Details.

changing soul verion of rlbtisx to:

test.20240223.zero-slope-firmware-and-repoint.

change the way to read the telemetry from the ocam.

parameters changes in ocam2k.conf.

LOOP_PERIOD 200 TO 20

LONG_WAIT 500 TO 50

SHORT_WAIT 50 TO 5

TELEMETRY_FACTOR 1 TO 1.

Fabio and Brandon made a new compilation.

A change in those parameters requires to power cycle the ocacrtl process.

The stop/start of the process shows some issues with the masterdiagnostic disconection.

Closing loop with Argos cal unt not a problem.

looking good so far.

Test done with drop every value to zero and the zero slope appears, but not "bad frames", it seems screw up the BCU.

back to factor 10 and no zero slopes events.

Close loop with the argos cal unit.

10:00 starting.

Doug will use the luciSX for NCPA at noon and he will use tyhe sw version that we are using.

Loop closed non issues.

Open loop

Ocam in safe mode.

12:00

Handover to doug for soul-luciSX.

The LBTSX still running in open loop with ocam in safe mode.

Move to soul-lbtidx

running the new software.

close loop OK.

No issues of bad pixels.

TN:20240229_202549.

-- JuanCarlosGuerra - 29 Feb 2024