UT 20130920 AO#2 DX

Observer:DMiller, JChristou (LBTO)
Telescope Operator: GBechetti
SW Support: JKraus
AO Support: JChristou
Telescope Support: KNewton
Instrument Support: ESolheid

Summary

Details

01:30 w_stop, w_start

01:32 Power on WFS via AOS GUI

01:55 All systems (wfs and adsec) up, running and ready.

01:55 Geno preset to pointing star

02:10 Active preset has collimated the telescope

02:12 preset, AO734, MODE=ACE, GS=18, PA=0

02:18 runao. CCD39 stop (Not Live) when the low order closed loop is closed during the acquireRef phase.

02:20 runao. Same freezing of CCD39

04:00 Talked to Simone and Alfio. Powered off and on the WUnit. Had problems with CCD47 equalization.
  • Click "Operate" button to get back to initial configuration
  • In Variable inspector GUI set slopecompctrl.R.FASTLINK.ENABLE.REQ for 0 to 1. Close loop button turns green on WFS Control GUI. WFSFrameCounter increment on AdSec Mirror GUI. Communication is working
  • While in Operate State can close the loop
    • setting bin = 1, Loop freq = 990, lambda=3.0 and Apply.
    • set gains to 0.0 on AdSec GUI.
    • close loop on wfs control gui
    • WFSFramecounter is changing
  • iif_preset and iif_runao and loop is closed.
  • Conclude that the power cycle of WUnit "fixed" it

Bin 1

04:30 preset AO728 bin 1

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20130920_043849 ??? 11.1 83 4 0.88 0.39 221257 2 153 991 1.2 0.5 0.5 3 81 ON 71.9 H 200 100 ONSKY zero.fits  
20130920_043918 ??? 11.1 83 4 0.85 0.40 221257 2 153 991 1.2 0.5 0.5 3 81 ON 72.9 H 200 100 ONSKY zero.fits  
20130920_043946 ??? 11.1 83 4 0.76 0.36 221257 2 153 991 1.2 0.5 0.5 3 81 ON 70.0 H 200 100 ONSKY zero.fits  
20130920_044106 ??? 11.1 82 4 0.78 0.43 221257 2 153 991 1.2 0.5 0.5 3 81 ON 69.4 H 200 100 ONSKY zero.fits  
20130920_044135 ??? 11.1 82 4 0.78 0.49 221257 2 153 991 1.2 0.5 0.5 3 81 ON 63.6 H 200 100 ONSKY zero.fits  
20130920_044203 ??? 11.1 82 4 0.79 0.48 221257 2 153 991 1.2 0.5 0.5 3 84 ON 62.2 H 200 100 ONSKY zero.fits  

SR 0.683229 (0.044224)
SEEING (WFS) 0.425914 (0.0519231)
SEEING (DIMM) 0.804418 (0.0475647)

04:45 preset AO731, Cat Mag = 12.2, WFS Mag = 11.8. Reset to bin 1, 990 hz. Want Bin 2

04:58 preset AO747, gs=4, pa=180, Cat Mag=13.5

05:02 runao. Failed because camera lens did not converge

05:10 preset and runao. Again failed because camera lens did not converge

Bin 2

05:20 preset AO747, gs=10, pa=0, Cat Mag=13.5

05:30 acq sky_irtc_psf

05:34 preset AO748, gs=3. runao. Closed, but bin2 Cat Mag = 14.4, WFS Mag = 12.2

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain modSorted ascending nph AntiDrift SR filter exp #frames disturb SN notes
20130920_052815 ??? 13.1 76 3 1.12 0.81 221257 2 153 400 1.6 0.8 0.5 3 32 ON 41.7 H 1000 10 ONSKY zero.fits  
20130920_052829 ??? 13.0 76 4 NaN 0.65 221257 2 153 400 1.6 0.8 0.5 3 35 ON 45.3 H 1000 10 ONSKY zero.fits  
20130920_052843 ??? 13.0 76 3 NaN 0.61 221257 2 153 400 1.6 0.8 0.5 3 37 ON 47.1 H 1000 10 ONSKY zero.fits  
20130920_052934 ??? 13.0 76 4 NaN 0.52 221257 2 153 400 1.6 0.8 0.5 3 36 ON 48.3 H 1000 10 ONSKY zero.fits  
20130920_052948 ??? 13.0 76 3 NaN 0.47 221257 2 153 400 1.6 0.8 0.5 3 35 ON 54.4 H 1000 10 ONSKY zero.fits  
20130920_053001 ??? 13.6 76 3 NaN 0.56 221257 2 153 400 1.6 0.8 0.5 3 21 ON 45.2 H 1000 10 ONSKY zero.fits  

