Details.

  • Pupils shift displacement.
  • Alessandro in teh mountain.
  • Work to clamp the ocam. Moved ~ 2 pixels down looking at the pupils using the internal source.
  • Check with argos cal unit.

Checking with the Argos cal unit.

Bin1 a shift of 3 pixels Y->-3 allows the pupils to be in the range with a small range where the CL Y goes to -100 with a pupil error of <0.3 pixels. This is inside of the Operation constraints.

Bin2 shift y->-2

Bin3 shifted->-1

TN:20240501_182411

Bin1

corrected 3 pixels and error for the Y axis below threshold 0.3.

First pass with a correction of 1 pixel and the second pass with 3 pixels.

Screen Shot 2024-05-01 at 3.31.22 PM.png

Bin2.

First pass looks good, stop moving forward at 210.

Screen Shot 2024-05-01 at 3.38.10 PM.png

bin3

pupil shift by y -1.

Screen Shot 2024-05-01 at 3.44.06 PM.png

Bin4

shifted one pixel and it goes in change quadrants.

default pupils shows to have an error or -0.5 to -0.7.

It can be increased the threshold for pupil error to 0.8 and the system it will not failed.

... leave default pupils--

Screen Shot 2024-05-01 at 1.00.44 PM.png

Screen Shot 2024-05-01 at 12.59.42 PM.png

Pupil wobble sequence:

Screen Shot 2024-05-01 at 3.29.10 PM.png

Check close loop.

Bin1.

pupil:20240501_153735 shifted y->-3

Wed May 1 21:07:14.852 2024 aos.Stop{FLAO}.complete - Stop{FLAO}[3541]
Wed May 1 21:07:58.897 2024 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3542] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed May 1 21:07:58.913 2024 aos.info - WFS source set to: FLAOWFS
Wed May 1 21:07:58.947 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq1700.00_sd3892.fits

OK.

TN:_211813,_212147.
performances looks OK the SR from slopes ~85"

Not images from Luci.

Loop stable.

Bin2.

pupil:20240501_171340

Wed May 1 21:34:57.219 2024 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3549] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:12.8,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed May 1 21:34:57.232 2024 aos.info - WFS source set to: FLAOWFS
Wed May 1 21:34:57.280 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq900.00_sd3892.fits

OK.

TN:_214235,_214258

Bin3.

Wed May 1 21:43:27.949 2024 aos.Stop{FLAO}.complete - Stop{FLAO}[3552]
Wed May 1 21:43:49.309 2024 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3553] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:14.7,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed May 1 21:43:49.328 2024 aos.info - WFS source set to: FLAOWFS
Wed May 1 21:43:49.375 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq900.00_sd3892.fits

optimize:

tn:_215052,_215102

bin4.

Wed May 1 21:56:05.148 2024 aos.Stop{FLAO}.complete - Stop{FLAO}[3556]
Wed May 1 21:57:03.426 2024 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3557] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.5,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed May 1 21:57:03.444 2024 aos.info - WFS source set to: FLAOWFS
Wed May 1 21:57:03.474 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq750.00_sd3892.fits

loop closed get stuck.

Also center pupils failed to converege, error y->-0.5

-- JuanCarlosGuerra - 01 May 2024
Topic revision: r2 - 01 May 2024, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback