04:50

Power on WFS SX and DX.

SX.

Ocam

ok

DX
  • Ocam not lve.
  • reconfigure from wfs control gui.
  • Error Attachung to shared buffer masterdiagnotic.R.OPTLOOPBUF (-3509) BUF_NOT_AVAIL_ERROR
  • MASTERDIAGNTIC process down.
  • four bad octants.
  • ocam not live
  • masterdi process down.
  • restart process, ocam live
  • afer a few second master down, ocam not live
badframe log.

[AOeng@soul-dxwfs bin]$ tail -f BadFramesTask.log
self._interpolatedData = self._interpolateMissingData()
File "/usr/lib/python2.6/site-packages/oaalib/timeSeries.py", line 387, in _interpolateMissingData
interpData = func( vector)
File "/usr/lib64/python2.6/site-packages/scipy/interpolate/polyint.py", line 80, in __call__
y = self._evaluate(x)
File "/usr/lib64/python2.6/site-packages/scipy/interpolate/interpolate.py", line 590, in _evaluate
below_bounds, above_bounds = self._check_bounds(x_new)
File "/usr/lib64/python2.6/site-packages/scipy/interpolate/interpolate.py", line 622, in _check_bounds
raise ValueError ("A value in x_new is above the interpolation "
ValueError: A value in x_new is above the interpolation range.
20231219 00:09:00 Exception in task BadFramesTask loop: A value in x_new is above the interpolation range.

Traceback (most recent call last):
File "/home/AOeng/live/aoroot/lib/python/auxloops/auxTask.py", line 387, in run
self.calcUser()
File "/home/AOeng/live/aoroot/lib/python/auxloops/tasks/badFramesTask/badFramesTask.py", line 105, in calcUser
pixels_data = self.pixels().data()
File "/usr/lib/python2.6/site-packages/oaalib/timeSeries.py", line 133, in data
data = self._index( self._data(), *args, **kwargs)
File "/usr/lib/python2.6/site-packages/oaalib/timeSeries.py", line 339, in _data
self._interpolatedData = self._interpolateMissingData()
File "/usr/lib/python2.6/site-packages/oaalib/timeSeries.py", line 387, in _interpolateMissingData
interpData = func( vector)
File "/usr/lib64/python2.6/site-packages/scipy/interpolate/polyint.py", line 80, in __call__
y = self._evaluate(x)
File "/usr/lib64/python2.6/site-packages/scipy/interpolate/interpolate.py", line 590, in _evaluate
below_bounds, above_bounds = self._check_bounds(x_new)
File "/usr/lib64/python2.6/site-packages/scipy/interpolate/interpolate.py", line 622, in _check_bounds
raise ValueError ("A value in x_new is above the interpolation "
ValueError: A value in x_new is above the interpolation range.
20231219 00:09:02 Exception in task BadFramesTask loop: A value in x_new is above the interpolation range.

Traceback (most recent call last):
File "/home/AOeng/live/aoroot/lib/python/auxloops/auxTask.py", line 387, in run
self.calcUser()
File "/home/AOeng/live/aoroot/lib/python/auxloops/tasks/badFramesTask/badFramesTask.py", line 105, in calcUser
pixels_data = self.pixels().data()
File "/usr/lib/python2.6/site-packages/oaalib/timeSeries.py", line 133, in data
data = self._index( self._data(), *args, **kwargs)
File "/usr/lib/python2.6/site-packages/oaalib/timeSeries.py", line 339, in _data
self._interpolatedData = self._interpolateMissingData()
File "/usr/lib/python2.6/site-packages/oaalib/timeSeries.py", line 387, in _interpolateMissingData
interpData = func( vector)
File "/usr/lib64/python2.6/site-packages/scipy/interpolate/polyint.py", line 80, in __call__
y = self._evaluate(x)
File "/usr/lib64/python2.6/site-packages/scipy/interpolate/interpolate.py", line 590, in _evaluate
below_bounds, above_bounds = self._check_bounds(x_new)
File "/usr/lib64/python2.6/site-packages/scipy/interpolate/interpolate.py", line 622, in _check_bounds
raise ValueError ("A value in x_new is above the interpolation "
ValueError: A value in x_new is above the interpolation range.
20231219 00:09:13 Exception in task BadFramesTask loop: A value in x_new is above the interpolation range.


Save TN: 20231219_001113

Reconfigure again.

two bad quadrants.

TN:_001231

powercycle bcu2k

ocam ok

05:20

Telescope closed. Could cover

10:39 still closed.

cloud cover.

11:18

reconfgure to LuciSX.

badframe reported code 2

TN:20231219_053845

20231219 06:21:43 rms: [34.433144900021091, 36.597217463086366, 38.936670493601753, 34.474194260530723, 32.876869404745818, 37.75961703411469, 36.974561873680898, 33.684687756779574]
20231219 06:21:43 std: [2.4209716100485381, 3.8837330184731211, 4.2355035115191368, 2.1969054076213168, 1.9447960073878248, 4.1873436636592389, 3.7380312389989077, 1.9271613466511983]
20231219 06:21:43 Alarm Code: 2
20231219 06:21:46 mean: [96.76314722222223, 96.704208333333298, 97.254186111111125, 96.76488333333333, 96.210547222222232, 95.650377777777763, 96.568508333333341, 96.552250000000001]
20231219 06:21:46 rms: [34.347799028964943, 36.493071014657474, 38.904864896092143, 34.41001187888871, 32.79478724107372, 37.826180299052673, 36.867343672948323, 33.718652293304913]
20231219 06:21:46 std: [2.419186005107933, 3.8797568347236311, 4.226975943781957, 2.2031923388254473, 1.946768013986395, 4.1844268344527826, 3.7240549347254994, 1.9253181044528833]
20231219 06:21:46 Alarm Code: 2
20231219 06:21:48 mean: [96.759822222222226, 96.750755555555571, 97.318291666666667, 96.776855555555542, 96.215036111111118, 95.738427777777758, 96.627875000000003, 96.552894444444448]
20231219 06:21:48 rms: [34.342041788642661, 36.598024925079578, 38.920505748320529, 34.445843584770529, 32.807884420294954, 37.630904791143287, 36.958896542892738, 33.719029560450402]
20231219 06:21:48 std: [2.4208812636383543, 3.8875831817635036, 4.2176414000753146, 2.196903533152383, 1.9472490798829594, 4.1692037888105444, 3.7282135341107083, 1.9205752633025746]
20231219 06:21:50 mean: [96.816577777777781, 96.743308333333317, 97.269311111111108, 96.791863888888898, 96.247255555555526, 95.700186111111108, 96.632005555555551, 96.599508333333304]
20231219 06:21:50 rms: [34.496625301041789, 36.582197426031918, 38.750268319014495, 34.51613615906183, 32.914760458646143, 37.827837344122244, 36.760337696022646, 33.853506755028967]
20231219 06:21:50 std: [2.4240491147421039, 3.8490088958209561, 4.2074854871973741, 2.200175867593698, 1.9479988869374834, 4.175204401723648, 3.7055925132592646, 1.925246240903336]

The ocam viewer didn;t show any sloperms zeros.

WfsArbitrator _79884|ERR| 60|2023-12-19 06:21:43.818626| MAIN > OCAM Bad Frames/Slopes Error events detected during AOSet, code 2
WfsArbitrator_79884|ERR| 61|2023-12-19 06:21:43.818680| MAIN > OCAM bad frame description: Code 2 - Slopes are all 0
WfsArbitrator_79884|ERR| 62|2023-12-19 06:21:43.818696| MAIN > OCAM suggested action to recover: Change Bin back and forth via WFS control GUI; if required, power cycle BCU2k
WfsArbitrator_79884|ERR| 63|2023-12-19 06:21:46.079634| MAIN > OCAM Bad Frames/Slopes Error events detected during AOSet, code 2
WfsArbitrator_79884|ERR| 64|2023-12-19 06:21:46.079666| MAIN > OCAM bad frame description: Code 2 - Slopes are all 0
WfsArbitrator_79884|ERR| 65|2023-12-19 06:21:46.079678| MAIN > OCAM suggested action to recover: Change Bin back and forth via WFS control GUI; if required, power cycle BCU2k

12:01

starting luciS foucs check

run auto mode.

Tue Dec 19 07:24:23.439 2023 aos.Stop{FLAO}.complete - Stop{FLAO}[494]
Tue Dec 19 07:25:28.221 2023 aos.PresetAO{FLAO}.started - PresetAO {FLAO}[495] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Tue Dec 19 07:25:28.239 2023 aos.info - WFS source set to: FLAOWFS
Tue Dec 19 07:25:28.275 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.6_L040.0_v15.0_ovfreq1700.00_sd3892.fits

AutofocusLuci

Acquire 3 set of data the and it showed 51.43,51.3,51.53.Run in FeII subffame 256x256

same run wth the N375 camera it showed values around 51.4 looking at the pixel intensity. FeII and ND2, subframe 256x256

The run was done at loop closed.

03:45

No perspective to observe.

WFs power off.

N30 set.

/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0296.fits
peak: 14987
real pos: 50.4838
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0297.fits
peak: 17392
real pos: 50.6922
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0298.fits
peak: 18090
real pos: 50.9525
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0299.fits
peak: 19867
real pos: 51.0764
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0300.fits
peak: 20926
real pos: 51.3364
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0301.fits
peak: 22308
real pos: 51.4231
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0302.fits
peak: 23057
real pos: 51.699
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0303.fits
peak: 20923
real pos: 51.8558
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0304.fits
peak: 20846
real pos: 52.064
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0305.fits
peak: 19519

/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0316.fits
peak: 17196
real pos: 50.7255
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0317.fits
peak: 17611
real pos: 50.9118
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0318.fits
peak: 18351
real pos: 51.1103
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0319.fits
peak: 20666
real pos: 51.3399
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0320.fits
peak: 21170
real pos: 51.4331
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0321.fits
peak: 22275
real pos: 51.7158
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0322.fits
peak: 20496
real pos: 51.9655"
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0323.fits
peak: 19908
real pos: 52.0654
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0324.fits
peak: 19067
real pos: 52.4148


/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0306.fits
peak: 17489
real pos: 50.73

/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0307.fits
peak: 18272
real pos: 50.8998

/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0308.fits
peak: 18752
real pos: 51.1082

/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0309.fits
peak: 20946
real pos: 51.3367
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0310.fits
peak: 21756
real pos: 51.4779
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0311.fits
peak: 21632
real pos: 51.7084
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0312.fits
peak: 22005
real pos: 51.8588
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0313.fits
peak: 22078
real pos: 52.008
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0314.fits
peak: 19161
commanded pos: 52.3058962264
real pos: 52.3305
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0315.fits
peak: 17960

N375.


/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0360.fits
real pos: 50.7252
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0361.fits
real pos: 50.8662
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0362.fits
real pos: 51.0728
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0363.fits
real pos: 51.3452
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0364.fits
real pos: 51.4387
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0365.fits
real pos: 51.6907
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0366.fits
real pos: 51.9416
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0367.fits
real pos: 52.0767
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0368.fits
real pos: 52.3146
/dataA/readout1/DATA/20231219/cubes/luci1.20231219.0369.fits

-- JuanCarlosGuerra - 18 Dec 2023