WFS Startup

Fri Nov 28 09:44:50.684 2014 aos.WfsOn.started - WfsOn [249] (wfs:FLAOWFS)

Fri Nov 28 09:44:50.712 2014 aos.info - WFS: starting power on sequence

Fri Nov 28 09:48:40.720 2014 aos.WfsOn.complete - WfsOn [249]

ADC wheel 2 is not stopping.It is an Operating stage all the time

9:49 DIMM 0.88"

9:56 DIMM: 0.76"

Fri Nov 28 10:40:20.909 2014 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[253] (aomode:TTM-

AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:-21.9697[-21.9697],roy:-8.29221,mag:15.43,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

The preset failed, but the WFS shows AOPrepared.

ri Nov 28 10:42:30.962 2014 aos.PresetAO{FLAO}.failed - PresetAO {FLAO}[253] - Command timeout - -13107: Arbitrator interface generic error

The CCD47 took long to equalize the quadrant and this it could be the cause of the failure of the preset.

Fri Nov 28 10:41:30.415 2014 aos.info - WFS: acquiring CCD darks

Fri Nov 28 10:41:42.491 2014 aos.warning - Variable iif.DIMM.seeing expired

Fri Nov 28 10:41:43.486 2014 aos.warning - Variable iif.DIMM.seeing expired

Fri Nov 28 10:41:43.501 2014 aos.warning - Variable iif.DIMM.seeing expired

Fri Nov 28 10:42:30.962 2014 aos.PresetAO{FLAO}.failed - PresetAO {FLAO}[253] - Command timeout - -13107: Arbitrator interface generic error

Fri Nov 28 10:42:34.084 2014 aos.info - WFS: preset done

IDL TERMINAL

> Opne IDL session from the AOeng account@obs3_..

>iif_register, LEFT

Fri Nov 28 10:44:57.439 2014 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[255] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:-21.9631[-21.9631],roy:-8.29145,mag:15.43,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

WfsArbitrator_2333 |DEB| 268|2014-11-28 10:47:39.199542| MAIN > arbFunc doAcquireRef (, 'FLAOWFS', 'LUCIFER')
WfsArbitrator_2333 |INF| 269|2014-11-28 10:47:39.201523| MAIN > source_acq.acquire(): starting source acquisition
WfsArbitrator_2333 |ERR| 270|2014-11-28 10:47:47.271960| MAIN > RETRY: No stars found in TV frame (-20001) WFSARB_STAR_NOTFOUND
WfsArbitrator_2333 |ERR| 271|2014-11-28 10:47:47.271993| MAIN > Traceback (most recent call last):
WfsArbitrator_2333 |ERR| 272|2014-11-28 10:47:47.272013| MAIN > File "/home/aoacct/AO/current/lib/python/AdOpt/WfsArbScripts.py", line 2508, in arbFunc
WfsArbitrator_2333 |ERR| 273|2014-11-28 10:47:47.272033| MAIN > errstr = func(*args)
WfsArbitrator_2333 |ERR| 274|2014-11-28 10:47:47.272051| MAIN > File "/home/aoacct/AO/current/lib/python/AdOpt/WfsArbScripts.py", line 1309, in doAcquireRef
WfsArbitrator_2333 |ERR| 275|2014-11-28 10:47:47.272070| MAIN > acqRet = source_acq.acquire(app, TVavg = app._acquireRefSettings.TVavg, TVflux = app._acquireRefSettings.TVflux, fwhm = app._acquireRefSettings.seeing)
WfsArbitrator_2333 |ERR| 276|2014-11-28 10:47:47.272089| MAIN > File "/home/aoacct/AO/current/lib/python/AdOpt/source_acq.py", line 235, in acquire
WfsArbitrator_2333 |ERR| 277|2014-11-28 10:47:47.272108| MAIN > raise AORetryException ( code = AOErrCode ['WFSARB_STAR_NOTFOUND'], errstr = errstr)
WfsArbitrator_2333 |ERR| 278|2014-11-28 10:47:47.272126| MAIN > AORetryException: No stars found in TV frame - code: -20001 WFSARB_STAR_NOTFOUND

ON-AXIS STAR

Check

Fri Nov 28 10:50:40.294 2014 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[258] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0.0633119[0.0633119],roy:0.20564,mag:11.63,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

The Star is not in the CCD47 field.

TO adjusted the pointing.

Fri Nov 28 11:03:22.663 2014 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[264] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:-21.9364[-21.9364],roy:-8.28784,mag:15.43,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

Move ADC wheel 2 to 130

Fri Nov 28 11:04:24.671 2014 aos.PresetAO{FLAO}.complete - PresetAO {FLAO}[264] - dx:nan dy:nan slNull: f1:Filter #0 f2:Filter #0 freq:200 mag:0 nBins:4 nModes:10 r0:0 snMode:0 strehl:0 ttMod:6

Fri Nov 28 11:10:21.677 2014 aos.AcquireRef{FLAO}.started - AcquireRef {FLAO}[265] Mode: Move stages, Automatic acquire

CCD39 VIEWER NOT LIVE, THE Adsec is not receiving frames.

slopecompctrl.L |ERR| 157|2014-11-28 11:17:32.572649| MAIN > Error in TestComm (): [AOException] Timeout error in
sendMultiBcuCommand() (code -5001) Timeout error File bcucommand.cpp line 16