SR 0.47007 (0.042724)
SEEING (WFS) 0.604661 (0.119783)
SEEING (DIMM) 1.11559 (NaN)

Bin 3

05:45 preset AO750, gs=3, pa=0, Cat Mag=15.1 runao WFS Mag = 14.0

05:52 acq sky_irtc_psf

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20130920_055305 ??? 14.2 71 4 0.82 0.44 170946 3 66 312 2.1 0.6 -1.0 6 36 ON 51.6 H 1500 10 ONSKY zero.fits - Pupils not on target
20130920_055324 ??? 14.2 71 4 0.81 0.39 170946 3 66 312 2.1 0.6 -1.0 6 36 ON 52.5 H 1500 10 ONSKY zero.fits  
20130920_055344 ??? 14.2 71 4 0.82 0.39 170946 3 66 312 2.1 0.6 -1.0 6 36 ON 51.2 H 1500 10 ONSKY zero.fits  
20130920_055447 ??? 14.2 71 3 0.82 0.44 170946 3 66 312 2.1 0.6 -1.0 6 37 ON 49.4 H 1500 10 ONSKY zero.fits - Pupils not on target
20130920_055507 ??? 14.2 71 4 0.79 0.36 170946 3 66 312 2.1 0.6 -1.0 6 37 ON 52.9 H 1500 10 ONSKY zero.fits  
20130920_055526 ??? 14.2 71 3 0.79 0.40 170946 3 66 312 2.1 0.6 -1.0 6 37 ON 51.7 H 1500 10 ONSKY zero.fits  

SR 0.51541 (0.0123732)
SEEING (WFS) 0.401677 (0.0300041)
SEEING (DIMM) 0.809668 (0.0145723)

06:05 runao failed No star found on TV frame

Bin 4

06:06 preset AO751, runao. Cat Mag = 15.7, WFS Mag = 16.4

06:12 acq sky_irtc_psf

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20130920_061710 ??? 16.6 67 4 0.76 0.36 214654 4 36 100 1.6 0.9 0.9 6 22 ON * H 1500 10 ONSKY zero.fits - Pupils not on target - SR may be in error
20130920_061752 ??? 16.6 67 4 0.85 0.33 214654 4 36 100 0.0 -1.0 -1.0 6 22 ON * H 1500 10 ONSKY zero.fits - SR may be in error
20130920_061835 ??? 16.5 67 4 0.86 0.37 214654 4 36 100 0.0 -1.0 -1.0 6 23 ON * H 1500 10 ONSKY zero.fits - Pupils not on target - Center of PSF estimation error

SR 0.666667 (0.57735)
SEEING (WFS) 0.353413 (0.0171778)
SEEING (DIMM) 0.825601 (0.0570215)

06:20 preset, runao

06:27 Shell Ripped during Gain optimization routine

06:33 preset, runao
  • Closed and auto gain success
  • Offset -1.5, 3.6 Failed when resumeAO. Too much tilt? Safe Skip and then shell ripped.

06:43 Recover fail and Set

06:44 preset, runao .... Humidity goes to 95% Close Dome

Bin 4

08:15 Chamber open. preset AO824, Bin 4 WFS Mag 14.9
  • Collected data set
  • Offset 5 times with no problems
  • Got a big tip and safe skip to ripped shell

