Summary
- Check of homing position of the slit, there are some angle at 300 degrees that the slit is not doing the homing.
- Acquisition of ASM_SX telemetry at different elevations, starting from 90 (step of 10 degrees).
- Light cirrus passing at the beginning of the night.
- ND 4 installed in the wfs.
- Working in a new procedure for the acquire reference star. This new procedure will be more accurate to measure the magnitude.
- Performances of the bin 4 at faint magnitudes up to mag.17.3 and bright star. mag 7.
- GOOD & STABLE SEEING AROUND 0.7" MEASURED FROM THE DIMM.
Details
- It has been checked the position of the slit during the homing, for a rotator position where the slit is on top, the stage has difficulty to perform the homing.
Start 2:00 UT
- Acquisition of ASM_SX telemetry data at different elevations. Time expended: 35minutes.
- 2:37 Pointing and collimation. Star nearby to the first target AO1016.
- Starting to send ACE mode presets:
iif_preset command key with the ND=4 installed < iif_preset,ao1016,mode=ace,gs=9,rs=0,pa=180,ND=4 >
IDL> iif_preset,ao734,mode=ace,gs=18,rs=0,pa=0,ND=4
# Target star: 0 AO734
20:56:03.388 +38:50:56.32 RADEC 2000.0 J2000 305.4 -5.2 10.70 R 2.81 B_V 1.65
# Guide star: 18
20:56:01.649 +38:53:42.46 RADEC 2000.0 J2000 0.0 0.0 12.31 R 3.47 B_V 0.63
# AO Reference star: 0
20:56:03.388 +38:50:56.32 RADEC 2000.0 J2000 305.4 -5.2 14.70 R 2.81 B_V 1.65
preset, AO734, MODE=ACE, GS=18, RS=0
Wed Sep 19 02:54:15.919 2012 aos.PresetAO.started -
PresetAO[694] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-1.01246e-06,ro2:-5.84156e-08,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 02:54:15.970 2012 aos.info - WFS source set to: FLAOWFS
Wed Sep 19 02:56:20.790 2012 aos.AcquireRef.started -
AcquireRef[695]
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error
err = Resmsg[1] = RETRY: Star too faint for mode ACE-AO. Maximum magnitude is 17.5 (-5002) VALUE_OUT_OF_RANGE_ERROR
err = Resmsg[2] =
AcquireRef[695] failed - RETRY: Star too faint for mode ACE-AO. Maximum magnitude is 17.5 (-5002) VALUE_OUT_OF_RANGE_ERROR
Debug the new procedure to estimate the flux at close loop
Wed Sep 19 03:19:51.225 2012 aos.PresetAO.started -
PresetAO[697] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:8.30112e-07,ro2:-1.49068e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 03:19:51.276 2012 aos.info - WFS source set to: FLAOWFS
Wed Sep 19 03:30:43.343 2012 aos.PresetAO.started -
PresetAO[701] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:3.70545e-06,ro2:-9.89285e-08,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 03:30:43.393 2012 aos.info - WFS source set to: FLAOWFS
Wed Sep 19 03:30:43.565 2012 aos.PresetAO.failed -
PresetAO[701] failed - Unknown exception
===> Error in irc_presetTelescope
err = Resmsg[0] =
PresetTelescope result status: Error err = Resmsg[1] = Stop[700] complete
err = Resmsg[2] = Unknown exception err = Resmsg[3] =
PresetAO[701] failed - Unknown exception
Still debugging
Wed Sep 19 03:32:25.464 2012 aos.PresetAO.started -
PresetAO[703] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:1.17856e-06,ro2:-6.00804e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 03:32:25.514 2012 aos.info - WFS source set to: FLAOWFS
3:36 SEEING:1.1"
Wed Sep 19 03:39:09.037 2012 aos.PresetAO.started -
PresetAO[707] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-8.59977e-07,ro2:-1.25431e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 03:39:09.088 2012 aos.info - WFS source set to: FLAOWFS
Wed Sep 19 03:41:02.563 2012 aos.AcquireRef.failed -
AcquireRef[708] failed - RETRY: source_acq error: Star centering procedure failed to converge after 5 iterations (-20003) WFSARB_STAR_NOTCONVERGED
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error
err = Resmsg[1] = RETRY: source_acq error: Star centering procedure failed to converge after 5 iterations (-20003) WFSARB_STAR_NOTCONVERGED
err = Resmsg[2] =
AcquireRef[708] failed - RETRY: source_acq error: Star centering procedure failed to converge after 5 iterations (-20003) WFSARB_STAR_NOTCONVERGED
Still debugging
Wed Sep 19 03:41:47.708 2012 aos.PresetAO.started -
PresetAO[710] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:2.11193e-06,ro2:-7.15286e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 03:41:47.758 2012 aos.info - WFS source set to: FLAOWFS
Wed Sep 19 03:43:28.321 2012 aos.info - COILS DISABLED
Wed Sep 19 03:43:28.372 2012 aos.error - Stopping WFS...
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error err = Resmsg[1] = err = Resmsg[2] =
AcquireRef[711] failed
SKIP+RIP The ASM is taken too much tilt?
3:54 SEEING: 0.9"
Wed Sep 19 03:54:42.448 2012 aos.PresetAO.started -
PresetAO[714] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-1.99741e-06,ro2:-1.99336e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 03:54:42.499 2012 aos.info - WFS source set to: FLAOWFS
IMPORTANT: THE PRESET FAILED STOPPING THE WFSARBITRATOR, THE WFS STATUS IS OFFLINE AND THE IIF AND THE AOS DON'T REPORT BACK THE ERROR. It took about 6 minutes to get a time out error from the arbitrator.
