Kaman SMU9000-51U Eddy current sensors

  • Caution! Channel B calibration has been invalidated. (Electronics recalibrated for smaller sensor head...)

  • Calibration_S0702274-01-01.pdf: 2007 calibration data for Kaman SMU9000-51U
    • Note: There is a difference in performance depending on which cables are used--(2-m+5-m versus 2-m only). Using the longer cables puts the offset at about 1-mm, consistent with the calibration data. Using the shorter cables puts the offset at about 1.5-cm.
    • Use the short and long cables together.

-- JorgeEsguerra - 08 Jan 2010

SMU9000 20N calibration data esed in OVMS accelerometer testing (Feb-March 2010)

  • LBT Kaman SMU9000 Ch B calibration with 20N sensor head:
    SMU9000_20N_calibration.jpg

Raw Calibration Data (ASCII, Matlab-ready)

-- JorgeEsguerra - 11 Mar 2010
I Attachment Action Size Date Who Comment
SMU9000_20N_calibration.jpgjpg SMU9000_20N_calibration.jpg manage 143 K 11 Mar 2010 - 18:52 UnknownUser LBT Kaman SMU9000 Ch B calibration with 20N sensor head
cal_1.txttxt cal_1.txt manage 557 bytes 11 Mar 2010 - 19:12 UnknownUser Calibration data (set 1)
cal_2.txttxt cal_2.txt manage 505 bytes 11 Mar 2010 - 19:12 UnknownUser Calibration data (set 2)
cal_3.txttxt cal_3.txt manage 2 K 11 Mar 2010 - 19:12 UnknownUser Calibration data (set 3)