Fri Nov 28 11:17:12.518 2014 aos.WfsOff.started - WfsOff [266] (wfs:FLAOWFS)

w_stop/start

Fri Nov 28 11:19:01.118 2014 aos.WfsOn.started - WfsOn [267] (wfs:FLAOWFS)

ADC Wheel 2 is stopped after the power on

Fri Nov 28 11:26:08.249 2014 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[269] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:-21.9014[-21.9014],roy:-8.28074,mag:15.43,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

Fri Nov 28 11:31:14.243 2014 aos.info - WFS: star found, estimated magnitude: 16.33

Fri Nov 28 11:31:48.554 2014 aos.AcquireRef{FLAO}.complete - AcquireRef {FLAO}[270] - dx:0 dy:0 slNull: f1:Filter #0 f2:Filter #0 freq:100 mag:16.325 nBins:4 nModes:10 r0:0 snMode:0 strehl:0 ttMod:6

Fri Nov 28 11:31:48.575 2014 aos.StartAO{FLAO}.started - StartAO {FLAO}[271]

TN:#20141128_113751,_113806,_113951,_114009,_114241

Luci 0077-

luci.20141128.0077,

IDL> dice5_collect, DIST=5, ITER=1, EXP=4.0, FRAMES=4

=> Collected luci.20141128.0077.fits

IRCL: offset, 0.00000, 0.00000, ROT=0, COORD=DETXY, MOVE=ABS (iif_offset)

===> Error in irc_offsetpointing
err = Resmsg[0] = OffsetPointing result status: Error
err = Resmsg[1] = Pause{FLAO}[267]
err = Resmsg[2] = OffsetXY {FLAO}[279]
err = Resmsg[3] = left PMC Mirror GoToPositionAbs Command Failed because Chase failed to make tolerance

The offset can not be finished, the primary went to the limit.

Fri Nov 28 11:46:59.348 2014 aos.OffsetXY{FLAO}.complete - OffsetXY {FLAO}[279]

RECOVERING THE PRIMARY SX.

The loop is in Pause.

11:51 DIMM:0.9"
Stop from:11:46:59 to 11:56:19

Fri Nov 28 11:57:52.166 2014 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[282] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:-21.848[-21.848],roy:-8.26628,mag:15.43,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

TN:#_120532,_120545,_120722,_121646

12:01 DIMM:0.91"

Luci 0078-83

12:13.DIMM:1.14"

ACTIVE 0084-89

Fri Nov 28 12:38:35.053 2014 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[323] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:12.1517[12.1517],roy:30.5129,mag:14.4,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

12:42 DIMM:0.99"

Fri Nov 28 12:42:53.984 2014 aos.info - WFS: star found, estimated magnitude: 15.41

Fri Nov 28 12:43:22.672 2014 aos.info - Adsec 672b-left loading reconstructor.

Fri Nov 28 12:43:24.338 2014 aos.info - Stages offset applied - dx: 0; dy: 0Fri Nov 28 12:43:24.352 2014 aos.AcquireRef{FLAO}.complete - AcquireRef {FLAO}[324] - dx:0 dy:0 slNull: f1:Filter #0 f2:Filter #0 freq:200 mag:15.406 nBins:4 nModes:10 r0:0 snMode:0 strehl:0 ttMod:6

tn:_124628,_124641,_

Luci 0090-

Shell RIP Skip FRames 100% fsm_status failure and it is in the loop of skipping frames

dsecarb.L |INF| 2004|2014-11-28 12:49:09.053737| COMMANDHANDLER > FSM (status: AORunning) has received command 2015 (Stop Ao)
adsecarb.L |INF| 2005|2014-11-28 12:49:09.053766| SERIALIZATION > StopAoImpl (): restoreShape = 1
adsecarb.L |WAR| 2006|2014-11-28 12:49:09.720568| MAIN > Skip frame detected!
adsecarb.L |WAR| 2007|2014-11-28 12:49:11.801003| MAIN > Skip frame detected!
adsecarb.L |ERR| 2008|2014-11-28 12:49:11.923673| IDL > Command returns error : IDL Safe params (current absolute or rms or delta, position absolute o delta o rms) out of bounds
Setting the Shell, elevation 37, wind speed 8m/s TSS activated and the shell can not be set. It build up a lot of forces.

Fri Nov 28 13:07:17.739 2014 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[337] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:12.2204[12.2204],roy:30.5419,mag:14.4,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

Fri Nov 28 13:11:12.899 2014 aos.AcquireRef{FLAO}.complete - AcquireRef {FLAO}[338] - dx:0 dy:0 slNull: f1:Filter #0 f2:Filter #0 freq:200 mag:14.914 nBins:4 nModes:10 r0:0 snMode:0 strehl:0 ttMod:6

Fri Nov 28 13:11:12.916 2014 aos.StartAO{FLAO}.started - StartAO {FLAO}[339]

13:09 DIMM:1.08"

Luci 0091-96

TN:_131753,_132443
Active 0097-0098

NO DITHER

13:26 dimm13:26

NIGHT NIGHT 13:34

Comments

 
Topic revision: r1 - 28 Nov 2014, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.