UT 20130324 Technical Observing

Observer: DMiller, JGuerra, JChristou (LBTO)
Telescope Operator: GBechetti (LBTO)
Additional AO Support: DMiller (LBTO)
Telescope Support: JUrban (LBTO)
Instrument Support: ESoliod (LBTO)

Summary

Details

Phase III Active Optics at RFBG

10:15 Edited /home/telescope/TCS/Configuration/GCS/LUCI_IRTC_R.cfg and LUCI_R_RM_sub11.cfg to have the same Act Opt parameters as LFBG. We are now using 9x9 Dynamic reconstructor. Also set gain factors for Z12-21 to 0.0 and Z22 to 1.5 with overall gain of 0.1

10:37 Set IDL parameter for RFBG to get exactly the same Act Opt corrections as GCS for Phase III Source Extractor Dynamic Reconstructor

Remove SH Pupil Wobble

11:01 Preset Active to on-axis star. collimate. Stop WFSing. pupil_collect, EXP=15.

0.0 385.1 277
60. 384.8 275.2
120. 381.5 276.3
180. 383.6 272.9
240. 385.0 274.8
300. 385.8 273.

11:08 Does not need a change

-- DougMiller - 24 Mar 2013
Topic revision: r1 - 24 Mar 2013, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.