Wed Sep 19 04:00:22.854 2012 aos.AcquireRef.failed -
AcquireRef[715] failed - Command timeout - -13107: Arbitrator interface generic error
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error
err = Resmsg[1] =
AcquireRef[715] failed - Command timeout - -13107: Arbitrator interface generic error
Still debugging
4:08. SEEING: 0.9"
Wed Sep 19 04:10:16.051 2012 aos.PresetAO.started -
PresetAO[717] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:9.41217e-07,ro2:-7.85245e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 04:10:16.102 2012 aos.info - WFS source set to: FLAOWFS
WFSARBITRATOR RIP
Still debugging
Ctrl+C from IDL and a new preset sent, the system needs to wait till for the arbitrator time out
Wed Sep 19 04:15:43.651 2012 aos.warning - Stop rejected. Busy serving
AcquireRef [718]
Wed Sep 19 04:15:45.466 2012 aos.warning -
PresetAO rejected. Busy serving
AcquireRef [718]
Wed Sep 19 04:15:57.539 2012 aos.AcquireRef.failed -
AcquireRef[718] failed - Command timeout - -13107: Arbitrator interface generic error
Wed Sep 19 04:16:18.923 2012 aos.PresetAO.started -
PresetAO[720] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:6.46211e-07,ro2:-1.2779e-06,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 04:16:18.974 2012 aos.info - WFS source set to: FLAOWFS
WFSARBITRATOR RIP
Still debugging
WFS SOFTWARE STOP.
Wed Sep 19 04:25:38.533 2012 aos.warning - NO connection. Reason:
MsgD not responding
4:20. SEEING: 0.92"
4:34 SEEING:0.92"
4:41 Still Debugging.
Wed Sep 19 04:51:19.290 2012 aos.PresetAO.started -
PresetAO[730] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-1.26388e-06,ro2:-9.56171e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 04:51:19.341 2012 aos.info - WFS source set to: FLAOWFS
Wed Sep 19 04:52:00.975 2012 aos.PresetAO.failed -
PresetAO[730] failed - Timeout waiting for techviewctrl.L.PIXELLUT.CUR (-5001) TIMEOUT_ERROR
===> Error in irc_presetTelescope
err = Resmsg[0] =
PresetTelescope result status: Error err = Resmsg[1] = Stop[729] complete
err = Resmsg[2] = Timeout waiting for techviewctrl.L.PIXELLUT.CUR (-5001) TIMEOUT_ERROR
err = Resmsg[3] =
PresetAO[730] failed - Timeout waiting for techviewctrl.L.PIXELLUT.CUR (-5001) TIMEOUT_ERROR
4:54 WFS SOFTWARE STOP/START
Wed Sep 19 05:01:33.973 2012 aos.PresetAO.started -
PresetAO[734] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-9.77532e-07,ro2:-2.99873e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 05:01:34.023 2012 aos.info - WFS source set to: FLAOWFS
Wed Sep 19 05:01:51.777 2012 aos.PresetAO.failed -
PresetAO[734] failed - Timeout waiting for techviewctrl.L.STATUS (-5001) TIMEOUT_ERROR
===> Error in irc_presetTelescope
err = Resmsg[0] =
PresetTelescope result status: Error err = Resmsg[1] = Stop[733] complete
err = Resmsg[2] = Timeout waiting for techviewctrl.L.STATUS (-5001) TIMEOUT_ERROR
err = Resmsg[3] =
PresetAO[734] failed - Timeout waiting for techviewctrl.L.STATUS (-5001) TIMEOUT_ERROR
Still Debugging.
Wed Sep 19 05:05:24.042 2012 aos.PresetAO.started -
PresetAO[736] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-3.00843e-06,ro2:-5.9096e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 05:05:24.092 2012 aos.info - WFS source set to: FLAOWFS
Wed Sep 19 05:07:24.170 2012 aos.PresetAO.failed -
PresetAO[736] failed - Command timeout - -13107: Arbitrator interface generic error
===> Error in irc_presetTelescope
err = Resmsg[0] =
PresetTelescope result status: Error err = Resmsg[1] = Stop[735] complete
err = Resmsg[2] =
PresetAO[736] failed - Command timeout - -13107: Arbitrator interface generic error
The stage in Y didn't do the homing. Rotator:248. The rotator is moved to 180 to allow the homing.
Wed Sep 19 05:11:44.212 2012 aos.PresetAO.started -
PresetAO[739] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:3.94854e-07,ro2:1.77365e-06,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 05:11:44.262 2012 aos.info - WFS source set to: FLAOWFS
5:15 Pointing Check
5:15. SEEING:0.9"
Wed Sep 19 05:17:11.640 2012 aos.PresetAO.started -
PresetAO[744] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:3.69841e-06,ro2:-2.90782e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspee
===> Error in irc_presetTelescope
err = Resmsg[0] =
PresetTelescope result status: Warning err = Resmsg[1] = Stop[743] complete
err = Resmsg[2] = WARNING: guide cam exposure time is less than 2000 ms which might cause guiding on atmospheric dispersions
err = Resmsg[3] = adjusted exposure time to 1456 ms
err = Resmsg[4] =
PresetAO[744] complete - dx:nan dy:nan slNull: f1:Filter #0 f2:Filter #0 freq:200 mag:0 nBins:4 nModes:36 r0:0 snMode:0 strehl:0 ttMod:6
d:nan,winddir:nan)
Wed Sep 19 05:17:11.690 2012 aos.info - WFS source set to: FLAOWFS
Star collimated and we can send the iif_runao. The wfs do the close loop procedure.
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error err = Resmsg[1] = Camera lens centering did not converge after 180 seconds
err = Resmsg[2] =
AcquireRef[745] failed - Camera lens centering did not converge after 180 seconds
Wed Sep 19 05:22:53.581 2012 aos.AcquireRef.failed -
AcquireRef[745] failed - Camera lens centering did not converge after 180 seconds.
*BACK TO THE ORIGINAL WFS CODE.*
Change the rotation angle by 90 degrees.
Wed Sep 19 05:24:02.469 2012 aos.PresetAO.started -
PresetAO[747] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:3.23595e-07,ro2:-1.58623e-06,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 05:24:02.520 2012 aos.info - WFS source set to: FLAOWFS
Taken measurements. Open and close loop
5:36 SEEING:0.85"
5:56. SEEING: 0.9"
Wed Sep 19 06:01:55.733 2012 aos.PresetAO.started -
PresetAO[775] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:1.7324e-07,ro2:8.55706e-07,mag:14.7,cindex:2.81,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 06:01:55.784 2012 aos.info - WFS source set to: FLAOWFS
Taken measurements at close loop.
6:11. SEEING:0.98"
6:18. SEEING:0.90"
Wed Sep 19 06:22:57.418 2012 aos.PresetAO.started -
PresetAO[791] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:1.19068e-06,ro2:-4.49027e-08,mag:15.1,cindex:2.24,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 06:22:57.469 2012 aos.info - WFS source set to: FLAOWFS
Check pointing
6:34. SEEING:0.9"
Wed Sep 19 06:28:50.358 2012 aos.PresetAO.started -
PresetAO[794] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-8.92065e-07,ro2:-1.52018e-07,mag:15.1,cindex:2.24,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 06:28:50.409 2012 aos.info - WFS source set to: FLAOWFS
Taken measurements at close loop
Wed Sep 19 06:52:34.361 2012 aos.PresetAO.started -
PresetAO[810] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:9.73974e-07,ro2:-3.14417e-07,mag:15.1,cindex:2.24,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 06:52:34.412 2012 aos.info - WFS source set to: FLAOWFS
Taken measurements at open and close loop
Wed Sep 19 07:17:21.566 2012 aos.PresetAO.started -
PresetAO[832] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-3.48446e-06,ro2:-3.41914e-07,mag:15.7,cindex:1,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 07:17:21.616 2012 aos.info - WFS source set to: FLAOWFS
< Check Pointing>
Wed Sep 19 07:22:14.541 2012 aos.PresetAO.started -
PresetAO[835] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:7.25211e-05,ro2:9.9019e-05,mag:15.7,cindex:1,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 07:22:14.592 2012 aos.info - WFS source set to: FLAOWFS
7:33 SEEING:0.85"
Wed Sep 19 07:34:14.230 2012 aos.PresetAO.started -
PresetAO[839] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-2.32504e-07,ro2:1.62936e-07,mag:16.4,cindex:2.2,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 07:34:14.281 2012 aos.info - WFS source set to: FLAOWFS
bcu39 offline
Wed Sep 19 07:42:30.277 2012 aos.PresetAO.started -
PresetAO[842] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:1.4503e-06,ro2:5.13074e-08,mag:16.4,cindex:2.2,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 07:42:30.328 2012 aos.info - WFS source set to: FLAOWFS
Star Not found. Check Pointing.
Wed Sep 19 07:51:48.159 2012 aos.PresetAO.started -
PresetAO[848] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-1.25859e-06,ro2:6.40826e-07,mag:16.4,cindex:2.2,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 07:51:48.209 2012 aos.info - WFS source set to: FLAOWFS
7:58. SEEING:1:05"
Taken measurements at close loop
Wed Sep 19 08:21:13.124 2012 aos.PresetAO.started -
PresetAO[864] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-8.02758e-07,ro2:-1.53707e-06,mag:16.9,cindex:2.8,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 08:21:13.175 2012 aos.info - WFS source set to: FLAOWFS
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error
err = Resmsg[1] = RETRY: Star too faint for mode ACE-AO. Maximum magnitude is 17.5 (-5002) VALUE_OUT_OF_RANGE_ERROR
err = Resmsg[2] =
AcquireRef[865] failed - RETRY: Star too faint for mode ACE-AO. Maximum magnitude is 17.5 (-5002) VALUE_OUT_OF_RANGE_ERROR
Wed Sep 19 08:23:50.800 2012 aos.AcquireRef.failed -
AcquireRef[865] failed - RETRY: Star too faint for mode ACE-AO. Maximum magnitude is 17.5 (-5002) VALUE_OUT_OF_RANGE_ERROR
Due to the configuration setting, the system rejects stars fainter than mag.17.5. The configuration file will be changed to allow to go to magnitude fainter.
Taken measurements at close loop
8:35. SEEING:0.8"
8:54. SEEING:0.88"
Wed Sep 19 08:55:14.077 2012 aos.PresetAO.started -
PresetAO[889] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-6.35938e-07,ro2:5.35873e-07,mag:17.1,cindex:2.19,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 08:55:14.128 2012 aos.info - WFS source set to: FLAOWFS
Preset Cancel. Change to another a fainter star.
Wed Sep 19 08:59:32.788 2012 aos.PresetAO.complete -
PresetAO[891] complete - dx:nan dy:nan slNull: f1:Filter #0 f2:Filter #0 freq:100 mag:0 nBins:4 nModes:36 r0:0 snMode:0 strehl:0 ttMod:6
Taken measurements at close and open loop.
Wed Sep 19 09:24:58.938 2012 aos.PresetAO.started -
PresetAO[909] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:4.12011e-07,ro2:5.54899e-08,mag:17.7,cindex:2.2,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 09:24:58.989 2012 aos.info - WFS source set to: FLAOWFS
9:36. SEEING: 0.89"
Taken measurements at close loop.
Wed Sep 19 09:58:16.991 2012 aos.PresetAO.started -
PresetAO[928] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-8.42681e-07,ro2:2.95e-07,mag:18.5,cindex:3.13,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 09:58:17.041 2012 aos.info - WFS source set to: FLAOWFS
SKIP during the optimize gain. The star is mag. 17.3
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error
err = Resmsg[1] =
AcquireRef[929] complete - dx:-0.808 dy:-1.682 slNull: f1:Filter #0 f2:Filter #0 freq:100 mag:17.479 nBins:4 nModes:36 r0:0 snMode:0 strehl:0 ttMod:6
err = Resmsg[2] = Safe skip active - cannot proceed (-10045) IDL_SAFE_SKIP_ACTIVE
err = Resmsg[3] =
StartAO[930] failed - Safe skip active - cannot proceed (-10045) IDL_SAFE_SKIP_ACTIVE
PRESET RESENT. MAG:17.3
10:07. SEEING:0.68"
SKIP+RIP. During the optimize gain.
