 ??? 7.5 90 2 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 559 ON 51.7 Br_gam clear 150000 1 ON NO - Background correction failed
IDL string: ['20150122_211044','20150122_211546','20150122_211852']

h14:20 Autofocus BraG con expT =60s e NCPA on

Dal plotjitter si vedono vibrazioni anche a 50Hz per un totale di ca 15mas RMS TN 211044

h15:10 test cryocooler LUCI in HeI (Tree House off)

Cryo ON TN221415

Cryo OFF TN221522 TN221554

Cryo On TN221659 TN221852

piccola differenza sulle vibraz a 20Hz

h15:20 ARGOS Chiller Test (TH off, LUCI CRyo ON)

*Argos chiller off TN221948 TN222049*

Nessuna impatto evidente

H15:30LUCI electronic OFF test

LUCI Electr on TN222920

LUCI electr Off TN222957 TN223012 (cooling line on)

LUCI electr Off and coolin Off TN223429 TN23449

Check vecchio TN

Controlliamo un vecchio TN20110604_172718 dove con 12mas di RMS facciamo il 92% SR in H


TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150122_221415 ??? 7.5 90 5 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 538 ON 35.5 HeI clear 3000 1 ON NO - Warning: Selected luci dark +5h old!
20150122_221522 ??? 7.5 90 4 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 529 ON 37.2 HeI clear 3000 1 ON NO - Warning: Selected luci dark +5h old!
20150122_221554 ??? 7.6 90 5 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 527 ON 35.5 HeI clear 3000 1 ON NO - Warning: Selected luci dark +5h old!
20150122_221659 ??? 7.6 90 8 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 528 ON 38.3 HeI clear 3000 1 ON NO - Background correction failed - Warning: Selected luci dark +5h old!
20150122_221852 ??? 7.5 90 7 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 544 ON 33.3 HeI clear 3000 1 ON NO - Background correction failed - Warning: Selected luci dark +5h old!
20150122_221948 ??? 7.5 90 7 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 549 ON 33.4 HeI clear 3000 1 ON NO - Warning: Selected luci dark +5h old!
20150122_222049 ??? 7.5 90 7 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 536 ON 30.9 HeI clear 3000 1 ON NO - Warning: Selected luci dark +5h old!
20150122_222920 ??? 7.5 90 4 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 545 ON 34.1 HeI clear 3000 1 ON NO - Warning: Selected luci dark +5h old!
20150122_222957 ??? 7.5 90 7 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 533 ON -1.0 ? -1 -1 ON NO  
20150122_223012 ??? 7.5 90 7 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 533 ON -1.0 ? -1 -1 ON NO  
20150122_223429 ??? 7.6 90 3 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 528 ON -1.0 ? -1 -1 ON NO  
20150122_223449 ??? 7.6 90 6 NaN -1.00 014135 1 400 990 0.4 0.2 -1.0 3 525 ON -1.0 ? -1 -1 ON NO  

-- EnricoPinna - 22 Jan 2015
Topic revision: r3 - 26 Jan 2015, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.