Staff:

LBTO: Xianyu, Brandon, Guido, Juan Carlos

Observer: NA

Summary.

Link of working plan.

https://docs.google.com/spreadsheets/d/1Zjn7o75mflXVb1JjjLzXRk_eAQRuzCGbWFc2LqCs5bo/edit#gid=0

Details.

08:56

telescope release for daytime testing.

switch soft versions for testing.

adsec sw: Currently running: test.20230302.offsetZ (soul)
wfs : test.20230302.offsetZ

argos connection: vncviewer -geometry 1920x1080 192.168.26.148:3


Set flat.

125322 |INF| 1346|2023-03-10 16:02:29.336256| MAIN > Error in low level procedure CMD #4
idl_125322 |INF| 1347|2023-03-10 16:02:29.554258| MAIN > Timeout error
idl_125322 |INF| 1348|2023-03-10 16:02:29.554362| MAIN > FSM_LOAD_SHAPE: -----------> Error in low level procedure CMD #4 -5001
AdSecArbitrator_125485|INF| 75|2023-03-10 16:01:36.306880| COMMANDHANDLER > FSM (status: Ready) has received command 2009 (Set Flat AO)

load shape again
  • 16:08:36.692875| MAIN > Maximum mode index used to apply flat -> # 639
Argos cal unit:
  • status input error.
  • It was required to restart the software from the MONIT
  • wait.
  • RESET from the swing arm.
  • OK
9:18

WFS power on

OCAM ok

stop LNfan47 and fan47 to warmup the bcu47

test preset.

Fri Mar 10 16:35:21.249 2023 aos.PresetAO{FLAO}.started - PresetAO{FLAO}[397] (aomode:ACE-AO,focst:bentGregorianFront,instr:LUCIFER,wfs:FLAOWFS,sox:0[0],soy:0,rox:0[0],roy:0,mag:10,cindex:0,,elev:nan,rotang:nan,grvang:nan,r0:0,skybr:nan,windsp:nan,winddir:nan)
Fri Mar 10 16:35:21.264 2023 aos.info - WFS source set to: FLAOWFS
Fri Mar 10 16:35:21.374 2023 aos.info - FLAO: WARNING: added atmospheric disturbance file dist_flao2_KL_v20_atm_s1.6_L040.0_v15.0_ovfreq1700.00_sd3892.fits

close loop.

there are a level of skip frames of aobut 5%. (?)

open loop

resend teh preset.

optimize: 20230310_170303

the ho2 can be reduce one step 0.8 can be the max.

Done

TN:20230310_170557

1 mm offsetz

Mar 10 17:08:46.172 2023 aos.OffsetZ{FLAO}.started - OffsetZ{FLAO}[402] (ofstz:1)
Fri Mar 10 17:08:48.473 2023 aos.info - TTF Offload request (FLAO) Z: -0.000522153 0.00990329 0.000833479
Fri Mar 10 17:08:51.408 2023 aos.info - TTF offload command end
Fri Mar 10 17:08:52.355 2023 aos.info - TTF offload: previous 3 requests ignored
Fri Mar 10 17:08:52.367 2023 aos.info - TTF Offload request (FLAO) Z: 0.0558935 0.104531 0.000768018
Fri Mar 10 17:08:54.909 2023 aos.info - TTF offload command end
Fri Mar 10 17:08:55.379 2023 aos.info - TTF offload: previous 2 requests ignored
Fri Mar 10 17:08:55.391 2023 aos.info - TTF Offload request (FLAO) Z: -0.0227674 -0.0306085 0.000646654
Fri Mar 10 17:08:57.207 2023 aos.OffsetZ{FLAO}.complete - OffsetZ{FLAO}[402]

TN(10k):_170839 start the acquire bedore to apply the command

move z-1

z=52.85

TN:_171138

Fri Mar 10 17:11:46.056 2023 aos.OffsetZ{FLAO}.started - OffsetZ{FLAO}[403] (ofstz:-1)
Fri Mar 10 17:11:47.785 2023 aos.info - TTF offload: previous 1 requests ignored
Fri Mar 10 17:11:47.806 2023 aos.info - TTF Offload request (FLAO) Z: -0.0194737 -0.0455048 -0.000420708
Fri Mar 10 17:11:51.109 2023 aos.info - TTF offload command end
Fri Mar 10 17:11:51.711 2023 aos.info - TTF offload: previous 3 requests ignored
Fri Mar 10 17:11:51.731 2023 aos.info - TTF Offload request (FLAO) Z: -0.0755586 -0.0904586 -0.000617261
Fri Mar 10 17:11:54.616 2023 aos.info - TTF offload command end
Fri Mar 10 17:11:54.802 2023 aos.info - TTF offload: previous 2 requests ignored
Fri Mar 10 17:11:54.820 2023 aos.info - TTF Offload request (FLAO) Z: -0.0971324 -0.0800561 -0.000500619
Fri Mar 10 17:11:57.393 2023 aos.OffsetZ{FLAO}.complete - OffsetZ{FLAO}[403]

