Calibration of Z4: FPIA vs EF results

On 20091014, Marco obtained a set of extra and intra focal images for which various amounts of Z4 were introduced: Z4 = -2000, -1500, -500, 0, 500, 1000, 1500, 2000. For each, dofpia was run to focus and collimate, then some Z4 was manually introduced, and then extra and intra focal images were obtained (chip 2, 1000 rows), and finally this Z4 was removed and dofpia was again run to focus and collimate before the next value of Z4.

Results for FPIA and EF will be compared.

Unfortunately the OB for obtaining the intra-focal pupil image did not move the DX mirror to the intra-focal position, +0.8mm up, and instead was simply at the -0.8mm extra-focal position. So we do not have red extra/intra focal pairs, only blue ones.

FPIA results:

Z4=-2000nm bckgnd=1269.3078 /d2/kuhn/LBCdata/20091014//lbcb.20091014.031708.fits[1] lbcb.20091014.032519.fits[1]
0 1215.5000 3886.5000 1821.49 -101.448 -113.507 -9.26767 -137.262 -165.474 7.83227 333.588 -14.4173
1 1587.5000 3881.5000 1843.87 -159.405 -204.632 57.1280 -94.6821 -288.918 -15.1849 328.343 12.2698
2 189.50000 3390.5000 1768.99 -192.874 101.428 27.9258 -10.9736 -403.645 8.76519 366.515 -88.1861
3 1546.5000 4010.5000 1914.99 -108.246 -259.665 -90.4539 -56.7807 -293.648 1.43525 296.789 3.95015
4 207.50000 4148.5000 1644.69 56.1934 -87.9452 -95.6134 -169.035 -246.180 16.6417 256.182 -150.541
5 1274.5000 4289.5000 1839.98 -55.5136 -201.165 -83.8488 -84.2234 8.12216 21.9575 239.588 -63.7480
6 409.50000 3890.5000 1801.17 -46.8705 92.9568 95.4303 -51.9589 -324.570 -0.367298 296.620 -111.532
7 1121.5000 4503.5000 1849.66 -27.3395 -166.885 -415.423 -24.1067 -106.619 37.6135 266.071 -289.665
8 1793.5000 3617.5000 2000.73 -205.083 -112.945 -22.5521 84.3022 -251.087 18.2660 317.483 1.38518
9 1026.5000 4478.5000 1838.34 -7.65227 -211.699 -323.220 -166.423 -126.132 40.2127 228.103 -95.9858
10 1566.5000 3312.5000 1769.67 -445.290 -57.2226 223.177 -167.190 -412.200 21.2248 754.827 22.8983
11 738.50000 4165.5000 1793.25 -155.575 -101.715 20.2201 128.197 -547.969 14.5869 252.456 -110.013
12 882.50000 4418.5000 1810.61 -16.7326 33.7300 -253.699 26.8895 -476.469 19.6401 314.793 -22.6088
13 744.50000 4362.5000 1838.96 266.963 140.364 -1271.69 2010.20 1082.21 6.24081 404.741 -193.911
14 1124.5000 3364.5000 1728.24 -336.039 113.665 -422.997 250.929 -18.3036 -13.3523 395.110 -38.7830
15 277.50000 3553.5000 1700.99 -17.8868 164.498 31.8962 -187.335 -427.654 16.0953 227.793 10.2784
16 1362.5000 4255.5000 1665.55 -91.8523 -552.262 -305.523 -121.026 -54.4377 -10.2132 270.873 -70.5184
                       
-1     1810.61 -91.8523 -101.715 -83.8488 -56.7807 -251.087 14.5869 296.789 -63.7480
-2     1801.83 -96.7442 -83.7060 -166.972 72.3245 -179.587 11.2585 326.463 -70.5369
-3     1810.61 -130.181 -167.117 -86.9512 -70.7391 -272.441 14.5869    
-4     2328.25 -130.181 -167.117 -86.9512 -70.7391 -272.441 14.5869    

Z4=-1500nm bckgnd=1272.3369 /d2/kuhn/LBCdata/20091014//lbcb.20091014.033023.fits[1] lbcb.20091014.033150.fits[1]
0 1675.5000 3916.5000 1001.71 -113.046 309.195 -89.0929 21.7587 -34.5992 -16.5377 353.089 116.426
1 123.50000 4445.5000 678.158 297.989 323.675 -167.910 -185.079 -159.163 -15.7522 243.762 -98.0502
2 276.50000 3426.5000 867.952 120.741 551.684 -0.489319 46.1923 -200.909 11.3532 377.314 -30.2442
3 1633.5000 4045.5000 1005.63 -85.8416 180.467 149.204 -128.401 123.800 0.164673 306.477 82.6791
4 294.50000 4184.5000 714.748 285.280 338.292 -10.5754 6.66707 -163.578 -6.43418 256.858 -72.8648
5 1362.5000 4324.5000 896.012 50.3983 259.130 23.6921 -145.520 -13.0102 -1.66105 243.782 61.1851
6 495.50000 3926.5000 838.158 161.389 435.533 -93.8716 -69.7829 -231.219 -14.4406 282.875 -35.7100
7 1209.5000 4539.5000 985.093 -21.8509 131.907 -274.199 42.1025 -119.370 16.2577 182.056 -192.079
8 1114.5000 4512.5000 975.431 187.642 63.8167 142.646 -38.1981 -119.962 -2.00927 227.008 31.2024
9 1652.5000 3348.5000 960.188 155.172 411.799 218.420 -70.0297 -122.029 35.2702 705.186 -12.7033
10 825.50000 4201.5000 871.950 243.936 305.694 -132.630 -126.743 35.4032 13.9973 262.656 -56.4110
11 969.50000 4454.5000 782.291 18.0766 236.578 -146.427 -77.3200 -343.904 18.1149 262.484 31.5373
12 1211.5000 3400.5000 683.840 -175.194 507.775 -164.931 163.320 -255.422 -15.6759 395.670 17.5354
13 831.50000 4398.5000 781.885 93.4387 325.809 -311.897 -64.5548 -232.337 -10.1779 228.610 -45.1364
14 364.50000 3589.5000 924.098 200.699 467.309 -354.663 39.9308 -298.410 0.514448 292.268 84.3072
15 1440.5000 3741.5000 733.810 119.465 106.991 -9.32141 11.7929 -472.705 -26.7840 334.205 68.6827
16 2014.5000 3941.5000 1074.42 -160.358 365.114 -98.1027 -64.9548 -197.941 -1.37246 240.584 90.8849
17 1421.5000 4060.5000 -3994.74 -219.426 606.132 -1970.96 1066.61 -838.150 215.913 166.145 45.2606
                       
-1     871.950 119.465 325.809 -93.8716 -38.1981 -163.578 -1.37246 262.656 31.2024
-2     598.924 64.3617 329.272 -182.839 23.7661 -202.417 11.1522 297.835 4.80567
-3     871.950 270.096 293.378 -96.9896 -30.4643 -145.792 -1.66105    
-4     1148.95 270.096 293.378 -96.9896 -30.4643 -145.792 -1.66105    

Z4=-1000nm bckgnd=1282.0769 /d2/kuhn/LBCdata/20091014//lbcb.20091014.033834.fits[1] lbcb.20091014.034001.fits[1]
0 161.50000 4475.5000 496.437 56.7316 351.068 -383.894 -123.609 154.075 38.4260 317.103 -112.230
1 1712.5000 3945.5000 797.881 -248.866 221.645 -9.99062 114.740 287.512 39.9339 394.066 44.2420
2 313.50000 3455.5000 693.952 -11.2556 473.970 137.904 99.3200 -38.9422 21.0485 457.770 -79.2938
3 124.50000 3653.5000 605.431 5.11094 577.039 120.351 0.309464 62.0740 29.9770 394.001 -122.402
4 1672.5000 4074.5000 738.179 -241.059 264.821 -165.272 22.5582 13.4683 6.81870 325.157 66.8922
5 332.50000 4214.5000 497.508 -38.3187 303.268 -255.412 -99.3283 16.2805 21.9656 262.412 -90.1395
6 1400.5000 4354.5000 786.077 -193.724 252.461 -32.6077 -7.08238 12.5712 14.8725 235.579 27.5490
7 1247.5000 4568.5000 746.094 -93.1486 166.495 -233.905 210.660 398.163 7.38041 186.066 -185.210
8 533.50000 3955.5000 604.518 75.0041 324.137 -65.0934 -6.58354 215.651 27.1518 331.137 -63.7989
9 1918.5000 3680.5000 1064.15 98.9065 1657.61 164.898 -224.686 55.3638 27.7671 398.738 28.3415
10 863.50000 4231.5000 627.064 -77.6461 314.809 -121.222 -12.9928 350.670 34.5433 226.474 -42.8091
11 1151.5000 4541.5000 682.075 -218.760 -13.7109 -182.558 54.4124 127.081 38.4491 248.564 -28.2088
12 1690.5000 3377.5000 736.339 -156.090 358.599 69.3709 -114.677 191.456 85.7796 717.613 -107.668
13 1248.5000 3429.5000 560.705 -38.1912 510.724 105.808 64.3331 -86.8269 15.3369 431.515 -5.71511
14 1008.5000 4483.5000 518.493 -143.386 76.3642 -141.896 -63.8992 220.712 27.0262 322.649 13.9090
15 869.50000 4427.5000 522.487 -66.2392 226.532 1.32235 62.6739 -43.3668 33.8640 276.800 -45.3732
16 402.50000 3618.5000 672.996 45.2922 267.202 -262.704 -13.4632 -440.221 62.8859 382.245 63.3340
17 1487.5000 4320.5000 742.141 -346.931 113.505 -218.455 180.472 192.846 35.0540 241.605 39.6509
18 1478.5000 3770.5000 722.202 -239.327 607.239 183.618 -27.2087 -232.199 18.2117 368.833 25.3719
19 129.50000 4311.5000 573.387 -86.8679 194.302 -83.5819 17.2433 -214.902 19.4332 260.383 -38.5820
                       
-1     682.075 -77.6461 303.268 -65.0934 0.309464 62.0740 27.7671 325.157 -28.2088
-2     669.406 -95.9382 362.404 -68.6660 6.65957 62.0732 30.2963 338.936 -30.6070
-3     682.075 -93.5298 319.864 -83.5819 17.2433 35.8204 27.1518    
-4     614.016 -93.5298 319.864 -83.5819 17.2433 35.8204 27.1518    

