conf/adsec/672b/configuration.txt: all null string except shell field. Old file saved in configuration.txt.copy
conf/adsec/672b/TS5/elec.txt: empty act_wo_curr, act_wo_pos, act_wo_icc. Old file saved in elec.txt.copy
conf/adsec/672b/processConf/fastdiagn/fastdiagn.param: all actuators enabled. Old file saved in fastdiagn.param.copy
make install-conf
restarted software SX
shell does not set: seems a software problem with act lists
conf#20111113_180000: act 32 wo_po and wo_curr to test
conf#20111113_182000: act inner ring wo_icc to test problems with empty list
conf#20111113_184000: act wo_curr acts without magnets (DSP#341,375)
la shell ora setta, sembra che fosse un problema di liste vuote, probabilmente ICC. Da controllare.
act#546 in warning because high CHDISTRMS level (38nm rms): to be increased the threshold in fasdiagn.param
elevation slew test without ICC: nothing relevant
elevation sew test with ICC enabled:
90deg: 2011-11-13 19:47:54
20deg: 2011-11-13 19:49:08
90deg: 2011-11-13 19:50:02
20deg: ??
90deg: 2011-11-13 19:52:28: #518; cluster of 2x2:295,232,233,280
20deg: 2011-11-13 19:52:30
90deg: 2011-11-13 19:57:09: #65 (50nm jump), #254
20deg: 2011-11-13 20:02:42:
90deg: 2011-11-13 20:02:45
conf#20111113_202000: reintegrato act 32 tolto per prova delle liste vuote
test setting the shell with AO gains, but without miniskirt and with shell-mounting poles
start test: 2011-11-13 21:06:08: start setting the shell
Setting successfull, but Roberto in the tent reported audible vibration. Fasdiagn reports lot of warnings of large position rms
end test: 2011-11-13 21:09:11
forced RIP
miniskirt mounting and shell-mounting poles removed
test setting the shell with AO gains
2011-11-13 22:00:23: settaggio shell (gia' iniziato)
comfermato no risonanza
drift correnti comtrollo: act#386, 500. Confermato #65 dominante
slewing test:
90:
20: 2011-11-13 22:33:36
90: 2011-11-13 22:35:32
20: 2011-11-13 22:36:08
90: 2011-11-13 22:36:20
90: 2011-11-13 22:38:35
20: 2011-11-13 22:39:31
90: 2011-11-13 22:40:25
20: 2011-11-13 22:41:00
90: 2011-11-13 22:42:01
20: 2011-11-13 22:43:32
90: 2011-11-13 22:44:19
20: 2011-11-13 22:45:05
etc, etc.
we had:
act 32 strange calibration
act#59, #66 20-30nm jump
act#341, #286, #183, #546 small drift or jump (<10nm)
act#513 10nm jump
2011-11-13 22:50:53.: inizio a scaldare (20C)
2011-11-13 23:05:20: fine misure e settato di nuovo a 10C
--
ArmandoRiccardi
- 13 Nov 2011
E
dit
|
A
ttach
|
P
rint version
|
H
istory
: r3
<
r2
<
r1
|
B
acklinks
|
V
iew wiki text
|
Edit
w
iki text
|
M
ore topic actions
Topic revision: r3 - 13 Nov 2011,
ArmandoRiccardi
FLAO
Log In
Toolbox
Create New Topic
Index
Search
Changes
Notifications
RSS Feed
Preferences
Users
Groups
Webs
ARGOS
AdaptiveOptics
Commissioning
Engineering
FLAO
Instrumentation
LBTI
Main
MountainOperations
Operators
PartnerObserving
ProjectManagement
Safety
Sandbox
Software
SummerShutdown
System
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki?
Send feedback