move z=3mm

z=55.85

TN(20k)_171414

it moved by steps of 1mm

move z=-3mm

z=52.83

TN(20k)=_171539

possition of Z move back and forward

z=52.875

TN for ref: _171846

move offset15mm

Fri Mar 10 17:19:17.832 2023 aos.OffsetZ{FLAO}.started - OffsetZ{FLAO}[406] (ofstz:15)
Fri Mar 10 17:19:17.850 2023 aos.OffsetZ{FLAO}.failed - OffsetZ{FLAO}[406] - [Exception] 'NoneType' object is not callable

move -15mm

no moved.

Fri Mar 10 17:22:02.809 2023 aos.OffsetZ{FLAO}.started - OffsetZ{FLAO}[407] (ofstz:-15)
Fri Mar 10 17:22:02.826 2023 aos.OffsetZ{FLAO}.failed - OffsetZ{FLAO}[407] - [Exception] 'NoneType' object is not callable

disable TT offload.

Fri Mar 10 17:23:42.188 2023 aos.OffsetZ{FLAO}.started - OffsetZ{FLAO}[408] (ofstz:-1)
Fri Mar 10 17:23:42.203 2023 aos.OffsetZ{FLAO}.failed - OffsetZ{FLAO}[408] - Illegal command for state LoopClosed

enable TT offload

z=-5mm

moving

z=47.87

the PSF moved to the edge of the CCD47

Clean the oofload

PSF back to pyramid

apply TT gains and there are not movements in the Adsec, the loop is closed but it is not correcting.

preset

ok

Tn:_174240

z=52.9

move z=-5

hexapos: z=-1.1982mm

move z=5

hexapod: -1.165

TN:_175335

z=52.92

move z=-10

hexapod z=-1.22

z=42.909

Tn:_175844

10:59am

Ready for luci images.

There is an error Glenn is checking it.

Reboot of the GEIRS

TN+luci:_180210.

OK defocused.

move z=10

moving in steps of 1mm

saving 20k frames during the movement.

TN:_180448

TN:_180616

tn+luci_180715

hexapod:-1.66

loop pause

luci dark

luci.20230310.0024-33

Checking PSF, there is jitter and oscillation , 9not good day for PSF performanves.

TN+luci:_182125

11:32

move combine x and y by 0.,1

Fri Mar 10 18:24:27.403 2023 aos.OffsetXY{FLAO}.started - OffsetXY{FLAO}[421] (ofsx:0.1[0.1],ofsy:0.1)
Fri Mar 10 18:24:29.441 2023 aos.info - TTF offload: previous 1 requests ignored
Fri Mar 10 18:24:29.457 2023 aos.info - TTF Offload request (FLAO) Z: 0.848964 -0.447014 9.35499e-06
Fri Mar 10 18:24:32.435 2023 aos.OffsetXY{FLAO}.complete - OffsetXY{FLAO}[421]

TN:_182516

11:26

move xy -0.1

stop ao

18:33:01.232913| MAIN > Maximum mode index used to apply flat -> # 402

one act seems off.

rest.

Xianyu debugging an error in the offsetxy command, zero command is not accepted.

18:39:28.653515| MAIN > Maximum mode index used to apply flat -> # 639

11:40

PresetAO

optimize

184340

gopt.

TN+luci:_184534

11:53

loop pause.

z=52.89

z=1mm

z=53.92

z=0.5

z=54.39

z=-1

z=53.44

z=-0.5

z=52.86

z=1.5

error send not command, stopped in the aos.. As expecrted.

12:02

Stop AO

12:09

set safe

luci blind,blind

change use_soul to stable both adsec/wfs

w_stop_all

[AOeng@soul-sxwfs source]$ use_soul
Usage: use_soul [version]
Currently running: test.20230302.offsetZ (soul)
Available versions in /home/AOeng/soul:
stable
stable.20221117
stable.20230201
stable.20230202
test.20230302.offsetZ
test.20230303.ncpa-threshold
[AOeng@soul-sxwfs source]$ use_soul stable
Switched live version to /home/AOeng/soul/stable.
[AOeng@soul-sxwfs source]$ use_soul
Usage: use_soul [version]
Currently running: stable (soul)
Available versions in /home/AOeng/soul:
stable
stable.20221117
stable.20230201
stable.20230202
test.20230302.offsetZ
test.20230303.ncpa-threshold
[AOeng@soul-sxwfs source]$

w_start_all

w_idle

-- JuanCarlosGuerra - 10 Mar 2023
Topic revision: r1 - 10 Mar 2023, JuanCarlosGuerra
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback