SHARK VIS

1:23
  • Had a failure with the BCU2K that was causing the loop to go not live.
  • Resetting the BCU2K was the fix.
  • This happened on an attempted offset
  • WfsArbitrator _26601|INF| 1605|2024-01-10 01:23:51.994200| WFS > Executing pauseLoop...
  • WfsArbitrator _26601|INF| 1606|2024-01-10 01:23:51.994277| WFS > Python command: WfsArbScripts.pauseLoop(self, sensor = 'LBTIWFS', instr = 'SHARK')
    WfsArbitrator_26601|ERR| 1607|2024-01-10 01:23:54.117302| WFS > Command returns error : Reading shared buffer masterdiagnostic.R:OPTLOOPBUF (-5001) TIMEOUT_ERROR
  • Took a while to correctly diagnose. Jenny helped. We don't see this on LUCI, that I'm aware of.
1:54
  • Target => Io
  • Baysides: -42.6 -32.2 -27.55
  • IE 75.0
  • CA 144.0
  • Seeing about 1.1"
  • 20240110_015455 25500 frames WITH OCAM2K
    • | seeing from OL modes | 0.655074 |
    • Dimm seeing [arcsec] 1.092200
    • Dimm seeing elevation [arcsec] 1.155610
  • 20240110_015602 '' '' WITHOUT OCAM2K
    • | seeing from OL modes | 0.849314 |
    • Dimm seeing [arcsec] 1.170370
    • Dimm seeing elevation [arcsec] 1.131880
2:10
  • Had to re-close the loop
  • New dataset
  • 20240110_021312 No OCAM frames, they don't add anything
    • | seeing from OL modes | 0.496276 |
    • Dimm seeing [arcsec] 1.072530
      Dimm seeing elevation [arcsec] 1.187570
  • 20240110_021822
    • | seeing from OL modes | 1.008350 |
    • Dimm seeing [arcsec] 1.403780
      Dimm seeing elevation [arcsec] 1.253090
  • Something really screwy is going on.
  • 20240110_022108
    • | seeing from OL modes | 0.736974 |
    • Dimm seeing [arcsec] 1.130420
      Dimm seeing elevation [arcsec] 1.132640
  • 20240110_024722
    • | seeing from OL modes | 1.111844 |
    • Dimm seeing [arcsec] 1.352330
      Dimm seeing elevation [arcsec] 1.665670
2:59
  • Moving to new target
  • LkCa 15
  • Bin2 993 Hz Mag 13
  • They want bin 1, I'm going to say 12.6 mag and skip checking flux
  • Bin 1 684 Hz
  • Shows Mag 12.9
  • IE 79.0
  • CA 135.0
  • Datasets
  • 20240110_032250 20000 frames no OCAM2K frames
    • | seeing from OL modes | 0.773008 |
    • Dimm seeing [arcsec] 1.100730
      Dimm seeing elevation [arcsec] 1.086110
  • 20240110_032528
    • | seeing from OL modes | 0.869186 |
    • Dimm seeing [arcsec] 0.925739
      Dimm seeing elevation [arcsec] 0.869432
  • 20240110_032843
    • | seeing from OL modes | 0.794509 |
    • Dimm seeing [arcsec] 1.168740
      Dimm seeing elevation [arcsec] 1.240990
  • 20240110_033101
    • | seeing from OL modes | 0.921358 |
    • Dimm seeing [arcsec] 1.083690
      Dimm seeing elevation [arcsec] 0.992499
3:35
  • They wanted fewer modes, we're going to go to bin2, but that should do it.
  • Bin2 1207 Hz
  • Camera lens is at a limit, but I think that we can still go.
  • Nope, switching back to the previous configuration.
  • 20240110_035638
    • | seeing from OL modes | 0.739724 |
    • Dimm seeing [arcsec] 0.976559
      Dimm seeing elevation [arcsec] 1.063220
  • 20240110_040151
    • | seeing from OL modes | 0.810011 |
    • Dimm seeing [arcsec] 1.171860
      Dimm seeing elevation [arcsec] 1.029390
  • 20240110_040545
    • | seeing from OL modes | 0.786650 |
    • Dimm seeing [arcsec] 1.029040
      Dimm seeing elevation [arcsec] 1.024540
  • Boosted Tip-Tilt gain a lot, during this observation. SHARK had a hard time keeping the image centered.
4:30
  • Switching to MODS

-- GregoryTaylor - 10 Jan 2024
Topic revision: r3 - 11 Jan 2024, GregoryTaylor
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.