RFBG performance verification: UTC 2011 March 15

Observer(s):D. Miller, J. Hill, A. Rakich, K. Boutsia (Tucson)
Telescope Operator:D. Huerta (LBTO)
Telescope Support: ?
Instrument Support: J. Morris

Summary


Taking transform data on Stone I.

08:00 Started to collect transformation data. Collected background and one data point before GCS had problems. See was from 0.7" to 0.9".

08:17 See issue 3241, GCS stopped WFS for unknown reason

08:40 Started to collect more transformation data. Collected another 6 data points before clouds came along. David shut the chamber at 09:30 or so.

10:40 Clouds have thinned. David open and collimated

11:00 slewing to Stone_K, seeing 0.6-0.8. COllected 42 data points. Preliminary result show ~27 microns rms in x and y. Good enough. We will install this transformation for tomorrow evening and start collectind Performance Verification data.

12:30 Moving on to lenslet transformation - CLOUDY and starting to be bright

12:40 End of night

-- AndrewRakich - 15 Mar 2011
Topic revision: r2 - 15 Mar 2011, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.