E/D Night - 20141230 UT
Observers: B. Rothberg, D. Miller, (LBTO, Remote from Tucson)
Telescope Operator: D. Huerta (LBTO)
AO Support: D. Miller (LBTO)
Telescope Support: John Little (LBTO)
Software Support: T. Golata
Instrument Support: J. Morris (LBTO)
Summary
Details
MODS2 AGw Commissioning
01:15 David set IE and CA, preset to ACT0325. Collimated in ~ 5 iterations
01:30 Note that the pupil offset rotation is not being applies. Should be 20 degrees for MODS2
02:30 Determined that the calculation of degrees to radians in GCS is incorrect. Currently the line in
WFSingThread.cpp is:
double pupil_rot_offset_rad = _pupil_rotoffset * 180.0 / M_PIl; // 5401 WFS_pupil_rotoffset
It should be:
double pupil_rot_offset_rad = _pupil_rotoffset * M_PIl / 180.0; // 5401 WFS_pupil_rotoffset
To correct for this I put, in MODS_R.cfg, for WFS_pupil_rotoffset = -20.2*Pi*Pi/180.0/180.0 so
WFS_pupil_rotoffset 0.0061532721 double
GCS now does the correct pupil rotation calculation. I will update IT#5401
01:45 03:15 Seeing varying from 2.0 -> 3.5" on the dim and 1.2->2.0 on the GCS FWHM. OK for some parameter determinations for MODS2 AGw.
03:15 Start determination of best guider hotspot position to have 0,0 x,y average offsets of SH spot positions
03:35 selectAGW right MODS_R.cfg stuck for 3 minutes on initializing WFS cam... David stop and start GCSR. OK
03:36 preset to BS9107. WFSC timeout. David stop GCSR and restart MODS2 azcam server.
03:45 OK. wfsc works now
03:51 getHotSpot right
- 267.0 240.0
- -0.2014 -0.1649
03:52 setHotSpot right -x 262.0 -y 245.0
03:55 setHotSpot right -x 272.0 -y 235.0
03:55 setHotSpot right -x 272.5 -y 236.0
- 0.0662 0.1573
- change of + in x hotspot give + in y average removed
- change fo + in y hotspot give - in x average removed
04:02 setHotSpot right -x 270.5 -y 237.0
04:04 setHotSpot right -x 269.5 -y 238.0
04:06 setHotSpot right -x 270.0 -y 237.5
- -0.0497 -0.035
- Seeing and has been from 1.5" to 1.9" FWHM on guider
04:15 preset to BS9122 @ 48 elevation. Seeing is so bad that that wfsc pinholes are filled an Sextractor cannot find enough SH spots.
04:23 Preset to BS9108 @ 85 elevation. Seeing is 1.7" FWHM on guider
04:38 Installed measured Gregorian Zernike Crosstalk values in to /lbt/tcs/2014D/psf/etc/psf.conf from the 2014C version of psf.conf
04:40 Try to measure new crosstalk coefficients for DX Gregorian
04:43 DIMM =2.34"
04:52 zernike_collect, 'z11_22.list'. DIMM seeing 2.38, guider 1.4". Directory /home/lbto/data/20141230/cross_1
05:01 Set done. Will analyze to see if I can get any information out of these data.
05:06 Surprising good signal for Z11, Z22 and crosstalk
05:24 modified pdf.conf. David stop and start PSFR
- Z11/4 4.0
- Z11/22 0.3
- Z22/4 5.2
- Z22/11 -2.3
05:30 zernike_collect, 'z11_22.list'. DIMM seeing 2.5, guider 1.4". Directory /home/lbto/data/20141230/cross_2
05:53 modified pdf.conf. David stop and start PSFR
- Z11/4 3.0
- Z11/22 0.2
- Z22/4 0.0
- Z22/11 -3.0
06:00 zernike_collect, 'z11_22.list'. DIMM seeing 1.6, guider 1.1". Directory /home/lbto/data/20141230/cross_3
06:11 modified pdf.conf. David stop and start PSFR
- Z11/4 3.0
- Z11/22 0.1
- Z22/4 8.0
- Z22/11 -4.0
06:20 zernike_collect, 'z11_22.list'. DIMM seeing 2.0, guider 1.3". Directory /home/lbto/data/20141230/cross_4
06:30 modified pdf.conf. David stop and start PSFR
- Z11/4 6.0
- Z11/22 0.1
- Z22/4 8.0
- Z22/11 0.0
06:43 zernike_collect, 'z11_22.list'. DIMM seeing 2.3, guider 1.4". Directory /home/lbto/data/20141230/cross_5
06:52 modified pdf.conf. David stop and start PSFR
- Z11/4 5.0
- Z11/22 0.2
- Z22/4 7.0
- Z22/11 -2.0
07:00 zernike_collect, 'z11_22.list'. DIMM seeing 2.3, guider 1.1". Directory /home/lbto/data/20141230/cross_6
07:15 Winds hit 20 m/s. David Closes
===============================================
12:20 - Winds at 15 m/s - However, twilight is at 12:53, thus by the time we open, DOFPIA - science time runs into twilight - CLOSING FOR NIGHT
-- %USERSIG{DougMiller - 2014-12-30}%