AO Commissioning at RFBG: 20-26 December 2011 UTC

Observer(s): SEsposito, F, LBusoni, EPinna, DMiller, GBrusa, JC (LBTO)

Telescope Operator: Geno (LBTO)

SW Support: TEdgin (Tucson)

Telescope Support: JUrban (LBTO)

Instrument Support: DOfficer (LBTO)

Summary

4:41 Send Preset ACE, success

4:43 runao, success

AO raw stage on-axis x -75, y = +35

PCS Log Xi = -0.207206 Eta = -1.89710
AOS GUI Preset values x = -0.207206 y = -1.89710
AOS GUI Acquire ref dx = -0.1 dy = 2.4

4:55 offset, 0.2, 0.

PCS log Xi = 0.115785, Eta = 0.911455
AOS GUI X_offet 0.323056 Y_offset 2.10117

5:05 preset ACE

PCS log Xi = 0.00059992 Eta = 0.897992
AOS GUI Preset values x = 0.00, y = 0.90
Raw stage x= -75 y = =35.9

5:09 runao, Success, but AO loop not closed. IIF says loop is closed. WFSC counter is static, but Adsec counter on Adsec actuator gui is incrementing

AOS GUI Acquire ref dx = -0.11 dy = 0.35

5:25 offset, 0.2, 0.0, failed AO wrong state

5:27 preset ACE

PCS log Xi = 0.0006334 Eta = 0.974238
AOS GUI x = 0.0, y = 0.97

5:29 runao

dx=-0.48 dy=-0.12
raw stage x = -75.5 y = 35.8

5:32 offset, 0.2, 0.0, Failed, command timeout from AOS OffsetXY

PCS log Xi = 0.178919 Eta = 1.049657
AOS GUI x_offset = 0.106426 y_offset = 0.75419

5:39 preset ACE success

5:42 runao, failed, command timeout, but x, y_offset were correct

5:57 preset ACE, success

Xi = 0.0006776, Eta 1.12422
AOS preset 0.0 1.12

6:00 runao

dx = -0.91 dy -0.32

6:05 offset, 5.0, 0, failed, timeout

Xi = 2.4285 Eta= 2.908304
AOS x_offset= 2.42784 y_offset=1.784 Correct