Creation of Pathfinder AdSec flat with Conic Correction

Preparation of AdSec

  • Start AdSec Control GUI
    • Disable Astigmatism LUT application
  • Start AdSec Mirror GUI
    • Click Take Offset on Position and Current panels
    • Click Offset on on Position and Current panels
    • See PF Flat minus Offset image

Application of + Z11 to AdSec

  • Start AdSec IDL terminal
  • AdOpt> a = fltarr(22)
  • AdOpt>a[10]=549.8e-9
  • AdOpt>print, fsm_correct_modes(a, /PASSED)
  • See PF plus Z11=549.8 nm image

Save current AdSec flat that include Z11 correction

  • Click folder button in Save Shape row
    • A chooser menu pops up
    • See PF Choose Flat File image
    • Enter in Location entry field "20131109_Pathfinder_plus_Z11_549.8.sav"
    • Click OK
  • Click Save button in Save Shape row
  • Flat file is "/towerdata/meas/adsec_calib/flat/Applicable/20131109_Pathfinder_plus_Z11_549.8.sav"

Application of - Z11 to AdSec

  • At the same IDL terminal as above
  • AdOpt>a[10]=-549.8e-9
  • AdOpt>print, fsm_correct_modes(a, /PASSED)
  • See PF minus Z11=549.8 nm image

Save current AdSec flat that include Z11 correction

  • Click folder button in Save Shape row
    • A chooser menu pops up
    • See PF Choose Flat File image
    • Enter in Location entry field "20131109_Pathfinder_minus_Z11_549.8.sav"
    • Click OK
  • Click Save button in Save Shape row
  • Flat file is "/towerdata/meas/adsec_calib/flat/Applicable/20131109_Pathfinder_minus_Z11_549.8.sav"

Apply AdSec flat that includes Z11 correction

  • Click folder button in Load Shape row of AdSec Control GUI
    • A chooser menu pops up
    • Click file "20131109_Pathfinder_plus(or minus)_Z11_549.8.sav"
    • Click OK
  • Click Load button in Load Shape row
  • See PF Flat with Conic applied image

  • PF Flat minus offset:
    PF_flat_minus_offset.jpg

  • PF Flat plus Z11=549.8 nm:
    PF_plus_Z11_549.8.jpg

  • PF Flat minus Z11=-549.8 nm:
    PF_minus_Z11_549.8.png

  • PF Choose Flat File:
    PF_Flat_Chooser.jpg

  • PF Flat with Conic applied:
    PF_Conic_Applied.jpg

-- DougMiller - 09 Nov 2013
I Attachment Action Size Date Who Comment
PF_Conic_Applied.jpgjpg PF_Conic_Applied.jpg manage 271 K 09 Nov 2013 - 21:06 DougMiller PF Flat with Conic applied
PF_Flat_Chooser.jpgjpg PF_Flat_Chooser.jpg manage 509 K 09 Nov 2013 - 20:53 DougMiller PF Choose Flat File
PF_flat_minus_offset.jpgjpg PF_flat_minus_offset.jpg manage 287 K 09 Nov 2013 - 20:50 DougMiller PF Flat minus offset
PF_minus_Z11_549.8.pngpng PF_minus_Z11_549.8.png manage 171 K 09 Nov 2013 - 20:52 DougMiller PF Flat minus Z11=-549.8 nm
PF_plus_Z11_549.8.jpgjpg PF_plus_Z11_549.8.jpg manage 302 K 09 Nov 2013 - 20:51 DougMiller PF Flat minus Z11=549.8 nm
Topic revision: r1 - 09 Nov 2013, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback