LIST BEFORE TO GO ON SKY.

Acronymous:

  • Adsec: Adaptive Secondary.
  • WFS. Wavefrontsensor. SOULSX/DX.
  • ACALU: Argos Calibration Unit.
  • SX: Left side.
  • DX: Right Side.

ADSEC ADSEC FLAT NUMBER OF MODES APPLIED

adsec log file:

grep " Maximum mode index used to apply flat" adsec.L.20*

Example of log line:

" MAIN > Maximum mode index used to apply flat -> # 638 "

ADSEC ADSEC SHAPE
  • VISUALIZE SHAPE.
  • open MIRROR CTRL GUI.
  • NO BUMPS.

  • TYPICAL SHAPE FOR SX
ADSEC_FLAT_SHAPE.png
WFS POWER ON
  • w_stop_all
  • w_start_all
  • ALL SYSTEM UP
WFSoperate.png
WFS OCAM GOOD BIAS.

OCAM-GOOD_BAD.png

ADSEC/WFS FASTLINK COMMUNICATION FASTLINK CHECK. [AOeng@sxadsec opsScriptsAdSec]$ ./fastlink.sh

Usage: ./fastlink.sh [option]
-f FLAO/LUCI
-l LBTI
-n Linc Nirvana

[AOeng@sxadsec opsScriptsAdSec]$ ./fastlink.sh -f

Error in check_routines() procedure: there is at least one procedure that NOT COMPILES.
ERROR ON COMPILING ROUTINES: init_dsp_map

ADSEC/WFS/AOS CLOSE LOOP
  • USING ACALU
  • AOS: PRESET 10 MAG.
    • inject atmo turbulence.
    • dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq750.00_sd3892.fits
  • FLAO GUI- intervention mode-
  • CLOSE LOOP
  • OPTIMIZE GAIN
  • GOPT.
  • Open Aux loop Gui.
    • Modal plot tab
    • STREHL FROM WFS.~ 82%
  • Take Optical loop Data+ Luci image
    • MODAL PLOT.
    • STREHL FROM WFS ~82%
    • LUCI IMAGE PSF.
    • Pause/Light off/
    • Luci Dark from WFs controil GUI.
    • light On/Resume
  • Take Optical loop Data+ Luci image
    • MODAL PLOT.
    • STREHL FROM WFS ~82%
    • LUCI IMAGE PSF.
    • strehl from Luci image.~70%
  • DAYTIME OSCO "DISCO". Not implemented yet.

BIN1_NODIST-MANUALGAIN_20211118.png bin1-gopt_20211119.png
WFS PUPIL POSITION bin1_check20211129.png
-- JuanCarlosGuerra - 05 Jan 2022
I Attachment Action Size Date Who Comment
ADSEC_FLAT_SHAPE.pngpng ADSEC_FLAT_SHAPE.png manage 152 K 05 Jan 2022 - 20:06 JuanCarlosGuerra  
BIN1_NODIST-MANUALGAIN_20211118.pngpng BIN1_NODIST-MANUALGAIN_20211118.png manage 46 K 05 Jan 2022 - 21:37 JuanCarlosGuerra  
OCAM-GOOD_BAD.pngpng OCAM-GOOD_BAD.png manage 271 K 05 Jan 2022 - 21:30 JuanCarlosGuerra  
OCAM_GOOD.pngpng OCAM_GOOD.png manage 109 K 05 Jan 2022 - 21:23 JuanCarlosGuerra  
Ocam_Bad_frame.pngpng Ocam_Bad_frame.png manage 85 K 05 Jan 2022 - 21:13 JuanCarlosGuerra  
WFSoperate.pngpng WFSoperate.png manage 163 K 05 Jan 2022 - 21:07 JuanCarlosGuerra  
bin1-gopt_20211119.pngpng bin1-gopt_20211119.png manage 19 K 06 Jan 2022 - 18:54 JuanCarlosGuerra  
bin1_check20211129.pngpng bin1_check20211129.png manage 60 K 06 Jan 2022 - 18:53 JuanCarlosGuerra  
Topic revision: r9 - 22 Dec 2022, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback