Collected LUCIFER data to build an M2 Collimation model.
Preparations
8:00 after reverting to previous software build, reloading and reconfiguring for LBC we get started now. Adjusting pointing and collimation.
Details
All Lucifer image named luci.20100521.*.fits
Time |
BS star |
Image number |
Elevation |
08:58 |
9162 |
33 |
72 |
09:08 |
9166 |
35 |
55 |
09:19 |
9160 |
38 |
48 |
09:26 |
9155 |
40 |
33 |
10:01 |
9145 |
44 |
27 |
10:13 |
9156 |
46 |
39 |
10:24 |
9169 |
48 |
62 |
10:42 |
FS 149 |
52 |
80 |
10:55 |
FS 151 |
55 |
69 |
11:06 |
FS 149 |
58 |
85 |
11:25 |
9183 |
62 |
61 |
11:35 |
9182 |
64 |
52 |
Follow-up
JMH wrote to telescope work the following afternoon:
Dear Telescope,
Following the Andrew et al. laser-tracker-based experiments of
the last three nights, I have restored the SX collimation and pointing
models to their traditional configuration.
Specifically:
1) The LUCIFER pointing model has been restored to
LUCIFERSX.20100501.ptmod
2) The LUCIFER PM collimation model has been restored to the contents
of SXPMLUCIFERCollimation.20100105.dat with the active file name
SXPMLUCIFERCollimation.dat .
(Yes, 0105 and 0501 are the correct dates, and not transposition typos.)
3) The LBC-Blue PM collimation model has been restored to the contents
of SXPMLBCCollimation.20100508.dat (which was the last update for the
hub shift to the summer position). The active filename is of course
SXPMLBCCollimation.dat .
and finally unrelated to Andrew's activities,
4) I have updated the LBC pointing model LBCSX.20100521.ptmod to have
the current azimuth zeropoint of IA=-393 arcsec. We saw a value
similar to this on 20100505, and I confirmed it was still true on
20100519 (although we were using Andrew's special M1 collimation model
at the time). This particular value of IA is chosen to match the
LUCIFER pointing model from 20100501.
The LBC pointing model does need to be remeasured. It has changed
since the azimuth absolute encoder has been fixed, and weight added to
the telescope has changed the elevation (tube flexure) term by several
(ten?) arcsec.
Ciao,
John
M2 Collimation Model
hselect luci*.fits M2-X,M2-Y,M2-Z,M2RX,M2RY,TELALT,M1CTEMP,$I yes > /home/LBTO/M2.20100521_collimation.txt
--