RFBG performance verification: UTC 2011 March 16

Observer(s):D. Miller, J. Hill, A. Rakich, K. Boutsia (Tucson)
Telescope Operator:S. Allanson (LBTO)
Telescope Support: ?
Instrument Support: J. Morris

Summary

Elevation Preset Guiding/Collimation Dice5 (1x1) Dice5 (10x10) Large Offsets (1'-3')
30 col_30a col_30b col_30a col_30b      
45 col_45a col_45a      
60 col_60a col_60a dice5_1_60a dice5_10_60a  
75 col_75a col_75b col_75a col_75b dice5_1_75a (bad trans.) dice5_1_75b dice5_10_75a  

Details

All times are MST

12:00 Soft group finish and Konstantina and I take over

12:10 preset BS9142 Active gs=0 Collimate

12:20 test new routine collimate_collect, bs9142, logname, retval, gs=9, /GCS. IRTC was in continuous mode. stop that and resend collimate_collect

12:40 collimate_collect, bs9145, logname, retval, gs=??, /GCS -> col_60a

12:55 collimate_collect, bs9126, logname, retval, gs=??, /GCS -> col_30a
  • we had a communication glitch - only 9 sets saved - we consider it done
  • seeing 1.3''-1.5''

01:10 ask TO to change the pointing model

01:20 collimate_collect, bs9155, logname, retval, gs=7, /GCS -> col_45a * seeing ~1.2''

01:40 collimate_collect, bs9148, logname, retval, gs=1, /GCS, PA=180 -> col_75b
  • preset failed because GS was on the edge of the field
  • ask TO to correct pointing
  • repeat collimate_collect - seeing ~0.9''

02:00 collimate_collect, bs9177, logname, retval, gs=11, /GCS -> col_30b
  • aborted because gs too faint (13.6)
  • change gs=16 (12.6) - seeing >2.2'' - but it gets better as elevation increases

02:10 Modified /lbt/oacontrol/etc/oacontrol.conf and copied this new configuration to oacontrol.conf.20110316. The new transformation numbers are:

Transformation:
ax1 + by1 + kx = x2
  -1.00253428404995       1.716537354280230E-002  0.299803153168764     
ax1 + by1 + ky = y2
 -1.716537354280230E-002  -1.00253428404995        612.709892527265     
RMS(X)=  0.02260  RMS(Y)=  0.02310

where we used 26 data points. The data are in /OldRepository/AGW_Data/20110315/transform_1/Stone_K

02:22 Restart AGW and OA control
  • iif_preset, BS9158, GS=4, MODE=ACTIVE
  • we go to a higher elevation and let it collimate - verify that transformation is reasonable

02:52 dice5_collect, dist=1, iter=5 we are on star BS9158.

02:55 (09:55UT) OffsetPointing failed with Invalid Buffer. Retryed offset and was successful. Submitted IssueTrak # 3244

03:04 dice5_collect, dist=1, iter=5 -> dice5_1_75a
  • the star seems to have a little offset... not exactly centered on the IRTC

Ra/X Dec/Y Single Image Cube Image
Background Image
30 0 irtc.20110316.095247.fits irtc.20110316.00065_cube.fits
Dice5 1.00000 arcsec RADEC
0.00000 0.00000 irtc.20110316.095341.fits irtc.20110316.00066_cube.fits
Ra/X Dec/Y Single Image Cube Image
Background Image
30 0 irtc.20110316.095750.fits irtc.20110316.00067_cube.fits
Dice5 1.00000 arcsec RADEC
0.00000 0.00000 irtc.20110316.095835.fits irtc.20110316.00068_cube.fits
-1.00000 1.00000 irtc.20110316.095928.fits irtc.20110316.00069_cube.fits
1.00000 -1.00000 irtc.20110316.100015.fits irtc.20110316.00070_cube.fits
-1.00000 -1.00000 irtc.20110316.100121.fits irtc.20110316.00071_cube.fits
1.00000 1.00000 irtc.20110316.100208.fits irtc.20110316.00072_cube.fits
0.00000 0.00000 irtc.20110316.100301.fits irtc.20110316.00073_cube.fits
-1.00000 1.00000 irtc.20110316.100429.fits irtc.20110316.00074_cube.fits
1.00000 -1.00000 irtc.20110316.100554.fits irtc.20110316.00075_cube.fits
-1.00000 -1.00000 irtc.20110316.100640.fits irtc.20110316.00076_cube.fits
1.00000 1.00000 irtc.20110316.100736.fits irtc.20110316.00077_cube.fits
0.00000 0.00000 irtc.20110316.100823.fits irtc.20110316.00078_cube.fits
-1.00000 1.00000 irtc.20110316.100909.fits irtc.20110316.00079_cube.fits
1.00000 -1.00000 irtc.20110316.100955.fits irtc.20110316.00080_cube.fits
-1.00000 -1.00000 irtc.20110316.101044.fits irtc.20110316.00081_cube.fits
1.00000 1.00000 irtc.20110316.101135.fits irtc.20110316.00082_cube.fits
0.00000 0.00000 irtc.20110316.101220.fits irtc.20110316.00083_cube.fits
-1.00000 1.00000 irtc.20110316.101307.fits irtc.20110316.00084_cube.fits
1.00000 -1.00000 irtc.20110316.101353.fits irtc.20110316.00085_cube.fits
-1.00000 -1.00000 irtc.20110316.101447.fits irtc.20110316.00086_cube.fits
1.00000 1.00000 irtc.20110316.101542.fits irtc.20110316.00087_cube.fits
0.00000 0.00000 irtc.20110316.101629.fits irtc.20110316.00088_cube.fits
-1.00000 1.00000 irtc.20110316.101754.fits irtc.20110316.00089_cube.fits
1.00000 -1.00000 irtc.20110316.101843.fits irtc.20110316.00090_cube.fits
-1.00000 -1.00000 irtc.20110316.101933.fits irtc.20110316.00091_cube.fits
1.00000 1.00000 irtc.20110316.102019.fits irtc.20110316.00092_cube.fits

