AGw Unit Status

AGw Unit location and fiber connectivity status

AGW3 was installed in place of AGW1 in front of LUCI1 at LFBG on 20180601. This temporary configuration is expected to remain until 20180716. See IT 7252 for configuration details.

Unit / Location AGW Name Status Unit / Location AGW Name Status
AGW1 @ TEL, LFBG "bentGregorianFront left" LUCI-1 (AIP) ON AGW2 @ TEL, RFBG "bentGregorianFront right" LUCI-2 (AIP) ON
AGW5 @ TEL, LDG "directGregorian left" MODS-1 (OSU) ON AGW6 @ TEL, RDG "directGregorian right" MODS-2 (OSU) ON
AGW3 @ SCR, LDG "directGregorian left" PEPSIPOL-1 (AIP) OFF AGW4 @ Tucson, RDG "directGregorian right" PEPSIPOL-2 (AIP) OFF
AGW7 @ TEL, "bentGregorianRearFiberFeed left" PEPSIPFU-1 (AIP) ON AGW8 @ TEL, "bentGregorianRearFiberFeed right" PEPSIPFU-2 (AIP) ON

  • AGW1 - LUCI-1 (AIP) Telescope - LFBG
  • AGW2 - LUCI-2 (AIP) Telescope - RFBG
  • AGW5 - MODS-1 (OSU) Telescope - LDG
  • AGW6 - MODS-2 (OSU) Telescope - RDG
  • AGW7 - PEPSIPFU-1 (AIP) Telescope - LPFU Nov2013
  • AGW8 - PEPSIPFU-2 (AIP) Telescope - RPFU Nov2013

  • AGW3 - PEPSIPOL-1 (AIP) Summit Lab
  • AGW4 - PEPSIPOL-2 (AIP) - RDG

AGw1

Unit Change Request Connected Details Current Location Current Status
AGW1 @ TEL          
AGW1 SCR->TEL 201308xx     TEL OK by TG
  20130522 by T.H.   fw.rev.129+XY    
 

TEL->BCR->20140812

by Eng.

20140813   B.CL.R  
           
           

AGw2

Unit Change Request Connected Details Current Location Current Status
AGW2 @ TEL          
AGW2 TEL -> SCR 20131011 RM 20131021   SCR OK by RM
  to JK from TG/RM CRB from LIU 0210 (TEL) fibers had not been switched over in CRB    
    CRB to LIU 0212 (SCR ) no network connectivity to MOXA    
      wrong Media Converter in the Small Clean Room    
AGW2 SCL->TEL 20131021 ENG. 20131022   TEL OK by RM/TG
  by Eng. CRB from LIU 0212 (SCR )      
    CRB to LIU 0210 (TEL)      
  20130522 by T.H.   fw.rev.129+XY    
 

TEL->BCR->20140812

by Eng.

20140813      

AGw3

Unit Change Request ConnectedSorted ascending Details Current Location Current Status
AGW3 @ SCR          
  to JK from RM        
  20130522 by T.H.   fw.rev.129+XY    
           
AGW3 Base -> SCR 20130701 RM 20130701     OK by RM

AGw4

Unit Change Request Connected Details Current Location Current Status
AGW4 @ Tucson          
AGW4 Base 20130610 TG 20130610   TEL OK by TIG
  20130522 by T.H.   fw.rev.129+XY    
  20140404 by T.H   fw: ver. 1001    
  20140815 by C.B./M.G.   fw: I65=40200    

MODS1

Unit Change Request Connected Details Current Location Current Status
MODS1 @ TEL          
MODS1       TEL OK by TG
           

MODS2

Unit Change Request Connected Details Current Location Current Status
MODS2 @ TEL          
MODS2       TEL OK
           

AGw7

Unit Change Request Connected Details Current Location Current Status
AGW7 @ TEL          
AGW7       TEL OK
           

AGw8

Unit Change Request Connected Details Current Location Current Status
AGW8 @ TEL          
AGW8       TEL OK
           

