UT 20130122 Technical Observing with AGW2
Observer: DMiller (LBTO) JHill (Tucson)
Telescope Operator: GBechetti (LBTO)
Additional AO Support: DMiller (LBTO)
Telescope Support: JUrban (LBTO)
Instrument Support: ESolheid (LBTO)
Summary
We worked most of the night on AGW2 alignment tasks @ RFBG (originally scheduled for 24UT). Seeing was excellent (0.5-0.6 arcsec on guider) except for 9-10 UT.
We have verified RFBG active optics (95% was done last month on 20121229).
We have aligned the AGw2 probe to the rotator center (and fixed errors in the instructions).
We have adjusted the DX M3 tip-tilt to minimize the pupil wobble for "w". Doug's new routine to measure the pupil position makes this faster.
We've checked the angle of dice5 offsets relative to the IRTC columns, and it is OK until LUCI2 arrives for a precise measurement.
We've collected a preliminary set of transformation data for AGw2.
We've collected one set of RFBG field aberration data at EL=58, and started on a second set at EL=30. There were lots of struggles with saturated wfs images, but the data have little atmospheric noise.
Details
02:40 Found that I had left the WFS_high_factor to 0.8 in December, and had not completed the task of finding the correct signs for the high order zernikes (Z12-22). Set factor to 0.0 and now stable convergence. FWHM = 0.5"
Align AGW2 probe to rotator center
02:48 Rotator at 132.6. Send ACQUIRE preset with PA=-130. Rot angle = 3. Change IE by +5".
02:55 Stop guiding. Geno send RFBG rotator to 360. take 70 second image. Image /home/AOeng/data/20130122/wfsc/right_guider000001.fits and 002.fits
- measured pixel center at 155,102
- getxy -u2 0.132, 613.38
- getxy -u2 -o 16 --> 9.0, 1005593.0
- LUCI_IRTC_R.cfg guidecam_hotspot_x, y = 310, 239 in bin 2
- want 155, 120 (in binned x2 coordinates)
- right_guider000002.fits:
03:00 ACTIVE Preset to collimate
03:09 Stop guiding, Geno send rotator to 0, 90 second guider image
03:24 Stop guiding. Send rotator to 0. Timeout guider image.
- setxy -u 2 -x 0 -y 612.5
- getxy -u 2 -o 16 --> -18.0, 1001001.0
- measured center pixel = 149.2, 87.8
- right_guider000003.fits
03:34 JMH joins the party .....
04:05 collimate, ACQUIRE preset, Stop guiding. Send rotator to 0.
- setxy -u 2 -x -y 613.1
- getxy -u 2 -o 16 --> 9.0, 1001001.0
- measured center pixel = 151, 113
- right_guider000004.fits:
04:10 collimate, ACQUIRE preset, Stop guiding. Send rotator to 0.
- setxy -u 2 -x 0.132 -y 613.3
- getxy -u 2 -o 16 --> 9.0, 1010364.0
- measured center pixel = 154.1, 116.4
- right_guider000005.fits
04:15 probe x,y = 0.132, 614.4. collimate, ACQUIRE preset, Stop guiding. Send rotator to 0.
- setxy -u 2 -x 0.132 -y 613.4
- getxy -u 2 -o 16 --> 8.0, 1010884.0
- measured center pixel = 153.4, 118
- right_guider000006.fits:
Based on the encoder counts above, we think we want to shift encoder counts by dX=-26 and dY=-9883. The wiki instructions (
BGTechProcedures) say to
change homeoffset on each axis. We tried that.
04:43 reboot oacserver, but "network error" after that causes another round of restarting. We see that this didn't make any change in the Y (RAD) position. Scratch our heads for a while. Finally decided to change ooffset rather than homeoffset. Need to check later with Jesper to see if this was kosher.
05:25 John change parameters in oacontrol.conf (RAD ooffset moved from 419000 to 409117). That was way too much - probe runs into limit. Try dividing by 16.....
05:38 John change parameters in oacontrol.conf (RAD ooffset moved from initial 419000 to 418382 by subtracting 9883/16). Somewhat better.....
06:10 collimate, ACQUIRE preset, Stop guiding. Send rotator to 360.
- setxy -u 2 -x -0.03 -y 613.45 (preset)
- getxy -u 2 -o 16 --> -2.0, 1009183.0
- measured center pixel = 161.3, 112.9
- right_guider000009.fits
06:15 collimate, ACQUIRE preset, Stop guiding. Send rotator to 0.
- setxy -u 2 -x 0.0 -y 612.5
- getxy -u 2 -o 16 --> -2.0, 1004219.0
- measured center pixel = 155.0, 98.4
- right_guider000010.fits
06:25 That shift was only 1/3 what we needed, so John changes oacontrol.conf to have 2x more change in Y direction (RAD homeoffset=417147)
06:41 collimate, ACQUIRE preset, Stop guiding. Send rotator to 360.
- setxy -u 2 -x -0.132 -y 613.5 (preset)
- getxy -u 2 -o 16 --> 8.0, 1015056.0
- measured center pixel = 161.2, 131.7
- right_guider0000011.fits
06:45 collimate, ACQUIRE preset, Stop guiding. Send rotator to 0.
- setxy -u 2 -x 0.0 -y 612.5
- getxy -u 2 -o 16 --> -2.0, 1010637.0
- measured center pixel = 158.9, 118
- right_guider000012.fits
07:00 John changed oacontrol.conf for AGW2
X_0=0.000, Y_0=612.500, x_0=x_guide=310.0, y_0=y_guide=239.0
ROT homeoffset=-33876 (net decrease 20), ROT lim_f=-13164, lim_n=13216 (net increase 20)
RAD ooffset=417147 (net decrease 1853), RAD lim_f=1052353, lim_n=1853 (net increase 1853)
07:14 collimate, ACQUIRE preset, Stop guiding. Send rotator to 360.
- -x -0.029 -y 612.8 (preset)
- setxy -u 2 -x -0.0 -y 612.5
- getxy -u 2 -o 16 --> -1.0, 1010637.0
- measured center pixel = 156.1, 117.64
- right_guider0000013.fits
Check AGW2 Pupil Wobble wrt/ little w
07:19 preset to new target
Adjusting the DX M3 Tip-Tilt Global Offsets, and using Doug's routine "pupil_outer_fit" to measure the pupil center.
08:15 Tip=+300 Tilt=0
08:27 Tip=+300 Tilt=300
08:32 Tip=+300 Tilt=-300
08:40 Tip=+450 Tilt=-300 (IE=-21 CA=-157)
08:54 Tip=+450 Tilt=-150
09:02 Tip=+600 Tilt=-150
09:05 Tip=+300 Tilt=-150
AbsTip=-450
AbsTilt=+350 Select=-26.042
09:11 Tip=+200 Tilt=-150
AbsTip=-550
AbsTilt=+350 Adopted this number for "w". Updated
DXM3IRTCFCollimation.dat.
Verify Position Angle Alignment on IRTC
Dice5 measurements on IRTC. The following table is composed of eyeball centroids of the cubes measured on the fly by JMH. Beware that the target star has a companion.
Image |
X |
Y |
095445 |
159 |
135 |
095543 |
256 |
232 |
095715 |
62 |
35 |
095807 |
251 |
31 |
095904 |
59 |
231 |
11:xx
GStar |
PCS_X |
PCS_Y |
AGW_X |
AGW_Y |
Rot Angle |
Pos Angle |
Image File |
Header File |
|
|
4 |
-24.676 |
29.636 |
-24.448 |
581.867 |
-102.402 |
-120.000 |
irtc.20130122.00025_cube.fits |
irtc.20130122.111116.fits |
GStar |
PCS_X |
PCS_Y |
AGW_X |
AGW_Y |
Rot Angle |
Pos Angle |
Image File |
Header File |
|
|
15 |
45.910 |
136.053 |
48.089 |
476.427 |
-105.171 |
-120.000 |
irtc.20130122.00027_cube.fits |
irtc.20130122.111855.fits |
13 |
120.895 |
34.925 |
121.458 |
579.134 |
-105.608 |
-120.000 |
irtc.20130122.00028_cube.fits |
irtc.20130122.111952.fits |
6 |
-65.602 |
42.720 |
-65.200 |
568.036 |
-106.565 |
-120.000 |
irtc.20130122.00029_cube.fits |
irtc.20130122.112103.fits |
16 |
-97.000 |
112.509 |
-95.434 |
497.520 |
-106.998 |
-120.000 |
irtc.20130122.00030_cube.fits |
irtc.20130122.112331.fits |
GStar |
PCS_X |
PCS_Y |
AGW_X |
AGW_Y |
Rot Angle |
Pos Angle |
Image File |
Header File |
|
|
5 |
39.822 |
4.070 |
39.706 |
608.632 |
-108.455 |
-120.000 |
irtc.20130122.00032_cube.fits |
irtc.20130122.112717.fits |
11 |
62.292 |
101.577 |
63.894 |
511.280 |
-108.896 |
-120.000 |
irtc.20130122.00033_cube.fits |
irtc.20130122.112825.fits |
4 |
-36.650 |
13.019 |
-36.715 |
598.318 |
-79.392 |
-90.000 |
irtc.20130122.00034_cube.fits |
irtc.20130122.112943.fits |
GStar |
PCS_X |
PCS_Y |
AGW_X |
AGW_Y |
Rot Angle |
Pos Angle |
Image File |
Header File |
|
|
15 |
-28.727 |
140.478 |
-26.596 |
470.673 |
-80.611 |
-90.000 |
irtc.20130122.00036_cube.fits |
irtc.20130122.113305.fits |
13 |
86.774 |
90.391 |
88.221 |
522.929 |
-81.043 |
-90.000 |
irtc.20130122.00037_cube.fits |
irtc.20130122.113355.fits |
16 |
-140.720 |
48.636 |
-140.330 |
560.799 |
-81.777 |
-90.000 |
irtc.20130122.00038_cube.fits |
irtc.20130122.113504.fits |
GStar |
PCS_X |
PCS_Y |
AGW_X |
AGW_Y |
Rot Angle |
Pos Angle |
Image File |
Header File |
|
|
5 |
31.992 |
23.136 |
32.198 |
589.380 |
-83.243 |
-90.000 |
irtc.20130122.00039_cube.fits |
irtc.20130122.113929.fits |
11 |
2.698 |
118.814 |
4.502 |
492.945 |
-83.686 |
-90.000 |
irtc.20130122.00040_cube.fits |
irtc.20130122.114036.fits |
10 |
112.507 |
31.423 |
112.980 |
582.500 |
-84.134 |
-90.000 |
irtc.20130122.00041_cube.fits |
irtc.20130122.114143.fits |
GStar |
PCS_X |
PCS_Y |
AGW_X |
AGW_Y |
Rot Angle |
Pos Angle |
Image File |
Header File |
|
|
15 |
-120.254 |
78.503 |
-119.327 |
531.210 |
-40.485 |
-45.000 |
irtc.20130122.00043_cube.fits |
irtc.20130122.114520.fits |
11 |
-82.721 |
85.412 |
-81.613 |
524.936 |
-40.903 |
-45.000 |
irtc.20130122.00044_cube.fits |
irtc.20130122.114610.fits |
13 |
-3.165 |
124.757 |
-1.260 |
486.884 |
-41.331 |
-45.000 |
irtc.20130122.00045_cube.fits |
irtc.20130122.114714.fits |
10 |
56.721 |
101.262 |
58.313 |
511.498 |
-41.779 |
-45.000 |
irtc.20130122.00046_cube.fits |
irtc.20130122.114821.fits |
14 |
130.807 |
32.922 |
131.387 |
581.311 |
-42.248 |
-45.000 |
irtc.20130122.00047_cube.fits |
irtc.20130122.114932.fits |
9 |
101.014 |
46.895 |
101.750 |
566.783 |
-42.693 |
-45.000 |
irtc.20130122.00048_cube.fits |
irtc.20130122.115039.fits |
3 |
29.302 |
15.060 |
29.364 |
597.430 |
-43.114 |
-45.000 |
irtc.20130122.00049_cube.fits |
irtc.20130122.115143.fits |
11:50 EL=58:09 (meridian) on M1 Z5=-152 Z6=-4134, on M2 Z6=+1.5 microns (are signs reversed?)
Field Aberration Data on Stone I at EL=58 deg
field_1/
12:00
12:40 changed wfs exposure time to 25000, but didn't restart GCS (ok). Is failing to find info file because of wfs saturation? That helped, but wasn't enough. Stars really are saturated in this good seeing.
GStar |
wfsc # |
0 |
636 |
GStar |
wfsc # |
0 |
639 |
GStar |
wfsc # |
0 |
642 |
13 |
643-645 |
GStar |
wfsc # |
0 |
648 |
8 |
649-651 |
GStar |
wfsc # |
0 |
654 |
14 |
655-657 |
GStar |
wfsc # |
0 |
661-662 |
15 |
663-665 |
GStar |
wfsc # |
0 |
670 |
15 |
671-673 |
GStar |
wfsc # |
0 |
676-677 |
16 |
678-680 |
0 |
681 |
10 |
682-684 |
12:50 Doug finally goes to bed after his "1/3 night" of observing.
Field Aberration Data on Stone G at EL=30 deg
field_2/
12:53 field_collect, start=1, limit=300 (Stone_G_field.list)
13:12 changed wfs exposure time to 20000, but didn't restart GCS (ok). The "saturation" preferentially recurs when we return on-axis. Why is that different than the initial collimation?
Seeing on guider is 0.7 arcsec FWHM at EL=27.5.
13:17 EL=26:25 on M1 Z5=-240 Z6=-10358
13:25 Shells RIP together as we go below EL=25 deg
GStar |
wfsc # |
0 |
689-692 |
10 |
693-695 |
GStar |
wfsc # |
0 |
698 |
10 |
699-701 |
GStar |
wfsc # |
0 |
705-706 |
11 |
707-709 |
GStar |
wfsc # |
0 |
712-713 |
8 |
714-716 |
GStar |
wfsc # |
0 |
719 |
12 |
720-722 |
GStar |
wfsc # |
0 |
726 |
13 |
727-729 |
GStar |
wfsc # |
Close
13:27 Close for impending dawn.
13:36 Put wfs exposure time back to 30000 in LUCI_IRTC_L.cfg.
--