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error
err = Resmsg[1] =
AcquireRef[933] complete - dx:-0.851 dy:-1.65 slNull: f1:Filter #0 f2:Filter #0 freq:100 mag:17.472 nBins:4 nModes:36 r0:0 snMode:0 strehl:0 ttMod:6
err = Resmsg[2] = thWaitMsg error(-5001): Timeout error
err = Resmsg[3] =
StartAO[934] failed - thWaitMsg error(-5001): Timeout error
IDL>
SKIP during the optimize gain.
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error
err = Resmsg[1] =
AcquireRef[937] complete - dx:-0.858 dy:-1.619 slNull: f1:Filter #0 f2:Filter #0 freq:100 mag:17.402 nBins:4 nModes:36 r0:0 snMode:0 strehl:0 ttMod:6
err = Resmsg[2] = Safe skip active - cannot proceed (-10045) IDL_SAFE_SKIP_ACTIVE
err = Resmsg[3] =
StartAO[938] failed - Safe skip active - cannot proceed (-10045) IDL_SAFE_SKIP_ACTIVE
MANUAL CLOSE LOOP
Taken measurements at close and open loop.
10:29. SEEING:0.68"
SKIP+RIP. during the movements of the stages to center better the star in the detector. movements x:0.1 and y:0.1
Wed Sep 19 10:39:27.699 2012 aos.Stop.failed - Stop[951] failed - Command timeout - -13107: Arbitrator interface generic error
DL> iif_preset,ao112,mode=ace,gs=2,rs=0,pa=0,ND=4
# Target star: 0 AO112
03:01:33.09 +32:51:22.7 RADEC 2000.0 J2000 333.0 -119.0 14.40 R 2.70 B_V 1.65
# Guide star: 2
03:01:27.335 +32:52:30.24 RADEC 2000.0 J2000 2.0 -20.0 12.85 R 1.38 B_V 0.63
# AO Reference star: 0
03:01:33.09 +32:51:22.7 RADEC 2000.0 J2000 333.0 -119.0 18.40 R 2.70 B_V 1.65
preset, AO112, MODE=ACE, GS=2, RS=0
===> Error in irc_presetTelescope
err = Resmsg[0] =
PresetTelescope result status: Error err = Resmsg[1] = Stop[952] complete
err = Resmsg[2] = couldn't find a star in image
CHECK POINTING
Wed Sep 19 10:51:24.340 2012 aos.PresetAO.started -
PresetAO[956] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:8.18372e-07,ro2:-2.8032e-08,mag:18.4,cindex:2.7,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 10:51:24.390 2012 aos.info - WFS source set to: FLAOWFS
SKIP during the acquire ref. Really faint star.
Wed Sep 19 10:58:17.919 2012 aos.info - COILS DISABLED
Wed Sep 19 10:58:17.970 2012 aos.error - Stopping WFS...
Wed Sep 19 10:58:18.242 2012 aos.error - Recovering
AdSec failure
Wed Sep 19 10:58:18.375 2012 aos.warning - Stop rejected. Busy serving
AcquireRef [959]
Wed Sep 19 11:02:06.190 2012 aos.PresetAO.started -
PresetAO[961] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-9.56055e-07,ro2:-3.20777e-08,mag:18.4,cindex:2.7,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 11:02:06.241 2012 aos.info - WFS source set to: FLAOWFS
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error err = Resmsg[1] = err = Resmsg[2] =
AcquireRef[962] failed
SKIP DURING THE ACQUIRE OF THE REFERENCE STAR. THE CAMERA LENS DIDN'T CONVERGE. REALLY FAINT. MAG.17.8
TELESCOPE TO ZENITH and TO REMOVE THE ND FILTER
11:09 ASM_SX IN A CONTINOUS SKIP FRAME. AFTER RECOVER THE SHELL InformationTechnology FOUND A FULL DISK SPACE IN THE ADSECSX. THE SYSTEM STOP TILL WE HAVE SOME SPACE IN DISK.
Wed Sep 19 11:30:29.393 2012 aos.PresetAO.started -
PresetAO[965] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:-1.79502e-07,ro2:3.62908e-09,mag:11.07,cindex:2.61,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 11:30:29.444 2012 aos.info - WFS source set to: FLAOWFS
HIGH FLUX
Wed Sep 19 11:42:57.955 2012 aos.PresetAO.started -
PresetAO[975] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:3.07381e-06,ro2:2.34155e-08,mag:11.07,cindex:2.61,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 11:42:58.006 2012 aos.info - WFS source set to: FLAOWFS
Wed Sep 19 11:45:17.721 2012 aos.AcquireRef.failed -
AcquireRef[976] failed - RETRY: No stars found in TV frame (-20001) WFSARB_STAR_NOTFOUND
IDL> iif_runao
===> Error in irc_runao
err = Resmsg[0] =
RunAO result status: Error
err = Resmsg[1] = RETRY: No stars found in TV frame (-20001) WFSARB_STAR_NOTFOUND
err = Resmsg[2] =
AcquireRef[976] failed - RETRY: No stars found in TV frame (-20001) WFSARB_STAR_NOTFOUND
The star is about mag7 and the ccd47 didn't cente the star
RESENT THE RUNAO. THE SYSTEM CLOSE THE LOOP. BIN 1 990 HZ.
11:54. SEEING:0.85"
Taken measurements at close loop.
Wed Sep 19 12:08:46.849 2012 aos.PresetAO.started -
PresetAO[986] (aomode:ACE-AO,focalst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,so1:0,so2:0,ro1:1.05649e-06,ro2:4.88158e-08,mag:11.07,cindex:2.61,elev:nan,rotang:nan,gravang:nan,r0:0,skybrgt:nan,windspeed:nan,winddir:nan)
Wed Sep 19 12:08:46.900 2012 aos.info - WFS source set to: FLAOWFS
Taken measurements at close loop.
12:24. SEEING.0.92"
Stop 12:44
DATA SUMMARY
TrackNo |
RefStar |
Mag |
El |
Wind |
DIMM/OL |
Rec |
bin |
#mod |
freq |
gain |
mod |
nph |
AntiDrift |
SR |
filter |
exp |
#frames |
disturb |
SN |
notes |
20120919_053832 |
??? |
15.4 |
73 |
3 |
0.95 0.62 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
40 |
ON |
16.1 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_053901 |
??? |
15.3 |
73 |
2 |
0.98 0.70 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
42 |
ON |
12.6 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_053923 |
??? |
15.3 |
73 |
2 |
0.99 0.68 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
42 |
ON |
15.0 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_053946 |
??? |
15.3 |
73 |
3 |
1.05 0.68 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
43 |
ON |
16.5 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_054048 |
??? |
15.3 |
73 |
2 |
0.96 0.59 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
43 |
ON |
18.6 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_054110 |
??? |
15.3 |
73 |
2 |
0.89 0.64 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
43 |
ON |
14.8 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_054132 |
??? |
15.3 |
73 |
2 |
0.89 0.63 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
43 |
ON |
18.7 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_054951 |
??? |
15.3 |
71 |
3 |
0.86 0.55 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
41 |
ON |
22.5 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_055013 |
??? |
15.4 |
71 |
3 |
0.93 0.56 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
41 |
ON |
23.7 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_055035 |
??? |
15.4 |
71 |
3 |
0.94 0.68 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
41 |
ON |
19.5 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_055137 |
??? |
15.4 |
71 |
2 |
0.88 0.56 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
41 |
ON |
23.2 |
H |
200 |
50 |
ONSKY |
zero.fits |
- Pupils not on target |
20120919_055159 |
??? |
15.4 |
71 |
2 |
0.90 0.58 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
41 |
ON |
19.6 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_055221 |
??? |
15.4 |
71 |
2 |
0.88 0.56 |
223524 |
4 |
36 |
200 |
1.8 0.5 -1.0 |
6 |
41 |
ON |
23.0 |
H |
200 |
50 |
ONSKY |
zero.fits |
|
20120919_055544 |
??? |
15.3 |
70 |
2 |
0.93 0.38 |
223524 |
4 |
36 |
200 |
0.1 -1.0 -1.0 |
6 |
42 |
ON |
1.0 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_055606 |
??? |
15.3 |
70 |
2 |
0.91 0.37 |
223524 |
4 |
36 |
200 |
0.1 -1.0 -1.0 |
6 |
43 |
ON |
0.8 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_055628 |
??? |
15.3 |
70 |
2 |
0.92 0.40 |
223524 |
4 |
36 |
200 |
0.1 -1.0 -1.0 |
6 |
43 |
ON |
0.8 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Background correction failed |
20120919_055729 |
??? |
15.3 |
70 |
2 |
1.00 0.39 |
223524 |
4 |
36 |
200 |
0.1 -1.0 -1.0 |
6 |
41 |
ON |
0.5 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_055751 |
??? |
15.3 |
70 |
1 |
1.00 0.35 |
223524 |
4 |
36 |
200 |
0.1 -1.0 -1.0 |
6 |
42 |
ON |
0.8 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_055813 |
??? |
15.3 |
70 |
1 |
0.96 0.36 |
223524 |
4 |
36 |
200 |
0.1 -1.0 -1.0 |
6 |
41 |
ON |
0.9 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Background correction failed |
20120919_061032 |
??? |
15.3 |
67 |
1 |
0.94 0.65 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
40 |
ON |
15.1 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061054 |
??? |
15.3 |
67 |
2 |
0.93 0.70 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
41 |
ON |
16.5 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061116 |
??? |
15.3 |
67 |
1 |
0.93 0.64 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
41 |
ON |
17.7 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061216 |
??? |
15.3 |
67 |
2 |
0.90 0.59 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
41 |
ON |
18.6 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061238 |
??? |
15.3 |
67 |
2 |
0.94 0.61 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
40 |
ON |
15.6 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061300 |
??? |
15.3 |
67 |
2 |
0.99 0.71 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
40 |
ON |
13.5 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061357 |
??? |
15.3 |
67 |
2 |
0.96 0.64 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
41 |
ON |
16.4 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061419 |
??? |
15.3 |
67 |
2 |
0.97 0.60 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
40 |
ON |
16.3 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061441 |
??? |
15.3 |
66 |
2 |
1.03 0.60 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
41 |
ON |
16.3 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061540 |
??? |
15.3 |
66 |
2 |
0.93 0.49 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
41 |
ON |
24.7 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061602 |
??? |
15.3 |
66 |
2 |
0.89 0.51 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
44 |
ON |
19.8 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_061624 |
??? |
15.3 |
66 |
2 |
0.82 0.52 |
200610 |
4 |
36 |
200 |
2.0 0.9 0.5 |
6 |
40 |
ON |
21.0 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_063651 |
??? |
16.0 |
82 |
2 |
NaN 0.64 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
46 |
ON |
13.0 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_063734 |
??? |
16.0 |
82 |
2 |
NaN 0.49 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
45 |
ON |
12.8 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_063816 |
??? |
15.9 |
82 |
3 |
NaN 0.54 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
46 |
ON |
16.2 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_063937 |
??? |
16.0 |
82 |
2 |
NaN 0.50 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