03:26 Check the tranformation coefficients in oacontrol.conf and found I had missed a (-) sign on sfpagw21. Fixed this and Steve restarted AGW2 and oacserver. Ooops.

03:36 sending the preset again after restarting AGW and OAC (BS9158)
  • for some reason we have no wfs image and we run acquisition manually - it worked before
  • this time it did not work - stop/start GCS - manual acqusition again - first time failed - second time worked - issue track 3245
  • repeat preset - preset cancelled because Telescope off source too long - issue track already exists 3124
  • repeat preset

03:51 dice5_collect, dist=1, iter=5 -> dice5_1_75b
  • seeing 1.0''-1.3''

Ra/X Dec/Y Single Image Cube Image
Background Image
30 0 irtc.20110316.105105.fits irtc.20110316.00093_cube.fits
Dice5 1.00000 arcsec RADEC
0.00000 0.00000 irtc.20110316.105155.fits irtc.20110316.00094_cube.fits
-1.00000 1.00000 irtc.20110316.105243.fits irtc.20110316.00095_cube.fits
1.00000 -1.00000 irtc.20110316.105332.fits irtc.20110316.00096_cube.fits
-1.00000 -1.00000 irtc.20110316.105422.fits irtc.20110316.00097_cube.fits
1.00000 1.00000 irtc.20110316.105511.fits irtc.20110316.00098_cube.fits
0.00000 0.00000 irtc.20110316.105559.fits irtc.20110316.00099_cube.fits
-1.00000 1.00000 irtc.20110316.105648.fits irtc.20110316.00100_cube.fits
1.00000 -1.00000 irtc.20110316.105737.fits irtc.20110316.00101_cube.fits
-1.00000 -1.00000 irtc.20110316.105827.fits irtc.20110316.00102_cube.fits
1.00000 1.00000 irtc.20110316.105916.fits irtc.20110316.00103_cube.fits
0.00000 0.00000 irtc.20110316.110005.fits irtc.20110316.00104_cube.fits
-1.00000 1.00000 irtc.20110316.110056.fits irtc.20110316.00105_cube.fits
1.00000 -1.00000 irtc.20110316.110145.fits irtc.20110316.00106_cube.fits
-1.00000 -1.00000 irtc.20110316.110235.fits irtc.20110316.00107_cube.fits
1.00000 1.00000 irtc.20110316.110324.fits irtc.20110316.00108_cube.fits
0.00000 0.00000 irtc.20110316.110413.fits irtc.20110316.00109_cube.fits
-1.00000 1.00000 irtc.20110316.110504.fits irtc.20110316.00110_cube.fits
1.00000 -1.00000 irtc.20110316.110553.fits irtc.20110316.00111_cube.fits
-1.00000 -1.00000 irtc.20110316.110644.fits irtc.20110316.00112_cube.fits
1.00000 1.00000 irtc.20110316.110732.fits irtc.20110316.00113_cube.fits
0.00000 0.00000 irtc.20110316.110821.fits irtc.20110316.00114_cube.fits
-1.00000 1.00000 irtc.20110316.110911.fits irtc.20110316.00115_cube.fits
1.00000 -1.00000 irtc.20110316.111000.fits irtc.20110316.00116_cube.fits

04:14 dice5_collect, dist=10, iter=5 -> dice5_10_75a
  • seing 1.2 -1.5''
Ra/X Dec/Y Single Image Cube Image
Background Image
30 0 irtc.20110316.111353.fits irtc.20110316.00119_cube.fits
Dice5 10.0000 arcsec RADEC
0.00000 0.00000 irtc.20110316.111443.fits irtc.20110316.00120_cube.fits
-10.0000 10.0000 irtc.20110316.111535.fits irtc.20110316.00121_cube.fits
10.0000 -10.0000 irtc.20110316.111626.fits irtc.20110316.00122_cube.fits
-10.0000 -10.0000 irtc.20110316.111716.fits irtc.20110316.00123_cube.fits
10.0000 10.0000 irtc.20110316.111806.fits irtc.20110316.00124_cube.fits
0.00000 0.00000 irtc.20110316.111856.fits irtc.20110316.00125_cube.fits
-10.0000 10.0000 irtc.20110316.111948.fits irtc.20110316.00126_cube.fits
10.0000 -10.0000 irtc.20110316.112038.fits irtc.20110316.00127_cube.fits
-10.0000 -10.0000 irtc.20110316.112127.fits irtc.20110316.00128_cube.fits
10.0000 10.0000 irtc.20110316.112217.fits irtc.20110316.00129_cube.fits
0.00000 0.00000 irtc.20110316.112345.fits irtc.20110316.00130_cube.fits
-10.0000 10.0000 irtc.20110316.112437.fits irtc.20110316.00131_cube.fits
10.0000 -10.0000 irtc.20110316.112530.fits irtc.20110316.00132_cube.fits
-10.0000 -10.0000 irtc.20110316.112622.fits irtc.20110316.00133_cube.fits
10.0000 10.0000 irtc.20110316.112711.fits irtc.20110316.00134_cube.fits
0.00000 0.00000 irtc.20110316.112803.fits irtc.20110316.00135_cube.fits
-10.0000 10.0000 irtc.20110316.112852.fits irtc.20110316.00136_cube.fits
10.0000 -10.0000 irtc.20110316.112940.fits irtc.20110316.00137_cube.fits
-10.0000 -10.0000 irtc.20110316.113030.fits irtc.20110316.00138_cube.fits
10.0000 10.0000 irtc.20110316.113120.fits irtc.20110316.00139_cube.fits
0.00000 0.00000 irtc.20110316.113210.fits irtc.20110316.00140_cube.fits
-10.0000 10.0000 irtc.20110316.113300.fits irtc.20110316.00141_cube.fits
10.0000 -10.0000 irtc.20110316.113348.fits irtc.20110316.00142_cube.fits
-10.0000 -10.0000 irtc.20110316.113438.fits irtc.20110316.00143_cube.fits
10.0000 10.0000 irtc.20110316.113532.fits irtc.20110316.00144_cube.fits

04:37 preset to BS9152 for dice5 at 60 elevation
  • iif_preset, BS9152, gs=8, MODE=ACTIVE

04:44 dice5_collect, dist=1.0, iter=5 -> dice5_1_60a
  • seeing 1.2-1.5''

