Staff:

LBTO: Juan CArlos, Brandon

Details

Switch version.

Adsec: Currently running: test.20240422.gopt-retry (soul)
WFS: Currently running: test.20240425.gopt-retry-and-rerotator (soul).

Test plan excel:

https://docs.google.com/spreadsheets/d/1VH0XTD1iLvGc9FTOtxbHuZ-DbHpNPwele5TIKEyvoyg/edit#gid=0

10:00

TEsting.

power on.

17:02:17.220792| MAIN > Turning WFS on

ocam frames OK

Done.

17:02:17.220792| MAIN > Turning WFS on

17:11:20 Done

Test.1

SetSafe.

Operating

OK

17:12:43.496593| MAIN > arbFunc doSetSafe done.

Thu Apr 25 17:13:28.274 2024 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[3066] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

AOpreperared

SETSAFe

OK

-25 17:15:08.007658| MAIN > arbFunc updateModulationInternal done.

center pupil.s

The Y axis goes to -100.

Diffx -3.5 pixles.

?

It should be for DX.

shift y+1

47: 20240425_173931 y+1
48: 20240425_174518 y-1
shift +3 gives and error of -0.5

shift +4 pxels

change ao table for bn1

[AOeng@soul-dxwfs ao]$ cp table_LUCIFER_ACE-AO_20230507_623m_v01.txt table_LUCIFER_ACE-AO_20230507_623m_v01_shiftpupil_y.txt
Modify for bn1 mag 10.5 and 11.4

rwxrwxrwx. 1 AOeng AOeng 55 Apr 25 18:16 table_LUCIFER_ACE-AO.txt -> table_LUCIFER_ACE-AO_20230507_623m_v01_shiftpupil_y.txt

Thu Apr 25 18:23:14.126 2024 aos.AcquireRef{FLAO}.complete - AcquireRef {FLAO}[3071] - dx:0.359054 dy:-0.244678 slNull: f1:0.0 f2:1.0 freq:1700 mag:0 nBins:1 nModes:630 r0:nan snMode:0 strehl:0 ttMod:0
Thu Apr 25 18:23:21.479 2024 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[3072] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Thu Apr 25 18:23:21.497 2024 aos.info - WFS source set to: FLAOWFS
Thu Apr 25 18:23:21.529 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s2.0_L040.0_v15.0_ovfreq1700.00_sd3892.fits

PresetAO

u Apr 25 18:26:30.340 2024 aos.PresetAO{FLAO}.failed - PresetAO {FLAO}[3072] - [AOException] presetAO: FATAL - Reading shared buffer mirrorctrl.R:TECHVIEWER_65536 (-5001) TIMEOUT_ERROR

This is long timeout. it should be reduced... Brandon looking into it.

reset bcu47, change bin to 16

ok.

test shift -4 offbin1 Y from 1 to 0

back to the default AO table.

12:20

Power off

t requires to check the M3 positions and algnment of the wobble. To be checked rotation test.

RESUME ACTIVITY

02:20

not found the PSF.

stop 03:10

-- JuanCarlosGuerra - 25 Apr 2024