45 |
ON |
16.7 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_064019 |
??? |
16.0 |
82 |
2 |
NaN 0.44 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
45 |
ON |
18.8 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_064101 |
??? |
16.0 |
82 |
2 |
NaN 0.52 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
45 |
ON |
18.6 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_064153 |
??? |
16.0 |
82 |
2 |
0.82 0.47 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
44 |
ON |
17.4 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_064235 |
??? |
16.0 |
82 |
2 |
0.83 0.46 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
44 |
ON |
18.8 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_064317 |
??? |
15.9 |
82 |
2 |
0.88 0.49 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
48 |
ON |
18.3 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_064439 |
??? |
15.9 |
82 |
2 |
0.86 0.44 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
47 |
ON |
19.3 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_064521 |
??? |
16.0 |
82 |
3 |
0.85 0.45 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
45 |
ON |
18.2 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_064603 |
??? |
16.0 |
82 |
2 |
NaN 0.41 |
200610 |
4 |
36 |
100 |
2.9 1.6 0.9 |
6 |
45 |
ON |
18.1 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_065937 |
??? |
16.0 |
82 |
2 |
NaN 0.59 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
45 |
ON |
11.8 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070019 |
??? |
16.0 |
81 |
2 |
0.85 0.53 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
44 |
ON |
18.7 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070101 |
??? |
16.0 |
81 |
3 |
0.77 0.51 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
44 |
ON |
18.5 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070221 |
??? |
16.0 |
81 |
2 |
0.93 0.52 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
45 |
ON |
20.2 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070303 |
??? |
16.0 |
81 |
2 |
0.90 0.53 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
45 |
ON |
18.4 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070348 |
??? |
16.0 |
81 |
3 |
0.85 0.49 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
46 |
ON |
21.1 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070523 |
??? |
16.0 |
81 |
2 |
0.00 0.52 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
44 |
ON |
19.1 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070605 |
??? |
16.0 |
81 |
2 |
0.85 0.46 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
43 |
ON |
23.9 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070647 |
??? |
16.0 |
81 |
1 |
0.85 0.46 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
45 |
ON |
21.1 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070807 |
??? |
16.0 |
81 |
2 |
0.86 0.48 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
46 |
ON |
20.1 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070849 |
??? |
16.0 |
81 |
1 |
0.85 0.59 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
44 |
ON |
20.3 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_070930 |
??? |
16.0 |
81 |
2 |
0.92 0.50 |
223524 |
4 |
36 |
100 |
2.7 0.5 0.9 |
6 |
44 |
ON |
20.6 |
H |
400 |
25 |
ONSKY |
zero.fits |
|
20120919_071144 |
??? |
16.0 |
80 |
2 |
0.00 0.40 |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
44 |
ON |
3.3 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_071226 |
??? |
16.0 |
80 |
2 |
0.00 0.40 |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
43 |
ON |
2.0 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_071309 |
??? |
16.0 |
80 |
2 |
NaN 0.38 |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
43 |
ON |
1.2 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_071428 |
??? |
16.0 |
80 |
2 |
0.93 0.34 |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
44 |
ON |
0.9 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Background correction failed |
20120919_071509 |
??? |
16.1 |
80 |
2 |
0.92 0.34 |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
43 |
ON |
0.8 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Background correction failed |
20120919_071551 |
??? |
16.1 |
80 |
2 |
0.87 0.36 |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
42 |
ON |
0.7 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_073216 |
??? |
16.0 |
76 |
2 |
0.76 0.49 |
223524 |
4 |
36 |
100 |
2.4 0.9 0.5 |
6 |
46 |
ON |
21.6 |
H |
500 |
20 |
ONSKY |
zero.fits |
- Warning: Selected IRTC dark +2547h old! |
20120919_073258 |
??? |
16.0 |
76 |
2 |
0.87 0.49 |
223524 |
4 |
36 |
100 |
2.4 0.9 0.5 |
6 |
47 |
ON |
19.6 |
H |
500 |
20 |
ONSKY |
zero.fits |
- Warning: Selected IRTC dark +2547h old! |
20120919_073340 |
??? |
NaN |
76 |
3 |
0.88 -1.00 |
223524 |
4 |
36 |
100 |
2.4 0.9 0.5 |
6 |
-1 |
-1 |
20.5 |
H |
500 |
20 |
ONSKY |
NO |
|
20120919_080310 |
??? |
16.4 |
84 |
3 |
1.05 0.67 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
33 |
ON |
16.8 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_080352 |
??? |
16.3 |
84 |
3 |
1.05 0.67 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
33 |
ON |
13.3 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_080434 |
??? |
16.3 |
84 |
3 |
0.98 0.65 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
33 |
ON |
15.7 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_080553 |
??? |
16.3 |
84 |
2 |
1.17 0.62 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
33 |
ON |
14.2 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_080635 |
??? |
16.3 |
85 |
3 |
0.99 0.64 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
33 |
ON |
15.4 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_080717 |
??? |
16.3 |
85 |
3 |
1.02 0.60 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
34 |
ON |
16.1 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_080811 |
??? |
16.3 |
85 |
3 |
1.00 0.90 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
34 |
ON |
15.2 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_080853 |
??? |
16.3 |
85 |
3 |
NaN 0.69 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
33 |
ON |
16.3 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_080936 |
??? |
16.4 |
85 |
3 |
0.93 0.66 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
31 |
ON |
15.2 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_081056 |
??? |
16.4 |
85 |
3 |
0.31 0.64 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
31 |
ON |
14.3 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_081138 |
??? |
16.4 |
86 |
3 |
0.93 0.81 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
31 |
ON |
12.4 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_081220 |
??? |
16.4 |
86 |
2 |
1.01 0.64 |
223524 |
4 |
36 |
100 |
1.8 1.2 0.5 |
6 |
31 |
ON |
12.6 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_083458 |
??? |
16.5 |
75 |
3 |
0.82 0.60 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
29 |
ON |
-1.0 |
? |
-1 |
-1 |
ONSKY |
zero.fits |
|
20120919_083540 |
??? |
16.5 |
76 |
2 |
0.85 0.59 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
29 |
ON |
21.2 |
H |
700 |
13 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_083622 |
??? |
16.5 |
76 |
3 |
0.86 0.63 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
29 |
ON |
21.3 |
H |
700 |
13 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_083740 |
??? |
16.5 |
76 |
3 |
0.81 0.58 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
28 |
ON |
22.2 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_083822 |
??? |
16.5 |
76 |
2 |
0.87 0.61 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
28 |
ON |
23.0 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_083905 |
??? |
16.5 |
76 |
3 |
0.85 0.59 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
28 |
ON |
21.9 |
H |
700 |
13 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_084042 |
??? |
16.5 |
76 |
3 |
0.85 0.55 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
29 |
ON |
25.4 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_084125 |
??? |
16.5 |
76 |
2 |
0.88 0.54 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
29 |
ON |
19.7 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_084210 |
??? |
16.5 |
76 |
3 |
0.86 0.55 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
29 |
ON |
22.0 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_084328 |
??? |
16.5 |
77 |
3 |
NaN 0.56 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
28 |
ON |
21.5 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_084410 |
??? |
16.5 |
77 |
3 |
NaN 0.99 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
28 |
ON |
20.0 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_084452 |
??? |
16.5 |
77 |
3 |
NaN 0.59 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
28 |
ON |
23.6 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_084817 |
??? |
16.5 |
77 |
3 |
NaN 0.54 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
30 |
ON |
17.8 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_084859 |
??? |
16.5 |
77 |
3 |
0.86 0.52 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
29 |
ON |
24.5 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_084941 |
??? |
16.5 |
77 |
3 |
0.85 1.47 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
29 |
ON |
24.0 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_085101 |
??? |
16.6 |
77 |
3 |
0.82 0.96 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
27 |
ON |
21.9 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_085143 |
??? |
16.6 |
77 |
2 |
0.82 0.58 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
27 |
ON |
20.4 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_085225 |
??? |
16.6 |
78 |
3 |
0.86 0.56 |
223524 |
4 |
36 |
100 |
1.6 0.5 -1.0 |
6 |
27 |
ON |
21.7 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_090909 |
??? |
16.9 |
84 |
4 |
0.82 0.69 |
223524 |
4 |
36 |
100 |
0.6 0.5 -1.0 |
6 |
21 |
ON |
26.9 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_090951 |
??? |
16.9 |
84 |
3 |
0.88 1.46 |
223524 |
4 |
36 |
100 |
0.6 0.5 -1.0 |
6 |
19 |
ON |
22.4 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_091033 |
??? |
16.8 |
84 |
5 |
0.80 0.66 |
223524 |
4 |
36 |
100 |
0.6 0.5 -1.0 |
6 |
22 |
ON |
22.4 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_091153 |
??? |
16.8 |
84 |
3 |
0.86 0.67 |
223524 |
4 |
36 |
100 |
0.6 0.5 -1.0 |
6 |
21 |
ON |
22.5 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_091236 |
??? |
16.7 |
84 |
3 |
0.83 1.08 |
223524 |
4 |
36 |
100 |
0.6 0.5 -1.0 |
6 |
23 |
ON |
20.7 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_091318 |
??? |
16.8 |
85 |
3 |
0.86 0.71 |
223524 |
4 |
36 |
100 |
0.6 0.5 -1.0 |
6 |
21 |
ON |
22.9 |
H |
700 |
13 |
ONSKY |
zero.fits |
|
20120919_091653 |
??? |
16.9 |
85 |
6 |
0.86 0.66 |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
20 |
ON |
1.9 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_091735 |
??? |
17.0 |
85 |
4 |
0.76 0.72 |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
18 |
ON |
1.7 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Pupils not on target |
20120919_091817 |
??? |
17.2 |
85 |
3 |
0.84 0.90 |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
15 |
ON |
2.0 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_092021 |
??? |
17.7 |
86 |
6 |
0.83 *** |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
10 |
ON |
32.8 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Pupils not on target |
20120919_092104 |
??? |
18.2 |
86 |
5 |
0.81 *** |
223524 |
4 |
36 |
100 |
0.1 -1.0 -1.0 |
6 |
6 |
ON |
34.3 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Pupils not on target - SR may be in error |
20120919_093455 |
??? |
17.0 |
78 |
3 |
0.92 0.85 |
223524 |
4 |
36 |
100 |
1.2 0.5 0.9 |
6 |
17 |
ON |
3.8 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_093537 |
??? |
17.1 |
78 |
3 |
0.85 0.89 |
223524 |
4 |
36 |
100 |
1.2 0.5 0.9 |
6 |
17 |
ON |
8.4 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_093619 |
??? |
17.1 |
78 |
3 |
0.90 0.91 |
223524 |
4 |
36 |
100 |
1.2 0.5 0.9 |
6 |
17 |
ON |
4.1 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_093740 |
??? |
17.1 |
78 |
3 |
0.77 0.88 |
223524 |
4 |
36 |
100 |
1.2 0.5 0.9 |
6 |
17 |
ON |
5.4 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_093822 |
??? |
17.1 |
79 |
3 |
NaN 0.86 |
223524 |
4 |
36 |
100 |
1.2 0.5 0.9 |
6 |
17 |
ON |
6.7 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Background correction failed |
20120919_093905 |
??? |
17.0 |
79 |
3 |
NaN 0.82 |
223524 |
4 |
36 |
100 |
1.2 0.5 0.9 |
6 |
18 |
ON |
9.6 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_094940 |
??? |
17.2 |
81 |
3 |
0.89 0.86 |
223524 |
4 |
36 |
100 |
0.3 0.5 -1.0 |
6 |
15 |
ON |
5.6 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_095022 |
??? |
17.1 |
81 |
3 |
0.85 1.38 |
223524 |
4 |
36 |
100 |
0.3 0.5 -1.0 |
6 |
17 |
ON |
5.5 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_095104 |
??? |
17.0 |
81 |
3 |
0.82 0.89 |
223524 |
4 |
36 |
100 |
0.3 0.5 -1.0 |
6 |
17 |
ON |
6.1 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_095223 |
??? |
17.1 |
82 |
3 |
0.79 0.85 |
223524 |
4 |
36 |
100 |
0.3 0.5 -1.0 |
6 |
17 |
ON |
11.7 |
H |
2000 |
5 |
ONSKY |
zero.fits |
|
20120919_095305 |
??? |
17.3 |
82 |
3 |
0.86 *** |
223524 |
4 |
36 |
100 |
0.3 0.5 -1.0 |
6 |
14 |
ON |
6.7 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Pupils not on target |
20120919_095348 |
??? |
-NaN |
75 |
3 |
0.62 *** |
223524 |
4 |
36 |
100 |
0.0 -1.0 -1.0 |
6 |
-4 |
ON |
330.3 |
H |
2000 |
5 |
ONSKY |
zero.fits |
- Pupils not on target - SR may be in error |
20120919_102654 |
??? |
17.6 |
64 |
3 |
0.64 0.81 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
10 |
ON |
13.7 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_102736 |
??? |
17.4 |
64 |
3 |
0.64 0.74 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
13 |
ON |
18.3 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_102818 |
??? |
17.3 |
64 |
3 |
0.64 0.71 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
13 |
ON |
13.7 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_102937 |
??? |
17.3 |
64 |
3 |
0.68 0.71 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
13 |
ON |
21.4 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_103019 |
??? |
17.6 |
63 |
3 |
0.62 0.77 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
10 |
ON |
16.4 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_103101 |
??? |
17.4 |
63 |
4 |
0.64 0.89 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
13 |
ON |
18.0 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- Pupils not on target - SR may be in error |
20120919_103310 |
??? |
17.5 |
63 |
3 |
0.62 0.84 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
11 |
ON |
12.7 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- Pupils not on target - SR may be in error |
20120919_103354 |
??? |
17.4 |
63 |
3 |
0.66 0.71 |
223524 |
4 |
36 |
100 |
0.2 -1.0 -1.0 |
6 |
13 |
ON |
6.4 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_103436 |
??? |
17.1 |
63 |
4 |
0.70 1.35 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
16 |
ON |
17.9 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- Pupils not on target - SR may be in error |
20120919_103554 |
??? |
17.4 |
63 |
3 |
0.66 1.15 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
13 |
ON |
17.8 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_103636 |
??? |
17.4 |
63 |
3 |
0.67 1.03 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
12 |
ON |
14.0 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_103718 |
??? |
17.0 |
63 |
3 |
0.74 0.97 |
223524 |
4 |
36 |
100 |
0.3 0.2 -1.0 |
6 |
17 |
ON |
14.1 |
H |
1000 |
10 |
ONSKY |
zero.fits |
- SR may be in error |
20120919_115757 |
??? |
7.4 |
71 |
2 |
0.92 0.39 |
201326 |
1 |
400 |
990 |
0.9 0.6 0.6 |
3 |
642 |
ON |
88.2 |
H |
2 |
1000 |
ONSKY |
zero.fits |
|
20120919_115913 |
??? |
7.4 |
71 |
3 |
0.84 0.33 |
201326 |
1 |
400 |
990 |
0.9 0.6 0.6 |
3 |
643 |
ON |
89.2 |
H |
2 |
1000 |
ONSKY |
zero.fits |
|
20120919_120030 |
??? |
7.4 |
71 |
4 |
0.82 0.32 |
201326 |
1 |
400 |
990 |
0.9 0.6 0.6 |
3 |
645 |
ON |
89.5 |
H |
2 |
1000 |
ONSKY |
zero.fits |
|
20120919_120327 |
??? |
7.4 |
72 |
3 |
0.84 0.33 |
201326 |
1 |
400 |
990 |
0.9 0.6 0.6 |
3 |
642 |
ON |
89.0 |
H |
2 |
1000 |
ONSKY |
zero.fits |
|
20120919_120444 |
??? |
7.4 |
72 |
3 |
0.94 0.34 |
201326 |
1 |
400 |
990 |
0.9 0.6 0.6 |
3 |
645 |
ON |
86.6 |
H |
2 |
1000 |
ONSKY |
zero.fits |
|
20120919_120600 |
??? |
7.4 |
73 |
3 |
0.79 0.33 |
201326 |
1 |
400 |
990 |
0.9 0.6 0.6 |
3 |
644 |
ON |
88.7 |
H |
2 |
1000 |
ONSKY |
zero.fits |
|
20120919_121939 |
??? |
7.4 |
75 |
3 |
0.85 0.34 |
182144 |
1 |
500 |
990 |
0.8 0.6 0.4 |
3 |
653 |
ON |
60.2 |
J |
4 |
1000 |
ONSKY |
zero.fits |
- Pupils mismatch |
20120919_122059 |
??? |
7.4 |
76 |
4 |
0.81 0.31 |
182144 |
1 |
500 |
990 |
0.8 0.6 0.4 |
3 |
651 |
ON |
60.7 |
J |
4 |
1000 |
ONSKY |
zero.fits |
- Pupils mismatch |
20120919_122216 |
??? |
7.4 |
76 |
3 |
0.89 0.41 |
182144 |
1 |
500 |
990 |
0.8 0.6 0.4 |
3 |
651 |
ON |
53.3 |
J |
4 |
1000 |
ONSKY |
zero.fits |
- Pupils mismatch |
20120919_122511 |
??? |
7.4 |
77 |
3 |
0.88 0.41 |
182144 |
1 |
500 |
990 |
0.8 0.6 0.4 |
3 |
649 |
ON |
53.2 |
J |
4 |
1000 |
ONSKY |
zero.fits |
- Pupils mismatch |
20120919_122627 |
??? |
7.4 |
77 |
2 |
0.89 0.44 |
182144 |
1 |
500 |
990 |
0.8 0.6 0.4 |
3 |
650 |
ON |
47.0 |
J |
4 |
1000 |
ONSKY |
zero.fits |
- Pupils mismatch |
SAMPLE PSF OF BIN 1
- Strehl Ratio in H band_90%:
- Strehl Ratio in J band_60%:
--
JuanGuerra - 18 Sep 2012