Summary.

Using the reset_ocam in open and loop closed. A change in the way to execute the pause/resume was done and it going to be tested.

Details.

10am

Power on Soul-luciDX

Working in parallel with soul-luciSX. Doug is performing ncpa.

Ocam not live.
  • change bin showed BUFF_
Screen Shot 2024-04-15 at 10.21.21 AM.png
  • restart of the masterdiagnostics + rebin
  • same error.
  • power cycle bcu2k.
  • ocam OK.

test preset.

Mon Apr 15 17:27:06.545 2024 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[235] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Apr 15 17:27:06.561 2024 aos.info - WFS source set to: FLAOWFS
Mon Apr 15 17:27:06.658 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.00_sd3892.fits

OK.

PSF centered.

test script open loop:

[AOeng@soul-dxwfs scripts]$ python reset_ocam.py
No configuration file found for AdSec arbitrator
Traceback (most recent call last):
File "reset_ocam.py", line 20, in
arbproxy = AOIce.AOARBProxy()
File "/home/AOeng/live/aoroot/lib/python/AdOpt/AOIce.py", line 107, in AOARBProxy
port = cfg.cfg['AOARB']['ListeningPort'].Value()
KeyError: 'AOARB

DEbug time.

12:00

scrpt runs wth the commands thorugh the AO arbitrator, the gain zero is applied.

Found that time to execute the resume at 10 seconds is fine, below 10 seconds the masterdiagnsotic get stuck an requires power cycle bcu2k.

testing.

Open loop.

Bin1.

  • State AOset:OK
  • State AO preraped:OK
  • State: RunAO. OK
20240415 19:33:49 mean: [412.49426666666676, 391.94063333333332, 196.19794444444449, 401.51539999999989, 407.71533333333338, 388.69112222222225, 387.04376666666673, 402.91455555555564]
20240415 19:33:49 rms: [168.72003158250561, 151.78190591034351, 196.59316091182623, 158.4569914696167, 164.25201374246851, 150.04654610255204, 141.09774477458311, 158.72069714899445]
20240415 19:33:49 std: [44.612730307108635, 13.088685586624324, 12.743296134478813, 29.861642673832151, 36.402047175436081, 14.824520233837577, 11.702377307397144, 31.447032243111011]
20240415 19:33:51 Suspended because Ocam is reconfiguring
20240415 19:33:53 Suspended because Ocam is reconfiguring
20240415 19:33:55 Suspended because Ocam is reconfiguring
20240415 19:33:57 valid_pixels shape: (101, 1)
20240415 19:33:57 pixels_data shape: (101, 40000)
20240415 19:33:57 decimated pixels data shape: (24, 28800)
20240415 19:33:57 mean: [413.60145833333348, 392.10221064814806, 196.60662037037036, 399.43194444444435, 410.76598379629627, 388.69625000000002, 386.59368055555552, 400.73121527777784]
20240415 19:33:57 rms: [168.40197378332087, 151.98062869173174, 196.99363630249806, 157.47394817722571, 164.95401073482014, 150.30190486176514, 139.92383869211469, 156.7270276369473]
20240415 19:33:57 std: [47.178489122672012, 12.990813781551248, 12.492215702116896, 25.794110376735194, 40.680291447644514, 15.071086700457466, 11.660296816054684, 28.879389971480361]
20240415 19:33:59 valid_pixels shape: (100, 1)

[AOeng@soul-dxwfs scripts]$ python reset_ocam.py
ocam2 interface loaded
Loop is Closed, Pausing
Loop was Closed, Re-Closing

Once the reset is executed the ocam viwere is closed.

three executions success.

bin2.

