Details.

08:30

SoulSx power on

OCAM bad frame.

TN:20230423_154417.

Power cycle BCU2k

OK

Argos cal unit right side. The hardware status shows input parameters error.

TReset the hardware; still the same.

REstsrt the process-> same results of error.

two resets and it took some time to have the SimotionOK.

WFSDX

Power on

Ocam with one bad quadrnat.

TN:20230423_155112.

Power cycle bcu2k

still the bad quadratn.

Talk with Doug about it.

Continous with SX.

09:00

SX test ACE preset.

OK

SoulDX.
  • BCU47
  • time out. too long
  • WfsArbitrator_119002|DEB| 205|2023-04-24 15:52:47.284584| MAIN > arbFunc updateModulationInternal done.
  • WfsArbitrator_119002|DEB| 206|2023-04-24 16:03:39.574774| MAIN > arbFunc calibTVDarkInternal (, 'FLAOWFS', '', None, 10, True)
  • WfsArbitrator_119002|INF| 207|2023-04-24 16:03:39.574811| MAIN > calibTVDarkInternal(): starting measure, 10 frames

test preset SoulSX TTM.
  • mag10
  • OK
  • mag13
  • ok
presetACE.

Mon Apr 24 16:11:41.834 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3142] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:na

ok

09:18am

rotation

Bin1.

test stability with Bin1 3l/d,1500Hz.

OK.

Open loop.

| 5269|2023-04-24 16:32:58.118303| MAIN > COMM MAX [m] 6.88342e-05
idl_103510 |INF| 5270|2023-04-24 16:32:58.118346| MAIN > FORCE MAX [N] 0.208455
idl_103510 |INF| 5271|2023-04-24 16:32:58.579889| MAIN > Error in low level procedure CMD #5
idl_103510 |INF| 5272|2023-04-24 16:32:58.650577| MAIN > Timeout error
20232404 16:32:58.6517 Alert: Warning: stopAo failed or syntax error occured
AdSecArbitrator_103627|ERR| 1872|2023-04-24 16:32:58.650980| IDL > Command returns error : Timeout error
idl_103510 |INF| 5273|2023-04-24 16:32:58.650654| MAIN > FSM_LOAD_SHAPE: -----------> Error in low level procedure CMD #5 -5001
AdSecArbitrator_103627|INF| 1873|2023-04-24 16:32:58.675396| COMMANDHANDLER > Command execution failed: Fsm forced to Failure
AdSecArbitrator_103627|ERR| 1874|2023-04-24 16:32:58.675591| MAIN > StopAo: FATAL - Timeout error

AdSecArbitrator_103627|INF| 1871|2023-04-24 16:32:55.539847| SERIALIZATION > StopAoImpl(): restoreShape = 1
AdSecArbitrator_103627|ERR| 1872|2023-04-24 16:32:58.650980| IDL > Command returns error : Timeout error
AdSecArbitrator_103627|INF| 1873|2023-04-24 16:32:58.675396| COMMANDHANDLER > Command execution failed: Fsm forced to Failure
AdSecArbitrator_103627|ERR| 1874|2023-04-24 16:32:58.675591| MAIN > StopAo: FATAL - Timeout error

restart adsecarb.

ok

09:43

bin2

Mon Apr 24 16:43:26.356 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3144] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:13,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
  • shutter 65 mag 11.7
  • shutter 60 mag 13

bin3

shutter 60

Mon Apr 24 17:08:56.902 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3146] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:15,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

bin4

  • shutter 54 mag 14.4
  • shutter 49 mag 15.1
  • shutter 44 mag 16.6
Mon Apr 24 17:27:42.285 2023 aos.AcquireRef{FLAO}.complete - AcquireRef{FLAO}[3147] - dx:-0.487832 dy:1.32748 slNull: f1:0.0 f2:1.0 freq:846 mag:0 nBins:3 nModes:90 r0:nan snMode:0 strehl:0 ttMod:0
Mon Apr 24 17:30:01.083 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3148] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.6,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)

shutter 49

done ok

11:05 am

argos cal units parked.

adsec sx/dx rest.

there is a request from Mark to have access to the telescope; lights on and telescope can be tilted.

1:14pm

resume DX and SX activity.

24 20:14:08.889129| MAIN > Maximum mode index used to apply flat -> # 639

Mon Apr 24 20:17:44.312 2023 aos.PresetAO{FLAO}.complete - PresetAO{FLAO}[3150] - dx:nan dy:nan slNull: f1:0.0 f2:1.0 freq:1700 mag:0 nBins:1 nModes:501 r0:nan snMode:0 strehl:0 ttMod:0
Mon Apr 24 20:18:14.192 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3151] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Apr 24 20:18:14.206 2023 aos.info - WFS source set to: FLAOWFS
Mon Apr 24 20:18:14.251 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.6_L040.0_v15.0_ovfreq1700.00_sd3892.fits

optimize:20230424_202330

TN+Luci: 20230424_202723,_202843,_202933

ncpaON:_203023,_203046,_203112

luci1.20230424.0001-0009

dark:0009,0010

resume.

ncpaon:_203331,_203353

Mon Apr 24 20:40:36.272 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3157] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.55,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Apr 24 20:40:36.289 2023 aos.info - WFS source set to: FLAOWFS
Mon Apr 24 20:40:36.354 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.0000_sd3892.fits

shutter 43 mag 17.1

shutter 45 mag 16.2

optimize: _204519

TN+luci(J):_204826,_204935,_205012,_205212

dark.0021,22

TTM preset.

Mon Apr 24 20:52:43.032 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3162] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16.55,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Apr 24 20:52:43.045 2023 aos.info - WFS source set to: FLAOWFS
Mon Apr 24 20:52:43.133 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.0000_sd3892.fits

CCD47 images for mag 16.5 but it shows 16.2

TN:_205536 (10 frames)

soul/tmp/ average and sequence of 10 images

to check..

manula optimize: 0.95, 0.10

TN+Luci:_210057(elongated),_210130,_

tt0.60

TN+luvi:_210222,_210249

Mon Apr 24 21:04:57.407 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3169] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Apr 24 21:04:57.438 2023 aos.info - WFS source set to: FLAOWFS
Mon Apr 24 21:04:57.508 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq500.0000_sd3892.fits

cancel preset.

change dist freq.

Mon Apr 24 21:07:37.611 2023 aos.AcquireRef{FLAO}.complete - AcquireRef{FLAO}[3170] - dx:0 dy:0 slNull: f1:2.0 f2:1.0 freq:273 mag:0 nBins:4 nModes:13 r0:nan snMode:0 strehl:0 ttMod:0
Mon Apr 24 21:08:22.750 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[3171] (aomode:TTM-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:16,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Mon Apr 24 21:08:22.765 2023 aos.info - WFS source set to: FLAOWFS
Mon Apr 24 21:08:22.819 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.5_L040.0_v15.0_ovfreq200.0000_sd3892.fits

0.35,0.10,0.05

TN+luci:_211242,_211318

new region:_211539,_211606

tt 0.85

_211715,_211758,_212002

luci darks:0036,37

tt0.6

_212053,_212115,_212135

Stop

02:30PM

-- JuanCarlosGuerra - 24 Apr 2023