Procedure:
remember to people on the summit to:
switch on the main cooling
switch on the Neslab
check 110V is already plugged in
plug in the 208V power
our software:
set to lab-mode
stop AO arbitrator
power on and load program
stop Adsec Arbitrator
set the shell using terminal: print, fsm_set_flat(/NO_FLAT)
start Adsec Arbitrator
disable HO Offload
send to teminal. print, setpistoncurrent(20.0)
Test
LV = little volcano (order of 10mN)
act 366 (es 2011-11-18 14:50:00) (LV: to be removed from ICC)
act 500 (14:59:00) (LV: to be removed from ICC)
start slewing: ???
bunch of 2x2 acts: 8, 9, 71, 56 (LV: to be removed from ICC)
end slewing: 2011-11-18 15:19:00
it seems that astigmatism is oscillating (2011-11-18 15:25:10)
act 515 (2011-11-18 15:34:21) (LV: to be removed from ICC)
inserted in wo_icc list: 366 , 500 , 8, 9, 71, 56 , 515 (new elec.txt configuration: 20111118_173000)
start slewing: 2011-11-18 17:55:47
nothing relevant
stop slewing: 2011-11-18 17:44:56
astigmatism problem is still there
next test session on Monday
--
ArmandoRiccardi
- 18 Nov 2011
E
dit
|
A
ttach
|
P
rint version
|
H
istory
: r2
<
r1
|
B
acklinks
|
V
iew wiki text
|
Edit
w
iki text
|
M
ore topic actions
Topic revision: r2 - 18 Nov 2011,
ArmandoRiccardi
FLAO
Log In
Toolbox
Create New Topic
Index
Search
Changes
Notifications
RSS Feed
Preferences
Users
Groups
Webs
ARGOS
AdaptiveOptics
Commissioning
Engineering
FLAO
Instrumentation
LBTI
Main
MountainOperations
Operators
PartnerObserving
ProjectManagement
Safety
Sandbox
Software
SummerShutdown
System
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki?
Send feedback