Staff:

LBTO: Juan Carlos

Observer:

Summary.

Close loop test using the zero slope firmware. Acquire performances images with Luci and test the ncpa measurements. SharkVis way.

No events of zero slopes observed. Close loop OK.

Details.

Telescope released.

-8:00

Power on WFS and setup telescope +_ Argos swim arm.

Ocam OK.

test preset.

Wed Feb 7 15:25:34.381 2024 aos.info - Doing PresetFlat (LUCIFER)
Wed Feb 7 15:31:01.657 2024 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1272] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 7 15:31:01.672 2024 aos.info - WFS source set to: FLAOWFS
Wed Feb 7 15:31:02.801 2024 aos.info - FLAO: WFS: loading setup for LUCIFER
Wed Feb 7 15:31:10.624 2024 aos.info - FLAO: WFS: configuring CCDs

OK

Wed Feb 7 15:32:43.146 2024 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1273] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 7 15:32:43.161 2024 aos.info - WFS source set to: FLAOWFS
Wed Feb 7 15:32:43.200 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.6_L040.0_v15.0_ovfreq1700.00_sd3892.fits

CEnter pupils.

FLAOGUI

shows the error

2024-02-07 15:35:42.328654| -13006 - [AOException] centerPupils: WARNING - RETRY: Camera lens temperature too low. T= 5.26, minimum allowed is 7.00 (0) NO_ERROR

AOS log message doen not show any err.

wfsarb:

shows the error.

WfsArbitrator _55004|DEB| 254|2024-02-07 15:35:39.157342| MAIN > arbFunc doCenterPupils (, 'FLAOWFS', 'LUCIFER')
WfsArbitrator_55004|ERR| 255|2024-02-07 15:35:39.158294| MAIN > RETRY: Camera lens temperature too low. T= 5

Repeat preset when temp bcu47 is above 7C

Wed Feb 7 15:48:51.537 2024 aos.StartAO{FLAO}.complete - StartAO {FLAO}[1275]

TN+luci:_155147,_155251,_

no loight in ocam taking dark.

Code #32.

pause/resume.

Wed Feb 7 15:53:43.741 2024 aos.Pause{FLAO}.complete - Pause{FLAO}[1271]
Wed Feb 7 15:55:36.119 2024 aos.Resume{FLAO}.started - Resume{FLAO}[1271]
Wed Feb 7 15:55:36.298 2024 aos.Resume{FLAO}.failed - Resume{FLAO}[1271] - [AOException] RETRY: Star not found - loop still paused (-20002) WFSARB_STAR_WRONGFLUX
Wed Feb 7 15:55:53.379 2024 aos.Stop{FLAO}.started - Stop{FLAO}[1278] (msg:Stopped from operator console)

It seems that the time of taking the pause to the resume the flux of the star was dfferent.

close loop again

TN:_161406

TN+luci:_162314(subframe)

ncpa:

TN:_162450.

change to bin2

code 128

bin2

close loop

change back to bin1.

Wed Feb 7 16:41:36.590 2024 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1291] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 7 16:41:36.607 2024 aos.info - WFS source set to: FLAOWFS
Wed Feb 7 16:41:36.650 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.6_L040.0_v15.0_ovfreq1700.00_sd3892.fits

pause/resume.

-15,10

TN:_171143

ncpa:

TN:_171354

no ncpa

TN:_171427(no good)_171548

ed Feb 7 17:24:39.581 2024 aos.PresetAO{FLAO}.complete - PresetAO {FLAO}[1301] - dx:nan dy:nan slNull: f1:0.0 f2:1.0 freq:835 mag:0 nBins:3 nModes:90 r0:nan snMode:0 strehl:0 ttMod:0
Wed Feb 7 17:25:30.792 2024 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1302] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:14.8,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 7 17:25:30.810 2024 aos.info - WFS source set to: FLAOWFS
Wed Feb 7 17:25:30.849 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq1000.00_sd3892.fits

close loop

done.

stoAO

Wed Feb 7 17:29:29.775 2024 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1306] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:14.8,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 7 17:29:29.792 2024 aos.info - WFS source set to: FLAOWFS
Wed Feb 7 17:29:29.830 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.0000_sd3892.fits

mag 15.5

TN:_174149.

TN:_174243(50frames),_174649(100),_175724

ncpa:

_175840,_180343

150,10

TN:_180552,_180739(200).

pause/resume/pause/resume

taking dark for 200

no ncpa

TN:_181553

ncpa:

TN:_181816

change TT gain

TN:_182246, better psf.reduced the elongation

ncpa off.

TN:_182625

-15,10

TN:_182904

Wed Feb 7 18:33:42.230 2024 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[1314] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Wed Feb 7 18:33:42.246 2024 aos.info - WFS source set to: FLAOWFS
Wed Feb 7 18:33:42.284 2024 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.6_L040.0_v15.0_ovfreq1700.00_sd3892.fits

TN:_185251

TN:-192132

TN:_192415

pause/resume X5

done analysie.

z#5 -8

TN:_195146

se of images.

Z5=-8,z6=11

TN:_203136, luc.0023

zero

TN:_203256 0024

ncpa:

_203412 0025

optimize

TN:_204104,_205539

[-8,10]

_205620

[ncpa]

_205732

02:00

power off

-- JuanCarlosGuerra - 07 Feb 2024