Binary Star

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20130920_082258 ??? 15.2 71 4 1.01 1.33 214654 4 36 200 2.8 3.0 0.9 6 39 ON * H 1500 10 ONSKY zero.fits - SR may be in error
20130920_082320 ??? 15.1 71 4 1.13 1.21 214654 4 36 200 2.8 3.0 0.9 6 41 ON * H 1500 10 ONSKY zero.fits - SR may be in error
20130920_082342 ??? 15.1 71 4 1.02 0.95 214654 4 36 200 2.8 3.0 0.9 6 41 ON * H 1500 10 ONSKY zero.fits - SR may be in error
20130920_082447 ??? 15.1 71 4 1.00 0.84 214654 4 36 200 2.8 3.0 0.9 6 41 ON * H 1500 10 ONSKY zero.fits - SR may be in error
20130920_082509 ??? 15.1 71 3 0.90 0.75 214654 4 36 200 2.8 3.0 0.9 6 41 ON * H 1500 10 ONSKY zero.fits - SR may be in error
20130920_082531 ??? 15.1 71 3 0.93 0.81 214654 4 36 200 2.8 3.0 0.9 6 41 ON * H 1500 10 ONSKY zero.fits - SR may be in error

SR 0.0855602 (0.0133905)
SEEING (WFS) 0.983134 (0.233531)
SEEING (DIMM) 0.999969 (0.0809216)

08:23 Recover Fail. Success

08:27 Set. Failed. Had Safe Skip and then rip and recover fail

08:31 Set. Successful

Bin 3

08:37 preset AO130, Cat Mag = 14.1 bin 3. Failed no guide star. Geno set pointing

08:44 preset AO130

08:50 acq sky_irtc_psf

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20130920_084921 ??? 13.5 68 3 0.88 0.43 221257 2 153 400 1.6 0.5 0.5 3 24 ON 50.2 H 1500 10 ONSKY zero.fits  
20130920_084940 ??? 13.5 68 3 0.86 0.40 221257 2 153 400 1.6 0.5 0.5 3 24 ON 51.1 H 1500 10 ONSKY zero.fits - SR may be in error
20130920_085000 ??? 13.4 68 3 0.87 0.39 221257 2 153 400 1.6 0.5 0.5 3 24 ON 49.4 H 1500 10 ONSKY zero.fits  
20130920_085103 ??? 13.5 68 3 1.18 0.54 221257 2 153 400 1.6 0.5 0.5 3 23 ON 37.3 H 1500 10 ONSKY zero.fits  
20130920_085121 ??? 13.5 68 3 0.96 0.67 221257 2 153 400 1.6 0.5 0.5 3 23 ON 34.0 H 1500 10 ONSKY zero.fits  
20130920_085139 ??? 13.5 68 3 0.95 0.66 221257 2 153 400 1.6 0.5 0.5 3 23 ON 38.8 H 1500 10 ONSKY zero.fits  

SR 0.43462 (0.0757836)
SEEING (WFS) 0.517156 (0.128701)
SEEING (DIMM) 0.951349 (0.119611)

Bin 2

08:55 preset AO199, Cat Mag = 12.4, WFS Mag = 11.6

09:05 acq sky_irtc_psf

Star not on IRTC

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20130920_090550 ??? 11.7 46 3 1.00 0.71 221257 2 153 991 2.1 0.6 0.5 3 50 ON * H 1500 10 ONSKY zero.fits - SR may be in error
20130920_090609 ??? 11.7 46 3 1.00 0.67 221257 2 153 991 2.1 0.6 0.5 3 49 ON * H 1500 10 ONSKY zero.fits - Could not test SR quality
20130920_090629 ??? 11.7 46 3 1.00 0.62 221257 2 153 991 2.1 0.6 0.5 3 50 ON * H 1500 10 ONSKY zero.fits - SR may be in error
20130920_090730 ??? 11.7 46 4 0.92 0.66 221257 2 153 991 2.1 0.6 0.5 3 49 ON * H 1500 10 ONSKY zero.fits - SR may be in error
20130920_090748 ??? 11.7 46 3 0.89 0.74 221257 2 153 991 2.1 0.6 0.5 3 50 ON * H 1500 10 ONSKY zero.fits - Center of PSF estimation failed
20130920_090807 ??? 11.7 46 3 1.01 0.88 221257 2 153 991 2.1 0.6 0.5 3 49 ON * H 1500 10 ONSKY zero.fits - SR may be in error

SR 0.833333 (0.408248)
SEEING (WFS) 0.712773 (0.0895572)
SEEING (DIMM) 0.971539 (0.0513105)

09:15 Closed Dome for Humidity

-- DougMiller - 20 Sep 2013
Topic revision: r4 - 21 Sep 2013, JohnHill
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback