WFS.

SOULSX.

  • OPEN THE LOOP.
    • From WFs control GUI open.
    • From AOS StopAO
  • Power off.
    • The power off will start with the warm up of the OCAM.

  • Manual power off.
  • SET SAFE OCAM.
    • Wait until tempearure reaches Ambient. The Ocam at ambient.
  • Power off

SOULDX.

  • SoulDX software has not been update (last check 20220111). The sequence to power off requires first to check for the cube translator and rotator.
    • The cube rotator and traslator has to be parked.
    • Set OCAM to ambient.
    • Power off.

  • setSAfe.png

ADSEC - FLAO GUI -.

  • CANCEL PRESET FROM FLAOGUI.
    FlaoGui2.png
  • REST THE SHELL.
  • DISABLE RR MODE.
    From the Adsec terminal type: >disable_rr_mode.

PSF GUI

  • ENABLE PSF LOOK UP TABLES AND TEMPERATURE COMPENSATION.

MCS GUI

  • IF ROTATOR USED SET TO DEFAULT PARK POSITION :
  • LFBG 341
  • RFBG 210
  • SPEED CHANGED TO 5deg/sec.

CHAMBER LIGHTS ON.

ARGOS CALIB UNIT.

  • PARK ARGOS CALIB UNIT.
  • TURN OFF THE LIGHT.

-- JuanCarlosGuerra - 10 Mar 2021