Z4=-500nm bckgnd=1289.0961 /d2/kuhn/LBCdata/20091014//lbcb.20091014.034514.fits[1] lbcb.20091014.034639.fits[1]
0 1362.5000 3969.5000 320.314 -3.26266 -3.98298 90.8224 34.4245 -106.288 38.0025 327.660 59.5950
1 1734.5000 3963.5000 399.681 -24.4202 -33.5650 110.298 75.0311 -24.0863 49.2408 330.814 110.471
2 183.50000 4495.5000 184.628 292.429 -75.8485 -309.749 22.3680 -153.187 20.6730 169.387 -25.3545
3 335.50000 3474.5000 286.741 148.551 141.750 20.7991 40.1688 -49.8479 56.5659 353.038 -6.39492
4 146.50000 3672.5000 274.921 269.784 152.789 -38.2327 -33.6422 -126.695 63.9535 289.228 -40.4368
5 1694.5000 4093.5000 464.006 -139.913 53.2300 -125.719 165.027 -20.7153 84.2209 317.890 131.212
6 353.50000 4232.5000 198.792 255.100 -107.207 -175.837 -118.308 -16.2311 61.0624 272.940 -39.3371
7 1421.5000 4372.5000 300.970 -103.777 -89.8092 -193.463 143.052 258.724 47.4574 228.911 112.067
8 554.50000 3974.5000 307.532 284.746 -47.0087 -116.119 117.144 -409.614 61.0984 274.332 -4.37535
9 1939.5000 3698.5000 355.363 -164.080 150.400 164.987 73.7538 -161.501 28.9430 376.382 70.7882
10 1173.5000 4560.5000 370.241 -94.0360 -237.344 -206.355 118.333 222.211 41.3375 217.861 79.7590
11 884.50000 4249.5000 346.213 64.7644 -149.797 -242.294 268.852 -369.907 66.1946 237.321 22.3122
12 1711.5000 3395.5000 261.333 -194.682 280.016 186.131 -23.7649 -620.851 90.0169 644.656 -113.445
13 1029.5000 4502.5000 395.558 -40.3099 -57.4677 -298.932 50.9529 -112.172 45.6059 283.261 71.6395
14 1269.5000 3447.5000 141.919 142.354 258.150 196.772 -123.291 -27.0033 25.5972 447.885 71.7570
15 891.50000 4446.5000 302.692 21.2332 -89.7856 -235.935 139.021 -162.587 52.0487 220.885 37.9716
16 423.50000 3637.5000 230.525 165.523 147.303 -240.245 278.967 -571.449 44.3315 344.660 118.869
17 1499.5000 3789.5000 250.796 -401.721 126.592 13.0618 -2.69914 -389.267 32.4342 302.843 78.7410
18 1509.5000 4338.5000 295.540 -430.530 -308.510 -39.6580 228.139 -134.944 56.9100 283.572 94.8295
                       
-1     300.970 -3.26266 -33.5650 -116.119 73.7538 -126.695 49.2408 289.228 70.7882
-2     299.356 2.51338 5.78442 -75.7719 76.5016 -156.601 50.8260 311.764 43.7194
-3     300.970 -127.818 -182.223 -105.887 96.7149 -123.714 49.2408    
-4     536.027 -127.818 -182.223 -105.887 96.7149 -123.714 49.2408    

Z4=0 bckgnd=1313.7468 /d2/kuhn/LBCdata/20091014//lbcb.20091014.035037.fits[1] lbcb.20091014.035202.fits[1]
0 208.50000 4520.5000 -383.550 292.006 187.918 -214.831 -17.2377 79.4919 13.0385 112.575 -18.8687
1 1759.5000 3988.5000 7.78949 -76.0805 360.477 26.9900 59.4233 11.9652 26.0889 252.295 126.218
2 359.50000 3499.5000 -208.445 317.575 396.023 141.205 43.5079 -133.268 42.5764 290.128 -27.2796
3 171.50000 3698.5000 -262.125 355.346 301.479 -24.6567 13.9909 -107.763 69.8200 239.768 -32.5035
4 377.50000 4257.5000 -287.285 325.939 78.6037 -243.119 -6.10049 34.7076 28.7747 153.794 -3.48519
5 1718.5000 4117.5000 -84.4703 -13.4005 390.552 31.8021 71.1949 -13.8219 52.5344 279.401 147.067
6 1446.5000 4397.5000 -170.576 32.6287 264.101 -195.676 154.333 27.1811 58.2721 176.257 131.674
7 1976.5000 4499.5000 -3.23815 53.3080 201.873 -322.804 241.891 121.553 37.3809 144.591 175.741
8 579.50000 3999.5000 -276.039 263.219 298.387 -104.634 -54.2537 21.2338 48.3111 224.547 11.3817
9 863.50000 3247.5000 -263.890 303.287 640.164 97.3240 71.1070 -62.3654 46.1459 360.974 -16.3335
10 1963.5000 3723.5000 -87.2650 -78.6426 386.216 111.532 256.860 -171.094 56.6224 265.293 74.5247
11 909.50000 4274.5000 -257.292 187.824 53.2128 30.7407 177.398 -345.662 34.0016 161.518 18.4477
12 1735.5000 3419.5000 -166.262 82.0633 321.588 228.660 -109.869 -154.260 86.5388 585.734 -164.744
13 1053.5000 4527.5000 -149.932 42.0668 126.702 -139.288 -81.4678 -130.194 74.0858 163.409 74.4141
14 916.50000 4471.5000 -356.508 6.42678 157.363 -430.145 200.106 -129.599 28.4245 148.262 70.8684
15 1293.5000 3472.5000 -301.247 -21.1139 147.955 -159.137 45.7899 -605.607 33.3728 354.663 26.6145
16 448.50000 3662.5000 -355.703 -76.9535 255.776 -371.881 217.009 -396.619 5.42811 280.637 95.5101
                       
-1     -257.292 53.3080 264.101 -104.634 59.4233 -107.763 42.5764 239.768 26.6145
-2     -212.120 117.382 268.729 -90.4659 75.5107 -114.948 43.6128 246.697 40.5439
-3     -257.292 -11.4974 335.086 -47.8950 77.7558 -107.199 42.5764    
-4     -53.6140 -11.4974 335.086 -47.8950 77.7558 -107.199 42.5764    

Z4=+500nm bckgnd=1302.3384 /d2/kuhn/LBCdata/20091014//lbcb.20091014.035502.fits[1] lbcb.20091014.035628.fits[1]
0 1774.5000 4006.5000 -485.331 -173.277 139.193 -102.146 99.7430 78.5301 12.3210 271.746 76.0568
1 374.50000 3518.5000 -662.429 258.055 230.911 -6.90210 -119.791 -143.630 19.3267 269.455 -60.1763
2 186.50000 3716.5000 -710.589 236.383 112.228 67.6988 -76.7085 -118.438 18.3681 204.296 -103.509
3 393.50000 4276.5000 -767.255 140.321 -39.3327 -182.658 -179.557 -11.9988 65.2996 194.440 -69.6361
4 1734.5000 4135.5000 -541.485 -176.450 99.0117 -133.385 -9.56135 4.50875 32.8530 274.194 105.328
5 1993.5000 4517.5000 -517.480 -51.4865 -33.7914 -302.161 95.1384 47.2518 16.9644 240.498 144.771
6 1462.5000 4415.5000 -616.310 -79.6397 118.798 -116.664 -8.47877 2.67411 30.7728 196.446 71.3364
7 595.50000 4017.5000 -674.603 293.789 7.96128 -92.8684 -236.112 -152.752 21.6233 202.839 -17.4959
8 878.50000 3265.5000 -787.116 95.6193 559.991 39.7782 -133.236 -60.9353 64.7969 347.746 -105.733
9 1979.5000 3740.5000 -548.677 -259.149 279.802 86.4497 130.762 -150.776 34.7960 269.286 15.9317
10 925.50000 4292.5000 -765.242 65.1611 -1.23337 -228.186 28.5878 63.0780 47.4971 183.162 -39.2553
11 1751.5000 3438.5000 -564.748 -99.7511 258.614 108.161 -86.4982 -279.050 127.938 533.241 -205.031
12 1070.5000 4544.5000 -726.920 -5.72307 13.2632 -235.596 164.266 -378.809 33.3216 209.731 48.5031
13 932.50000 4489.5000 -682.003 29.4057 92.2864 -626.631 304.805 -340.481 41.4008 172.707 1.71031
14 1309.5000 3490.5000 -689.976 190.399 141.549 124.619 -126.136 -184.533 32.0196 354.788 -22.5276
15 463.50000 3680.5000 -674.731 410.862 -69.3310 -198.977 165.674 -437.030 29.8351 328.669 -0.469499
16 1549.5000 4381.5000 -674.987 -10.0683 312.103 -127.348 -47.2331 134.103 46.6391 202.163 11.7424
17 192.50000 4373.5000 -741.276 127.219 120.127 -8.14372 -47.9556 -238.526 17.3343 152.181 -34.5823
18 1540.5000 3831.5000 -471.385 -114.565 233.927 129.707 -230.691 -370.336 98.0846 305.348 67.0052
                       
-1     -674.731 29.4057 118.798 -102.146 -47.2331 -143.630 32.8530 240.498 -0.469499
-2     -647.502 46.1634 135.583 -95.0133 -16.4727 -133.534 41.6417 258.576 -6.10693
-3     -674.731 -132.109 197.066 -96.7134 -12.3628 -145.115 32.8530    
-4     -399.012 -132.109 197.066 -96.7134 -12.3628 -145.115 32.8530    

Z4=+1000nm bckgnd=1319.1395 /d2/kuhn/LBCdata/20091014//lbcb.20091014.040113.fits[1] lbcb.20091014.040253.fits[1]
0 242.50000 4562.5000 -1525.49 117.006 -10.4682 -379.871 -42.4549 254.333 16.5407 199.605 -47.4501
1 1793.5000 4028.5000 -1122.76 -206.502 195.410 120.295 -30.4273 196.844 18.9889 398.153 62.7523
2 393.50000 3541.5000 -1281.12 271.040 126.521 210.128 35.2127 -79.5479 25.6414 339.441 -96.8359
3 204.50000 3739.5000 -1385.25 261.393 39.8960 50.5087 -110.339 191.164 10.3756 299.970 -95.0072
4 1914.5000 3257.5000 -1464.75 156.255 44.4590 -3.18274 -102.219 -329.410 4.53054 435.840 -64.7186
5 412.50000 4299.5000 -1459.11 128.625 -48.6872 31.4561 -104.612 63.3003 22.8224 220.253 -26.9051
6 1481.5000 4438.5000 -1279.07 -47.6160 67.0450 -99.5640 140.615 24.1516 21.1210 263.015 94.9173
7 2012.5000 4540.5000 -1258.20 5.99957 148.494 -196.118 211.606 109.423 12.5888 284.471 142.645
8 613.50000 4040.5000 -1351.96 176.787 -11.3615 -18.6946 -172.248 54.5815 5.99343 276.014 -36.0519
9 897.50000 3288.5000 -1449.86 121.792 430.331 10.5690 39.7956 -61.2221 1.64103 394.168 -54.1652
10 1998.5000 3763.5000 -1147.60 -127.636 330.906 85.8420 82.5025 65.6446 35.1104 338.232 39.4684
11 944.50000 4315.5000 -1438.27 -97.5058 -11.3582 -185.162 112.067 64.2936 12.9209 204.100 14.0907
12 1770.5000 3460.5000 -1243.94 8.46158 440.631 127.472 -90.1756 -315.032 70.7885 591.658 -198.098
13 1088.5000 4568.5000 -1445.36 -154.580 -184.808 44.8464 -8.37899 -16.4989 24.6012 272.651 15.7995
14 950.50000 4512.5000 -1378.22 82.9501 -178.610 -213.126 102.969 -96.3025 20.8152 192.606 18.2405
15 1327.5000 3512.5000 -1386.72 124.983 182.649 101.546 10.8976 32.5670 -13.1686 431.538 -58.4893
16 481.50000 3703.5000 -1376.45 177.587 76.6037 223.864 55.3426 -64.6140 1.69966 393.102 -37.5393
17 1568.5000 4403.5000 -1185.85 -370.438 -21.0559 99.9545 41.8373 181.153 14.9928 289.505 48.9012
18 1558.5000 3853.5000 -1320.38 18.5777 -58.9362 312.944 134.098 -682.200 8.39243 330.738 36.8877
                       
-1     -1376.45 82.9501 44.4590 44.8464 35.2127 32.5670 14.9928 299.970 -26.9051
-2     -1342.12 34.0621 81.9821 17.0372 16.1099 -21.4406 16.6524 323.951 -12.7136
-3     -1376.45 -126.349 -151.057 73.1379 -0.693579 33.7592 14.9928    
-4     -1440.59 -126.349 -151.057 73.1379 -0.693579 33.7592 14.9928    

Z4=+1500nm bckgnd=1447.7092 /d2/kuhn/LBCdata/20091014//lbcb.20091014.040824.fits[1] lbcb.20091014.040950.fits[1]
0 1439.5000 4059.5000 -1924.12 258.355 -145.277 -59.8795 126.092 -29.0487 17.9096 222.018 58.9782
1 1809.5000 4053.5000 -1828.40 110.161 -165.778 -83.8537 38.7663 5.34111 7.79242 231.538 53.2242
2 410.50000 3565.5000 -1900.82 380.974 -308.911 -131.380 -13.8007 -47.9357 42.8665 205.911 -62.5646
3 221.50000 3763.5000 -1974.08 368.808 -304.996 -49.3897 -109.832 17.1750 32.1330 159.171 -56.0184
4 1931.5000 3280.5000 -2029.43 575.093 -25.2110 -80.7997 -139.547 -273.893 2.64539 329.723 -76.3659
5 429.50000 4322.5000 -2075.75 271.790 -406.858 -312.119 47.4267 -7.30151 33.2117 101.684 -17.0811
6 1769.5000 4180.5000 -1929.55 171.169 -77.0500 -5.32261 47.1171 -40.6988 21.1379 275.958 113.311
7 1498.5000 4460.5000 -1899.64 108.328 -105.875 -326.521 361.912 -169.669 -16.8882 132.576 138.696
8 2029.5000 4562.5000 -1976.33 237.906 -197.158 -190.839 187.166 -156.943 -17.5159 195.965 157.633
9 631.50000 4063.5000 -2033.82 426.644 -286.519 23.0445 -77.8512 -19.7040 38.0189 154.914 -22.8185
10 914.50000 3311.5000 -1975.79 223.233 -145.375 84.6892 -92.4689 -202.847 19.4451 280.592 -91.9353
11 2015.5000 3786.5000 -1799.09 70.7175 -59.6690 -140.748 170.300 -361.742 28.4836 255.533 43.3084
12 961.50000 4338.5000 -2122.49 310.368 -341.910 -139.762 173.892 237.889 46.8310 127.896 -32.2904
13 1057.5000 3257.5000 -2223.60 194.694 -160.397 291.317 271.079 -783.663 -17.8509 319.722 -71.3306
14 968.50000 4534.5000 -2057.93 99.6721 -221.098 -175.995 104.935 -530.082 38.2102 137.923 58.9027
                       
-1     -1975.79 237.906 -165.778 -83.8537 47.4267 -47.9357 21.1379 205.911 -17.0811
-2     -1983.39 253.861 -196.805 -86.5039 73.0125 -157.541 18.4287 208.742 12.9099
-3     -1975.79 228.413 -188.892 -88.5378 42.3407 -57.2270 21.1379    
-4     -1867.06 228.413 -188.892 -88.5378 42.3407 -57.2270 21.1379    

