Staff:

LBTO: Juan Carlos

Arcetri: Fabio and Alfio.

Details.

Discussion with Fabio the results from last tests. Bin1 to bin3 all good.

bin4 debug. This is not clear the reason of the hard failure of the adsec, it could be related to change of setup, etc..

Telescope is running other high priority test, so no tme for a close loop test.

08:45am

Power on WFS

OCAm not live.
  • Reset bn didnt work.
  • Reset masterdiagnostc didn not work.
  • Power cycle bcu
  • OK

Testing the ocam Ron Test.

Gui has the RON test only.

thaoshell.py -wait -dev ocam2 test_ocam_ron.py

The script runs in the terminal and goes OK but the GUI ios not displying anything. Alfio also connected.

Debug the script in the GUI.

Resolved the output to the GUI. The Guis shows when the script was launched and the outputs wth a final report.

Screen Shot 2024-04-17 at 9.39.45 AM.png


SUMMARY ------------------------------------
Mode: 1x1 Cropped - Framerate 1700 Hz - emGain 1 - RON (ADUs): 5.14 - RON (e-): 171.49 (estimated)
Mode: 1x1 Cropped - Framerate 1500 Hz - emGain 1 - RON (ADUs): 5.16 - RON (e-): 171.89 (estimated)
Mode: 1x1 Cropped - Framerate 1000 Hz - emGain 1 - RON (ADUs): 5.14 - RON (e-): 171.39 (estimated)
Mode: 1x1 Cropped - Framerate 500 Hz - emGain 1 - RON (ADUs): 5.14 - RON (e-): 171.21 (estimated)
Mode: 1x1 Cropped - Framerate 200 Hz - emGain 1 - RON (ADUs): 5.14 - RON (e-): 171.28 (estimated)
Mode: 1x1 Cropped - Framerate 100 Hz - emGain 1 - RON (ADUs): 5.15 - RON (e-): 171.68 (estimated)
Mode: 2x2 Full - Framerate 1700 Hz - emGain 1 - RON (ADUs): 6.05 - RON (e-): 201.57 (estimated)
Mode: 2x2 Full - Framerate 1500 Hz - emGain 1 - RON (ADUs): 6.06 - RON (e-): 202.00 (estimated)
Mode: 2x2 Full - Framerate 1000 Hz - emGain 1 - RON (ADUs): 6.16 - RON (e-): 205.21 (estimated)
Mode: 2x2 Full - Framerate 500 Hz - emGain 1 - RON (ADUs): 6.19 - RON (e-): 206.35 (estimated)
Mode: 2x2 Full - Framerate 200 Hz - emGain 1 - RON (ADUs): 6.25 - RON (e-): 208.20 (estimated)
Mode: 2x2 Full - Framerate 100 Hz - emGain 1 - RON (ADUs): 6.27 - RON (e-): 209.05 (estimated)
Mode: 4x4 Full - Framerate 1700 Hz - emGain 1 - RON (ADUs): 7.65 - RON (e-): 254.89 (estimated)
Mode: 4x4 Full - Framerate 1500 Hz - emGain 1 - RON (ADUs): 7.76 - RON (e-): 258.51 (estimated)
Mode: 4x4 Full - Framerate 1000 Hz - emGain 1 - RON (ADUs): 7.73 - RON (e-): 257.73 (estimated)
Mode: 4x4 Full - Framerate 500 Hz - emGain 1 - RON (ADUs): 7.77 - RON (e-): 259.11 (estimated)
Mode: 4x4 Full - Framerate 200 Hz - emGain 1 - RON (ADUs): 7.86 - RON (e-): 261.88 (estimated)
Mode: 4x4 Full - Framerate 100 Hz - emGain 1 - RON (ADUs): 7.84 - RON (e-): 261.28 (estimated)
>>>

Done

09:50

Power off WFS.

-- JuanCarlosGuerra - 17 Apr 2024