  • State AOset:OK
  • State AO prepared:OK
  • State RunAO: OK
20240415 19:41:39 decimated pixels data shape: (25, 14400)
20240415 19:41:39 mean: [478.60246666666666, 410.15222222222224, 211.17731111111112, 452.79475555555558, 474.56382222222231, 406.27291111111111, 397.72439999999995, 454.30331111111116]
20240415 19:41:39 rms: [183.43195930248501, 191.76729832648221, 211.10994767906723, 180.18615364640974, 190.65373531395576, 182.44737808357914, 165.59736592648511, 180.87002379795737]
20240415 19:41:39 std: [218.04592043400658, 16.835850020970831, 14.450210382049796, 138.78576075059522, 173.86714765001108, 17.079540357755544, 14.458222599360438, 145.98193682041304]
20240415 19:41:41 Suspended because Ocam is reconfiguring
20240415 19:41:43 Suspended because Ocam is reconfiguring
20240415 19:41:45 Suspended because Ocam is reconfiguring
20240415 19:41:47 valid_pixels shape: (100, 1)
20240415 19:41:47 pixels_data shape: (100, 40000)
20240415 19:41:47 decimated pixels data shape: (25, 14400)
20240415 19:41:47 mean: [483.70157777777786, 410.76404444444438, 211.09102222222219, 458.6279555555555, 473.04657777777788, 407.08933333333317, 397.91755555555545, 454.51620000000003]
20240415 19:41:47 rms: [184.27274961654919, 192.82305853822433, 210.96816462001624, 180.76973221091041, 191.66893139279222, 183.8982764962463, 165.69273125609024, 180.90045314289387]
20240415 19:41:47 std: [226.51537035057584, 16.672915253915839, 14.596253743039988, 157.39856087282283, 167.94493629626481, 16.851834845966252, 14.223868364966163, 147.37032958067587]
20240415 19:41:49 valid_pixels shape: (100, 1)

[AOeng@soul-dxwfs scripts]$ python reset_ocam.py
ocam2 interface loaded
Loop is Closed, Pausing
Loop was Closed, Re-Closing

20240415 19:43:08 mean: [476.67039999999992, 410.2109111111111, 211.64177777777772, 456.17120000000011, 470.80266666666671, 406.63706666666667, 398.25844444444454, 456.5980222222222]
20240415 19:43:08 rms: [183.59295109286325, 191.85798594898267, 211.4330379151489, 180.67980460913296, 191.08962719996686, 183.2335311640494, 166.56438333202797, 181.43054915117699]
20240415 19:43:08 std: [215.04215297096343, 16.718217111011526, 14.793176415773107, 152.80257015996926, 165.88812874843632, 16.927986491265358, 14.320289676571457, 152.12643163927942]
20240415 19:43:10 Suspended because Ocam is reconfiguring
20240415 19:43:12 Suspended because Ocam is reconfiguring
20240415 19:43:14 Suspended because Ocam is reconfiguring
20240415 19:43:16 valid_pixels shape: (100, 1)
20240415 19:43:16 pixels_data shape: (100, 40000)
20240415 19:43:16 decimated pixels data shape: (25, 14400)
20240415 19:43:16 mean: [481.40682222222233, 411.31066666666663, 211.90893333333332, 448.51900000000001, 481.39433333333341, 408.02215555555551, 398.79455555555546, 457.91522222222221]
20240415 19:43:16 rms: [184.01214692480193, 193.78374235109041, 211.70658179867584, 181.43397992777068, 191.91532602573182, 185.46889873385547, 167.50857120088074, 182.62146519355167]
20240415 19:43:16 std: [224.2317685878713, 16.611596515885985, 14.633379342413814, 127.32382167720225, 186.42542952129403, 16.882311445349483, 14.541568398316398, 154.33122142909872]
20240415 19:43:18 valid_pixels shape: (101, 1)

20240415 19:44:59 mean: [462.52807692307692, 402.58480769230778, 207.97809829059827, 449.13177350427344, 473.06376068376056, 399.17237179487188, 390.86993589743599, 443.75715811965802]
20240415 19:44:59 rms: [185.08484963672828, 192.86577142320681, 207.89562684469104, 182.21867436270728, 191.74968412491859, 184.67791949879145, 168.49915218790667, 183.11287881433284]
20240415 19:44:59 std: [200.63111304030525, 16.347849115210487, 14.44503926239317, 147.71403559818592, 183.97722625609262, 16.511367523306646, 14.056433247426806, 135.98241232056208]
20240415 19:45:01 Suspended because Ocam is reconfiguring
20240415 19:45:05 Suspended because Ocam is reconfiguring
20240415 19:45:07 valid_pixels shape: (100, 1)
20240415 19:45:07 pixels_data shape: (100, 40000)
20240415 19:45:07 decimated pixels data shape: (25, 14400)
20240415 19:45:07 mean: [476.33177777777775, 411.18004444444438, 211.69697777777779, 465.59782222222208, 471.42202222222221, 407.16084444444442, 398.0240888888888, 456.46871111111102]
20240415 19:45:07 rms: [184.80094378022767, 193.55447068827183, 211.59606026797636, 181.48557158782546, 192.37448404190312, 183.84373893858904, 166.00289454618979, 181.49771504155902]
20240415 19:45:07 std: [219.43029872457598, 16.992613416316583, 14.561328909638705, 171.62103503776191, 166.31307810973885, 17.248454956158753, 14.601081547553658, 151.67367787244126]
20240415 19:45:09 valid_pixels shape: (100, 1)

bin3

  • State AOSet: OK
  • State AOPrepared: OK
  • State RunAO: OK
20240415 19:53:11 mean: [449.63922499999984, 462.0069499999999, 237.95064999999991, 460.81992500000001, 457.54817500000013, 478.896075, 461.74377499999986, 457.30392500000016]
20240415 19:53:11 rms: [192.74559053379966, 159.89383170068479, 219.17496869748476, 154.18486559702475, 176.47026285424275, 177.19963273819266, 158.33049543402126, 163.12731275036322]
20240415 19:53:11 std: [40.409189002367491, 23.359419561838767, 16.392602800040859, 27.317317845455065, 31.770878951431037, 20.601219348589453, 16.651417478228176, 29.323482884750362]
20240415 19:53:13 Suspended because Ocam is reconfiguring
20240415 19:53:17 Suspended because Ocam is reconfiguring
20240415 19:53:20 valid_pixels shape: (100, 1)
20240415 19:53:20 pixels_data shape: (100, 40000)
20240415 19:53:20 decimated pixels data shape: (50, 6400)
20240415 19:53:20 mean: [451.61032500000016, 463.04279999999994, 239.30897500000006, 460.66347499999995, 458.97822500000001, 479.386325, 462.23624999999998, 458.07639999999992]
20240415 19:53:20 rms: [191.04629702636058, 160.0626362487275, 216.83501618158039, 152.3488003673761, 175.43581249385949, 178.0942211766876, 158.80754990762759, 161.8974696785738]
20240415 19:53:20 std: [44.52836408092881, 23.412055073980497, 16.681446724936272, 28.06453988135495, 36.192127786644676, 20.780153425835351, 16.693905722481542, 36.324845693348315]
20240415 19:53:22 valid_pixels shape: (100, 1)

20240415 19:54:13 pixels_data shape: (100, 40000)
20240415 19:54:13 decimated pixels data shape: (50, 6400)
20240415 19:54:13 mean: [449.07065, 462.4479750000001, 239.30985000000004, 461.43402499999979, 456.48622499999999, 478.15717499999994, 461.94089999999994, 457.70422500000001]
20240415 19:54:13 rms: [193.23433089115039, 159.41693016193349, 217.05464021145301, 153.33929449399108, 178.71289651566886, 177.1550443072835, 158.56550841252169, 162.6836638373824]
20240415 19:54:13 std: [40.255523725955385, 23.279457898236647, 16.495988921183649, 27.524586746053561, 31.920302091572356, 20.636210868097525, 16.730070005405373, 29.633315976903127]
20240415 19:54:15 Suspended because Ocam is reconfiguring
20240415 19:54:17 Suspended because Ocam is reconfiguring
20240415 19:54:21 valid_pixels shape: (100, 1)
20240415 19:54:21 pixels_data shape: (100, 40000)
20240415 19:54:21 decimated pixels data shape: (50, 6400)
20240415 19:54:21 mean: [450.66294999999991, 463.28677500000003, 237.58000000000007, 459.55077499999987, 461.30262499999981, 480.28620000000001, 461.57029999999992, 461.66109999999992]
20240415 19:54:21 rms: [188.02519799537018, 158.09972501327869, 218.88351887255101, 153.3000323156021, 171.22266310434202, 179.86866061796846, 158.63546299855841, 162.50906621142067]
20240415 19:54:21 std: [38.726863259504995, 23.306801865769515, 16.567759171404148, 27.793258161815505, 38.209147393101951, 20.640259981644476, 16.692734929870884, 45.940700986285044]
20240415 19:54:23 valid_pixels shape: (100, 1)
20240415 19:54:23 pixels_data shape: (100, 40000)
20240415 19:54:23 decimated pixels data shape: (50, 6400)
20240415 19:54:23 mean: [447.8573750000001, 462.46762499999988, 237.88174999999995, 459.43339999999995, 458.64729999999997, 478.3122249999999, 461.10187500000006, 461.46979999999996]
20240415 19:54:23 rms: [193.50502181405693, 157.28793534267638, 219.0582396621854, 153.47667968601348, 176.07118545843565, 177.72995927155284, 158.05590355336085, 162.98145614473813]
20240415 19:54:23 std: [39.683478943245078, 23.1137766335481

20240415 19:55:18 decimated pixels data shape: (50, 6400)
20240415 19:55:18 mean: [449.33184999999997, 462.62129999999991, 238.68562499999993, 461.73447500000009, 457.16334999999998, 478.73632499999991, 461.90250000000009, 457.66272500000014]
20240415 19:55:18 rms: [192.24563413776349, 159.87153081367131, 217.85641483540232, 153.93880375159125, 177.35184200979859, 177.07463999695585, 159.16325444167126, 163.72330681026713]
20240415 19:55:18 std: [39.412194506619791, 23.46925530108059, 16.590600556619933, 28.461926997528153, 31.703639970827048, 20.75180431257624, 16.853081795839739, 30.193026640601794]
20240415 19:55:20 Suspended because Ocam is reconfiguring
20240415 19:55:22 Suspended because Ocam is reconfiguring
20240415 19:55:26 Suspended because Ocam is reconfiguring
20240415 19:55:28 valid_pixels shape: (100, 1)
20240415 19:55:28 pixels_data shape: (100, 40000)
20240415 19:55:28 decimated pixels data shape: (50, 6400)
20240415 19:55:28 mean: [449.65485000000001, 463.21872500000006, 240.64417499999996, 461.30780000000004, 458.07877499999995, 478.53765000000004, 462.86697500000014, 458.67067500000007]
20240415 19:55:28 rms: [193.62619893002613, 160.20339411194314, 214.95927668982807, 151.62430161474722, 178.12721061072043, 177.47703339960259, 159.94334387310676, 161.38197496495224]
20240415 19:55:28 std: [43.761157294624745, 23.401898743674383, 16.653776907906266, 27.50150815033367, 37.534345728018728, 20.954054377363573, 16.878443112286128, 37.68935267215177]
20240415 19:55:30 valid_pixels shape: (100, 1)
20240415 19:55:30 pixels_data shape: (100, 40000)
20240415 19:55:30 decimated pixels data shape: (50, 6400)
20240415 19:55:30 mean: [449.60117500000001, 463.10730000000012, 240.86172500000001, 461.16142500000007, 456.99867500000005, 478.57889999999992, 462.97105000000005, 458.10332500000004]
20240415 19:55:30 rms: [194.07248965953681, 160.47362544351461,

ocam viewer closed.

20240415 19:56:28 decimated pixels data shape: (50, 6400)
20240415 19:56:28 mean: [448.75957499999998, 463.58749999999998, 241.26142500000006, 462.42502499999995, 457.01532499999996, 478.80094999999994, 463.23872499999999, 457.36159999999995]
20240415 19:56:28 rms: [193.21854848452367, 160.60244995720751, 214.16670337336589, 152.10112133007547, 178.2312573365264, 177.94436889823993, 159.53487945448822, 161.3719399698052]
20240415 19:56:28 std: [39.692867473579099, 23.40244615737171, 16.733255760388491, 26.894135191534101, 33.015976238982432, 20.692657702708789, 16.872600408758579, 28.206291758605545]
20240415 19:56:30 Suspended because Ocam is reconfiguring
20240415 19:56:34 Suspended because Ocam is reconfiguring
20240415 19:56:36 valid_pixels shape: (100, 1)
20240415 19:56:36 pixels_data shape: (100, 40000)
20240415 19:56:36 decimated pixels data shape: (50, 6400)
20240415 19:56:36 mean: [451.04170000000022, 463.64820000000009, 239.71510000000001, 461.52500000000009, 458.34905000000003, 478.47007500000007, 461.43585000000002, 457.97407500000008]
20240415 19:56:36 rms: [191.28908986650177, 158.68166545915648, 216.32156417454036, 152.09497914942455, 176.6829729893802, 177.43919276507194, 159.05868284612632, 163.71667502842777]
20240415 19:56:36 std: [43.47797568647524, 23.393089062739126, 16.633080830589513, 28.974736510022147, 36.611595358803427, 21.023484105923462, 16.806603666671904, 37.320912742211782]
20240415 19:56:38 valid_pixels shape: (100, 1)

20240415 19:57:17 mean: [449.3049749999999, 462.96827500000006, 239.81652499999996, 461.92240000000015, 456.15814999999998, 477.36122500000005, 461.20067499999988, 457.09809999999993]
20240415 19:57:17 rms: [193.08940531998778, 160.59987372098306, 215.8269143389038, 153.32162804959941, 178.84665274426058, 175.94102736355234, 159.58369018530522, 165.02351361777994]
20240415 19:57:17 std: [40.70069664790838, 23.588022763301584, 16.974813402677785, 28.21093851874987, 31.9252309300477, 21.138286634694659, 17.053538712077327, 30.462740235428036]
20240415 19:57:19 Suspended because Ocam is reconfiguring
20240415 19:57:23 Suspended because Ocam is reconfiguring
20240415 19:57:25 valid_pixels shape: (100, 1)
20240415 19:57:25 pixels_data shape: (100, 40000)
20240415 19:57:25 decimated pixels data shape: (50, 6400)
20240415 19:57:25 mean: [449.61382500000002, 462.70290000000017, 239.58295000000001, 460.27282500000001, 457.82555000000008, 478.00449999999989, 461.33269999999999, 457.69162499999982]
20240415 19:57:25 rms: [194.27690625649214, 159.00942037247896, 216.52409992461861, 153.04946846802059, 178.82347240698644, 177.09371629584084, 159.23312012514461, 164.92253130773832]
20240415 19:57:25 std: [44.233750109284927, 23.34363392594685, 16.679135936693832, 27.153017112801489, 37.769663916123228, 20.945216155232764, 16.880090589883931, 37.04196673404816]
20240415 19:57:27 valid_pixels shape: (100, 1)

bin4.

State AOSet: OK.

State AOPrepeared: OK.

State RunAO:

[AOeng@soul-dxwfs scripts]$ python reset_ocam.py
ocam2 interface loaded
Loop is Closed, Pausing
Loop was Closed, Re-Closing
Traceback (most recent call last):
File "reset_ocam.py", line 69, in
arbproxy.Resume()
File "/home/AOeng/live/aoroot/lib/python/aoArb_ice.py", line 1823, in Resume
return _M_aoArb.Loop._op_Resume.invoke(self, ((), _ctx))
AODefs.AOException: exception ::AODefs::AOException
{
code = -13006
what = [AOException] ResumeAo: WARNING - adsecarb:Illegal command for state Failure (ResumeAo Ao)
}

Adsec went in failure during the pause.

20240415 20:03:57 mean: [459.67933333333343, 465.75497777777798, 243.10315555555556, 485.95062222222214, 470.66684444444434, 475.85373333333331, 465.41706666666664, 480.14746666666673]
20240415 20:03:57 rms: [195.27240359130099, 166.11310413882637, 208.75012666374317, 158.19547176079701, 190.43423944886672, 174.15333931961229, 165.87591924410353, 160.10556578396972]
20240415 20:03:57 std: [89.930791796094184, 29.287621287406747, 18.643878618744619, 66.495273892425274, 73.375823564617789, 24.827882943968969, 19.813268484849729, 69.779810393200265]
20240415 20:03:59 Suspended because Ocam is reconfiguring
20240415 20:04:01 Suspended because Ocam is reconfiguring
20240415 20:04:03 Suspended because Ocam is reconfiguring
20240415 20:04:05 valid_pixels shape: (100, 1)
20240415 20:04:05 pixels_data shape: (100, 40000)
20240415 20:04:05 decimated pixels data shape: (50, 3600)
20240415 20:04:05 mean: [445.74031111111123, 465.61542222222198, 241.84880000000015, 472.50879999999967, 451.76231111111099, 477.58684444444441, 465.5207111111111, 467.6622666666666]
20240415 20:04:05 rms: [197.0290341929867, 167.9635161932483, 210.28364643505836, 155.9983300237335, 190.59404509859866, 175.90828686536869, 166.18595738661213, 157.76570555411735]
20240415 20:04:05 std: [65.828023805377981, 29.452165845689304, 18.618174079406121, 47.942603731751745, 36.751329154259679, 24.922130621773348, 19.728530530277236, 53.996358463609397]
20240415 20:04:07 valid_pixels shape: (100, 1)

AOPrepared.

20240415 20:16:10 mean: [374.57088888888887, 382.58880000000005, 156.93373333333332, 398.91497777777772, 391.02288888888887, 392.67457777777776, 376.29382222222216, 394.21311111111106]
20240415 20:16:10 rms: [185.43931748214601, 166.04716201308551, 158.59208565850625, 167.9008724343507, 145.39758004394187, 157.30741598768805, 175.45550475069504, 157.61480633151993]
20240415 20:16:10 std: [36.389579118504955, 33.575761955324069, 22.870147205561157, 13.965933085544153, 28.289558024867134, 32.772239327172464, 24.54378821719099, 16.012507303447151]
20240415 20:16:10 Alarm Code: 128 (Pixel means for each octant: 374.570888889, 382.5888, 156.933733333, 398.914977778, 391.022888889, 392.674577778, 376.293822222, 394.213111111)
20240415 20:16:12 Suspended because Ocam is reconfiguring
20240415 20:16:14 Suspended because Ocam is reconfiguring
20240415 20:16:18 valid_pixels shape: (100, 1)
20240415 20:16:18 pixels_data shape: (100, 40000)
20240415 20:16:18 decimated pixels data shape: (50, 3600)
20240415 20:16:18 mean: [376.75040000000007, 385.89297777777773, 161.1817777777778, 401.19733333333335, 392.35004444444445, 394.08968888888899, 377.85782222222224, 395.8930666666667]
20240415 20:16:18 rms: [178.30927705989555, 161.16

OK

runAO.Closed loop with zero gan.

same error in the adsec.Adsec to falure.

adsc_stop/start OK.

15 20:23:14.418578| MAIN > Received cmd: LoadGainTag("/local/aomeas/adsec_calib/M2C/KL_v29/gain/gain0.fits", 0)
AdSecArbitrator_27622|INF| 151|2024-04-15 20:23:14.418625| COMMANDHANDLER > FSM (status: AORunning) has received command 2018 (SetGain)
AdSecArbitrator_27622|WAR| 152|2024-04-15 20:23:14.418766| IDL > Tip-tilt gain is set to zero. Offload disabled until next setGain command
AdSecArbitrator_27622|INF| 153|2024-04-15 20:23:14.439285| COMMANDHANDLER > Command SetGain (code 2018) successfully completed
AdSecArbitrator_27622|INF| 154|2024-04-15 20:23:54.074872| MAIN > Received cmd: Offloading(0)
AdSecArbitrator_27622|INF| 155|2024-04-15 20:23:54.074947| COMMANDHANDLER > FSM (status: AORunning) has received command 2029 (TTOffload)
AdSecArbitrator_27622|INF| 156|2024-04-15 20:23:54.086250| COMMANDHANDLER > Command TTOffload (code 2029) successfully completed
AdSecArbitrator_27622|INF| 157|2024-04-15 20:23:54.086465| MAIN > Received cmd: PauseAo()
AdSecArbitrator_27622|INF| 158|2024-04-15 20:23:54.086490| COMMANDHANDLER > FSM (status: AORunning) has received command 2027 (Pause Ao)
AdSecArbitrator_27622|WAR| 159|2024-04-15 20:23:54.086758| IDL > Tip-tilt gain is set to zero. Offload disabled until next setGain command
AdSecArbitrator_27622|INF| 160|2024-04-15 20:23:54.099452| COMMANDHANDLER > Command Pause Ao (code 2027) successfully completed
AdSecArbitrator_27622|ERR| 161|2024-04-15 20:23:59.515181| ALERTHANDLER > Received an ERROR alert! Forcing Failure state.
AdSecArbitrator_27622|ERR| 162|2024-04-15 20:23:59.515221| ALERTHANDLER > Alert message: COILS DISABLED from housekeeper.R

python_27694 |DEB| 73|2024-04-15 20:23:54.241892| MAIN > Status after command: LoopSuspended
python_27694 |ERR| 74|2024-04-15 20:23:59.515693| MAIN > Alert from AdSec [0] - Error
python_27694 |DEB| 75|2024-04-15 20:23:59.515742| MAIN > Sending to 2 clients: Error

02:26PM

Stop testing.

running SETSAFE.

Mon Apr 15 21:27:47.386 2024 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[278] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.7,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Apr 15 21:27:47.403 2024 aos.info - WFS source set to: FLAOWFS
Mon Apr 15 21:27:47.983 2024 aos.info - FLAO: WFS: loading setup for LUCIFER

bin4 AO prepared.

WfsArbitrator_23585|DEB| 184|2024-04-15 21:32:58.018044| MAIN > arbFunc doSetSafe done.
WfsArbitrator_23585|DEB| 185|2024-04-15 21:32:58.018223| MAIN > arbFunc updateModulationInternal (, 'FLAOWFS', 'LUCIFER', 1000)
WfsArbitrator_23585|INF| 186|2024-04-15 21:32:58.018413| MAIN > Tip-tilt frequency is not synced, skipping update
WfsArbitrator_23585|DEB| 187|2024-04-15 21:32:58.018455| MAIN > arbFunc updateModulationInternal done.

The gui changes to ambient.

and starts to warm up.

Screen Shot 2024-04-15 at 2.33.43 PM.png

02:38

Power in off.

Adsec rest.

disable rr mode.

Ckear active optics.

-- JuanCarlosGuerra - 15 Apr 2024
Topic revision: r2 - 15 Apr 2024, JuanCarlosGuerra