Staff:

LBTO: Juan Carlos, Jenny

Observer:

Summary.

Power ON WFS, check functionality.

PDU web page:

http://lbti-web.lbti.lbto.org:8080/indi_pdus.html?dx_wfs_rack_pdu,llth_pdu1,llth_pdu2,llth_pdu3,nic_pdu1,nic_pdu2,sx_wfs_rack_pdu

Details.

Power on WFS.

Before the power on it requies to bring up the PDU back to On state.

Found all PDU off.

lbt_dx_pdu20230616.png

Jenny sent instruction for power on sequence.

"

Power up:

  • #3 DX cooling and Heat Exchanger
  • #4 DX Network Switch
  • #6 Flow Meters
Next:

  • #1 DX Port Server
  • #2 DX Perle Switch
  • #13 DX ADAM and Port Server #2
Wait at least 5 min (10 is better)

Next:

  • #14 DX ADC/Filter wheel Power
    followed by
  • #11 DX Bayside Zenus Motor Power
  • #12 DX JENA Camera Lens
Note that because the FW and Baysides were powered off they MUST be homed when you start up.

Lastly, the usual:

  • #15 DX CCD47 Tray Power
  • #8 DX PI WFS Tip Tilt Controller

  • #16 DX BCU2K 24 Volt Power Supply
  • #5 DX OCAM 24 Volt Power Supply

Do not power up #7 DX PI Tip-Tilt SPC Controller. This is for the SPC's in the UBC and should not be on if the UBC is warm!

  • #10 DX Cube stage and Translation stage
  • #9 DX Arcetri Source
You can power this up if you are using the retro but not needed otherwise. You can home the translation stage but do not home the Cube Stage. Homing the Cube Stage does work (for now) for DX but will fail for SX! I'm not sure which source is connected right now but best to leave off.

"

Port server On 10:40am

(wait 10 minutes)

Done all the PDUs.

[dxwunit@lbti-dxwfs source]$ w_start_all
stop: Unknown instance:
Side: R
Subsystem: WFS
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/msgdrtdb
clean_ts8 script is disabled
-- Stage mirrorctrl pinger slopecompctrl techviewctrl ocam2 ccd47 ttctrl adc1 adc2 filterwheel1 filterwheel2 mercury cubestage baysidex baysidey baysidez cameralensctrl optloopdiag pupilcheckocam2
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/MirrorCtrl -i mirrorctrl
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/Pinger -i pinger
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/SlopeCompCtrl2K -i slopecompctrl
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/FrameGrabberCtrl -i techviewctrl
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/OCAM2Ctrl -i ocam2
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/JoeCtrl -i ccd47
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/TTCtrl -i ttctrl
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/SimpleMotorCtrl -i adc1
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/SimpleMotorCtrl -i adc2
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/SimpleMotorCtrl -i filterwheel1
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/SimpleMotorCtrl -i filterwheel2
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/SimpleMotorCtrl -i mercury
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/SimpleMotorCtrl -i cubestage
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/CopleyCtrl -i baysidex
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/CopleyCtrl -i baysidey
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/CopleyCtrl -i baysidez
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/CameraLensCtrl -i cameralensctrl
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/OptLoopDiagnostic -i optloopdiag
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/pupilcheck.py -i pupilcheckocam2 ocam2
-- Stage masterdiagnostic housekeeperWFS dbfiller
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/MasterDiagnostic -i masterdiagnostic
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/housekeeperWFS -i housekeeperWFS
startProcessByName(): starting process /home/dxwunit/live/aoroot/bin/DBFiller -i dbfiller
Traceback (most recent call last):
File "/home/dxwunit/live/aoroot/bin/w_start", line 256, in <module>
if name == '__main__': main()
File "/home/dxwunit/live/aoroot/bin/w_start", line 144, in main
processControl.waitProcessStart( app, process)
File "/home/dxwunit/live/aoroot/lib/python/AdOpt/processControl.py", line 243, in waitProcessStart
raise AdOptError( AOErrCode['TIMEOUT_ERROR'], "Timeout error waiting for process %s start" % process)
AdOpt.AOExcept.AdOptError: Timeout error waiting for process housekeeperWFS start - TIMEOUT_ERROR
keepUp start/running, process 23485

[dxwunit@lbti-dxwfs source]$ w_check

msgdrtdb running
mirrorctrl running and connected to Msgd
pinger running and connected to Msgd
slopecompctrl running and connected to Msgd
techviewctrl running and connected to Msgd
ocam2 running and connected to Msgd
ccd47 running and connected to Msgd
ttctrl running and connected to Msgd
adc1 running and connected to Msgd
adc2 running and connected to Msgd
filterwheel1 running and connected to Msgd
filterwheel2 running and connected to Msgd
mercury running and connected to Msgd
cubestage running and connected to Msgd
baysidex running and connected to Msgd
baysidey running and connected to Msgd
baysidez running and connected to Msgd
cameralensctrl running and connected to Msgd
optloopdiag running and connected to Msgd
pupilcheckocam2 running and connected to Msgd
masterdiagnostic running and connected to Msgd
housekeeperWFS NOT running
dbfiller running and connected to Msgd
wfsarb NOT running
od_gopt NOT running
gopt NOT running

do another w_stop/start_all

OK.

BCU2K not connected.

Jenny power cycle the Perle ...
  • Connected.
All ON.....

FW and bayside stages INIT

CCD47 and OCAM live running ok.

12:32PM

Power off WFS.

-- JuanCarlosGuerra - 16 Jun 2023
Topic revision: r2 - 16 Jun 2023, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.