Ra/X Dec/Y Single Image Cube Image
Background Image
30 0 irtc.20110316.114335.fits irtc.20110316.00145_cube.fits
Dice5 1.00000 arcsec RADEC
0.00000 0.00000 irtc.20110316.114425.fits irtc.20110316.00146_cube.fits
-1.00000 1.00000 irtc.20110316.114514.fits irtc.20110316.00147_cube.fits
1.00000 -1.00000 irtc.20110316.114603.fits irtc.20110316.00148_cube.fits
-1.00000 -1.00000 irtc.20110316.114652.fits irtc.20110316.00149_cube.fits
1.00000 1.00000 irtc.20110316.114741.fits irtc.20110316.00150_cube.fits
0.00000 0.00000 irtc.20110316.114831.fits irtc.20110316.00151_cube.fits
-1.00000 1.00000 irtc.20110316.114924.fits irtc.20110316.00152_cube.fits
1.00000 -1.00000 irtc.20110316.115014.fits irtc.20110316.00153_cube.fits
-1.00000 -1.00000 irtc.20110316.115103.fits irtc.20110316.00154_cube.fits
1.00000 1.00000 irtc.20110316.115153.fits irtc.20110316.00155_cube.fits
0.00000 0.00000 irtc.20110316.115242.fits irtc.20110316.00156_cube.fits
-1.00000 1.00000 irtc.20110316.115335.fits irtc.20110316.00157_cube.fits
1.00000 -1.00000 irtc.20110316.115424.fits irtc.20110316.00158_cube.fits
-1.00000 -1.00000 irtc.20110316.115515.fits irtc.20110316.00159_cube.fits
1.00000 1.00000 irtc.20110316.115605.fits irtc.20110316.00160_cube.fits
0.00000 0.00000 irtc.20110316.115655.fits irtc.20110316.00161_cube.fits
-1.00000 1.00000 irtc.20110316.115744.fits irtc.20110316.00162_cube.fits
1.00000 -1.00000 irtc.20110316.115834.fits irtc.20110316.00163_cube.fits
-1.00000 -1.00000 irtc.20110316.115923.fits irtc.20110316.00164_cube.fits
1.00000 1.00000 irtc.20110316.120012.fits irtc.20110316.00165_cube.fits
0.00000 0.00000 irtc.20110316.120103.fits irtc.20110316.00166_cube.fits
-1.00000 1.00000 irtc.20110316.120152.fits irtc.20110316.00167_cube.fits
1.00000 -1.00000 irtc.20110316.120241.fits irtc.20110316.00168_cube.fits
-1.00000 -1.00000 irtc.20110316.120333.fits irtc.20110316.00169_cube.fits
1.00000 1.00000 irtc.20110316.120425.fits irtc.20110316.00170_cube.fits

05:06 dice5_collect, dist=10.0, iter=5 -> dice5_10_60a
  • seeing >1.5''

Ra/X Dec/Y Single Image Cube Image
Background Image
30 0 irtc.20110316.120608.fits irtc.20110316.00171_cube.fits
Dice5 10.0000 arcsec RADEC
0.00000 0.00000 irtc.20110316.120658.fits irtc.20110316.00172_cube.fits
-10.0000 10.0000 irtc.20110316.120747.fits irtc.20110316.00173_cube.fits
10.0000 -10.0000 irtc.20110316.120837.fits irtc.20110316.00174_cube.fits
-10.0000 -10.0000 irtc.20110316.120926.fits irtc.20110316.00175_cube.fits
10.0000 10.0000 irtc.20110316.121016.fits irtc.20110316.00176_cube.fits
0.00000 0.00000 irtc.20110316.121105.fits irtc.20110316.00177_cube.fits
-10.0000 10.0000 irtc.20110316.121157.fits irtc.20110316.00178_cube.fits
10.0000 -10.0000 irtc.20110316.121245.fits irtc.20110316.00179_cube.fits
-10.0000 -10.0000 irtc.20110316.121334.fits irtc.20110316.00180_cube.fits
10.0000 10.0000 irtc.20110316.121425.fits irtc.20110316.00181_cube.fits
0.00000 0.00000 irtc.20110316.121514.fits irtc.20110316.00182_cube.fits
-10.0000 10.0000 irtc.20110316.121604.fits irtc.20110316.00183_cube.fits
10.0000 -10.0000 irtc.20110316.121652.fits irtc.20110316.00184_cube.fits
-10.0000 -10.0000 irtc.20110316.121742.fits irtc.20110316.00185_cube.fits
10.0000 10.0000 irtc.20110316.121832.fits irtc.20110316.00186_cube.fits
0.00000 0.00000 irtc.20110316.121922.fits irtc.20110316.00187_cube.fits
-10.0000 10.0000 irtc.20110316.122014.fits irtc.20110316.00188_cube.fits
10.0000 -10.0000 irtc.20110316.122102.fits irtc.20110316.00189_cube.fits
-10.0000 -10.0000 irtc.20110316.122152.fits irtc.20110316.00190_cube.fits
10.0000 10.0000 irtc.20110316.122242.fits irtc.20110316.00191_cube.fits
0.00000 0.00000 irtc.20110316.122332.fits irtc.20110316.00192_cube.fits
-10.0000 10.0000 irtc.20110316.122420.fits irtc.20110316.00193_cube.fits
10.0000 -10.0000 irtc.20110316.122509.fits irtc.20110316.00194_cube.fits
-10.0000 -10.0000 irtc.20110316.122558.fits irtc.20110316.00195_cube.fits
10.0000 10.0000 irtc.20110316.122648.fits irtc.20110316.00196_cube.fits

05:28 End of night

-- DougMiller - 16 Mar 2011

This topic: Commissioning > Main > WebHome > Com20110316
Topic revision: 25 Mar 2011, KonstantinaBoutsia
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.