AGw Unit move procedure from the Telescope to the Small Clean Room

  • Reconnect request to move AGw unit from Telescope to Clean Room
  • Fibers disconnected at the telescope
    1. Network Fibers disconnected from the telescope to the AGw
    2. G fibers disconnected at the telescope from the AGw
    3. W fibers disconnected at the telescope from the AGw
  • CRB fibers reconnect for this AGw from Telescope location to Clean Room
  • AGw Unit moved to the clean room
    1. Connect G and W channels at AGw to Small Clean Room fibers from LIU connected to G,W channels
    2. Connect "E-BOX" AGw connector to Small Clean Room Media Convertor #1 (for small AGw) connected to Switch 102 (port 3, 192.168.2.*) and small AGw Network fiber connectors
    3. Do not use "WFS-NET" AGW connector at Small Clean Room Media Convertor #2 (for BIG AGW) connected to Switch 102 (port 4 192.168.12.* ) and BIG AGW Network fiber connectors
  • Update records at this page

AGw Unit move procedure from the Small Clean Room to the Telescope

  • Reconnect request to move AGw unit from the Clean Room to the Telescope
  • Fibers disconnected at the Small Clean Room
    1. Disconnect G and W channels at AGw to Small Clean Room fibers from LIU connected to G,W channels
    2. Disconnect "E-BOX" AGw connector to Small Clean Room Media Convertor #1 (for small AGw) connected to Switch 102 (port 3, 192.168.2.*) and small AGw Network fiber connectors
    3. Do not use "WFS-NET" AGW connector at Small Clean Room Media Convertor #2 (for BIG AGW) connected to Switch 102 (port 4 192.168.12.* ) and BIG AGW Network fiber connectors
  • AGw Unit moved to the Telescope
    1. Network Fibers connected at the telescope
    2. G fibers connected at the telescope to the AGw
    3. W fibers connected at the telescope to the AGw
  • CRB fibers reconnect for this AGw from the Small Clean Room to the Telescope location
  • Update records at this page

AGw Unit test procedure after reconnect to the Telescope

  • Test request to verify AGw unit is ready for Observation
    1. Verify the Power Cable is connected to AGw, when power cable was connected
    2. Test if power is available the the AGw unit (use Voltmeter)
    3. Verify Network fibers polarity at the AGw unit
    4. Check if AGw moxa is reachabe over the network
    5. Verify if G and W channels connected correctly and polarity of the fibers for G and W channels
    6. Use Fiber tester to verify polarity and light availability at the G channel
    7. Use Fiber tester to verify polarity and light availability at the W channel
    8. Use image acquisition to confirm G and W channels a) Turn the Dome Lights OFF (Call summit for clearance) b) For MODS only: (Open hatch, cal.unit out)
  • E-mail to telescope after AGw unit is ready for Observation
  • Update records at this page

Network

  • Small Clean Room, Switch 102
    1. port 3, 192.168.2.* "E-BOX" AGw connector to Small Clean Room Media Convertor #1 (for small AGw)
    2. port 4, 192.168.12.* "WFS-NET" AGW connector at Small Clean Room Media Convertor #2 (for BIG AGW)
  • Big Clean Room, Switch 103
    1. (port 2 192.168.10.* ) "WFS-NET" (for BIG AGW)
    2. (port 5 192.168.10.* ) "WFS-NET" (for BIG AGW Pyramid WFS)
    3. (port 7 192.168.10.* ) "WFS-NET" (for BIG AGW DX WFS)
    4. (port 8 192.168.10.* ) "WFS-NET"
    5. (port 9 192.168.10.* ) "WFS-NET"
  • Big Clean Room, Switch 104
    1. (port 3 192.168.10.* ) "WFS-NET" (for BIG AGW)

- LBTO_Wiki search

-- TarasGolota - 21 Oct 2013
Topic revision: r15 - 02 Jun 2018, JohnHill
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.