Z4=+2000nm bckgnd=1339.2657 /d2/kuhn/LBCdata/20091014//lbcb.20091014.041514.fits[1] lbcb.20091014.041653.fits[1]
0 817.50000 3248.5000 -2671.82 432.635 254.841 -332.557 151.775 -539.973 -8.31888 318.761 -138.802
1 1823.5000 4072.5000 -2389.58 105.418 245.784 -143.945 504.565 -200.957 -52.6287 248.197 40.4079
2 422.50000 3585.5000 -2532.37 499.113 57.4386 -420.617 75.6827 -520.130 -13.4959 242.716 -169.814
3 1773.5000 3249.5000 -2626.18 550.087 236.131 -99.6446 151.754 -757.921 27.3279 400.529 -147.874
4 234.50000 3783.5000 -2616.61 375.323 64.5385 -318.701 -32.2044 -520.558 -19.8081 201.085 -163.846
5 1944.5000 3300.5000 -2619.77 513.793 211.961 234.726 -79.4403 -720.768 12.7972 380.211 -159.118
6 442.50000 4342.5000 -2664.61 353.452 -82.5072 -443.154 339.709 -397.267 -39.5645 125.272 -80.3748
7 1510.5000 4480.5000 -2587.84 116.102 167.393 201.636 414.549 -645.182 -38.8575 185.811 26.0364
8 643.50000 4083.5000 -2583.46 316.794 32.6860 -521.637 248.840 -399.318 -59.9019 199.119 -57.0393
9 2027.5000 3806.5000 -2441.27 92.5050 486.542 -115.285 323.901 -570.131 -7.22495 262.121 -36.9283
10 926.50000 3330.5000 -2675.11 433.184 308.539 -221.164 246.228 -81.5002 -99.2088 350.428 -146.609
11 974.50000 4358.5000 -2571.86 132.772 117.863 -648.197 325.287 -681.077 -3.11482 156.357 7.41318
12 1069.5000 3277.5000 -2791.56 689.266 306.740 3.22109 9.44827 -794.305 -74.4504 362.579 -164.820
13 1799.5000 3503.5000 -2464.74 257.193 494.669 64.4934 229.758 -468.388 84.3273 441.523 -256.019
14 981.50000 4555.5000 -2645.32 204.816 85.1424 104.589 454.503 -172.260 -91.9045 157.526 27.6540
15 1357.5000 3555.5000 -2724.56 331.340 131.302 319.996 93.4401 -931.779 -25.0401 334.122 -67.4299
16 510.50000 3746.5000 -2782.26 504.248 -140.178 -524.498 -2.68404 -534.104 -110.038 301.764 -153.558
17 1598.5000 4447.5000 -2581.13 340.475 241.266 -143.097 233.184 -312.061 -100.936 203.684 31.8182
18 1587.5000 3896.5000 -2573.26 -57.2784 -15.5661 -553.748 233.015 -434.413 -73.5001 240.257 -51.5200
                       
-1     -2616.61 340.475 167.393 -143.945 233.015 -520.558 -38.8575 248.197 -80.3748
-2     -2607.54 325.855 168.662 -187.241 206.385 -509.584 -36.5021 269.056 -87.3906
-3     -2616.61 312.070 277.029 -200.312 227.782 -520.558 -38.8575    
-4     -1627.55 312.070 277.029 -200.312 227.782 -520.558 -38.8575    

Z4 = -2000 nm Int D: 12.1117; Ext D: 67.7525 lbcb.20091014.031708.fits 1224 -130 -167 -86 -70 308 14 0.87 0.53

Z4 = -1500 nm IntD: 11.0489; ExtD: 64.9777

lbcb.20091014.033023.fits 45 270 293 -96 -30 435 -1 0.91 0.31

Z4 = -1000 nm IntD: 10.1477; ExtD: 64.4164

lbcb.20091014.033834.fits -489 -93 319 -83 17 616 27 1.04 0.45

Z4 = -500 nm IntD: 10.6656; ExtD: 63.2897

lbcb.20091014.034514.fits -567 -127 -182 -105 96 457 49 1.06 0.38

Z4 = +500 nm IntD: 10.2686; ExtD: 60.4054

lbcb.20091014.035502.fits -1502 -132 197 -96 -12 435 32 1.05 0.66

Z4 = +1000 nm IntD: 9.19733; ExtD: 58.3309

lbcb.20091014.040113.fits -2544 -126 -151 73 0 614 14 0.96 1.08

Z4 = +1500 nm IntD: 9.27238; ExtD: 56.5592

lbcb.20091014.040824.fits -2970 228 -188 -88 42 523 21 0.93 1.22

Z4 = +2000 nm IntD: 10.7062; ExtD: 54.6648

lbcb.20091014.041514.fits -2731 312 277 -200 227 60 -38 0.86 1.09

-- OlgaKuhn - 14 Oct 2009

  • z4plot.eps: Plot of z4(FPIA), inner and outer diameters x10 against the Z4 aberration input to the SX primary
I Attachment Action Size Date Who Comment
z4plot.epseps z4plot.eps manage 10 K 16 Oct 2009 - 17:55 OlgaKuhn Plot of z4(FPIA), inner and outer diameters x10 against the Z4 aberration input to the SX primary
Topic revision: r4 - 06 Jan 2010, OlgaKuhn